Study of Shunt Flow Under Different Valve Settings Using a Wireless Sensor

2026年6月25日 更新者:Rhaeos, Inc.

A Prospective Study Comparing CSF Shunt Flow Between "Virtual Off" and "Flow-Permissive" Shunt Valve Settings Using a Wireless, Noninvasive Shunt Flow Sensor

Hydrocephalus is commonly treated with a cerebrospinal fluid (CSF) shunt, a device that drains excess fluid from the brain. However, shunts can stop working over time, and determining whether fluid is flowing through a shunt can be difficult. Current diagnostic methods may require imaging studies or invasive procedures, which can be time-consuming, costly, or uncomfortable for patients.

The purpose of this study is to evaluate the performance of the Rhaeos Shunt Monitoring System (RSMS), an investigational wireless, noninvasive device designed to detect flow in implanted CSF shunts. The study will compare RSMS directly with ShuntCheck III, an FDA-cleared noninvasive device that is also used to assess CSF shunt flow. Both devices are placed on the skin over the shunt and use thermal sensing technology to evaluate whether fluid is flowing through the shunt.

Approximately 85 pediatric and adult participants with an existing ventricular CSF shunt will be enrolled. Participants must have a functioning shunt and no recent symptoms suggesting shunt malfunction. Some participants with programmable shunt valves will have their valve temporarily adjusted to a manufacturer-provided "virtual off" setting, which is intended to substantially restrict shunt flow. These participants may also undergo measurements at their normal valve setting. Other participants will only undergo measurements at their normal valve setting.

During the study visit, participants will undergo shunt flow assessments with both RSMS and ShuntCheck III. The order in which the devices are used will be randomized. Device measurements are collected while participants are seated upright, and participants and their treating clinicians will remain blinded to the measurement results. The study devices will not be used to make treatment decisions, and participation will not alter standard medical care. Adverse events associated with device use or valve adjustments will be monitored and recorded.

The primary goal of the study is to compare the ability of RSMS and ShuntCheck III to correctly identify shunts with flow-permissive ("valve normal") and restricted-flow ("virtual off") settings. Results from this study may help support the future development of a noninvasive tool to aid neurosurgeons in assessing CSF shunt flow. Participants are not expected to receive a direct medical benefit from taking part in the study.

研究概览

详细说明

Hydrocephalus is commonly treated with surgically implanted cerebrospinal fluid (CSF) shunts that divert excess CSF from the brain to another part of the body. Shunt malfunction is a frequent complication of treatment and can result in significant morbidity if not identified and treated promptly. However, diagnosis of shunt malfunction remains challenging because patient symptoms are often nonspecific and currently available diagnostic tools have important limitations. Existing noninvasive assessments generally rely on indirect indicators of shunt function, while invasive procedures may expose patients to discomfort, infection risk, or other complications.

The Rhaeos Shunt Monitoring System (RSMS) is an investigational, wireless, noninvasive device designed to aid in the detection of flow in implanted CSF shunts. RSMS uses thermal sensing technology to evaluate CSF flow through shunt tubing by measuring temperature changes associated with fluid movement. The device is placed on the skin overlying a palpable shunt segment, typically near the clavicle, and communicates wirelessly with a tablet application to perform measurements and record results.

This study is a prospective, multicenter, open-label, blinded, observational study designed to compare the performance of RSMS with ShuntCheck III, an FDA-cleared noninvasive device indicated as an aid to the detection of flow in implanted CSF shunts. Both devices use transcutaneous thermal sensing methods to assess shunt flow, but they differ in implementation. RSMS uses a controlled heating element and wireless electronics, while ShuntCheck III uses a disposable thermosensor and cooling from an instant cold pack.

Approximately 85 pediatric and adult participants with existing ventricular CSF shunts will be enrolled. Eligible participants must have a palpable shunt segment suitable for device placement and must not have recent symptoms suggestive of shunt malfunction.

Participants with programmable shunt valves capable of being adjusted to a manufacturer-defined "virtual off" setting (opening pressure greater than 400 mm H₂O) will undergo device assessments after the valve has been adjusted to the virtual off setting. These participants may also undergo optional measurements at their prescribed flow-permissive ("valve normal") setting prior to valve adjustment. Participants whose valves are not adjusted will undergo measurements only at their prescribed valve setting. For each assessment, the order of RSMS and ShuntCheck III testing will be randomized. Participants and treating clinicians will remain blinded to device outputs throughout the study.

The primary objective of the study is to compare the sensitivity and specificity of RSMS and ShuntCheck III for detecting flow-restricted and flow-permissive shunt conditions. Device performance will be evaluated using the known valve setting as the reference condition. Measurements obtained while the valve is set to the virtual off setting will serve as flow-restricted conditions, while measurements obtained at valve-normal settings will serve as flow-permissive conditions.

Secondary analyses will evaluate additional performance characteristics, including accuracy, positive predictive value, negative predictive value, receiver operating characteristic (ROC) curve analysis, measurement completion rates, and the number of measurement attempts required to obtain a successful result. Exploratory analyses will assess device performance across clinical and demographic subgroups and study-specific variables such as body position, shunt depth, valve setting, and study site.

All study procedures are conducted in conjunction with standard clinical care. Device measurements are used solely for research purposes and will not be available to treating clinicians for patient management decisions. Participants undergoing temporary valve adjustment will be monitored following restoration of their prescribed valve setting, and all participants will be monitored for adverse events associated with device use. Adverse events, device deficiencies, and protocol deviations will be documented throughout the study.

The results of this study are intended to provide clinical evidence regarding the performance of a next-generation, wireless, noninvasive shunt flow monitoring technology and to support future regulatory evaluation of the RSMS device.

研究类型

介入性

注册 (估计的)

85

阶段

  • 不适用

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习联系方式

  • 姓名:Anna L Somera
  • 电话号码:855-814-3569
  • 邮箱:info@rhaeos.com

学习地点

    • Illinois
      • Chicago、Illinois、美国、60642
        • Rhaeos, Imc
        • 接触:

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

  • 孩子
  • 成人
  • 年长者

接受健康志愿者

不

描述

Inclusion Criteria:

Existing ventricular cerebrospinal fluid (CSF) shunt. Region of intact skin overlying an unambiguously identifiable, palpable, chronically indwelling ventricular shunt crossing the clavicle and appropriate in size for application of the study device.

Written informed consent provided by the participant or, when applicable, by a parent, legal guardian, health care agent, or surrogate decision maker according to local regulations.

For participants with planned "virtual off" measurements:

Existing ventricular CSF shunt with a programmable valve that has an available manufacturer-reported "virtual off" setting and can be programmed noninvasively.

For participants whose valve will be adjusted to the "virtual off" setting during the study:

Subject-reported or documented history of successful shunt valve adjustment(s).

Exclusion Criteria:

Presence of more than one distal shunt catheter in the study device measurement region.

Presence of an open wound or edema in the study device measurement region that would interfere with device placement.

Shunt difficult to palpate or shunt depth greater than 5 mm from the skin surface at the measurement location as determined by ultrasound.

Subject-reported or documented history of serious adverse skin reactions to silicone-based adhesives.

New or worsening symptoms suggestive of shunt malfunction within the previous 7 days.

Use of the study device would interfere with standard patient care, emergency surgery cannot be delayed, or study participation would interfere with optimal medical care.

Participation in another investigational procedural, pharmaceutical, or device study that may affect data collected in this study.

Previous participation in a Rhaeos clinical study using the current-generation RH05-B1 study device.

For participants with planned "valve normal" measurements only:

Investigator judges the participant is likely to be shunt independent (e.g., does not require a functional shunt for management of hydrocephalus).

This format follows ClinicalTrials.gov conventions by separating general criteria from those that apply only to specific participant subgroups.

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 主要用途:诊断
  • 分配:不适用
  • 介入模型:单组作业
  • 屏蔽:无(打开标签)

武器和干预

参与者组/臂
干预/治疗
其他:Participants Undergoing Paired CSF Shunt Flow Assessment
Participants undergo paired, noninvasive assessment of cerebrospinal fluid (CSF) shunt flow using both the investigational Rhaeos® Shunt Monitoring System (RSMS) and the FDA-cleared predicate device, ShuntCheck III. The order of device use is randomized for each participant. All participants are evaluated at their prescribed flow-permissive ("valve normal") shunt setting. Participants with programmable shunt valves that can be adjusted to a manufacturer-reported "virtual off" setting also undergo measurements after temporary adjustment to the "virtual off" setting, followed by return of the valve to the prescribed setting. Device results are blinded to participants and treating clinicians and are not used for clinical decision-making. Adverse events are monitored throughout study participation.
The Rhaeos Shunt Monitoring System (RSMS) is an investigational, wireless, noninvasive thermal flow detection device that is applied to the skin overlying an implanted cerebrospinal fluid (CSF) shunt. The device uses controlled thermal energy and temperature sensors to detect thermal patterns associated with CSF flow. Measurements are obtained using a tablet application, and device results are encoded to maintain blinding. The device is used for research purposes only and is not used to guide clinical care.
ShuntCheck III is an FDA-cleared, noninvasive thermal flow detection device used as the predicate comparator. The device uses transcutaneous thermal dilution produced by an instant cold pack and a disposable thermosensor array applied over the implanted CSF shunt to detect shunt flow. Measurements are obtained using the manufacturer's software, and results are encoded to maintain blinding. The device is used for comparison with the investigational device and is not used to guide clinical care during the study.

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Sensitivity of the Rhaeos Shunt Monitoring System Compared With ShuntCheck III
大体时间:Day 1 (study visit)
Sensitivity of the investigational Rhaeos Shunt Monitoring System (RSMS) compared with the FDA-cleared predicate device (ShuntCheck III) for identifying non-flowing cerebrospinal fluid shunts during the "virtual off" valve setting. Sensitivity is calculated as the proportion of "virtual off" measurements correctly identified as "flow not confirmed."
Day 1 (study visit)
Specificity of the Rhaeos® Shunt Monitoring System Compared With ShuntCheck III
大体时间:Day 1 (study visit)
Specificity of the investigational Rhaeos Shunt Monitoring System (RSMS) compared with the FDA-cleared predicate device (ShuntCheck III) for identifying flow-permissive cerebrospinal fluid shunts during the prescribed "valve normal" setting. Specificity is calculated as the proportion of flow-permissive measurements correctly identified as "flow confirmed."
Day 1 (study visit)

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

出版物和有用的链接

负责输入研究信息的人员自愿提供这些出版物。这些可能与研究有关。

一般刊物

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (估计的)

2026年6月30日

初级完成 (估计的)

2026年11月30日

研究完成 (估计的)

2026年12月31日

研究注册日期

首次提交

2026年6月25日

首先提交符合 QC 标准的

2026年6月25日

首次发布 (实际的)

2026年7月1日

研究记录更新

最后更新发布 (实际的)

2026年7月1日

上次提交的符合 QC 标准的更新

2026年6月25日

最后验证

2026年6月1日