SynCardia Companion 2 Driver System Post Approval Study Protocol With INTERMACS™-Based Data Collection (C2 PAS)
Study Purpose
The purpose of this PAS is to confirm that patients implanted with the temporary Total Artificial Heart (TAH-t) and supported with the Companion 2 Driver System have outcomes similar to those patients implanted with the TAH-t and supported with the CSS Console. The data collection mechanism for this study is the INTERMACS Registry.
Study Objectives
Primary Objective 1:
To confirm that patients implanted with the TAH-t and supported by the Companion 2 Driver System have outcomes similar to those patients implanted with the TAH-t and supported by the CSS Console.
Primary Objective 2:
To compare positive outcome rates for all TAH-t patients entered in the Intermacs Registry initially supported with the Companion 2 Driver System to positive outcome rates for all TAH-t patients initially supported by the CSS Console.
Primary Objective 3:
To review additional data (patient survival, adverse event rates and bridge to transplant rates) to support the conclusion that the use of an artificial heart is reasonable and necessary for the treatment of irreversible biventricular failure in Medicare/Medicaid patients.
Secondary Objective:
- To evaluate the incidence of adverse events (AEs). The AEs will follow the definitions in the INTERMACS Registry.
Study Hypothesis: Primary Objective 1 The primary hypothesis is that the positive outcome rate (defined as transplant , transfer to the Freedom Driver System support, or continuing on implant driver support, whichever occurs first; assessed at three and six months post-implant) for patients supported with the Companion 2 Driver System is non-inferior to the survival rate for patients supported with the Circulatory Support System (CSS) Console.
Data Collection The data will be collected as defined in the INTERMACS Protocol and the patients will be followed at one week, one month, three months and six months or until their study endpoint is reached.
Data Analysis
Primary Objective 1:
Data analysis will provide descriptive statistics on enrollment, adverse events, and outcomes for TAH-t patients who were implanted while supported with the Companion 2 Driver System compared to TAH-t patients who were implanted while supported with the CSS Console.
Primary Objective 2:
Data analysis will provide descriptive statistics on enrollment, adverse events, and outcomes for all TAH-t patients enrolled in the INTERMACS Registry.
Phase I as described above was completed; Phase II of Study design is under review by FDA with anticipated initiation in early 2021.
調査の概要
状態
研究の種類
入学 (予想される)
参加基準
適格基準
就学可能な年齢
- 子
- 大人
- 高齢者
健康ボランティアの受け入れ
受講資格のある性別
サンプリング方法
調査対象母集団
説明
Inclusion Criteria:
- Implanted with the TAH-t
- Patient or their legal representative has signed an informed consent (ICF) form for INTERMACS Registry participation or waiver of consent has been approved at the implanting institution.
Exclusion Criteria:
- Patient or legal guardian has not signed an ICF for INTERMACS Registry participation, if required at the institution.
- Patient is incarcerated.
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 観測モデル:ケースコントロール
- 時間の展望:断面図
コホートと介入
グループ/コホート |
介入・治療 |
|---|---|
|
CSS Console TAH-t Patients
All TAH-t patients implanted while supported with the CSS Console who were enrolled in the INTERMACS Registry and implanted with the TAH-t on or after June 20, 2012.
|
Comparison of two pneumatic drivers supporting the SynCardia TAH-t.
他の名前:
|
|
C2 Driver System TAH-t Patients
200 TAH-t patients implanted while supported by the Companion 2 (C2) Driver System who were enrolled in the INTERMACS Registry and implanted with the TAH-t on or after June 20, 2012.
|
Comparison of two pneumatic drivers supporting the SynCardia TAH-t
他の名前:
|
|
All TAH-t Patient Records
All records for TAH-t patients enrolled in the INTERMACS Registry will be reviewed to support Objective 2 of the protocol.
|
Comparison of two pneumatic drivers supporting the SynCardia TAH-t.
他の名前:
Comparison of two pneumatic drivers supporting the SynCardia TAH-t
他の名前:
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Positive outcome rates at three months and six months Post TAH-t Implant (Concurrent Cohort)
時間枠:Three and Six Months Post-TAH-t Implant
|
The positive outcome rates, defined as transplant, transfer to the Freedom Driver System, or continuing on implant driver support (whichever occurs first) for patients implanted on or after June 20, 2012 and supported with the Companion 2 Driver System will be compared to the positive outcome rates at three and six months for patients implanted on or after June 20, 2012 and supported with the Circulatory Support System (CSS) Console. The analysis will include a lower one-sided 95% confidence bound on the observed survival rate for patients supported with the Companion 2 Driver system which will be compared to the CSS Console survival rate to evaluate non-inferiority. |
Three and Six Months Post-TAH-t Implant
|
|
Positive outcome rates at three months and six months Post TAH-t Implant (Cumulative Cohort)
時間枠:Three and Six Months Post-TAH-t Implant
|
The positive outcome rates, defined as transplant, transfer to the Freedom Driver System, or continuing on implant driver support (whichever occurs first) for patients supported with the Companion 2 Driver System will be compared to the positive outcome rates at three and six months for patients supported with the Circulatory Support System (CSS) Console. The analysis will include a lower one-sided 95% confidence bound on the observed survival rate for patients supported with the Companion 2 Driver system which will be compared to the CSS Console survival rate to evaluate non-inferiority. |
Three and Six Months Post-TAH-t Implant
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Review of all TAH-t patient records in the INTERMACS Registry to confirm acceptability of treatment for bi-ventricular failure.
時間枠:Three and Six Months Post-TAH-t Implant
|
To review additional data (patient survival, adverse event rates and bridge to transplant rates) to support the conclusion that the use of an artificial heart is reasonable and necessary for the treatment of irreversible biventricular failure as ordered by the CMS National Coverage Decisions which requires enrollment of Medicare patients in approved studies for evidence development.
|
Three and Six Months Post-TAH-t Implant
|
その他の成果指標
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Comparison of Adverse Event Rates
時間枠:Three and Six Months Post-TAH-t Implant or Study Endpoint, whichever occurs first.
|
To evaluate the incidence of adverse events (AEs).
The AEs will follow the definitions in the INTERMACS Registry.
|
Three and Six Months Post-TAH-t Implant or Study Endpoint, whichever occurs first.
|
協力者と研究者
スポンサー
捜査官
- 主任研究者:Francisco Arabia, MD、Banner University Medical Center
研究記録日
主要日程の研究
研究開始
一次修了 (予想される)
研究の完了 (予想される)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (見積もり)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。