2012年1月26日 更新者:Jung-Sun Kim、Yonsei University

ComparisiOn of Neointimal coVerage betwEen zotaRolimus Eluting Stent and Everolimus Eluting Stent Using Optical Coherence Tomography at 3 Months

This study try to 1) evaluate the neointimal coverage and malapposition at 3 month after new zotarolimus eluting stent (Endeavor resolute) and everolimus eluting stent (Xience) implantation and 2) compare them between ZES resolute and EES at 3 months (early period) after stent implantation.

調査の概要

詳細な説明

Recent autopsy study showed the most important histological and morphometric predictors of stent thrombosis were endothelial coverage and the ratio of uncovered to total strut after drug eluting stent (DES) implantation. Although there have been some controversies, incomplete stent apposition might be possible cause of very late stent thrombosis after DES implantation. Therefore, theoretically dual antiplatelet therapy should be continued to prevent the stent thrombosis until complete reendothelialization. But, detection of endothelial coverage over stent is not possible with available intravascular devices in clinical practice. Among currently available intravascular devices, intravascular optical coherence tomography (OCT) could give a more clear identification for a thin layer of neointimal hyperplasia (NIH) and malapposition with high-resolution (10 -20 μm) compared to intravascular ultrasound (100-150 μm). Previous OCT studies showed the significant different pattern of neointimal coverage between baremetal stent (BMS) and DES. In our experience, there were also some differences in neointimal coverage among the DESs, especially in zotarolimus eluting stent (ZES). ZES have been known to be associated with significantly more neointimal coverage than SES at 8 months intravascular untrasound (IVUS). This finding might be related with strut thickness and drug elution property. In our data, 99.7 % of stent struts were covered with neointima in ZES, but 93.2 % in sirolimus eluting stent (SES) at 9 month follow-up OCT. This difference was more distinct in patients with acute myocardial infarction (99.3 % in ZES vs.79.9 % in SES).

Both everolimus eluting stent (EES) was recently introduced and ZES resolute also will be available in the near future in Korea. Previous phosphorylcholine polymer was replaced with new, proprietary polymer which can provide extended elution kinetics in ZES resolute. Therefore, efficacy to suppress the neointimal growth might be improved but, safety for neointimal coverage needs to be evaluated in human coronary artery. Both new DESs are cobalt chrominum stent with thin stent strut and reported similar efficacy for suppressing the neointima growth. Also, endothelial coverage over stent strut occurred earlier compared to previous DESs (SES or PES) in animal experiments. But, there has been no data for neointimal coverage and malapposition of ZES resolute and EES using OCT in human coronary artery.

Recent paper according to Kubo et al reported that serial OCT examinations demonstrated markedly different vascular response up to 9 months after SES implantation between unstable angina and stable angina patients. These findings implied that initial native vessel status, which covered with stent, might be important and related with neoinitmal coverage and malapposition. But, there is little data evaluated the relationship between initial lesion status and neointimal coverage or late malapposition with OCT.

Therefore, this study will 1) evaluate the neointimal coverage and malapposition at 3 month after ZES resolute and EES implantation and 2) compare them between ZES resolute and EES at 3 months after stent implantation.

2. STUDY OBJECTIVE This study is an prospective open labeled randomized study to compare the neointimal coverage of ZES resolute (Endeavor® resolute) and EES (Xience®) in 3 month after stent implantation by OCT.

2.1. Primary end point The primary endpoint is to compare the pattern of neointimal coverage between ZES resoute (Endeavor® resolute) and EES (Xience®) at 3 months after stent implantation by OCT.

2.2. Secondary end point The secondary objective of this study is to evaluate neointimal growth between ZES resolute and EES at 3 months and 2) to compare the difference in early and late malapposition between ZES resolute and EES at 3 months.

3. Methods and Design

3.1 Patient enrollment Patients with stable angina pectoris who present to the cath lab for non-emergent PCI are eligible for participation. A patient is considered to be enrolled if they meet all of the inclusion and have none of the exclusion criteria.

Inclusion criteria

  • Significant coronary de novo lesion (> 70% by quantitative angiographic analysis)
  • Patients with stable angina pectoris who are considered for coronary revascularization.
  • Reference vessel diameter of 2.5 to 3.5 mm by operator assessment

Exclusion criteria

The criteria for exclusion were

  • Acute coronary syndrome
  • Contraindication to anti-platelet agents
  • Proximal leison within 15 mm from ostium
  • Treated with any DES within 1 year at other vessel
  • Creatinine level 2.0 mg/dL or ESRD
  • Severe hepatic dysfunction (3 times normal reference values)
  • Pregnant women or women with potential childbearing
  • Complex lesion morphologies (aorto-ostial, unprotected Left main, chronic total occlusion)
  • Target lesion is vein graft lesion
  • Reference vessel < 2.5 mm or > 4.0mm diameter by visual

3.2. Sample Size and Statistical analysis

Calculation of the sample size could not be done because of observation study. Therefore, we decided to enroll 40 patients for 20 ZESs resolute and 20 EESs considering sufficient stent strut (over 3000 stent struts each group).

Results are expressed as the mean ±SD. Comparisons of discrete variables were made using the chi-square test or Fisher's Exact Test. Student t-tests were used for continuous variables. Because the distribution was skewed due to a small study sample, non-parametric tests (Mann-Whitney test for unpaired variables and Wilcoxon-signed rank test for paired variables) were used. All statistical analysis was performed with SPSS 13.0. (SPSS Inc., Chicago, IL, USA). A value of p<0.05 was considered statistically significant.

4. STUDY PROCEDURE

The cath lab will follow site standard operating procedures for PTCA and stenting. Stent overlapping is not allowed. Treatment of multiple vessels is acceptable. However, information will be collected on only one study lesion per patient.

Information will be collected and recorded on the treated study vessel including angiographic, IVUS and OCT parameters. The IVUS and OCT study should be checked in the procedure and follow up at 3 months angiography.

4.1. PROCEDURAL MATERIALS Endeavor ® resolute 2.75 to 3.5 mm All available length Drug Eluting Stent Xience ® 2.75 to 3.5 mm All available length

4.2 RANDOMIZATION Patients with native coronary arteries who fulfill all enrollment criteria will be randomized to receive either ZES (Endeavor®) resolute or EES (Xience or Promus®). Randomization will be performed in a 1:1 ratio of ZES or EES according to each clinical presentation.

4.3 STENT SELECTION The vessel size will be determined either by quantitative angiography or by IVUS for appropriate stent size selection.

4.4 IVUS ASSESSMENT IVUS assessment was performed using a commercially available IVUS system (Boston Scientific/ Scimed, Natick, MA). Studies were recorded on Index system (Echoplaque 2; INDEC Systems Inc, Mountain View, CA) for off-line analysis. Motorized transducer pullback permitted cross-sectional area measurements at 1mm axial increments throughout the length of the stent. All IVUS studies were performed after the intracoronary administration of 200 µg of nitroglycerin. The reference segment was the most visually normal cross section (largest lumen with least plaque burden) within 10mm proximal or distal to the lesion. Using the software, external elastic membrane (EEM) CSA (mm2), lumen CSA (mm2), minimal stent cross-sectional area (MSA); identified as the CSA at the tightest segment within stent, stent MLD were determined. The CSA at the proximal and the distal reference segments and at the most stenotic lesion as well as the CSA at the proximal and the distal stent edge was determined. A determination of the percent expansion was made by measuring the MSA and the average reference segment CSA. Average reference segment CSA is calculated as [Proximal Reference segment CSA +Distal Reference segment CSA)/2]. Plaque volume was calculated as vessel minus stent volume. Neointimal volume was calculated as stent minus lumen volume, and percentage (%) of neointimal obstruction was calculated as neointimal volume divided by stent volume. Neointimal coverage over stent surface and mean physical neointimal thickness were obtained by measuring circumferential stent length covered with neointimal hyperplasia (LN) and stent perimeter (LS) at every 1-mm cross-sectional image throughout the stented segment: (a) percentage of neointimal coverage over stent surface, calculated as total LN divided by total LS, and (b) mean physical neointimal thickness, calculated as mean neointimal area divided by mean LN for only the cross sections with detectable neointima.

4.5. OCT ASSESSMENT OCT measurements will be made as following methods: (Takano M, et al. Evaluation by Optical Coherence Tomography of Neointimal Coverage of Sirolimus-Eluting Stent Three Months After Implantation. Am J Cardiol 2007;99: 10331038)

Cross-sectional OCT images were analyzed at 1-mm intervals (every 15 frames). Neointimal hyperplasia (NIH) thickness inside stent struts was measured. Stent area and lumen area in every image were measured by manual trace and percent NIH area was calculated as percent NIH area = ([stent area - lumen area]/ stent area) / 100. Measured NIH thickness equal to 0 micro meter was defined as exposure, and maximum distance > 110 micro meter between the strut surface and adjacent vessel surface (thickness of strut and polymer) was defined as malapposition. On an image of stent malapposition, lumen area was divided into an in-stent lumen area and an extra-stent lumen area. . Intracoronary thrombus was defined as a protruding mass beyond the stent strut into the lumen. Bifurcation lesions with major sidebranches were excluded from this analysis. Inter- and intraobserver variabilities were assessed by evaluation of 20 random cross- sectional images by 2 independent readers and by the same reader at 2 separate time points, respectively.

4.6. Follow-Up

All patients will be followed after hospital discharge. When patients come back in 3 and 9 months after the discharge due to clinical follow-up, the patient record will be updated.

1. Clinical follow-up at 3, 9 months and 2 years after the procedure.

研究の種類

介入

入学 (実際)

40

段階

  • フェーズ 4

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究場所

      • Daegu、大韓民国、700-712
        • Dongsan Medical Center, Keimyung University College of Medicine
      • Seoul、大韓民国、120-752
        • Division of Cardiology, Cardiovascular Hospital, Yonsei University

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

20年~90年 (大人、高齢者)

健康ボランティアの受け入れ

いいえ

受講資格のある性別

全て

説明

Inclusion Criteria:

  • Significant coronary de novo lesion (> 70% by quantitative angiographic analysis)
  • Patients with stable angina pectoris who are considered for coronary revascularization.
  • Reference vessel diameter of 2.5 to 3.5 mm by operator assessment

Exclusion Criteria:

  • Acute coronary syndrome
  • Contraindication to anti-platelet agents
  • Proximal leison within 15 mm from ostium
  • Treated with any DES within 1 year at other vessel
  • Creatinine level 2.0 mg/dL or ESRD
  • Severe hepatic dysfunction (3 times normal reference values)
  • Pregnant women or women with potential childbearing
  • Complex lesion morphologies (aorto-ostial, unprotected Left main, chronic total occlusion)
  • Target lesion is vein graft lesion
  • Reference vessel < 2.5 mm or > 4.0mm diameter by visual

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 主な目的:処理
  • 割り当て:ランダム化
  • 介入モデル:並列代入
  • マスキング:なし(オープンラベル)

武器と介入

参加者グループ / アーム
介入・治療
実験的:ZES resolute (Endeavor Resolute)
2nd generation drug eluting stent, which is coated with zotarolimus
他の名前:
  • たゆまぬ努力
アクティブコンパレータ:EES (Xience)
2nd generation drug eluting stent, which is coated with everolimus
他の名前:
  • ザイエンス

この研究は何を測定していますか?

主要な結果の測定

結果測定
時間枠
The rate of neointimal coverage between ZES resolute and EES at 3 months after stent implantation by OCT.
時間枠:3 months after stent implantation
3 months after stent implantation

二次結果の測定

結果測定
時間枠
Neointimal growth between ZES resolute and EES at 3 months after stent by OCT.
時間枠:3 months after stent implantation
3 months after stent implantation
Early malapposition just after stent implantation and late malapposition at 3 months after stent implantaiton between ZES resolute and EES by OCT.
時間枠:3 months after stent implantation
3 months after stent implantation

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

協力者

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始

2009年8月1日

一次修了 (実際)

2010年8月1日

研究の完了 (実際)

2010年8月1日

試験登録日

最初に提出

2010年3月22日

QC基準を満たした最初の提出物

2010年3月23日

最初の投稿 (見積もり)

2010年3月24日

学習記録の更新

投稿された最後の更新 (見積もり)

2012年1月31日

QC基準を満たした最後の更新が送信されました

2012年1月26日

最終確認日

2012年1月1日

詳しくは

本研究に関する用語

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する