このページは自動翻訳されたものであり、翻訳の正確性は保証されていません。を参照してください。 英語版 ソーステキスト用。

Surgical Revascularisation and Nerve Stimulation Trial (SRANS)

A Randomised Control Trial to Establish if Use of the gekoTM (Trademark) Post Infra-inguinal Surgical Vein Revascularisation Increases Flow Through the Graft

A study to investigate if the gekoTM device improves flow through vascular bypass grafts

調査の概要

状態

引きこもった

介入・治療

詳細な説明

The geko™ device, (CE 558928;British Standards Institute notified body 0086), wearable, discrete, self-contained stimulation unit that adheres to the skin (see www.gekodevices.com). The, geko™ produces small electrical impulses that gently activate the common peroneal nerve within the popliteal fossa, behind the knee, in turn activating the venous muscle pumps of the calf and foot. Substantive increases in lower limb blood flow have been demonstrated to improve arterial, venous and microcirculatory blood flow with no reported morbidity.

RATIONALE Early infra-inguinal graft failure, i.e. occurring within 30 days of surgery occurs in approximately 5% of patients16. It is known that the predominant cause of this is thrombosis. The geko™ is an electrical device placed over the common peroneal nerve alongside the knee. The device activates muscle contraction in the peronei using low intensity, low frequency (1Hz) stimulation. The muscle contraction expresses blood from the deep venous system. Application of the stimulation device over the peroneal nerve is suitable for practical purposes. Previous studies have shown that the geko™ substantively increases arterial, venous and microcirculatory blood flow in healthy volunteers. A significant decrease in Tissue Plasminogen Activator antigen has also been demonstrated which indicates increased fibrinolytic activity3. Application of the geko™ device should therefore decrease stasis and increase flow though the graft and as such reduce the risk of thrombosis and as such graft failure.

Study:

Environmental Conditions The examinations will be carried out in a designated temperature controlled clinical room in the Vascular Lab at Hull Royal Infirmary. Participants will be required to abstain from vigorous exercise, nicotine, caffeine and fatty foods prior to their examination. They will be required to wear light garments with their legs exposed.

The geko™ device will be applied to outer/ posterior aspect of the knee whilst the subjects are semi-recumbent, with their knees flexed. The participants will be given 30 minutes in the quiet, environmentally controlled room prior to commencement of measurement to enable equilibration.

Assessments Participants will complete the McGill Pain questionnaire prior to device activation Participants will have the geko™ device applied whilst semi-recumbent with their knees flexed. The intensity will be increased until visible muscle stimulation is evident (concentric isotonic contractions) or until patient tolerance is reached

  • Laser Doppler imaging will be used to assess microcirculation flow on the dorsum of both feet at:

    1. Baseline (rest)
    2. 5 minute intervals while the geko™ device is active
    3. 30 minutes following removal of the device Laser Doppler flowmetry is a reproducible, objective, non-invasive measurement which can be used to evaluate real time cutaneous blood flow. The low powered light is conducted via glass fibres to a probe attached to the skin and is scattered by the movement of blood cells inducing a Doppler shift. This allows information on flux and cell velocity to be recorded. The system to be used for this trial is the Moor Instruments DT-4 mains unit with integrated bilateral lasers.
  • Ultrasound Doppler arterial flow velocity and blood volume flow within the femoral artery (mid thigh) bilaterally at:

    1. Baseline (rest)
    2. 30 minutes while the device is active
  • Ultrasound Doppler flow velocity in femoral vein (mid thigh) bilaterally at:

    1. Baseline (rest)
    2. 40 minutes while the geko is active
  • Vicorder measurements of arterial stiffness, pulse wave velocity and stroke volume at:

    1. Baseline (rest)
    2. Immediately following removal/ deactivation of the device
  • Vascular Endothelial Growth Factor (VEGF), Tissue Plasminogen Activator antigen (TPA antigen), Full blood count (FBC) and Plasminogen activator inhibitor-1 (PAI-1): 15 mls of blood to be taken from both femoral veins at:

    1. Baseline (rest)
    2. 45 minutes while the geko™ device is active ELISA analysis will be carried out on samples (see appendix). Vascular Endothelial Growth Factor - Therapies which can lead to the development of new, collateral blood vessels, are of particular interest in the field of peripheral arterial disease where narrowing or blockages in existing blood vessels result in tissue ischaemia. Vascular endothelial growth factor (VEGF) is a signal protein produced by cells that stimulates the growth of new blood vessels from pre-existing vessels. Several previous studies have shown that electrical stimulation increases VEGF levels and in turn increases vessel density in muscles.

Tissue plasminogen activator (tPA) is a protein involved in the breakdown of blood clots. As a result it has therapeutic uses in clinical medicine to treat thromboses and emboli. Electrical stimulation has been previously shown to provoke an acute release of tPA into the circulation, although it appears that this response decreases with repeated stimulations. Use of the geko™ device has been shown to result in a significant decrease in Tissue Plasminogen Activator (tPA) antigen which is indicative of increased fibrinolytic activity.

Plasminogen activator inhibitor- 1 (PAI-1) is the primary inhibitor of plasminogen activators and inactivates tissue plasminogen activator (t-PA) and urokinase-type plasminogen activator (u-PA). PAI-1 is an important inhibitor of the fibrinolytic system, and elevated levels could suppress fibrinolysis and result in an increased risk of thrombosis. Increased PAI-1 levels have been shown to be associated with a number of atherosclerotic risk factors, PAI-1 has been shown to act as a prothrombic factor in both arterial and venous thromboembolic disorders.

  • Ankle Brachial Pressure Index at baseline and post removal of geko™- utilising both Posterior Tibial Artery and Anterior Tibial Artery
  • Maximum calf circumference at baseline and post removal of geko™
  • Three blood pressure and heart rate measurements will be taken bilaterally at baseline and @ 30 minutes while the device is active. The mean of the three values will be used for comparison.
  • Patient tolerance/ acceptability + Nurse tolerance/ ease of application Questionnaire using Visual Analogue Scores and Verbal Rating Scores to be completed following removal of geko™ device (see appendix)
  • The McGill Pain Questionnaire to be completed following device removal

研究の種類

介入

段階

  • 適用できない

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究場所

    • EastRiding Of Yorkshire
      • Hull、EastRiding Of Yorkshire、イギリス、HU3 2JZ
        • Hull Royal Infirmary

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年歳以上 (大人、高齢者)

健康ボランティアの受け入れ

いいえ

受講資格のある性別

全て

説明

Inclusion Criteria:

  1. Age ≥18 years,
  2. Undergoing infra-inguinal vein bypass graft.
  3. Intact cutaneous sensations to nociception in the lower limb, as determined by the investigator.
  4. Intact healthy skin at site of application.
  5. On effective contraception if sexually active - oral contraceptive pill (> 3 months use), condoms, intrauterine contraceptive device, depot injection.
  6. Able to understand the Patient Information Sheet and capable and willing to give informed consent and follow the protocol requirements

Exclusion Criteria:

  1. History of haematological disorder or DVT within the preceding 6 months,
  2. Pregnant or planning to become pregnant during study duration,
  3. Pacemakers or implantable defibrillators,
  4. Use of any other neuro-modulation device,
  5. Current use of TENS in pelvic region, back or legs,
  6. Use of investigational drug/device therapy within past 4 weeks that may interfere with this study,
  7. Significant varicose veins or lower limb ulceration.
  8. Recent surgery (such as abdominal, gynaecological, hip knee replacement).
  9. Recent trauma to lower limbs.
  10. Chronic Obesity (BMI Index >34).
  11. Any medication judged to be significant by the Principal Investigator.
  12. Any significant illness during the four (4) weeks preceding the screening period of the study.

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 主な目的:防止
  • 割り当て:ランダム化
  • 介入モデル:並列代入
  • マスキング:なし(オープンラベル)

武器と介入

参加者グループ / アーム
介入・治療
介入なし:介入なし
アクティブコンパレータ:GekoTM
geko applied to bypass limb
electrical neuromuscular stimulation

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Change in arterial blood volume flow in the bypass graft measured by ultrasound.
時間枠:1 day
Measured by ultrasound duplex
1 day

二次結果の測定

結果測定
メジャーの説明
時間枠
To assess effect on blood flow in the superficial femoral vein
時間枠:1 day
Measured by ultrasound doppler
1 day
To assess effect on microcirculation
時間枠:1 day
Measured using laser Doppler flowmetry
1 day
To assess effect on pulse wave velocity, cardiac output and stroke volume
時間枠:1 day
Measured using using the Vicorder device.
1 day
To assess effect on levels of VEGF, PAI-1 and tPA antigen.
時間枠:1 day
Blood tests and ELISA
1 day
To assess the impact of gekoTM device on pain score utilising the McGill Pain score
時間枠:1 day
Validated pain questionnaire
1 day
Patient tolerance/ acceptability
時間枠:1 day
Questionnaire using Visual Analogue Scores and Verbal Rating Scores to be completed following removal of geko™ device
1 day
To assess effect on calf circumference
時間枠:1 day
Measurement pre and post device
1 day

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

捜査官

  • スタディチェア:Rachel Barnes, MRCS, MA、Hull Royal Infirmary

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始

2013年9月1日

一次修了 (実際)

2014年12月1日

研究の完了 (実際)

2014年12月1日

試験登録日

最初に提出

2013年9月5日

QC基準を満たした最初の提出物

2013年9月9日

最初の投稿 (見積もり)

2013年9月10日

学習記録の更新

投稿された最後の更新 (実際)

2019年7月15日

QC基準を満たした最後の更新が送信されました

2019年7月11日

最終確認日

2019年7月1日

詳しくは

本研究に関する用語

追加の関連 MeSH 用語

その他の研究ID番号

  • 12/YH/0480

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する