此页面是自动翻译的,不保证翻译的准确性。请参阅 英文版 对于源文本。

Design and Rationale of the Assessment of the MindMics Recording System

2021年11月1日 更新者:MindMics Inc

Design and Rationale of the Assessment of the Mindmics Recording System to Collect Data to Facilitate the Development of an Algorithm to Discriminate Atrial Fibrillation From Sinus Rhythm

MindMics earbuds is a new technology designed enhance an individual's quality of life by continuously monitoring the heart rate, stress, and blood pressure allowing patients to make decisions regarding their health. MindMics has a working prototype of the ear buds which will perform all the conventional tasks of earbuds (listen to music, etc.), but, in addition, will measure heart rate, stress, blood pressure. The earbuds capture the sounds made by the various cardiac structures pulsing and moving blood. The sound is caused by the acceleration and deceleration of blood and the turbulence developed during rapid blood flow.

研究概览

地位

完全的

干预/治疗

详细说明

Introduction:

This study will evaluate the efficacy of the MindMics earbud recording apparatus to record data from patients with known atrial fibrillation as well as those in sinus rhythm. Patients in the clinic and patients with atrial fibrillation scheduled to undergo elective direct-current cardioversion will be the chosen subjects. This process will select a variety of patients to allow evaluation of the Mindmics apparatus's ability to collect data from patients with normal as well as abnormal cardiac rhythms.

Hypotheses

The MindMics earbud device is capable of recording normal cardiac rhythms in patients noninvasively using earbud device in a noninvasive manner. The same apparatus may be capable of discerning atrial fibrillation compared with normal sinus rhythm. This will be evaluated by collecting information from patients with atrial fibrillation in the office as well as those scheduled to undergo an elective direct-current cardioversion procedure. A timed recording of the clinical rhythm while wearing the apparatus will allow simultaneous recording of data from an ECG recording (documenting the patient's clinical arrhythmia) at the same time the MindMics earbud records its data obtained during the ECG recording of the patient's rhythm. For those patient scheduled for direct current cardioversion, the patient will undergo the direct-current cardioversion and a second recording will be obtained during sinus rhythm afterwards allowing an appropriate comparison of the differences in the recording sets.

The objective is to demonstrate differences between recordings made in atrial fibrillation and those in sinus rhythm that can be used to develop an algorithm allow discernment between atrial fibrillation and sinus rhythm.

OVERALL DESIGN

This study is designed to collect sequential patient's in a nonrandomized unblinded procedure designed to collect information using the apparatus to document its capability of recording cardiac rhythm data, then using this recorded data to determine whether the patient has an abnormal cardiac rhythm. Initial efforts will be directed to develop an algorithm to detect atrial fibrillation.

Patients will be recruited from inpatients at Palmetto Tuomey Medical center or in the outpatient office during routine office visits for patients with clinical arrhythmias for elective recording for group 1. Group 2 patients will be recruited prior to direct-current cardioversion. Both groups will be explained the study, what is their participation, potential risk and expectations for their participation in the study.

Once selected each patient will be required to sign a study protocol consent form before actual entry into the study. Once the consent is completed, the patients will be connected to the Mindmics apparatus as well as an electrocardiographic recording system. The study participant will lie quietly collecting simultaneous electrocardiographic data and data from the Mindmics earbuds. Group 1 patients will be identified as outpatients during their routine office visits will have 15 minutes of recording in a patient examination room during the office visit. Group 2 patients undergoing Direct-Current Cardioversion will have 15 minutes of recording prior to the cardioversion procedure and 15 minutes of recording after completion of the cardioversion procedure.

NUMBER OF PARTICIPANTS

Approximately 25 patients will be recruited for group 1 from inpatient or outpatient office visits. These patients will have persistent atrial fibrillation during the office visit. If necessary would consider increasing the total number of patients to 50 in this group Approximately 25 patients will be recruited for group 2 prior to direct-current cardioversion allowing collection of data from before and after cardioversion to document differences in recordings during sinus rhythm after having successful cardioversion from atrial fibrillation. While certainly some patients will revert back to atrial fibrillation and not have periods of sinus rhythm is anticipated the vast majority of patients will have successful cardioversions long enough to record at least the 15 minutes post cardioversion in sinus rhythm. If necessary would consider increasing the total number of patients to 50 in this group

DURATION The total duration of involvement for each study participant is for the duration of the recording interval. Group 1 participants will have a 15-minute recording interval while those in group to will have 2 sessions of 15 minutes each separated by the direct current cardioversion. No further participation will be required.

All data collected during the study will be kept by Mindmics. All data collected in the study will be de-identified

OUTCOME MEASURES and STATISTICAL ANALYSIS

Each recording will be evaluated by Mindmics to evaluate the ability of the recording apparatus to provide data that would lead to clear discernment of the patient's cardiac rhythm as well as rate related phenomena. At this juncture, it is anticipated the collected data will be used to develop and test algorithms yet to be developed and to make clinical determinations of the cardiac rhythms.

Initial data will be evaluated to determine if the Mindmics earbuds are capable of discriminating atrial fibrillation from normal sinus rhythm. For that reason, it is not possible to suggest a statistical method until it is determined the device data can be used to evaluate differences in cardiac rhythm.

SAFETY

There is no potential harm or injury associated with the research. The probability and magnitude of harm or discomfort anticipated in the research are not greater than those ordinarily encountered in daily life or during the performance of routine physical or psychological examinations or tests. There are no potential risks with respect to psychological, sociological, economic, or legal. The overall risk of the device in minimum.

Potential Benefit to Participants This study is performed in such a manner that it is not anticipated there would be an ability to provide diagnostic data that may be helpful for the study participants.

研究类型

观察性的

注册 (实际的)

29

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习地点

    • South Carolina
      • Sumter、South Carolina、美国、29150
        • Prisma Health

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

  • 孩子
  • 成人
  • 年长者

接受健康志愿者

不

有资格学习的性别

全部

取样方法

非概率样本

研究人群

Approximately 25 patients will be recruited for group 1 from inpatient or outpatient office visits.

Approximately 25 patients will be recruited for group 2 prior to direct-current cardioversion allowing collection of data from before and after cardioversion to document differences in recordings during sinus rhythm after having successful cardioversion from atrial fibrillation.

描述

Inclusion Criteria:

  • AFib

Exclusion Criteria:

  • N/A

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 观测模型:队列
  • 时间观点:横截面

队列和干预

团体/队列
干预/治疗
Patients with AFib
Patients rhythms was recorded using MindMics device
Normal patients
Normal patients without AFib
Patients rhythms was recorded using MindMics device

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Performance of the MindMics device for detecting Atrial Fibrillation based on inter-beat intervals
大体时间:30 minutes
Sensitivity and specificity for detecting Atrial Fibrillation using inter-beat intervals and statistical measures based on inter-beat intervals
30 minutes

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

赞助

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (实际的)

2020年3月24日

初级完成 (实际的)

2021年1月31日

研究完成 (实际的)

2021年1月31日

研究注册日期

首次提交

2021年10月18日

首先提交符合 QC 标准的

2021年11月1日

首次发布 (实际的)

2021年11月2日

研究记录更新

最后更新发布 (实际的)

2021年11月2日

上次提交的符合 QC 标准的更新

2021年11月1日

最后验证

2021年11月1日

更多信息

与本研究相关的术语

其他研究编号

  • Pro00096535

计划个人参与者数据 (IPD)

计划共享个人参与者数据 (IPD)?

不

药物和器械信息、研究文件

研究美国 FDA 监管的药品

不

研究美国 FDA 监管的设备产品

不

在美国制造并从美国出口的产品

不

此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.