Evaluation of Ultrasound Probe for Use in Bladder Radiotherapy

2012년 9월 28일 업데이트: Dr Anjali Zarkar, University Hospital Birmingham NHS Foundation Trust

Evaluation of Ultrasound Probe for Use in Adaptive Bladder Radiotherapy

Muscle-invasive bladder cancer affects over 3,000 new patients in the UK each year and radiotherapy comprises a key part of the treatment pathway for many of them. Bladder radiotherapy in the UK involves delivery of radiation to the entire bladder volume and aims to deliver a high dose to the bladder while sparing nearby healthy organs (e.g. the bowel). Unfortunately, large variations in bladder shape and volume are observed throughout treatment that compromise the probability of a positive outcome for the patient, either by increasing the risk of side-effects or by reducing the likelihood of adequately treating the disease.

In recent years, the implementation of 'adaptive' bladder radiotherapy strategies has increased. These techniques allow modification of the treatment based on the observed variations in bladder shape and volume. However, these strategies often rely on daily X-ray imaging of the patient prior to treatment, which is time-consuming and associated with a small risk of inducing a cancer in the patient.

This study will evaluate the potential of a commercially available ultrasound probe for use as an alternative to X-ray scanning during treatment for these patients. The study will also aim to compare the likelihood of reducing side-effects and successfully treating the disease for the different adaptive strategies that are used around the world.

연구 개요

상태

알려지지 않은

정황

상세 설명

All experiments and data analysis will be performed in the Magnetic Resonance Imaging (MRI) and Radiotherapy departments respectively, at the Queen Elizabeth Hospital.

Five volunteers will be recruited to this study. They will each be required to follow a documented drinking protocol prior to each of three scan sessions. The drinking protocol is specific to this study and will not be intended for use with patients. To address the primary endpoint of the study (i.e. the precision of the ultrasound probe for measurement of bladder volume), measurement of a wide range of bladder volumes is required. In order to address this aim while making efficient use of the limited available MRI scanner time, the drinking protocol has been designed such that the volunteer will begin each scan session with an empty bladder and finish it with a full bladder. To this end, each volunteer will be required to void their bladder upon arrival and then consume 500ml of water immediately prior to the start of the scan session.

Each volunteer will then be positioned on the scanner couch according to the same protocol used for patients and alternate MRI and ultrasound scans will then be acquired. Given the time required for each scan it is anticipated that five MRI scans and fifteen ultrasound scans (three scans at a time to reduce the noise on the results, i.e. ultimately giving five ultrasound bladder volume measurements) will be acquired in each session. Each volunteer will be asked to attend for three sessions over the course of a number of weeks so that the repeatability of bladder filling patterns can be determined.

Data processing and analysis

The volume measurements from the ultrasound probe will be recorded and each of the MRI images will be exported to the Radiotherapy department where the bladder volume will be determined. The precise timings of all scan acquisitions will be recorded during the imaging session. Bladder volumes measured by the ultrasound probe, which can be read instantly, will be recorded during the session. To evaluate the primary endpoint, the precision and accuracy of the ultrasound probe will be quantified by comparing the measured values to those expected from the MRI volumes. Each volume measurement from the MRI scans will be checked by a second observer to minimise any associated researcher error.

To evaluate the secondary endpoint, the imaging protocol described above will provide up to 15 MRI images for each volunteer that can be used to inform a comparison of the different adaptive strategies. However, since patients are always asked to empty their bladder prior to radiotherapy treatment (in order to minimise the volume that receives radiation), the deliberately wide-ranging bladder volumes that will have been measured in this study will not be representative of those that would be observed in clinical practice. To address this problem, a suitable dataset that quantifies the distribution of bladder volumes encountered in a typical patient cohort will be used (Christie dataset, permission obtained.Software will be written to randomly assign a bladder volume to each day of each volunteer's 'treatment'. The MRI scan from that volunteer that most closely corresponds to the selected volume will then be assigned for that day's treatment. This will be repeated for the full 20 days of treatment and the importance of each MRI scan can then be factored into the results analysis accordingly.

The treatment planning system within the Radiotherapy department will then be used with the MRI scans to simulate a range of different 'treatments' according to each of the different adaptive bladder radiotherapy strategies to be compared. The different strategies are summarised here:

  • Conventional strategy: This is the current standard treatment, in which the patient plan is prepared based on an initial CT scan of the patient with an empty bladder and no account is made of subsequent daily variations in bladder shape and volume.
  • Plan-of-the-day strategy: Based on an initial CT scan of the patient with an empty bladder, three different treatment plans are created based on an assumption of how the bladder might expand. These are intended to simulate empty, partially full and full bladders. The patient is scanned before treatment each day and the most appropriate plan is selected for treatment.
  • Dynamic strategy: Three different treatment plans are created based on a series of initial CT scans of the patient with empty, partially full and full bladder. These would be expected to directly represent the empty, partially full and full bladders for the individual patient. The patient is scanned before treatment each day and the most appropriate plan is selected for treatment.
  • Composite strategy: The patient is CT scanned with an empty bladder and treated according to the Conventional strategy for the first 7 days of treatment with daily imaging. The first 5 daily scans of the patient are used to generate plans based on small, partially full and full bladder. From the 8th day of treatment the daily image is used to select which of these plans is appropriate for treatment.

Data processing of the MRI scans will be carried out retrospectively. The MRI scans will be anonymised and exported to the Radiotherapy department. The data will then be used to (i) acquire bladder volume information to compare with that acquired from the ultrasound probe and (ii) prepare and evaluate radiotherapy treatment simulations to investigate the benefits of different adaptive bladder radiotherapy strategies

Comparison of different adaptive techniques will focus on how much radiation is received by the bowel and the extent to which the radiation misses the disease on any day of treatment.

연구 유형

관찰

등록 (예상)

5

연락처 및 위치

이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.

연구 장소

    • west Midlands
      • Birmingham, west Midlands, 영국, B15 2TH
        • University Hospital Birmingham

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

  • 어린이
  • 성인
  • 고령자

건강한 자원 봉사자를 받아들입니다

연구 대상 성별

모두

샘플링 방법

비확률 샘플

연구 인구

Healthy volunteers from within the staff of the Radiotherapy, Medical Physics and Oncology departments at the Queen Elizabeth Hospital

설명

Inclusion Criteria:

  • Healthy volunteers

Exclusion Criteria:

  • Pre-existing bladder conditions

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
Accuracy of ultrasound measurement of bladder volume in comparision with gold standard MRI measurement
기간: Three weeks
The principal aim of this study is to quantify the accuracy of ultrasound measurement of bladder volume in comparision with gold standard MRI measurement. To do this, each of the five volunteers will have 5 repeat measurements of each imaging modality (ultrasound and MRI) within each 45 minute scan session as the bladder volume increases in response to the specified drinking protocol. Each volunteer will undergo three scan sessions over three weeks. Study participants will not receive follow-up. An accuracy of +/-10% would be considered sufficient to inform future studies.
Three weeks

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

수사관

  • 수석 연구원: Anjali Zarkar, FRCR, UHB

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작

2012년 11월 1일

기본 완료 (예상)

2013년 7월 1일

연구 완료 (예상)

2013년 10월 1일

연구 등록 날짜

최초 제출

2012년 9월 17일

QC 기준을 충족하는 최초 제출

2012년 9월 28일

처음 게시됨 (추정)

2012년 10월 3일

연구 기록 업데이트

마지막 업데이트 게시됨 (추정)

2012년 10월 3일

QC 기준을 충족하는 마지막 업데이트 제출

2012년 9월 28일

마지막으로 확인됨

2012년 9월 1일

추가 정보

이 연구와 관련된 용어

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .

방광암에 대한 임상 시험

구독하다