Wireless EMG-Based Monitoring of Paraspinal Muscle Activation in Adolescent Idiopathic Scoliosis (AIS-EMG)
Design of a Wireless and Non-Invasive EMG-Based Embedded Measurement System for Monitoring Muscle Activation Asymmetries in Scoliosis
研究概览
详细说明
Adolescent idiopathic scoliosis (AIS) is a three-dimensional spinal deformity typically diagnosed and monitored through radiographic assessment of the Cobb angle. While imaging remains the clinical gold standard, it involves radiation exposure and is typically performed at limited intervals, making it less suitable for frequent, short-term monitoring of treatment response. Prior research has shown that paraspinal muscle activation asymmetries - particularly increased convex-side activation near the apex of the curve - are consistently observed in AIS and may serve as early, objective markers of curve progression and treatment response, as demonstrated in prospective cohort and machine-learning-based studies.
This study will use a custom-developed, low-power, wireless, embedded surface EMG (sEMG) measurement system, consisting of skin-surface EMG electrodes, an analog front-end (AFE) for signal conditioning and digitization, and a microcontroller-based wireless communication module. EMG signals will be sampled at a minimum of 1000 Hz using a differential (bipolar) electrode configuration to minimize noise and common-mode interference.
Electrodes will be placed bilaterally on the paraspinal muscles at the level of the apical vertebra of the major spinal curve, approximately 3 cm lateral to the spinous process, oriented parallel to the muscle fibers. Standard skin preparation procedures (shaving, alcohol cleaning) will be used to reduce skin impedance prior to electrode placement.
EMG recordings will be performed under two task conditions: a static task, in which participants sit upright with hips and knees flexed at 90°, arms at their sides, and feet on the floor; and a dynamic task, in which participants perform active trunk extension in the prone position for a defined number of repetitions. These tasks are designed to capture both static postural and dynamic movement-related muscle activation patterns.
Recorded EMG signals will be processed to calculate root mean square (RMS) and mean absolute value (MAV) parameters, with frequency-domain (FFT) analysis performed as needed. Convex- and concave-side RMS values at each vertebral level will be used to derive muscle activation ratios and asymmetry indices. Measurements will be repeated at different time points over the course of physiotherapy treatment to evaluate longitudinal changes in muscle activation patterns, and associations between EMG-based indices and clinical treatment outcomes will be examined using appropriate statistical methods.
The study is being conducted in two phases: an initial technical validation phase, in which system performance is verified using commercial EMG sensor evaluation boards and microcontroller-based development platforms; and a subsequent phase in which a custom-designed wireless sEMG system (incorporating a proprietary analog front-end and embedded control unit) is evaluated for feasibility and usability in monitoring muscle activation patterns in scoliosis patients during physiotherapy. Only the human-subjects (clinical) component of this work is registered here; the underlying hardware/firmware engineering and bench validation are not within the scope of this trial record.
研究类型
注册 (估计的)
阶段
- 不适用
联系人和位置
学习联系方式
- 姓名:Alpaslan Ersöz, Ph.D
- 电话号码:+905312561935
- 邮箱:alpaslan.ersoz@amasya.edu.tr
学习地点
-
-
Amasya
-
Amasya、Amasya、土耳其(türkiye)、05100
- Amasya University
-
副研究员:
- Ertuğrul Deniz Köse, Ph.D
-
接触:
- Alpaslan Ersöz, Ph.D
- 电话号码:+905312561935
- 邮箱:alpaslan.ersoz@amasya.edu.tr
-
首席研究员:
- Alpaslan Ersöz, Ph.D
-
副研究员:
- Eylem Küçük, Ph.D
-
-
参与标准
资格标准
适合学习的年龄
- 孩子
- 成人
接受健康志愿者
描述
Inclusion Criteria:
- Diagnosed with adolescent idiopathic scoliosis
- Cobb angle between 10° and 25°
- No prior surgical or conservative treatment for scoliosis
Exclusion Criteria:
- Neurological disease
- Significant concomitant orthopedic problems
- Back pain severe enough to interfere with the measurement process
学习计划
研究是如何设计的?
设计细节
- 主要用途:诊断
- 分配:不适用
- 介入模型:单组作业
- 屏蔽:无(打开标签)
武器和干预
参与者组/臂 |
干预/治疗 |
|---|---|
|
实验性的:Experimental: Wireless EMG Monitoring
All enrolled participants (adolescents with idiopathic scoliosis, Cobb angle 10°-25°) will undergo surface EMG monitoring using the study's custom-developed wireless, embedded measurement system.
Bilateral paraspinal EMG recordings will be obtained at the level of the apical vertebra during standardized static (seated) and dynamic (prone trunk extension) tasks.
Measurements will be repeated at multiple time points over the course of each participant's physiotherapy treatment to longitudinally assess muscle activation asymmetry.
|
A custom-developed, low-power, wireless, non-invasive surface EMG system consisting of skin-surface electrodes, an analog front-end (AFE) for signal conditioning and digitization, and a microcontroller-based wireless communication module.
Electrodes are placed bilaterally on the paraspinal muscles at the apical vertebra level, approximately 3 cm lateral to the spinous process.
EMG signals are sampled at ≥1000 Hz using a differential (bipolar) configuration and analyzed for root mean square (RMS), mean absolute value (MAV), and frequency-domain (FFT) parameters to quantify muscle activation asymmetry between the convex and concave sides of the spinal curve.
|
研究衡量的是什么?
主要结果指标
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Change in Paraspinal Muscle Activation Asymmetry Index
大体时间:Baseline and through study completion, an average of 1 year
|
An asymmetry index will be calculated from bilateral surface EMG RMS (root mean square) values recorded at the apical vertebra level during static and dynamic tasks, comparing convex- and concave-side paraspinal muscle activation.
Change in this index over the course of treatment will be evaluated.
|
Baseline and through study completion, an average of 1 year
|
次要结果测量
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Correlation Between EMG-Based Muscle Activation Asymmetry Index and Clinical Treatment Outcome (Cobb Angle Change)
大体时间:Baseline and through study completion, an average of 1 year
|
The relationship between change in EMG-derived muscle activation asymmetry and change in Cobb angle (measured via standard clinical radiography) will be evaluated using appropriate correlation analysis.
|
Baseline and through study completion, an average of 1 year
|
|
Device Signal Acquisition Success Rate
大体时间:through study completion, an average of 1 year
|
The proportion of EMG recording sessions in which signal quality meets predefined criteria for analysis (e.g., acceptable signal-to-noise ratio, no excessive artifact), assessing the technical feasibility of the wireless embedded measurement system in a clinical setting.
|
through study completion, an average of 1 year
|
合作者和调查者
出版物和有用的链接
一般刊物
- Weinstein SL, Dolan LA, Cheng JC, Danielsson A, Morcuende JA. Adolescent idiopathic scoliosis. Lancet. 2008 May 3;371(9623):1527-37. doi: 10.1016/S0140-6736(08)60658-3.
- Liang R, Yip J, Fan Y, Cheung JPY, To KM. Electromyographic Analysis of Paraspinal Muscles of Scoliosis Patients Using Machine Learning Approaches. Int J Environ Res Public Health. 2022 Jan 21;19(3):1177. doi: 10.3390/ijerph19031177.
- Fan Y, To MK, Yeung EHK, Kuang GM, Liang R, Cheung JPY. Electromyographic Discrepancy in Paravertebral Muscle Activity Predicts Early Curve Progression of Untreated Adolescent Idiopathic Scoliosis. Asian Spine J. 2023 Oct;17(5):922-932. doi: 10.31616/asj.2023.0199. Epub 2023 Sep 11.
- Chan WWY, Fu SN, Chong TF, Singh G, Tsai DSJ, Wong MCY, Zheng YP, Parent EC, Cheung JPY, Wong AYL. Associations between paraspinal muscle characteristics and spinal curvature in conservatively treated adolescent idiopathic scoliosis: a systematic review. Spine J. 2024 Apr;24(4):692-720. doi: 10.1016/j.spinee.2023.11.008. Epub 2023 Nov 24.
研究记录日期
研究主要日期
学习开始 (估计的)
初级完成 (估计的)
研究完成 (估计的)
研究注册日期
首次提交
首先提交符合 QC 标准的
首次发布 (实际的)
研究记录更新
最后更新发布 (实际的)
上次提交的符合 QC 标准的更新
最后验证
更多信息
与本研究相关的术语
其他研究编号
- Amasya Uni-EMG
计划个人参与者数据 (IPD)
计划共享个人参与者数据 (IPD)?
IPD 计划说明
药物和器械信息、研究文件
研究美国 FDA 监管的药品
研究美国 FDA 监管的设备产品
此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.