Intraoral Aging of 3D Printed Orthodontic Aligners

2026年11月5日 更新者:Azza Gamal Kamel Wafa、Alexandria University

The Effect of Intraoral Aging on 3D-Printed Aligners: A Randomized Split-Mouth Study

This study aims to bridge the gap in literature by conducting a combined evaluation of the mechanical, chemical, optical, and morphological changes of 3D-printed clear aligners manufactured from TC-85 and TA-28 resins that occur after clinical use. By analyzing aligners retrieved after two weeks of intraoral use, this research will offer critical insights into the material stability and aging behavior of the two types of 3D-printed clear aligners.

研究概览

详细说明

Background: Direct 3D-printing of clear aligners is emerging as a promising alternative to conventional thermoformed systems, offering enhanced precision, digital workflow efficiency, and improved material control. Among the photopolymer resins used for 3D-printed aligners, Tera Harz TC-85 and Tera Harz TA-28 have emerged as leading materials due to their reported shape-memory behavior, dimensional stability, and mechanical performance. However, despite increasing clinical adoption, limited evidence exists regarding how these two materials respond to intraoral aging, where fluctuating pH, temperature changes, salivary enzymes, and occlusal forces may alter their mechanical, optical, chemical, and surface characteristics. Understanding these changes is essential for determining the clinical reliability and long-term performance of each resin.Therefore, a demanding evaluation of post-aging changes in 3D-printed aligners is essential to determine their suitability for sustained orthodontic force delivery and to guide material selection in clinical practice. Assessing how key properties such as mechanical behavior, dimensional accuracy, surface topography, and optical characteristics evolve after intraoral exposure will provide valuable insight into material durability and clinical reliability.

研究类型

介入性

注册 (实际的)

12

阶段

  • 不适用

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习地点

    • Alexandria Governorate
      • Alexandria、Alexandria Governorate、埃及、21526
        • Department of Orthodontics, Faculty of Dentistry, Alexandria University

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

18年 至 35年 (成人)

接受健康志愿者

不

有资格学习的性别

全部

描述

Inclusion Criteria:

  • Patients aged 18-35 years.
  • Mild to moderate crowding.
  • Good oral hygiene and healthy periodontium.
  • Permanent dentition with no active caries.

Exclusion Criteria:

  • Patients with systemic diseases affecting saliva or oral mucosa.
  • Smoking or alcohol consumption.
  • History of parafunctional habits.

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 主要用途:基础科学
  • 分配:随机化
  • 介入模型:交叉作业
  • 屏蔽:双倍的

武器和干预

参与者组/臂
干预/治疗
实验性的:TC-85 maxillary / TA-28 mandibular

Participants received a directly 3D-printed TC-85 aligner in the maxillary arch and a directly 3D-printed TA-28 aligner in the mandibular arch. Both aligners were worn simultaneously for 2 weeks (22 hours/day).

Aligner directly 3D-printed from Tera Harz TC-85 shape-memory resin [add printer, layer thickness and post-curing protocol]. Worn 22 hours/day for 2 weeks in the maxillary or mandibular arch, according to a computer-generated randomization sequence.

实验性的:TA-28 maxillary / TC-85 mandibular

Participants received a directly 3D-printed TA-28 aligner in the maxillary arch and a directly 3D-printed TC-85 aligner in the mandibular arch. Both aligners were worn simultaneously for 2 weeks (22 hours/day).

Aligner directly 3D-printed from Tera Harz TA-28 shape-memory resin [same printer, layer thickness and post-curing protocol as above]. Worn 22 hours/day for 2 weeks in the maxillary or mandibular arch, according to a computer-generated randomization sequence.

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Change in indentation hardness (HIT), measured by instrumented nanoindentation
大体时间:Baseline (unused aligner) and after 2 weeks of intraoral use

Indentation hardness (HIT, GPa) of directly 3D-printed TC-85 and TA-28 aligners, measured by instrumented nanoindentation using a Nano Indenter XP (MTS Systems Corporation, USA) with a Berkovich diamond indenter. Tests were displacement-controlled to a maximum depth of 2000 nm at a constant strain rate of 0.05 s-¹, with 25 indentations per specimen spaced 30 μm apart. Testing followed ISO 14577-1:2015 and data were analyzed by the Oliver and Pharr method. Retrieved aligners are compared with unused aligners printed from the same digital design.

Baseline (unused aligner) and after 2 weeks of intraoral use

次要结果测量

结果测量
措施说明
大体时间
Change in indentation elastic modulus (EIT), measured by instrumented nanoindentation
大体时间:Baseline (unused aligner) and after 2 weeks of intraoral use

Indentation elastic modulus (EIT, GPa) of directly 3D-printed TC-85 and TA-28 aligners, measured by instrumented nanoindentation using a Nano Indenter XP (MTS Systems Corporation, USA) with a Berkovich diamond indenter. Tests were displacement-controlled to a maximum depth of 2000 nm at a constant strain rate of 0.05 s-¹, with 25 indentations per specimen spaced 30 μm apart. A Poisson's ratio of 0.4 was assumed. Testing followed ISO 14577-1:2015 and data were analyzed by the Oliver and Pharr method. Retrieved aligners are compared with unused aligners printed from the same digital design.

Baseline (unused aligner) and after 2 weeks of intraoral use
Change in color (ΔE₀₀, CIEDE2000), measured by UV-visible spectrophotometry
大体时间:Baseline (unused aligner) and after 2 weeks of intraoral use

Color of directly 3D-printed TC-85 and TA-28 aligners measured with a UV-visible spectrophotometer (V-770, JASCO Corporation, Tokyo, Japan). Each aligner was sectioned along the right central incisor-canine region, and the labial surface was positioned in a standardized manner facing the incident-light window. CIE Lab* coordinates were recorded according to ISO 11664-4:2019, and the color change between unused and retrieved aligners was calculated as ΔE₀₀ using the CIEDE2000 equation (parametric factors set to 1). Values are interpreted against 0.8 (50:50% perceptibility) and 1.8 (50:50% acceptability).

Baseline (unused aligner) and after 2 weeks of intraoral use
Change in light transmittance and absorbance (400-700 nm), measured by UV-visible spectrophotometry
大体时间:Baseline (unused aligner) and after 2 weeks of intraoral use

Light transmittance and absorbance of TC-85 and TA-28 aligners recorded across the visible range (400-700 nm) with a UV-visible spectrophotometer (V-770, JASCO Corporation, Tokyo, Japan) using Spectra Manager II software, after calibration according to the manufacturer's instructions. Values at 400 and 700 nm were extracted from each spectrum and compared between materials and between unused and retrieved aligners.

Baseline (unused aligner) and after 2 weeks of intraoral use
Change in aligner thickness (μm), measured by confocal laser scanning microscopy
大体时间:Baseline (unused aligner) and after 2 weeks of intraoral use

Thickness of TC-85 and TA-28 aligners measured by confocal laser scanning microscopy (Leica DMi8, Leica Microsystems, Wetzlar, Germany) with a 10× objective and 1.5× digital zoom under standardized imaging conditions. Standardized 4 × 4 mm specimens were taken from the labial surface of the left central incisor region. Z-stack images were reconstructed in 3D using Leica LAS AF software (version 1.1.0.12420), and thickness was calculated as the perpendicular distance between the upper and lower specimen surfaces. Each specimen was measured twice and the mean was used.

Baseline (unused aligner) and after 2 weeks of intraoral use
Change in surface roughness (Sa, Sq, Sp, Sv), measured by confocal laser scanning microscopy
大体时间:Baseline (unused aligner) and after 2 weeks of intraoral use

Surface roughness of the inner surface of directly 3D-printed TC-85 and TA-28 aligners, assessed on the same 4 × 4 mm specimens used for thickness measurement, imaged by confocal laser scanning microscopy (Leica DMi8, Leica Microsystems, Wetzlar, Germany). Height maps generated from the z-stacks with Leica LAS AF software were analyzed in ImageJ (National Institutes of Health, USA) with the SurfCharJ-1q plugin to calculate Sa (arithmetical mean height), Sq (root mean square height), Sp (maximum peak height) and Sv (maximum valley depth), in µm. Analysis was repeated twice per specimen and the mean was used. Identical acquisition and analysis settings were applied to all specimens.

Baseline (unused aligner) and after 2 weeks of intraoral use
Change in degree of conversion (DC), measured by ATR-FTIR spectroscopy
大体时间:Baseline (unused aligner) and after 2 weeks of intraoral use

Degree of conversion (%) of TC-85 and TA-28 aligners, calculated from attenuated total reflectance Fourier-transform infrared (ATR-FTIR) spectra as the reduction of the aliphatic C=C band near 1638 cm-¹ relative to, referenced to the corresponding unset resin. Retrieved aligners are compared with unused aligners.

Baseline (unused aligner) and after 2 weeks of intraoral use
Change in chemical structure, assessed by ATR-FTIR spectral analysis
大体时间:Baseline (unused aligner) and after 2 weeks of intraoral use

Qualitative comparison of ATR-FTIR spectra of unused and retrieved TC-85 and TA-28 aligners, focusing on the O-H/N-H stretching bands (3400-3200 cm-¹), aliphatic C-H bands (about 2920 and 2850 cm-¹), the carbonyl (C=O) band (about 1700 cm-¹) and the C-O/C-O-C region (1200-1000 cm-¹).

Baseline (unused aligner) and after 2 weeks of intraoral use

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

调查人员

  • 首席研究员:Azza Gamal Wafa, PHD、Alexandria University

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (实际的)

2026年2月1日

初级完成 (实际的)

2026年4月1日

研究完成 (实际的)

2026年9月1日

研究注册日期

首次提交

2026年10月31日

首先提交符合 QC 标准的

2026年11月5日

首次发布 (实际的)

2026年11月9日

研究记录更新

最后更新发布 (实际的)

2026年11月9日

上次提交的符合 QC 标准的更新

2026年11月5日

最后验证

2026年1月1日

更多信息

与本研究相关的术语

计划个人参与者数据 (IPD)

计划共享个人参与者数据 (IPD)?

未定

IPD 计划说明

Individual participant data will not be shared publicly because of participant confidentiality and privacy considerations. The study involves a relatively small sample, and sharing individual-level data may pose a risk of participant identification. De-identified aggregate results will be reported in the published manuscripts. Data may be made available from the corresponding author upon reasonable request, subject to institutional ethical approval and applicable data-protection requirements.