이 페이지는 자동 번역되었으며 번역의 정확성을 보장하지 않습니다. 참조하십시오 영문판 원본 텍스트의 경우.

Safety and Efficacy Clinical Trial of the Vessel-X® Bone Filling Container System (CMT-VX)

2026년 6월 22일 업데이트: Juin-Hong Cherng

The Clinical Investigation to Evaluate the Safety and Efficacy of Vessel-X® Bone Filling Container System

Osteoporotic vertebral compression fractures (OVCFs) are a common and serious complication of osteoporosis, particularly among elderly and postmenopausal patients. OVCFs may result in severe pain, functional impairment, spinal deformity, and reduced quality of life. Conventional conservative treatments, including bed rest, analgesics, and bracing, may provide limited symptom relief. Minimally invasive vertebral augmentation procedures, such as vertebroplasty and kyphoplasty, have been widely used to improve clinical outcomes; however, risks including bone cement leakage and incomplete vertebral restoration remain concerns.

The Vessel-X® Bone Filling Container System, manufactured by Central Medical Technologies Inc. (CMT), is a third-generation vesselplasty technology designed for percutaneous vertebral augmentation procedures. The system utilizes an implantable biocompatible polyethylene terephthalate (PET) container with a microporous structure for controlled bone cement delivery. The implant remains within the vertebral body after cement injection and is designed to reduce cement leakage while maintaining vertebral height restoration and pain relief.

This post-market clinical study evaluates the safety and clinical effectiveness of the Vessel-X® Bone Filling Container System at two medical centers in Taiwan with a total target enrollment of 146 subjects:

Tri-Service General Hospital (TSGH): 86 subjects randomized in a 1:1 ratio to the experimental and control groups.

Taoyuan General Hospital, Ministry of Health and Welfare (TYGH): 60 subjects randomized in a 1:1 ratio to the experimental and control groups.

The primary objective is to evaluate the safety of the device by assessing the incidence of unanticipated serious adverse device effects (USADEs). Secondary objectives include evaluation of pain reduction measured by the Visual Analogue Scale (VAS), functional recovery assessed by the Oswestry Disability Index (ODI), and radiographic outcomes including vertebral height restoration and kyphotic deformity correction.

연구 개요

상세 설명

Background and Rationale:

Osteoporotic vertebral compression fractures (OVCFs) are a common and serious complication of osteoporosis, particularly among elderly and postmenopausal patients. OVCFs may result in severe pain, functional impairment, spinal deformity, and reduced quality of life. Conventional conservative treatments, including bed rest, analgesics, and bracing, may provide limited symptom relief. Minimally invasive vertebral augmentation procedures, such as vertebroplasty and kyphoplasty, have been widely used to improve clinical outcomes; however, risks including bone cement leakage and incomplete vertebral restoration remain concerns.

The Vessel-X® Bone Filling Container System, manufactured by Central Medical Technologies Inc. (CMT), is a third-generation vesselplasty technology designed for percutaneous vertebral augmentation procedures. The system utilizes an implantable biocompatible polyethylene terephthalate (PET) container with a microporous structure for controlled bone cement delivery. The implant remains within the vertebral body after cement injection and is designed to reduce cement leakage while maintaining vertebral height restoration and pain relief.

This is a prospective, open-label, randomized, parallel-controlled, non-inferiority clinical trial conducted in Taiwan to evaluate the safety and clinical effectiveness of this system (Model: BVFT-UP01-D20, Approved No. 005889) compared with conventional vertebroplasty performed using manual orthopedic surgical instruments (Approved No. 005698), both manufactured by CMT. Both groups will utilize the same radiopaque bone cement ("Tecres" Osteopal V) for the therapeutic procedures.

Hypothesis:

The study is designed as a non-inferiority trial. The primary hypothesis is that the incidence of unanticipated serious adverse device effects (USADEs) in the Vessel-X® group is not higher than that in the control group. For efficacy, the experimental group is hypothesized to be non-inferior to the control group regarding pain relief, disability improvement, and radiographic parameters, with a pre-specified non-inferiority margin of -15% for the difference in fused effective rates between the two groups.

Patient Enrollment and Site Allocation:

To achieve the clinical evaluation objectives, a total of up to 146 subjects will be recruited across participating medical centers in Taiwan. The target sample size is allocated specifically as 86 subjects from Tri-Service General Hospital (TSGH) and 60 subjects from Taoyuan General Hospital (TYGH). Subjects at each site will be allocated to the experimental and control groups using a 1:1 randomization scheme.

Surgical Procedure Protocol:

All surgical procedures will be performed under real-time fluoroscopic (C-arm) guidance using a standard unilateral transpedicular approach:

Experimental Group: The Vessel-X® PET biocompatible microporous container will be deployed into the vertebral body, and bone cement will be injected in a controlled manner, allowing partial cement interdigitation through the micropores. The container remains as a permanent implant.

Control Group: Conventional vertebroplasty will be performed using CMT's standard manual orthopedic surgical instruments to deliver the bone cement directly.

In both groups, fluoroscopic monitoring is maintained throughout the injection, and cement delivery will be discontinued immediately if any cement leakage or breach is observed.

Clinical Evaluation Timelines:

Subjects will participate in the study for a total duration of 13 months, consisting of a 1-month screening period, the surgical intervention (Day 1), and a 12-month post-operative follow-up period. Follow-up visits are scheduled at Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days).

Outcome Assessments:

Safety Assessment (Primary Endpoint, ITT Population): Evaluated by the incidence of adverse events, complications, bone cement leakage, and new-onset neurological deficits.

Efficacy Assessment (Secondary Endpoints, PP & FAS Populations): Evaluated by changes from baseline to each follow-up timepoint in Visual Analogue Scale (VAS) scores for pain, Oswestry Disability Index (ODI) scores for functional disability, and radiographic parameters (anterior vertebral height, midline vertebral height, and kyphotic Cobb's angle)

연구 유형

중재적

등록 (추정된)

146

단계

  • 해당 없음

연락처 및 위치

이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.

연구 연락처

  • 이름: Juin-Hong Cherng, Ph.D.
  • 전화번호: 18681, 18699 886-2-87923100
  • 이메일: i72bbb@gmail.com

연구 연락처 백업

  • 이름: Gang-Yi Fan, Ph.D.
  • 전화번호: 18699 886-2-87923100
  • 이메일: u9310318@gmail.com

연구 장소

      • Taipei, 대만, 114202
        • 모병
        • Tri-Service General Hospital, National Defense Medical Univesity
        • 연락하다:
        • 연락하다:
          • Juin-Hong Cherng, Ph.D.
          • 전화번호: 18698, 18699 886-2-87923100
          • 이메일: i72bbb@gmail.com
      • Taipei, 대만, 11490
        • 모병
        • National Defense Medical University, Taiwan
        • 연락하다:
        • 연락하다:
          • Juin-Hong Cherng, Ph.D.
          • 전화번호: 18698, 18699 886-2-8792-3100
          • 이메일: i72bbb@gmail.com
        • 수석 연구원:
          • Chih-Chern Wang, Ph.D.
        • 수석 연구원:
          • Yu-Hsu Chen, Ph.D.
        • 부수사관:
          • Yi-Chou Chen, M.D.
        • 부수사관:
          • Chien-Ning Hsu, M.D.

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

  • 성인
  • 고령자

건강한 자원 봉사자를 받아들입니다

아니

설명

Inclusion Criteria:

Subjects must meet all of the following criteria to be eligible for enrollment in this study:

  1. Male or female subjects aged ≥40 and ≤95 years.
  2. Subjects admitted with vertebral compression fractures (VCFs) caused by osteoporosis or trauma.
  3. Surgical site within the range of T6 to L5.
  4. Subjects with normal vital signs and hepatic and renal function values within 1.5 times the upper limit of normal (ULN), as determined by the investigator to be suitable for study participation.
  5. Subjects willing to comply with the study schedule and all required assessment procedures.
  6. Subjects who are conscious, cognitively intact, and able to provide written informed consent.

Exclusion Criteria:

Subjects who meet any of the following criteria will be excluded from participation in this study:

  1. Pathological fractures caused by bone diseases, including benign or malignant tumors, tuberculosis, osteomyelitis, endocrine or metabolic bone disorders, or severe degenerative bone diseases.
  2. Active or severe systemic infections.
  3. Metabolic disorders (e.g., calcium metabolism disorders), immune system disorders, substance abuse, or alcoholism.
  4. Severe primary diseases involving the hematopoietic or endocrine systems, or psychiatric disorders.
  5. History of allergy to implant materials, hypersensitivity reactions, or allergies to multiple medications.
  6. Non-viable bone surrounding the surgical site or insufficient bone quality to support the implant.
  7. Active infection at or adjacent to the surgical site.
  8. Acute spinal instability.
  9. Unwillingness or inability to restrict physical activity or comply with medical instructions.
  10. Considered unsuitable for study participation by the investigator, or unable to provide independent informed consent.

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

  • 주 목적: 치료
  • 할당: 무작위
  • 중재 모델: 병렬 할당
  • 마스킹: 하나의

무기와 개입

참가자 그룹 / 팔
개입 / 치료
실험적: CMT Vessel-X ® Bone Filling Container System
Subjects undergo percutaneous vertebral augmentation using the CMT Vessel-X® Bone Filling Container System (Model: BVFT-UP01-D20). Under fluoroscopic guidance, a unilateral transpedicular approach is used to access the fractured vertebral body. The microporous PET Vessel-X® Bone Filling Container is deployed within the vertebral body and subsequently filled with TEKNIMED OPACITY+ Radiopaque Bone Cement (Model: T040320Z) to stabilize the vertebral compression fracture.

Model: BVFT-UP01-D20 (TFDA No. 005889, GMP0106), manufactured by Central Medical Technologies Inc. (marketed since Aug 2017). An EO-sterilized, single-use, implantable kit for percutaneous vertebral augmentation (store <= 25°C, shelf-life: 5 years). The kit includes:

  1. Vessel-X® Bone Filling Container (Model: BVFX-D20): An implantable, microporous PET container to enclose cement and minimize leakage.
  2. Front-Opening Cement Injection Device with Extension Tube (Model: T-C308, max 20 ml): For controlled hydraulic delivery.
  3. Percutaneous Access Tools: Includes 1pc Bone Access Needle (T-N201T), 1pc Guide Pin (T-P602), and 1pc Precision Drill (T-D401).

    • Intervention Cement: TEKNIMED OPACITY+ Radiopaque Bone Cement (Model: T040320Z; TFDA License No. 026888), used during vertebral augmentation procedures for injection into the affected vertebra to stabilize vertebral compression fractures.
활성 비교기: CMT Conventional Vertebroplasty
Subjects undergo percutaneous vertebroplasty using the CMT Manual Orthopedic Surgical Instruments (Model: T-C308). Under real-time fluoroscopic guidance, a unilateral transpedicular approach is used to access the fractured vertebral body, and TEKNIMED OPACITY+ Radiopaque Bone Cement (Model: T040320Z) is directly injected into the vertebral body to stabilize the vertebral compression fracture.
The CMT Manual Orthopedic Surgical Instruments (Model: T-C308; TFDA License No. 005698) are nonimplantable manual surgical instruments intended for conventional vertebroplasty procedures. The device functions as a manual cement delivery system for the injection of TEKNIMED OPACITY+ Radiopaque Bone Cement (Model: T040320Z; TFDA License No. 026888) into the vertebral body to stabilize vertebral compression fractures.

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
Incidence of Adverse Events, Serious Adverse Events (SAEs), and Unanticipated Serious Adverse Device Effects (USADEs)
기간: Day 1 (post-op), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.
The primary safety endpoint is to evaluate the safety profile of the "Central Medical" Vessel-X® Bone Filling Container System. Safety assessments will analyze the incidence, specific manifestations, severity, and device-causality of all adverse events (AEs), serious adverse events (SAEs), and unanticipated serious adverse device effects (USADEs) following percutaneous vertebroplasty (including bone cement leakage, nerve root injury, vascular injury, or subsequent vertebral collapse). Proportions of abnormal post-operative cases and complication incidence with 95% Confidence Intervals (CI) will be calculated using the Intent-to-Treat (ITT) population.
Day 1 (post-op), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.

2차 결과 측정

결과 측정
측정값 설명
기간
Change From Baseline in Visual Analogue Scale (VAS) Pain Score
기간: Baseline (within 1 week pre-operatively), Day 1 (post-operatively), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.
Pain intensity will be assessed using the 100-mm Visual Analogue Scale (VAS), where 0 indicates no pain and 100 indicates the worst imaginable pain. Changes from baseline in VAS scores will be compared between the experimental and control groups to evaluate pain relief following treatment.
Baseline (within 1 week pre-operatively), Day 1 (post-operatively), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.
Change From Baseline in Oswestry Disability Index (ODI) Score
기간: Baseline (within 1 week pre-operatively), Day 1 (post-operatively), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.
Functional disability will be evaluated using the Oswestry Disability Index (ODI) questionnaire. Changes from baseline in ODI scores will be compared between the experimental and control groups to assess functional recovery following treatment.
Baseline (within 1 week pre-operatively), Day 1 (post-operatively), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.
Change From Baseline in Radiographic Anterior Vertebral Height
기간: Baseline (within 1 week pre-operatively), Day 1 (post-operatively), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.
Radiographic assessment of anterior vertebral height (measured in millimeters) will be performed using X-ray, computed tomography (CT), or magnetic resonance imaging (MRI). Changes from baseline will be compared between the experimental and control groups to evaluate vertebral body restoration.
Baseline (within 1 week pre-operatively), Day 1 (post-operatively), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.
hange From Baseline in Radiographic Mid-Vertebral Height
기간: Baseline (within 1 week pre-operatively), Day 1 (post-operatively), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.
Radiographic assessment of mid-vertebral height (measured in millimeters) will be performed using X-ray, computed tomography (CT), or magnetic resonance imaging (MRI). Changes from baseline will be compared between the experimental and control groups to evaluate vertebral body restoration.
Baseline (within 1 week pre-operatively), Day 1 (post-operatively), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.
Change From Baseline in Kyphotic Cobb's Angle
기간: Baseline (within 1 week pre-operatively), Day 1 (post-operatively), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.
Radiographic assessment of the kyphotic Cobb's angle (measured in degrees) will be performed using sagittal X-ray, computed tomography (CT), or magnetic resonance imaging (MRI) to evaluate spinal alignment correction. Changes from baseline will be compared between the experimental and control groups.
Baseline (within 1 week pre-operatively), Day 1 (post-operatively), Day 8 (8±2 days), Month 1 (31±2 days), Month 3 (91±15 days), Month 6 (181±15 days), Month 9 (271±15 days), and Month 12 (361±15 days) post-operatively.

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

스폰서

수사관

  • 수석 연구원: Dr. Chih-Chien Wang, Ph.D., Trial-Service General Hospital, National Defense Medical University
  • 수석 연구원: Yu-Hsu Chen, Ph.D., Taoyuan General Hospital, Ministry of Health and Welfare

간행물 및 유용한 링크

연구에 대한 정보 입력을 담당하는 사람이 자발적으로 이러한 간행물을 제공합니다. 이것은 연구와 관련된 모든 것에 관한 것일 수 있습니다.

일반 간행물

유용한 링크

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작 (실제)

2021년 4월 20일

기본 완료 (추정된)

2026년 12월 31일

연구 완료 (추정된)

2026년 12월 31일

연구 등록 날짜

최초 제출

2026년 6월 3일

QC 기준을 충족하는 최초 제출

2026년 6월 22일

처음 게시됨 (실제)

2026년 6월 24일

연구 기록 업데이트

마지막 업데이트 게시됨 (실제)

2026년 6월 24일

QC 기준을 충족하는 마지막 업데이트 제출

2026년 6월 22일

마지막으로 확인됨

2026년 5월 1일

추가 정보

이 연구와 관련된 용어

개별 참가자 데이터(IPD) 계획

개별 참가자 데이터(IPD)를 공유할 계획입니까?

아니요

IPD 계획 설명

IPD will not be shared to protect participant privacy and comply with institutional review board (IRB) restrictions and local data protection regulations.

약물 및 장치 정보, 연구 문서

미국 FDA 규제 의약품 연구

아니

미국 FDA 규제 기기 제품 연구

아니

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .

구독하다