Feasibility Study: Continuous ECG and Blood Pressure Monitoring With the M2VS Sensor System

2026年7月3日 更新者:University Hospital, Basel, Switzerland

Feasibility Study for Continuous Monitoring of Electrocardiogram (ECG) and Blood Pressure at the Chest and Upper Arm Using the M2VS System.

In this study, the investigators want to find out whether a new wearable sensor system, called the M2VS system, can reliably measure heart activity, blood pressure, and other vital signs over 24 hours during normal daily life.

The investigators also want to compare the M2VS system with the standard devices currently used for long-term heart and blood pressure monitoring, such as a Holter ECG and an ambulatory blood pressure monitor (ABPM). This will provide an estimation of the accuracy of the M2VS system's ECG and blood pressure measurements compared to the standard devices.

The aim of the study is to assess how reliable the M2VS system is in terms of data completeness and how comfortable it is to wear compared with the usual monitoring devices.

In the future, the results of this study may help make the monitoring of heart and blood vessel diseases more comfortable and easier for patients.

調査の概要

詳細な説明

Cardiovascular diseases remain a leading cause of morbidity and mortality worldwide. Long-term ambulatory monitoring of electrocardiography (ECG) and blood pressure is essential for the detection and management of conditions such as paroxysmal atrial fibrillation and masked hypertension. However, current standard monitoring devices, including Holter ECG monitors and ambulatory blood pressure monitors, may be limited by reduced wearing comfort, skin irritation from adhesive electrodes, motion artefacts, and discontinuous blood pressure measurements.

This feasibility study evaluates a novel wearable sensor system, the M2VS system, designed as a flexible textile-based band integrating dry electrodes and multiple sensors for continuous multimodal vital signs monitoring. The system records ECG, photoplethysmography (PPG), inertial measurement unit (IMU) data, body temperature, bioimpedance, acoustic respiratory signals, and piezo sensor data over a 24-hour period in daily life. In this study, the M2VS system will be used strictly as a non-diagnostic data logger.

This study will provide critical insights into the signal quality and technical reliability of a non-invasive, multi-sensor wearable compared to current standard of care (Holter monitor and ABPM). By evaluating the wearable, the research aims to overcome the comfort and compliance barriers of existing monitoring systems. These findings may represent a significant step toward developing more reliable, continuous diagnostic tools for the long-term management of conditions like atrial fibrillation and hypertension.

研究の種類

観察的

入学 (推定)

20

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

研究場所

    • Canton of Basel-City
      • Basel、Canton of Basel-City、スイス、4031
        • University Hospital Basel
        • コンタクト:

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

  • 大人
  • 高齢者

健康ボランティアの受け入れ

いいえ

サンプリング方法

非確率サンプル

調査対象母集団

The project population consists of adult participants (≥18 years) recruited at the USB who have a clinical indication for either 24-hour Holter monitoring or 24-hour ABPM. A total of N=20 participants will be enrolled, N=10 for the Holter cohort and N=10 for the ABPM cohort.

説明

Inclusion Criteria:

  • Written informed consent obtained prior to any project-specific procedures.
  • Age 18 years or older at the time of enrolment.
  • Ability to provide valid informed consent.
  • Clinical indication for 24-hour Holter monitoring (Holter cohort) or 24-hour ABPM (ABPM cohort).

Exclusion Criteria:

  • Insufficient knowledge of the German language, preventing adequate understanding of the project procedures and informed consent.
  • Enrolment of the project leader, project staff, their family members, employees, or other dependent persons.

Dermatological and Sensor Placement:

  • Pre-existing skin conditions at the application sites (medial upper arm or chest), such as extensive scarring, tattoos, open wounds, active inflammation, or severe oedema that could interfere with sensor coupling or signal quality.
  • Excessive hair at the application site that could interfere with sensor coupling or signal quality.
  • Presence of other concurrent wearable devices at the same anatomical application site that could interfere with the M2VS system.
  • Known allergy to skin-contacting materials utilized in the M2VS system.

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

コホートと介入

グループ/コホート
介入・治療
Holter cohort
Participants receiving standard of care Holter monitoring will concurrently wear the M2VS system for a 24-hour recording period.
The study features a dual-device monitoring approach integrated into the clinical routine: Participants receiving standard of care will concurrently wear the M2VS system for a 24-hour recording period.The M2VS system is designed for multimodal vital sign monitoring. It comprises a wearable and an associated belt system. The wearable consists of a battery pack and a recorder unit, that are connected via a sliding mechanism
Questionnaire regarding the usability and wearability of both the M2VS system and the standard of care device.
ABPM cohort
Participants undergoing standard-of-care ABPM will concurrently wear the M2VS system during the same 24-hour monitoring period.
The study features a dual-device monitoring approach integrated into the clinical routine: Participants receiving standard of care will concurrently wear the M2VS system for a 24-hour recording period.The M2VS system is designed for multimodal vital sign monitoring. It comprises a wearable and an associated belt system. The wearable consists of a battery pack and a recorder unit, that are connected via a sliding mechanism
Questionnaire regarding the usability and wearability of both the M2VS system and the standard of care device.

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Strong signal-to-noise ratio
時間枠:24 hours
A strong signal-to-noise ratio (dB), ensuring signals received lie within a range that is physiologically feasible and are usable for further vital sign analysis, and the percentage of continuous, gap-free recording time.
24 hours

二次結果の測定

結果測定
メジャーの説明
時間枠
Similarity of ECG signal compared to Standard of Care
時間枠:24 hours
Agreement of the ECG signal between the investigational M2VS system and the standard of care reference devices. This evaluates the mean heart rate derived from ECG RR-intervals compared to the 24-hour Holter monitor. The overall similarity is quantified using the Intraclass Correlation Coefficient. This is a Unitless index ranging from 0 to 1.
24 hours
Similarity of Systolic Blood Pressure compared to Standard of Care
時間枠:24 hours
Agreement of systolic blood pressure between the investigational M2VS system and the standard of care reference devices. This evaluates the estimated systolic blood pressure derived from Pulse Arrival Time (using synchronous M2VS ECG and PPG signals) compared to the 24-hour Ambulatory Blood Pressure Monitor (ABPM). The overall similarity for these metrics is quantified using the Intraclass Correlation Coefficient. This is a Unitless index ranging from 0 to 1.
24 hours
Similarity of Diastolic Blood Pressure compared to Standard of Care
時間枠:24 hours
Agreement of diastolic blood pressure between the investigational M2VS system and the standard of care reference devices. This evaluates the estimated diastolic blood pressure derived from Pulse Arrival Time (using synchronous M2VS ECG and PPG signals) compared to the 24-hour Ambulatory Blood Pressure Monitor (ABPM). The overall similarity for these metrics is quantified using the Intraclass Correlation Coefficient. This is a Unitless index ranging from 0 to 1.
24 hours
Reliability compared to Standard of Care
時間枠:24 hours
Quantification of the signal reliability and technical performance of the investigational M2VS system compared to the standard of care reference devices. This evaluates the technical data continuity and signal quality, measured as the percentage of the total recording period where the ECG and PPG data packets are continuous, gap-free, and meet acceptable Signal-to-Noise Ratio (SNR) and physiological thresholds to be considered usable.
24 hours
Wearability of M2VS
時間枠:After the 24-hour monitoring period
Participant-reported wearability of the M2VS system compared to standard of care devices, assessed via a structured questionnaire administered upon device removal. Responses are measured on a Likert scale ranging from 1 (strongly disagree) to 5 (strongly agree). Scores from reverse-coded items will be inverted before the mean score is calculated. A higher mean score indicates better usability and wearability.
After the 24-hour monitoring period
Localization of Physical Discomfort
時間枠:24 hours (Upon device removal)
Participant-reported areas of discomfort marked on a body map and supplemented by qualitative notes on the corresponding area.
24 hours (Upon device removal)
Acceptance compared to Standard of Care
時間枠:24 hours (Upon device removal)
Participant acceptance of the M2VS device compared to standard of care devices, assessed via the Net Promoter Score (NPS). The NPS is calculated using a single question ("How likely are you to recommend...?") on a scale from 0 (not at all likely) to 10 (extremely likely). The final NPS ranges from 0 to 10, where a higher score indicates a higher level of participant acceptance.
24 hours (Upon device removal)

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

捜査官

  • 主任研究者:Jens Eckstein, Prof. Dr.、University Hospital, Basel, Switzerland

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (推定)

2026年7月1日

一次修了 (推定)

2026年8月1日

研究の完了 (推定)

2026年8月31日

試験登録日

最初に提出

2026年6月25日

QC基準を満たした最初の提出物

2026年7月3日

最初の投稿 (実際)

2026年7月10日

学習記録の更新

投稿された最後の更新 (実際)

2026年7月10日

QC基準を満たした最後の更新が送信されました

2026年7月3日

最終確認日

2026年7月1日

詳しくは

本研究に関する用語

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

いいえ

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する