- ICH GCP
- 미국 임상 시험 레지스트리
- 임상시험 NCT07940361
Ultrashort Echo Time (UTE) MRI for Applicator Reconstruction in Gynaecologic Cancer Brachytherapy (UTE)
The goal of this clinical study is to learn whether an ultrashort echo time (UTE) MRI scan can show the position of brachytherapy applicators and needles more accurately than standard MRI scans in adults with gynaecologic cancer (cervical, endometrial, or vaginal cancer) who are receiving MRI-guided internal radiotherapy (brachytherapy).
The main questions it aims to answer are:
- Does UTE MRI locate the applicator and needle tips more accurately than standard T2-weighted (T2W) and proton-density-weighted (PDW) MRI, when compared with CT?
- Does UTE MRI show a clearer contrast between the applicator and the surrounding tissue?
Researchers will compare applicator positions reconstructed from UTE, T2W, and PDW MRI with CT taken on the same day as the reference standard. They will also check whether differences in applicator reconstruction change the planned radiation dose, and whether image features (radiomics) from UTE MRI before treatment and before brachytherapy help predict cancer recurrence.
Participants will:
- Receive their usual standard brachytherapy treatment, which is not changed by this study
- Have one additional MRI sequence (UTE, about 8 minutes) added to their standard MRI and CT scans at each brachytherapy planning session, with the applicator in place
- Have an additional UTE sequence added to their routine MRI before treatment and before brachytherapy
- Be followed during routine clinical visits for disease outcomes
연구 개요
상세 설명
Background MRI-guided adaptive brachytherapy is the standard of care for locally advanced gynaecologic cancer. T2-weighted MRI is the reference sequence for target delineation, but applicator and interstitial needle reconstruction on T2W MRI is limited by low contrast between the uterus and titanium applicators and by susceptibility artefacts, with reported geometric distortion of up to about 7 mm. Proton-density-weighted MRI (TE about 5.5 ms) improves tandem visualisation but remains limited for interstitial needles and residual blooming artefact. Ultrashort echo time (UTE) MRI (TE below 1 ms) captures signal from very short-T2 structures and is expected to reduce susceptibility artefact around metallic and plastic applicators. No prospective study has compared UTE with PDW and T2W MRI for intracavitary/interstitial applicator reconstruction in gynaecologic brachytherapy.
Hypothesis UTE MRI will localise the applicator tip more accurately than PDW and T2W MRI, with a mean tip-localisation error of no more than 1.2 mm relative to same-day CT, and will provide a higher uterus-to-applicator contrast-to-noise ratio (CNR), with about 8 minutes of additional scan time.
Study design Single-centre, prospective, single-group study at the Division of Radiation Oncology, Faculty of Medicine Siriraj Hospital, Mahidol University. Standard treatment is not altered.
Phase 1 - Phantom validation An MR- and CT-compatible phantom holding tandem-ring, tandem-ovoid, vaginal cylinder, intracavitary/interstitial applicators and interstitial needles in multiple orientations will be imaged with CT and MRI (T1W, T2W, UTE) without moving the applicator. Image quality (SNR, CNR, apparent applicator diameter, tip/channel visibility, geometric distortion, susceptibility artefact) will be assessed. Source dwell positions will be verified by autoradiography using an Ir-192 source. Dosimetric impact will be assessed with AAPM TG-43 calculations on simulated targets and organs at risk.
Phase 2 - Clinical validation After applicator insertion, at each brachytherapy planning session participants will undergo standard CT and MRI (T2W, PDW) with an additional 3D UTE sequence (1.5 mm isotropic, TE below 0.1 ms) on a Philips 1.5T Ingenia Ambition X system, with the applicator in situ and minimal time between scans. Three experienced medical physicists will independently reconstruct applicators on each image set, blinded and in randomised order. Applicator-based rigid registration between CT and MRI with a registration quality score will be used to separate reconstruction error from applicator displacement; only registrations meeting predefined acceptance criteria will be analysed. Comparison plans (CT reference, T2W, UTE) will use identical contours, dwell positions, and dwell times so that dose differences reflect reconstruction alone.
Phase 3 - Radiomics An additional UTE sequence will be acquired with routine T2W and DWI at pre-treatment and pre-brachytherapy time points. Radiomic features will be extracted from tumour ROIs (MIM software) and evaluated for prognostic value for locoregional and distant recurrence, in addition to existing T2W- and DWI-based models, using Kaplan-Meier and Cox proportional hazards analyses.
Sample size and analysis Target enrolment is 50 adult patients (about 150 imaging sessions). Assuming tip error of 1.2 +/- 1.0 mm for UTE versus 1.8 +/- 1.2 mm for PDW, a paired t-test with two-sided alpha 0.05 and 80% power (d = 0.46) was used. Agreement with CT will be assessed by Bland-Altman analysis; sequences will be compared with repeated-measures ANOVA or linear mixed-effects models; inter- and intra-observer agreement by ICC; and dosimetric parameters (CTV-HR D90/D98, OAR D2cc and others) with correction for multiple comparisons.
Support Philips (Thailand) Ltd. provides research equipment support without funding. The company has no role in study design, participant selection, data collection, analysis, interpretation, writing, or the decision to publish.
연구 유형
등록 (추정된)
단계
- 해당 없음
연락처 및 위치
연구 연락처
- 이름: Tissana Prasartseree, M.D.
- 전화번호: 66625246101
- 이메일: tissana.p@gmail.com
연구 연락처 백업
- 이름: Pittaya Dankulchai, M.D.
- 전화번호: 66966233606
- 이메일: pittaya.dan@mahidol.ac.th
연구 장소
-
-
Bangkok
-
Bangkok, Bangkok, 태국, 10700
- Faculty of Medicine Siriraj Hospital, Mahidol University
-
연락하다:
- Tissana Prasartseree, M.D.
- 전화번호: 66625246101
- 이메일: tissana.p@gmail.com
-
-
참여기준
자격 기준
공부할 수 있는 나이
건강한 자원 봉사자를 받아들입니다
연구 대상 성별
설명
Inclusion Criteria:
- Gynaecologic malignancy (cervix/endometrium/vagina) requiring BT with intracavitary or hybrid IC/IS applicator
- Locally advanced cervical, endometrial, or vaginal cancer undergoing MRI-guided BT
- Age ≥ 18 years
- ECOG 0-2
- Written informed consent
Exclusion Criteria:
- MRI contraindication (incompatible implant, severe claustrophobia)
- non-MRI-compatible applicator required
- inability to tolerate ~8 mins additional scan time
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
- 주 목적: 특수 증상
- 할당: 해당 없음
- 중재 모델: 단일 그룹 할당
- 마스킹: 없음(오픈 라벨)
무기와 개입
참가자 그룹 / 팔 |
개입 / 치료 |
|---|---|
|
실험적: UTE MRI
Gynaecologic malignancy (cervix/endometrium/vagina) requiring BT with intracavitary or hybrid IC/IS applicator. At every planning/fraction session: T2W, PDW, 3D UTE (1.5 mm isotropic) and same day CT (reference), all with applicator in situ on a Philips 1.5T system. An additional UTE sequence will be acquired alongside the standard T2W and DWI sequences at pre-treatment and pre-brachytherapy time points. |
Ultrashort Echo Time sequence (EFO 033), TE < 0.1 ms, investigational Philips sequence, on Philips 1.5T MR Systems Ingenia Ambition X release 5.7. 3D UTE (1.5 mm isotropic) acquired at every planning/fraction session with applicator in situ, in addition to T2W and PDW (~8 mins additional scan time).
다른 이름들:
|
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Applicator/needle tip-localisation error
기간: Periprocedural: after applicator insertion, at every planning/fraction session (~150 imaging sessions)
|
Applicator/needle tip-localisation error of UTE among the T2W, PDW MRI sequences, using same day CT as reference (geometric gold standard). Reported in millimetres. |
Periprocedural: after applicator insertion, at every planning/fraction session (~150 imaging sessions)
|
|
Contrast-to-noise ratio (CNR)
기간: Periprocedural: after applicator insertion, at every planning/fraction session (~150 imaging sessions)
|
Contrast-to-noise ratio (CNR) between uterus and applicator of UTE among the T2W and PDW MRI sequences. |
Periprocedural: after applicator insertion, at every planning/fraction session (~150 imaging sessions)
|
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Percentage dose difference of brachytherapy plan from UTE among the T2W, PDW MRI sequences and CT
기간: Periprocedural: after applicator insertion, at every planning/fraction session (~150 imaging sessions)
|
Dosimetric agreement of brachytherapy plan from UTE among the T2W, PDW MRI sequences and CT, including CTV-HR D90, CTV-HR D98, CTV-HR V100, CTV-IR D90, CTV-IR D98, and Bladder, Rectum, Sigmoid Colon and Small Bowel D0.1cc, D1cc and D2cc. |
Periprocedural: after applicator insertion, at every planning/fraction session (~150 imaging sessions)
|
|
Radiomics prognostic value for locoregional and distant recurrence
기간: Pre-treatment and pre-brachytherapy timepoints, followed up to 3 years
|
Radiomics with UTE at pre-treatment and pre-brachytherapy timepoints for locoregional and distant recurrent prognostic value in addition to the existing T2W- and DWI-based model (Kaplan-Meier survival analysis and Cox regression analysis). Reported as Hazard Ratio (HR) with 95% CI. |
Pre-treatment and pre-brachytherapy timepoints, followed up to 3 years
|
공동 작업자 및 조사자
연구 기록 날짜
연구 주요 날짜
연구 시작 (추정된)
기본 완료 (추정된)
연구 완료 (추정된)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
키워드
추가 관련 MeSH 약관
기타 연구 ID 번호
- Si 784/2569(IRB1)
개별 참가자 데이터(IPD) 계획
개별 참가자 데이터(IPD)를 공유할 계획입니까?
약물 및 장치 정보, 연구 문서
미국 FDA 규제 의약품 연구
미국 FDA 규제 기기 제품 연구
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .