Assessment of Binocular Vision Function With Neos™ (ROCKET) (ROCKET)
Clinical Performance of Neos™ for the Assessment of Binocular Vision Function: A Prospective Multicentre Cross-Sectional Study (ROCKET)
ROCKET evaluates the clinical performance of an investigational version of neos™, a non-invasive device that uses a virtual-reality headset and eye tracking to assess binocular vision-how the two eyes work together.
The study compares neos™ results with standard clinical examinations for three functions: sensory fusion, the ability to combine images from both eyes into a single image; stereopsis, the perception of depth from differences between the images seen by each eye; and suppression, reduced awareness of visual input from one eye during binocular viewing.
Approximately 184 patients aged 3 years and older who require a clinical binocular vision assessment will participate at ophthalmology centres in Switzerland. Each participant will undergo standard clinical testing and a neos™ examination during a single study visit. During the neos™ examination, visual stimuli are presented through the headset while the device records eye movements and pupil responses.
The main objective is to determine how accurately neos™ identifies the presence or absence of sensory fusion, stereopsis and central suppression compared with the clinical reference assessments. Additional analyses will compare fusion-related measurements, characterize stereopsis and suppression, and explore relationships between neos™ measurements and clinical findings.
研究概览
详细说明
The investigation uses a prospective, multicentre, cross-sectional, paired within-participant design. Each participant receives both assessments on the same day and serves as their own comparator, reducing the influence of differences between participants. There is no random allocation. Although the investigation is not formally blinded, clinicians performing the reference examination do not have access to automated neos™ results when determining the clinical classifications. Investigational results do not influence treatment decisions.
The investigation comprises two sequential phases. An initial tuning phase includes 10 selectively recruited participants: five with preserved fusion and stereopsis without suppression, and five with absent fusion and stereopsis and suppression present. These data are used to refine algorithm parameters and thresholds. The algorithm configuration is then locked and remains unchanged throughout the validation phase. Tuning-phase participants are excluded from all analysis sets.
During validation, eligible patients are recruited consecutively when study personnel and procedures are available. Recruitment includes patients with and without suspected or confirmed binocular dysfunction, reflecting the population undergoing clinical binocularity assessment. Approximately 174 participants are expected to enter validation to obtain around 157 evaluable participants. The final enrolment depends on the observed distribution of clinical reference classifications and the proportion of evaluable examinations.
The investigational configuration uses neos™ software version 4.0.0-rocket, incorporating additional binocular visual stimuli and algorithms. The examination typically takes less than 15 minutes. It includes gaze holding and ocular alignment, stimuli for assessing suppression, fusion and stereopsis, pupillary function, saccades, smooth pursuit and visual-field assessment.
The clinical reference examination is performed by an appropriately trained healthcare professional using established methods selected according to clinical findings and participant cooperation. These may include prism bar testing, Bagolini striated lenses, Worth 4-Dot testing and standardized stereotests. Not every test is required for every participant; suitable reference information must be obtained for each primary classification. When specified test results are unavailable, unusable or omitted for clinical reasons, the protocol allows the orthoptist's clinical diagnosis to determine the reference classification.
Prespecified rules define the reference outcomes. Sensory fusion is present when binocular single perception is demonstrated using Bagolini striated lenses or Worth 4-Dot testing. Stereopsis is present when stereoscopic disparity of 600 arcseconds or finer is detected. Central suppression is classified using a hierarchical rule: it is absent when stereopsis or sensory fusion is present. When both are absent, suppression is present if clinical testing demonstrates failure to perceive input from one eye; otherwise, it is classified as absent.
Diagnostic performance is assessed separately for each primary endpoint using a two-by-two comparison of neos™ and clinical reference classifications. Sensitivity and specificity are reported with two-sided 95% exact Clopper-Pearson confidence intervals. An endpoint meets its prespecified success criterion when the lower confidence bounds for both sensitivity and specificity exceed 65%. There is no composite success criterion across the three endpoints and no multiplicity adjustment. Agreement on the affected eye is summarized descriptively among participants classified as suppression-present by both methods.
Validation recruitment ends when all prespecified reference-class requirements are met: at least 41 evaluable participants without sensory fusion, 73 with stereopsis, 73 without stereopsis and 22 with suppression. An operational review at approximately 80 evaluable participants, or earlier if indicated, examines reference-class prevalence to update enrolment projections. It does not examine agreement between neos™ and the reference assessment. No formal interim efficacy or futility analysis is planned.
Secondary analyses compare neos™ fusion break-off angles with clinical prism-bar measurements using Bland-Altman analysis. Agreement in stereopsis severity categories is assessed using weighted Cohen's kappa. Suppression-related outputs and other eye-movement, pupil and visual-field measurements are characterized descriptively, with exploratory comparisons across clinically defined groups. No formal performance goals apply to these secondary analyses.
Primary analyses include participants with conclusive results from both methods for the endpoint concerned. Missing classifications are not imputed. Participants may contribute to one endpoint analysis even if another endpoint is not evaluable. Evaluability rates and reasons for missing or inconclusive results are reported. Safety assessments include validation-phase participants who begin the neos™ examination. No routine follow-up visit is planned; ongoing adverse events are followed as clinically appropriate until resolution or stabilization.
研究类型
注册 (估计的)
阶段
- 不适用
联系人和位置
学习联系方式
- 姓名:Pia Massatsch, PhD
- 电话号码:800-555-5555
- 邮箱:clinical@machinemd.com
参与标准
资格标准
适合学习的年龄
接受健康志愿者
有资格学习的性别
描述
Inclusion Criteria:
- Patients scheduled to undergo, or having undergone, a clinical binocularity assessment at the investigational site on the same day.
- Age ≥3 years at the time of consent.
- Ability to understand and comply with the examination, in the investigator's opinion.
- Written informed consent provided by the participant and/or legal representative.
Exclusion Criteria:
- Best-corrected visual acuity worse than 0.2 decimal (0.7 logMAR) in either eye.
- Refractive error beyond the range that can be corrected by the system, including lens inserts and the use of contact lenses.
学习计划
研究是如何设计的?
设计细节
- 主要用途:诊断
- 分配:不适用
- 介入模型:单组作业
- 屏蔽:无(打开标签)
武器和干预
参与者组/臂 |
干预/治疗 |
|---|---|
|
实验性的:neos™ Assessment and Clinical Reference Examination
All participants undergo the investigational neos™ assessment and a standard clinical binocular vision examination during the same visit. The clinical examination serves as the reference standard for comparing classifications of sensory fusion, stereopsis and central suppression, and for assessing agreement in fusion-related measurements. |
Standard clinical binocular vision examination performed by a trained healthcare professional. Assessments may include prism bar testing, Bagolini striated lenses, Worth 4-Dot testing and standardized stereotests, selected according to clinical findings and participant cooperation. Suitable reference information is obtained for each primary outcome. The examination lasts approximately 30 minutes. |
研究衡量的是什么?
主要结果指标
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Sensitivity and Specificity of neos™ for Detecting Sensory Fusion
大体时间:During the single study visit (approximately 60 minutes).
|
Sensitivity and specificity of the neos™ sensory fusion classification (present/absent) compared with the clinical reference classification. Reference sensory fusion is present when binocular single perception is demonstrated using Bagolini striated lenses or Worth 4-Dot testing. Performance is calculated from paired classifications and reported with two-sided 95% exact Clopper-Pearson confidence intervals. |
During the single study visit (approximately 60 minutes).
|
|
Sensitivity and Specificity of neos™ for Detecting Stereopsis
大体时间:During the single study visit (approximately 60 minutes).
|
Sensitivity and specificity of the neos™ stereopsis classification (present/absent) compared with standard clinical stereotesting. For both methods, stereopsis is classified as present when stereoscopic disparity of 600 arcseconds or finer is detected. Performance is calculated from paired classifications and reported with two-sided 95% exact Clopper-Pearson confidence intervals. |
During the single study visit (approximately 60 minutes).
|
|
Sensitivity and Specificity of neos™ for Detecting Central Suppression
大体时间:During the single study visit (approximately 60 minutes).
|
Sensitivity and specificity of the neos™ central (foveal) suppression classification (present/absent) compared with the clinical reference classification. The reference uses a prespecified hierarchical rule: suppression is absent when stereopsis or sensory fusion is present. When both are absent, suppression is present if clinical testing demonstrates failure to perceive input from one eye; otherwise, it is absent. Performance is reported with two-sided 95% exact Clopper-Pearson confidence intervals. |
During the single study visit (approximately 60 minutes).
|
合作者和调查者
赞助
调查人员
- 首席研究员:Muriel Dysli, MD PhD、Kantonsspital Aarau
研究记录日期
研究主要日期
学习开始 (估计的)
初级完成 (估计的)
研究完成 (估计的)
研究注册日期
首次提交
首先提交符合 QC 标准的
首次发布 (实际的)
研究记录更新
最后更新发布 (实际的)
上次提交的符合 QC 标准的更新
最后验证
更多信息
与本研究相关的术语
其他研究编号
- CI-002
- 2026-D0075 (其他标识符:Swissethics)
计划个人参与者数据 (IPD)
计划共享个人参与者数据 (IPD)?
药物和器械信息、研究文件
研究美国 FDA 监管的药品
研究美国 FDA 监管的设备产品
在美国制造并从美国出口的产品
此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.