Operator-Assisted Oscilla Audiometry as a Basis for Hearing Aid Fitting (The Oscilla Validation Study)

2026年11月15日 更新者:Lene Dahl Siggaard、Aalborg University Hospital

Validation of Operator-Assisted Oscilla Audiometry as a Diagnostic Tool and Basis for Hearing Aid Fitting in Adults With Uncomplicated Hearing Loss: A Blinded Non-Inferiority Randomised Controlled Trial

The purpose of this study is to find out whether hearing tests performed with the Oscilla A60 audiometer, where the patient largely tests themselves with minimal staff assistance, are accurate enough to be used for diagnosing hearing loss and as the sole basis for fitting hearing aids. Adults referred for their first hearing aids at Aalborg University Hospital will have their hearing tested both with the Oscilla audiometer and by a trained audiology assistant at the same visit. Participants are then randomly assigned to receive hearing aids fitted based on either the Oscilla hearing test or the audiology assistant's hearing test. The main questions are: (1) Do the Oscilla results agree with the audiology assistant's results for both air-conduction and bone-conduction hearing thresholds? (2) Do participants fitted based on the Oscilla test report a hearing benefit that is not worse than that of participants fitted based on the standard test, two months after fitting?

研究概览

详细说明

Demand for audiometric testing in public hearing clinics is increasing due to population ageing. Operator-assisted automated audiometry may expand capacity, but its use as the sole basis for hearing aid prescription, and its accuracy for bone-conduction (BC) thresholds, has not been established. This single-site study at the Department of Audiology, Aalborg University Hospital (locations in Aalborg and Hobro) enrols 268 first-time hearing aid users with uncomplicated, symmetric sensorineural hearing loss.

At Visit 1, all participants undergo both Oscilla A60 audiometry (air conduction [AC] 250-8000 Hz; BC 250-4000 Hz) and standard manual audiometry by an audiology assistant (reference standard), in randomised order. Agreement between methods is assessed with Bland-Altman analysis against pre-specified acceptance criteria (95% CI of bias within ±5 dB; 95% limits of agreement within ±15 dB for AC and ±20 dB for BC). After confirmation of eligibility, participants are randomised 1:1, stratified by hearing loss severity (PTA-4 ≤45 vs >45 dB HL) and location, to hearing aid fitting based on the Oscilla audiogram or the audiology assistant audiogram. All participants are fitted bilaterally with BTE/RIC hearing aids to the NAL-NL2 target, verified with automated real-ear measurements (within ±5 dB). Volume control and app connectivity are disabled until the two-month follow-up. The primary patient-reported outcome is the change in SSQ12 total score from baseline to two months post-fitting, with a non-inferiority margin of -0.81 points. Secondary outcomes are IOI-HA, daily hearing aid use (data logging) and SSQ12 subscales. The study is a clinical investigation under Article 82 of Regulation (EU) 2017/745.

研究类型

介入性

注册 (估计的)

268

阶段

  • 不适用

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习联系方式

  • 姓名:Lene D Siggaard, MD, PhD
  • 电话号码:+45 20633492
  • 邮箱:lene.dahl@rn.dk

研究联系人备份

学习地点

      • Aalborg、丹麦、9000
        • Department of Audiology, Aalborg University Hospital
        • 接触:
        • 接触:
          • Lene D Siggaard, MD, PhD
          • 电话号码:+4520633492
          • 邮箱:lene.dahl@rn.dk
      • Hobro、丹麦、9500
        • Hearing Clinic, Aalborg University Hospital
        • 接触:
        • 接触:
          • Lene D Siggaard, MD, PhD
          • 电话号码:+4520633492
          • 邮箱:lene.dahl@rn.dk

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

18年 及以上 (成人、年长者)

接受健康志愿者

不

有资格学习的性别

全部

描述

Inclusion Criteria:

  • Adults (≥18 years) referred for first-time hearing aid fitting at the Department of Audiology, Aalborg University Hospital
  • Uncomplicated sensorineural hearing loss as defined by the Danish national clinical guidelines
  • Hearing thresholds exceeding 20 dB HL at two or more frequencies in the range 0.5-4 kHz
  • Pure-tone average (PTA-4; 0.5, 1, 2, 4 kHz) ≥30 dB HL in the better ear
  • Symmetric hearing loss (PTA-4 difference between ears ≤15 dB)
  • Word discrimination score ≥25% (DANTALE I at PTA + 40 dB HL)
  • Danish native speaker
  • No prior hearing aid experience
  • Able to complete questionnaires online or on paper
  • Written informed consent

Exclusion Criteria:

  • Air-conduction thresholds exceeding 80 dB HL at two or more frequencies
  • Air-bone gap >15 dB at one or more frequencies below 1 kHz
  • Fluctuating hearing loss (e.g., Ménière's disease) or ongoing ototoxic medication
  • Ear surgery (ear canal, middle ear or inner ear) within the past 12 months
  • Hearing loss classified as complicated according to national clinical guidelines
  • Visual or motor impairment precluding independent use of the Oscilla system
  • Candidates for in-the-ear hearing aids only (all participants are fitted with BTE or RIC hearing aids)

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 主要用途:诊断
  • 分配:随机化
  • 介入模型:并行分配
  • 屏蔽:单身的

武器和干预

参与者组/臂
干预/治疗
实验性的:Oscilla-based fitting

Hearing aids fitted bilaterally to NAL-NL2 targets based on the Oscilla A60 audiogram, verified with automated real-ear measurement. Volume control and app connectivity disabled for two months (linked interventions: Paired audiometry and Oscilla-based hearing aid fitting).

At Visit 1, all participants undergo both operator-assisted Oscilla A60 audiometry (AudioConsole v4.6.0; AC and BC thresholds; operator places the BC headband only) in a quiet room (<50 dB SPL) and manual pure-tone audiometry (AC and BC, ISO 8253-1) by a qualified audiology assistant in a sound-attenuating booth, in randomised order.

Bilateral BTE/RIC hearing aid fitting to NAL-NL2 targets using the Oscilla A60 audiogram as the sole audiometric basis, verified with auto-REM (within ±5 dB of target). Other names: Oscilla A60 Diagnostic Audiometer.

有源比较器:Manual audiometry-based fitting

Hearing aids fitted bilaterally to NAL-NL2 targets based on the audiogram obtained by a qualified audiology assistant, verified with automated real-ear measurement. Volume control and app connectivity disabled for two months (linked interventions: paired audiometry and manual audiometry-based hearing aid fitting).

At Visit 1, all participants undergo both operator-assisted Oscilla A60 audiometry (AudioConsole v4.6.0; AC and BC thresholds; operator places the BC headband only) in a quiet room (<50 dB SPL) and manual pure-tone audiometry (AC and BC, ISO 8253-1) by a qualified audiology assistant in a sound-attenuating booth, in randomised order.

Bilateral BTE/RIC hearing aid fitting to NAL-NL2 targets using the audiology assistant audiogram, verified with auto-REM (within ±5 dB of target). Standard clinical practice.

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Agreement between Oscilla A60 and manual audiometry for air-conduction thresholds
大体时间:Visit 1 (Week 0)

Bland-Altman mean difference (Oscilla minus manual) and 95% limits of agreement in dB, plus mean absolute difference, at 0.5, 1, 2 and 4 kHz separately. Agreement is acceptable if the 95% CI of the bias is within ±5 dB and the 95% limits of agreement are within ±15 dB. Both ears included, clustering handled with a mixed-effects model.

Visit 1 (Week 0)
Agreement between Oscilla A60 and manual audiometry for bone-conduction thresholds
大体时间:Visit 1 (Week 0)

As above for bone-conduction thresholds at 0.5, 1, 2 and 4 kHz. Agreement is acceptable if the 95% CI of the bias is within ±5 dB and the 95% limits of agreement are within ±20 dB.

Visit 1 (Week 0)
Change in SSQ12 total score
大体时间:Baseline (Week 0) and 2 months after hearing aid fitting

Speech, Spatial and Qualities of Hearing Scale, 12-item version; total score = mean of 12 items, range 0-10, higher = better. Non-inferiority if the lower bound of the two-sided 95% CI for the between-group difference in change (Oscilla minus control) is above -0.81 points.

Baseline (Week 0) and 2 months after hearing aid fitting

次要结果测量

结果测量
措施说明
大体时间
IOI-HA total score
大体时间:2 months after hearing aid fitting

International Outcome Inventory for Hearing Aids, Danish version; 7 items scored 1-5, total range 7-35, higher = better outcome. Compared between groups

2 months after hearing aid fitting
Daily hearing aid use
大体时间:From hearing aid fitting to 2 months after fitting

Mean daily hearing aid usage time in hours/day from the hearing aid data log, compared between groups

From hearing aid fitting to 2 months after fitting
Change in SSQ12 subscale scores
大体时间:Baseline (Week 0) and 2 months after hearing aid fitting

Change in the speech, spatial and qualities subscale scores (each range 0-10, higher = better)

Baseline (Week 0) and 2 months after hearing aid fitting

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

调查人员

  • 首席研究员:Lene D Siggaard, MD, PhD、Department of Otolaryngology, Head- and Neck Surgery and Audiology, Aalborg University Hospital

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (估计的)

2026年12月1日

初级完成 (估计的)

2027年10月31日

研究完成 (估计的)

2027年10月31日

研究注册日期

首次提交

2026年11月15日

首先提交符合 QC 标准的

2026年11月15日

首次发布 (实际的)

2026年11月19日

研究记录更新

最后更新发布 (实际的)

2026年11月19日

上次提交的符合 QC 标准的更新

2026年11月15日

最后验证

2026年11月1日

更多信息

与本研究相关的术语

其他研究编号

  • F2026-115
  • Regional Innovation Fund (其他赠款/资助编号:North Denmark Region)
  • Danish Medicines Agency (注册表标识符:16-0302-225)

计划个人参与者数据 (IPD)

计划共享个人参与者数据 (IPD)?

未定

IPD 计划说明

Pseudonymised individual participant data may be shared upon reasonable request after publication, subject to approval under the GDPR and Danish data protection law and a data processing agreement.

药物和器械信息、研究文件

研究美国 FDA 监管的药品

不

研究美国 FDA 监管的设备产品

不