このページは自動翻訳されたものであり、翻訳の正確性は保証されていません。を参照してください。 英語版 ソーステキスト用。

Low-Level EMF Stimulation for Paroxysmal Atrial Fibrillation

2019年12月18日 更新者:University of Oklahoma

Effect of Low-Level Electromagnetic Field Stimulation on Patients With Paroxysmal Atrial Fibrillation

This research has two main specific aims. First, a commercially-available stimulator will be utilized to assess the effect of low-level electromagnetic field (EMF) stimulation on atrial fibrillation (AF) inducibility in patients with paroxysmal AF presenting for catheter ablation as compared to sham stimulation. We hypothesize a reduction in AF inducibility in patients treated with EMF stimulation. Second, we aim to assess for the effect of low-level EMF on level of systemic inflammatory mediators. We hypothesize a reduction in the level of the inflammatory mediator tissue necrosis factor (TNF)-α in patients treated with EMF stimulation as compared to sham stimulation. The long-term objective of this research is to develop low-level EMF as a therapeutic alternative for patients with AF, and this research will help to determine the efficacy of low-level pulsed EMF in this regard.

調査の概要

詳細な説明

This is a prospective, randomized, sham-controlled trial. Patients will be enrolled upon presentation to our electrophysiology (EP) lab for ablation of paroxysmal AF. After informed consent is obtained, the patient will be randomized to one of two protocols that have been pre-programmed into the magnetic field stimulator. As described below, one of the protocols will be an active stimulation protocol, and one will be sham (i.e. no stimulation delivered). These will be named "Stim 1" and "Stim 2" in the computer. Because the specifics of the protocol are programmed into the stimulator by the device manufacturer, investigators will not know which of these two protocols is active vs. sham. The patient will then be brought to the EP lab for the procedure and will be placed on the lab table with their head positioned in the magnetic coil. Vascular access will then be obtained and diagnostic catheters placed as is standard practice for AF ablation.

Because the study is limited to patients with paroxysmal AF, patients will begin the study in normal sinus rhythm (NSR). Before the experimental protocol is begun, baseline electrophysiologic intervals will be recorded as is standard practice for all patients presenting for EP study. This will include measurement of the atrio-ventricular nodal effective refractory period (AVNERP) using programmed stimulation from the right atrial (RA) appendage, atrio-his (AH) interval, and his-ventricular (HV) interval per standard practice. Burst atrial pacing will then be utilized to induce AF. Measurements of the number of burst pacing attempts required to induce AF, as well as the AF cycle length, and duration of pacing-induced AF will be recorded. Attempts at AF induction will occur for 15 minutes. Prior to induction of AF, 5ml of venous blood will be drawn from the central venous sheaths to measure baseline levels of the inflammatory mediator (TNF)-α. These samples will be stored for analysis.

After baseline measurements are obtained, the coil will be turned on and whichever protocol the patient has been assigned to will be run. The system to be used contains a signal-generator coupled to a Helmholtz coil capable of producing a magnetic field in response to an electric current. The Helmholtz coil itself is has 9 inches of separation in between the two coils that allows for placement of the coil around the participant's head and upper neck. The choice of the head and neck as the site of stimulation is intended to target the vagus nerve as it exits the jugular foramen to course caudally in the carotid sheath. The system is designed to create a homogenous, isotropic magnetic field with a field strength from 1 to 99 pico-Tesla (pT) with a frequency range of 0.01 Hz to 50 Hz.

Active Stimulation will involve application of a pulsed-EMF with the parameters 0.032 micro-gauss (µG) at 0.89 Hz for 60 minutes. In participants randomized to sham stimulation, the coil will be positioned around the neck as previously described, but no current will be applied to the stimulator.

During the 60 minute stimulation time, trans-septal puncture and mapping of the left atrium will be performed as is standard practice for this procedure, but no ablation will be performed. After the 60 minute protocol is complete, venous blood samples will again be drawn (within 5 minutes of protocol completion), stored, and analyzed as previously described. If the patient remains in AF after the end of the 60 minute session, they will be electrically cardioverted to sinus rhythm. The same stimulation protocol as described previously will again be employed, again noting number of attempts required to induce AF, duration of pacing-induced AF, AF cycle length, and measurements of AVNERP, AH, HV intervals and levels of TNF-α.

研究の種類

介入

入学 (実際)

18

段階

  • 適用できない

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究場所

    • Oklahoma
      • Oklahoma City、Oklahoma、アメリカ、73120
        • OU Health Sciences Center

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

21年~85年 (大人、高齢者)

健康ボランティアの受け入れ

いいえ

受講資格のある性別

全て

説明

Inclusion Criteria:

  • Adult patients aged 21-85 presenting for ablation of paroxysmal atrial fibrillation (episodes of atrial fibrillation lasting < 7 days with or without prior cardioversion)

Exclusion Criteria:

  • Left ventricular ejection fraction (LVEF) < 40%
  • Stroke or myocardial infarction within the past 6 months
  • Greater than moderate valvular stenosis or regurgitation as assessed by pre-procedure transthoracic echocardiogram (TTE)
  • Presence of a prosthetic heart valve.

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 主な目的:処理
  • 割り当て:ランダム化
  • 介入モデル:並列代入
  • マスキング:4倍

武器と介入

参加者グループ / アーム
介入・治療
実験的:Active Stimulation
Participants will receive active EMF stimulation through the study device for 60 minutes during the procedure.
Low-level pulsed electromagnetic field stimulator delivered through a Helmholtz coil positioned around the participant's head and neck. Stimulation parameters are 0.032 µG at 0.89 Hz for 60 minutes.
偽コンパレータ:Sham Stimulation
The participant will be positioned in the stimulator, but no EMF stimulation will be delivered.
The stimulator will be placed around the participant's head and neck, but no EMF will be delivered.

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Change in duration of pacing-induced atrial fibrillation
時間枠:60 minutes
Change in the duration of atrial fibrillation that is induced by burst-pacing compared between the two groups before and after the stimulation period.
60 minutes

二次結果の測定

結果測定
メジャーの説明
時間枠
Number of pacing attempts required to induce AF
時間枠:60 minutes
Number of burst-pacing attempts that are required to successfully induced atrial fibrillation, compared between the two groups before and after the stimulation period.
60 minutes
AF Cycle length
時間枠:60 minutes
Cycle length of induced atrial fibrillation, compared between the two groups before and after the stimulation period.
60 minutes
TNF-alpha levels
時間枠:60 minutes
Levels of the inflammatory mediator TNF-alpha measured in central venous blood, compared between the two groups before and after the stimulation period.
60 minutes

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

捜査官

  • 主任研究者:Sunny Po, MD, PhD、University of Oklahoma

出版物と役立つリンク

研究に関する情報を入力する責任者は、自発的にこれらの出版物を提供します。これらは、研究に関連するあらゆるものに関するものである可能性があります。

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (実際)

2018年9月27日

一次修了 (実際)

2019年11月26日

研究の完了 (実際)

2019年11月26日

試験登録日

最初に提出

2018年6月24日

QC基準を満たした最初の提出物

2018年7月9日

最初の投稿 (実際)

2018年7月20日

学習記録の更新

投稿された最後の更新 (実際)

2019年12月20日

QC基準を満たした最後の更新が送信されました

2019年12月18日

最終確認日

2019年12月1日

詳しくは

本研究に関する用語

その他の研究ID番号

  • EMF for Atrial Fibrillation

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

いいえ

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

いいえ

米国で製造され、米国から輸出された製品。

いいえ