Novel Non-Invasive Monitoring Parameter in a Hospital Setting
A Novel Cardio-Respiratory Parameter in a Hospital Setting
調査の概要
詳細な説明
研究の種類
入学 (実際)
連絡先と場所
研究場所
-
-
Colorado
-
Denver、Colorado、アメリカ、80262
- The University of Colorado Health Sciences Center (UCHSC)
-
-
Ohio
-
Columbus、Ohio、アメリカ、43210
- The Ohio State University Medical Center (OSUMC)
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
サンプリング方法
調査対象母集団
説明
Inclusion Criteria:
- Subjects 18 years old or older
- Written informed consent from the patient or their legally authorized representative before initiation of any study-related procedures
- Subjects on general care floor
Exclusion Criteria:
- Severe contact allergies to standard adhesive materials
- Abnormalities that may prevent proper application of the device
- Women who are pregnant or lactating
- Subjects with significant arrhythmia
研究計画
研究はどのように設計されていますか?
デザインの詳細
コホートと介入
グループ/コホート |
---|
In-patient Volunteers
In-patients from the hospital who choose to participate in the study
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
---|---|---|
The Covidien Nellcor Respiration Rate Software Shall Determine Respiration Rate Measured as Mean and Standard Deviation With Accuracy That is Non-inferior to Predicate Device.
時間枠:Participants were monitored for average of 30 minutes
|
Mean and standard deviations of respiration rates collected from patients in the hospital settings were compared between Covidien Respiration Rate Software, Transthoracic Impedance and End-Tidal Carbon Dioxide Waveforms.
Each patient served as its own control.
|
Participants were monitored for average of 30 minutes
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
---|---|---|
The Covidien Nellcor Respiration Rate Software Shall Calculate Respiration Rate With a Root Mean Square Difference (RMSD) of < 3 Breaths Per Minute Compared With a End-Tidal Carbon Dioxide Waveforms, With 95% Confidence.
時間枠:Participants were monitored on average for 30 minutes
|
The data used for analysis consisted of multiple simultaneous measures of RR_TTI, RR_V1.0 and RR_EtCO2 for each subject. Given N sets of simultaneous RR values for each of P patients, the simultaneous RR values were used to calculate a pair of RMSD values for each subject. The two RMSD values, RMSDRR_V1.0_vs_EtCO2, and RMSDTTI_vs_EtCO2, quantify the mean absolute difference in simultaneous estimates of respiratory rate between RR_V1.0 compared with RR_EtCO2 and RR_TTI compared with RR_EtCO2, respectively for the subjects with 95% confidence of mean ± 3.38 BrPM. The Measure type is Number and represents the RMSD of Covidien Nellcor Respiration Rate Software |
Participants were monitored on average for 30 minutes
|
協力者と研究者
スポンサー
捜査官
- 主任研究者:Sergio Bergese, MD、The Ohio State Medical Center, Dept of Anesthesia
- 主任研究者:Robert McIntyre, MD、University of Colorado Denver, School of Medicine dept of Surgery Division of GI, Tumor and Endocrine Surgery