Diathermy On Diabetes Glucose Monitoring Effectiveness (DODGE)

2026年5月6日 更新者:University of Nottingham

Assessing the Impact of Intraoperative Diathermy on Accuracy and Functioning of Glucose Monitoring Systems

The goal of this observational study is to investigate if diathermy (a surgical tool that uses electrical energy to control bleeding) has any affect on the accuracy and functioning of continuous glucose monitoring systems in adults and young people with Type 1 Diabetes. The main questions this study aims to answer is -

- Does the accuracy of continuous glucose monitoring systems change after use of diathermy?

Participants will:

  • Have their height and weight checked.
  • Provide information about their medical history including type of diabetes, other medical conditions and any current medications they take.
  • Have paired glucose meter and sensor glucose measurements taken every 15-75 minutes from up to 4 hours before surgery until up to 4 hours after the end of surgery.
  • Have two blood samples taken to measure glucose levels, The first one will be before the use of diathermy and the second will be after the use of diathermy.

研究概览

地位

尚未招聘

详细说明

Background and rationale:

The incidence of type 1 diabetes is rising and technology is an increasingly important part of monitoring and responding to glucose levels. Contemporary monitoring of glucose levels using continuous glucose monitoring (CGM) systems provide users with information about their glucose levels in real time without finger-prick capillary glucose testing. Crucially, these new systems provide information about the trend in blood glucose levels and is now standard in type 1 diabetes. Use of CGM systems is also needed for patients who use "closed loop" systems which adjust insulin at, typically, 5-minute intervals in response to glucose levels. Use of these strategies in glucose level monitoring and management improves long-term outcomes for persons living with diabetes and is now standard of care in type 1 diabetes. Their importance during inpatient admissions and surgery is increasingly recognised but data is limited.

During surgery, diathermy is often used to control bleeding by using high-frequency electrical energy to seal small blood vessels. Available literature supports the hypothesis that CGM systems may not be affected by diathermy, but there is limited information on the actual impact of diathermy on the accuracy of CGM systems. At present, CGM systems warranties are invalidated by diathermy use. Consequently, patients undergoing surgery require a period of blood glucose testing to monitor and respond to glucose levels which are often rendered unstable in the period around surgery. This exposes patients to potentially unnecessary interventions and, for those patients who are needle-phobic or have learning difficulties, the need to renew repeated finger-prick capillary blood testing around surgery may be a significant barrier to effective and optimal glucose monitoring and control. In addition, there may be a delay in responding to hypoglycaemia and hyperglycaemia, due to the time between measurements.

This study aims to assess the impact of intraoperative diathermy use on the accuracy and functioning of CGM systems. Demonstrating that glucose monitoring systems could be effectively used during surgery would:

  • Prevent/reduce the need for the anaesthetic team to access the patients' fingers/toes and venous/arterial lines for blood glucose testing during surgery.
  • Provide easy, convenient and accurate glucose monitoring to the anaesthetic team, with data transmitted from monitoring systems which can remain underneath surgical coverings.
  • Provide prompt, real-time information on glucose trends during surgery, leading to better control of glucose levels including earlier recognition of low and high glucose levels (hypoglycaemia and hyperglycaemia, respectively).

Study design:

This is a single-centre, prospective, single-arm, observational, non-inferiority study of adults and young people with diabetes using CGM systems. 126 participants will be recruited. Sample size was calculated using simulation.

Participants will undergo paired glucometer and sensor glucose measurements every 15-75 minutes from up to 4 hours before the planned time of surgery until up to 4 hours after the completion of surgery, for not more than 24 hours. They will also have two venous blood samples taken for glucose concentration analysis.

Patient recruitment:

Participants will be recruited from patients due to have an elective surgical procedure at Nottingham University Hospital NHS Trust. The initial approach may be from a member of the patient's usual care team with consent to share contact details with the study team. In addition, the Trust's Theatre Management system (Nervecentre) will be reviewed to identify patients with type 1 diabetes who are due to attend for surgery.

Data management and analysis:

Data will be collected by members of the research team using study specific case report forms (CRFs). An ANOVA mixed effects model will be used with participant as a random effect and pre/post diathermy as a fixed effect. An equivalence F-test with an ARD equivalence value of 0.05 will be performed to determine effect of diathermy on ARD.

Quality assurance plan:

Monitoring of study data shall include confirmation of informed consent; source data verification; data storage and data transfer procedures; local quality control checks and procedures, back-up and disaster recovery of any local databases and validation of data manipulation. Entries on CRFs will be verified by inspection against the source data. A 10% sample of CRFs will be checked on a regular basis for verification of all entries made. In addition, the subsequent capture of the data on the study database will be checked. Where corrections are required, these will carry a full audit trail and justification. Study data and evidence of monitoring and systems audits will be made available for inspection by the REC as required.

Reporting adverse effects:

Intraoperative glucose monitoring is currently undertaken using blood glucose meters. The same monitoring will be continued through this study and this study involves no deviation from standard care. Sensor glucose measurements are passively recorded from the participant's receiving device and eligible participants would already be routinely using sensors as part of their diabetes care.

Equipment failure issues, where a sensor or blood glucose measurement is not able to be obtained, will be recorded on the CRF. Adverse events as a result of participation within this study are not expected but will be recorded on an Adverse Event Form if any occurs. Clinical deterioration and surgical complications, not related to the study, will be considered expected in terms of classification.

研究类型

观察性的

注册 (估计的)

126

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习地点

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

  • 孩子
  • 成人
  • 年长者

接受健康志愿者

不

取样方法

非概率样本

研究人群

Patients aged 4 or older with type 1 diabetes mellitus scheduled to undergo elective surgery at Nottingham Universities Hospitals NHS Trust.

描述

Inclusion Criteria:

  • Diagnosis of type 1 diabetes mellitus.
  • Planned admission for elective surgery at Nottingham University Hospitals NHS Trust.
  • Currently routinely using a prescribed Abbott Libre-2, Abbott Libre-3, Dexcom G6 or Dexcom G7 glucose monitoring system to monitor glucose concentrations.
  • Aged 4 years or older on the day of surgery.
  • Ability to give informed consent / ability of parent/carer to give informed consent (as appropriate).

Exclusion Criteria:

  • Use of paracetamol above maximum dose within 7 days before the scheduled date of surgery (in adults with body weight under 51kg - maximum of 60mg/kg per day, in adults with body weight 51kg and above - maximum of 4g per day, in children - maximum dose as per their age as stated in the British National Formulary for Children)
  • Any use of hydroxyurea within 7 days before the scheduled date of surgery
  • Use of more than 500mg per day of ascorbic acid / vitamin C within 7 days before the scheduled date of surgery
  • Currently taking medications as part of a clinical trial.

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

队列和干预

团体/队列
Patients with T1DM aged 4 or older routinely using CGM systems undergoing elective surgery
Patients (aged 4 or above) with a diagnosis of type 1 diabetes mellitus who routinely use a prescribed Abbott Libre-2, Abbott Libre-3, Dexcom G6 or Dexcom G7 glucose monitoring system to monitor glucose concentrations. Patients will scheduled to have a planned admission for elective surgery.

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
To assess if the Absolute Relative Differences (ARDs) of continuous glucose monitoring systems change after use of diathermy.
大体时间:From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. The null hypothesis is that there is no difference (non-inferiority) following diathermy (where equivalence is defined as less than a 0.05 difference in ARD).
From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

次要结果测量

结果测量
措施说明
大体时间
To assess whether ARD changes following diathermy depending on type of diathermy used (bipolar/monopolar).
大体时间:From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.
From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
To assess whether ARD changes following diathermy when the model is adjusted for glucose trend (rate of change).
大体时间:From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.
From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
To assess whether ARD changes following diathermy when the model is adjusted for duration of surgery (minutes).
大体时间:From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.
From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
To assess whether ARD changes following diathermy when the model is adjusted for site of surgery.
大体时间:From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.
From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
To assess whether ARD changes following diathermy according to participant characteristic of age (years).
大体时间:From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.
From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
To assess whether ARD changes following diathermy according to participant characteristic of weight (kg).
大体时间:From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.
From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
To assess whether ARD changes following diathermy according to participant characteristic of BMI (or BMI SDS in under 18s).
大体时间:From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.
From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
To assess sensor accuracy before and after diathermy using error grid analysis.
大体时间:From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
Error grid analysis will be used to assess the effect of CGM systems accuracy after diathermy use on clinical decision making.
From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
To assess whether the ARD of each system changes after use of diathermy using laboratory glucose concentration results as the reference readings.
大体时间:From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.
The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (laboratory glucose concentration) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.
From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

调查人员

  • 研究主任:James M Law, PhD, BMBS、University Hospitals of Derby and Burton NHS Foundation Trust, University of Nottingham
  • 首席研究员:James Armstrong, BMBS、Nottingham University Hospitals NHS Trust

出版物和有用的链接

负责输入研究信息的人员自愿提供这些出版物。这些可能与研究有关。

一般刊物

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (估计的)

2026年5月1日

初级完成 (估计的)

2027年2月1日

研究完成 (估计的)

2027年2月1日

研究注册日期

首次提交

2026年4月13日

首先提交符合 QC 标准的

2026年5月6日

首次发布 (实际的)

2026年5月13日

研究记录更新

最后更新发布 (实际的)

2026年5月13日

上次提交的符合 QC 标准的更新

2026年5月6日

最后验证

2026年3月1日

更多信息

与本研究相关的术语

计划个人参与者数据 (IPD)

计划共享个人参与者数据 (IPD)?

不

药物和器械信息、研究文件

研究美国 FDA 监管的药品

不

研究美国 FDA 监管的设备产品

不

此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.