Kosmos Guidance Performance Evaluation Data Annotation

2026年8月13日 更新者:EchoNous Inc.
The purpose of this study is to supplement prior Kosmos Trio Guidance performance testing conducted under ECHO-020 Kosmos Guidance, Grading, and Capture Performance Evaluation Protocol. The Kosmos Trio Guidance algorithm is designed to provide guidance to help novice scanners acquire images and videos during ultrasound scanning of three cardiac views (Parastenal Long Axis, Apical 4 Chamber, and Apical 2 Chamber) with Kosmos Trio. The study will divide the guidance cues not evaluated sufficiently in the ECHO-020 study into four categories: 1) corrective guidance, 2) incorrect scanning window, 3) incorrect probe orientation, and 4) low confidence prediction. For each guidance cue evaluated in this study, expert annotators will review two clips: one suboptimal view, collected when the probe is positioned at an incorrect location; and one guided view, captured after the sonographer followed the probe motion cue displayed by the system. Annotators will review each pair of clips and respond indicating either i) Yes, the guided clip demonstrates the appropriate correction from the suboptimal view; ii) No, the guided clip does not demonstrate the appropriate correction from the suboptimal view; or iii) Unevaluable, the guided clip is technically inadequate for assessment. For each category of guidance cues, the outcome measure will be the rate at which annotators respond "Yes." Finally, this study will also evaluate the Kosmos Trio Auto-Capture algorithm's ability to detect and capture a diagnostic-quality image. Annotators will rate clips captured by Kosmos Trio Auto-Capture on an ACEP-based image quality scale. The outcome measure will be the image quality score. The study design involves the compilation of de-identified retrospective cardiac clips acquired using the Kosmos diagnostic ultrasound system. This data was collected for two previous studies; no additional data will be collected from previous study participants for the purposes of this study.

研究概览

研究类型

观察性的

注册 (估计的)

81

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习地点

    • Washington
      • Redmond、Washington、美国、98052
        • EchoNous, Inc.

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

  • 成人
  • 年长者

接受健康志愿者

是的

取样方法

非概率样本

研究人群

The study used retrospective datasets from 2 sources: ECHO-020 Kosmos Guidance, Grading, and Capture Performance Evaluation study, and ECHO-025 Kosmos Guidance Performance Evaluation study. Both datasets comprise fully de-identified, retrospective, echocardiographic clips acquired using the Kosmos diagnostic ultrasound system.

Data collection for both studies was performed at EchoNous headquarters in Redmond, Washington. Source data was collected targeting a balanced representation of demographics and clinical characteristics (including gender and BMI) to ensure the generalizability of study findings.

描述

Inclusion Criteria:

  • All data used in this study must have been collected per 45 CFR § 26,
  • Data must have been collected per source protocol requirements with no deviation for that data set,
  • Adequate demographic information available from original study participation (sex, BMI, ethnicity, and race)
  • Exam acquisition performed by a qualified echocardiographer using the Kosmos system

Exclusion Criteria:

  • Inadequate demographic information corresponding to the available data, and
  • Non-echocardiographer acquired images.

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

队列和干预

团体/队列
干预/治疗
Project 1
Participants included in Project 1, Category 1: Corrective Guidance
When the system detects that the probe is near the correct cardiac window but requires a specific directional adjustment, it displays a Category 1 - Corrective Guidance message. Each case in this project consists of two clips: Clip A - Suboptimal View: captured after the sonographer has intentionally moved the probe to a prescribed off-target position; and Clip B - Guided View: captured after the sonographer followed the Category 1 probe motion cue displayed by the system. For each pair of clips, annotators will be asked when Comparing Clip A and Clip B whether following the Category 1 corrective guidance instruction produces a discernible improvement in overall image quality. They will provide an answer of Yes, No, or Unevaluable.
Project 2
Participants included in Project 2, Category 2: Incorrect Scanning Window
When the system detects that the probe is positioned at a fundamentally incorrect anatomical location relative to the target cardiac view, it displays a Category 2 - Incorrect Scanning Window message. Each case in this project consists of two clips: Clip A - Suboptimal View: captured after the sonographer has intentionally a clearly incorrect anatomical location; and Clip B - Guided View: captured after the sonographer followed the Category 2 window-navigation cue and moved to the correct scanning window displayed by the system. For each pair of clips, annotators will be asked when Comparing Clip A and Clip B whether following the Category 2 guidance instruction produces a clip in the correct scanning window. They will provide an answer of Yes, No, or Unevaluable.
Project 3
Participants included in Project 3, Category 3: Incorrect Probe Orientation
When the system detects that the probe has been rotated 180° around its longitudinal axis, producing a laterally mirrored image, it displays a Category 3 - Incorrect Probe Orientation message. Each case in this project consists of two clips: Clip A - Suboptimal View: with the probe rotated 180° at the correct cardiac window, producing a mirrored image; and Clip B - Guided View: sonographer corrected the probe orientation in response to the Category 3 message displayed by the system. For each pair of clips, annotators will be asked when Comparing Clip A and Clip B whether following the Category 3 guidance instruction produced a clip that is oriented correctly. They will provide an answer of Yes, No, or Unevaluable.
Project 4
Participants included in Project 4, Category 4: Low Confidence Prediction
When the system's confidence in the current image is insufficient to issue a specific directional correction, it displays a Category 4 - Low Confidence Prediction message. Each case in this project consists of two clips: Clip A - Suboptimal View: with the probe rotated 180° at the correct cardiac window, producing a mirrored image; and Clip B - Guided View: sonographer corrected the probe orientation in response to the Category 3 message displayed by the system. For each pair of clips, annotators will be asked when Comparing Clip A and Clip B whether following the Category 3 guidance instruction produced a clip that is oriented correctly. They will provide an answer of Yes, No, or Unevaluable.
Project 5
Participants included in Project 5, Auto-Capture Frame-Level Re-annotation
The Kosmos Trio Auto-Capture feature automatically triggers clip recording when it detects that a diagnostic quality image has been achieved. Each case in this project is a single captured clip. No comparison clip is provided. The view (PLAX, A4C, or A2C) is indicated for each case. For each frame in this clip, annotators will be asked to assign a quality grade using the ACEP 1-5 image quality scale (where 1 represents a frame in which the relevant anatomy is not visualized or identifiable, and 5 represents a frame in which the image is sharp and the relevant anatomy is well visualized and present in the frame) to each frame.

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Guidance Algorithm Accuracy
大体时间:Post data acquisition (2 months)
Guidance algorithm accuracy will be evaluated using Positive Predictive Value (PPV), a calculation that compares the diagnostic quality between the suboptimal view following an indicated probe movement and the guided view following the Guidance correction of the probe location. PPV will be calculated for each cardiac view included in this study (PLAX, A4C, and A2C), across all four Guidance categories. PPV is calculated by dividing the number of True Positives (the number of trial pairs in which the post-action clip demonstrates that following the guidance instruction produced the intended improvement for the applicable guidance category) by the number of True Positives plus the number of False Positives (the number of trial pairs in which the post-action clip does not demonstrate the expected improvement for the applicable category). Success is defined as a PPV > 0.80 and 95% CI lower bound > 0.80 for each view. All three views must independently satisfy this criteria.
Post data acquisition (2 months)
Performance of Auto-Capture
大体时间:Post data acquisition (2 months)
The accuracy of Auto-Capture will be evaluated using Positive Predictive Value (PPV), a calculation that provides the proportion of clips captured using Auto-Capture that contain a sufficient proportion of diagnostic-quality frames to be considered clinically acceptable. PPV is calculated by dividing the number of True Positives (the number of clips with at least one full cardiac cycle of frames is determined to be diagnostic by expert reviewers, in accordance with the ACEP-based grading scale) by the number of True Positives plus the number of False Positives (the number of clips that do not have at least one full cardiac cycle of frames is determined to be diagnostic by expert reviewers, in accordance with the ACEP-based grading scale). Success is defined as a PPV > 0.80 and 95% CI lower bound > 0.80.
Post data acquisition (2 months)

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

赞助

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (实际的)

2026年6月5日

初级完成 (估计的)

2026年8月30日

研究完成 (估计的)

2026年9月30日

研究注册日期

首次提交

2026年8月13日

首先提交符合 QC 标准的

2026年8月13日

首次发布 (实际的)

2026年8月18日

研究记录更新

最后更新发布 (实际的)

2026年8月18日

上次提交的符合 QC 标准的更新

2026年8月13日

最后验证

2026年8月1日

更多信息

与本研究相关的术语

其他相关的 MeSH 术语

其他研究编号

  • ECHO-302

计划个人参与者数据 (IPD)

计划共享个人参与者数据 (IPD)?

未定

IPD 计划说明

Publication decisions will be made at a later date.