3D Printer Modified CPAP Mask for Obstructive Sleep Apnea (3DCPAP)
3-D Printer Custom Modified Continuous Positive Airway Pressure Masks to Improve Efficacy and Patient Compliance in Obstructive Sleep Apnea Management
調査の概要
詳細な説明
Purpose:
The purpose of this protocol is to perform a prospective pilot clinical study to customize patient CPAP masks to reduce air leak in patients with obstructive sleep apnea (OSA). Using 3-D printer-based technology modeling of patient facial features, patient FDA-approved CPAP masks will be modified by strategically applying silicone pads on the CPAP mask to mold to the specific facial features to optimize the CPAP mask-patient interface, improve CPAP mask fit and comfort, reduce air leak, limit air pressure loss and improve compliance with CPAP treatment strategy (as measured over a 30 day trial period).
Protocol:
Potential subjects for this study will be identified at the outpatient Sleep Disorders Clinic at BIDMC. Patients will have documented OSA, prescribed CPAP as standard-of-care, and demonstrate consistent problems with CPAP mask-patient interface with documented high air loss, and poor adherence/ compliance. All potential subjects will have used and will continue to use an FDA-approved CPAP mask selected by the patient.
Potential subjects (along with their FDA-approved CPAP mask) will then be referred to the BIDMC Specialties Clinic and screened, including review of the medical history of OSA, review of documented reports of excessive air leak and suboptimal adherence to CPAP, brief physical exam (esp. focusing on the head-and-neck exam) including assessment of the CPAP mask-patient interface, review of the purpose and goals of the pilot clinical study, a dynamic video of the study subject face will be performed, video downloaded to a computer which will then be used to generate a 3-dimensional plastic model of the study subject face using state-of-the-art 3D printer technology (see details below). The patient's FDA-approved CPAP mask will then be applied to his plastic model of the study subject face, and the CPAP mask modified with the application of silicon pads along the mask pillow cushion to optimize the CPAP mask-face model interface, primarily by adjustments to improve CPAP mask contact with the face model and limit the air leak during CPAP ventilation.
Once the patient's FDA-approved CPAP mask has been modified, the patient will then return to the BIDMC Specialties Clinic for a final fitting of the modified CPAP mask, with additional adjustments (as necessary) made to optimize the CPAP mask-patient interface. The study subject will then return home with his/her custom-modified CPAP mask to use as prescribed. Standard home monitoring and recording of use of the modified CPAP mask will be performed per usual case over a 30-day period. Data obtained during this 30-day period will focus on the duration of use of the modified CPAP mask, measurement of air leak and pressure loss, report of patient comfort and experience with the modified CPAP mask, and compare to data over a 30 day period using the CPAP mask prior to modification.
研究の種類
入学 (予想される)
段階
- 適用できない
連絡先と場所
研究場所
-
-
Massachusetts
-
Boston、Massachusetts、アメリカ、02215
- Beth Israel Deaconess Medical Center
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Diagnosis of obstructive sleep apnea prescribed CPAP
- High air leak:1. Absolute high leak duration more than 15% of the use period and / or 2. Absolute high leak over 5% associated with periodic breathing.
Exclusion Criteria:
- Unable to wear full face CPAP
- Unable to provide informed consent
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:処理
- 割り当て:なし
- 介入モデル:単一グループの割り当て
- マスキング:なし(オープンラベル)
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
実験的:Modified CPAP
Single arm cohort study of modified mask
|
Modified CPAP mask based on 3D image capture
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Air leak time
時間枠:30 day
|
Percent of air leak time over use of mask
|
30 day
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
CPAP mask leak volume
時間枠:30 day
|
Volume of leak recorded on CPAP machine
|
30 day
|
|
Epworth Sleepiness Scale
時間枠:30 day
|
Sleepiness score - Likelihood of falling asleep in various daytime activities.
Total Score Range 0-24.
|
30 day
|
|
Apnea Hypopnea Index
時間枠:30 day
|
Apnea Hypopnea Index (events/hour)
|
30 day
|
|
CPAP Inspiratory pressure
時間枠:30 day
|
Inspiratory pressure setting on CPAP machine (cm H2O)
|
30 day
|
協力者と研究者
捜査官
- 主任研究者:Chee ChunMin, MD、Beth Israel Deaconess Medical Center
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (予想される)
研究の完了 (予想される)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。