Evaluation of Calcified Coronary Lesion Preparation With the Naviscore Scoring Balloon (MILOU)

2026年6月10日 更新者:Stephane Carlier、University of Mons

MILOU: Minimize Intra Luminal Obstructive Underexpansion: A Prospective, Multi-Center Randomized Trial for the Evaluation of Calcified Coronary Lesion Preparation With the Naviscore Scoring Balloon

The purpose of this clinical study is to evaluate the effectiveness and safety of a specialized medical device, the Naviscore scoring balloon, in preparing calcified coronary artery narrowings before the implantation of a drug-eluting stent. During percutaneous coronary interventions, the presence of calcified plaques in the heart arteries represents a major challenge because it can prevent stents from expanding fully. When a stent remains under-expanded, it significantly increases the long-term risk of arterial re-narrowing or blood clot formation. To optimize stent expansion, appropriate preparation of the diseased vessel section before stent insertion is a critical phase.

This study is a prospective, multi-center randomized trial designed to test the hypothesis that treating calcified coronary lesions with the Naviscore scoring balloon will achieve a better stent expansion and a larger final minimal stent area compared to standard lesion preparation using regular non-compliant balloons. Eligible participants will be randomized in a one-to-one ratio to one of these two lesion preparation strategies. For all included patients, standard drug-eluting stents will be deployed. The study will use intravascular ultrasound imaging to evaluate the final minimum area of the stent directly inside the treated artery at the site of the highest initial calcium burden. Participant health and clinical outcomes will be monitored for up to twelve months following the procedure.

調査の概要

詳細な説明

Percutaneous transluminal coronary angioplasty (PTCA) is the main technique in percutaneous coronary intervention (PCI). It aims to restore coronary blood flow by mechanically enlarging a stenotic arterial segment through controlled balloon inflation within the vessel lumen. The acute luminal gain with balloon angioplasty is the result of multiple mechanisms, ranging from percolation and longitudinal redistribution of atheroma, to plaque fracture, and overstretching of the vessel wall, often accompanied by intimal tearing and medial dissection. While contemporary drug-eluting technologies have significantly improved long-term clinical outcomes, fibrocalcified atheromatous intracoronary plaques still present a significant challenge for stent crossing and optimal deployment.

Research has demonstrated that suboptimal stent expansion and malapposition are significant contributing factors to restenosis and stent thrombosis rates, regardless of the type of stent used. Intravascular imaging registries have identified that twenty to thirty percent of deployed stents remain under-expanded or malapposed in daily clinical practice. Severe coronary calcification can make it difficult to deliver devices and can limit the expansion of stents, which can result in suboptimal improvement in blood flow and an increased risk of adverse events, including target lesion failure. Furthermore, heavily calcified lesions have been found to be associated with higher rates of vessel dissection, perforation and impaired anti-proliferative drug delivery, resulting from the mechanical barrier of the calcium burden.

In order to achieve successful stent deployment, optimal preparation of fibro-calcific lesions is essential and involves a number of plaque modification strategies. Focused force angioplasty is a procedure that targets specific locations of the plaque, focusing inflation forces on the area to induce localized stress and facilitate cracking. It is crucial to ensure that the pre-dilatation balloon is correctly sized to maximise safety by minimising the risk of advanced coronary perforation or severe medial damage. It is imperative that sizing compliance chart data is followed rigorously, as compliant balloons can display an unpredictable non-linear expansion behaviour at higher pressures, whereas non-compliant balloons maintain a stable volume, concentrating their dilating force directly at the calcified site up to their rated burst pressure. Semi-compliant pre-dilatation balloons often expand asymmetrically in resistant lesions, increasing the risk of edge dissections. The sizing and performance of scoring devices and non-compliant balloons remains limited in contemporary clinical literature. This randomised trial is designed to evaluate the post-market clinical performance, safety and comparative effectiveness of the Naviscore scoring balloon catheter manufactured by iVascular.

研究の種類

介入

入学 (推定)

200

段階

  • 適用できない

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

研究場所

      • Barcelona、スペイン、08025
        • Hospital de Sant Pau
        • コンタクト:
        • コンタクト:
          • Elena Jiménez Xarrié
          • 電話番号:+34628839525
        • 主任研究者:
          • Estefanía Fernández-Peregrina, MD
      • Barcelona、スペイン、08901-08908
      • Cadiz、スペイン、11009
        • Hospital Puerta del Mar
        • コンタクト:
          • Alejandro Gutiérrez-Barrios, MD
          • 電話番号:+34 620688877
          • メール:aleklos@hotmail.com
        • 主任研究者:
          • Alejandro Gutiérrez-Barrios, MD
      • Cáceres、スペイン、10003
        • Hospital San Pedro De Alcantara
        • コンタクト:
        • 主任研究者:
          • Juan Carlos Rama, MD
      • Valencia、スペイン、46026
        • Hospital La Fe
        • コンタクト:
          • Jorge Sanz Sanchez
          • 電話番号:+34 660360598
          • メール:sjorge4@gmx.com
        • コンタクト:
        • 主任研究者:
          • Jorge Sanz Sanchez, MD
      • Valladolid、スペイン、47003
        • Hospital Clínico Universitario de Valladolid
        • 主任研究者:
          • Ignacio J Amat Santos, MD
        • コンタクト:
          • Ignacio J Amat Santos, MD
          • 電話番号:86571 +34983420000
          • メール:ijamat@gmail.com
        • コンタクト:
      • Bonheiden、ベルギー、2820
        • Imelda
      • Brussels、ベルギー、1070
        • Hôpital Erasme - Cliniques universitaires de Bruxelles
        • コンタクト:
        • 主任研究者:
          • Gabriela Flores Vivian, MD
      • Libramont、ベルギー、6800
        • Hôpital Vivalia
        • コンタクト:
        • 主任研究者:
          • Chadi Ghafari, MD, PhD
      • Liège、ベルギー、4000
        • CHR de la Citadelle
        • コンタクト:
        • コンタクト:
        • 主任研究者:
          • Georges Saad, MD
      • Lodelinsart、ベルギー、6042
        • Humani CHU Charleroi - Chimay
        • コンタクト:
        • 主任研究者:
          • Mohamed Ben Yedder, MD
    • Hainaut
      • Mons、Hainaut、ベルギー、7000
        • CHU - HELORA site Kennedy
        • コンタクト:
        • 主任研究者:
          • Stéphane Carlier, MD, Phd
      • Vila Nova de Gaia、ポルトガル、4430-999
        • Centro Hospitalar de Vila Nova de Gaia/Espinho
        • コンタクト:
        • 主任研究者:
          • Alberto Rodrigues, MD
      • Luxembourg、ルクセンブルク、1210
        • INCCI Haerz Zenter
        • コンタクト:
          • Bruno Pereira, MD
          • 電話番号:+ 352 691 86 31 86
          • メール:dr.pereira@pt.lu
        • 主任研究者:
          • Bruno Pereira, MD

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

  • 大人
  • 高齢者

健康ボランティアの受け入れ

いいえ

説明

Inclusion Criteria:

  • Patient written consent is given, ≥18 years.
  • A de novo lesion to be treated in a vessel between 2.5 and 4.0 mm.
  • Moderate to heavily calcified lesions detected by coronary angiography on two orthogonal views, and confirmed by IVUS if the lesion can be crossed with the catheter.

Exclusion Criteria:

  • Patient <18 years old.
  • Pregnant female.
  • Contraindication to dual antiplatelet therapy.
  • Thrombocytopenia (under 100 000).
  • Major surgical intervention planned within one year.
  • Significant left main lesion.
  • Chronic total occlusion.
  • Lesion in a graft.
  • In-stent restenosis lesion.
  • Lesion responsible of a ST elevated Myocardial Infarction (STEMI)

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 主な目的:処理
  • 割り当て:ランダム化
  • 介入モデル:並列代入
  • マスキング:なし(オープンラベル)

武器と介入

参加者グループ / アーム
介入・治療
アクティブコンパレータ:Standard NC group
Lesion preparation of calcified coronary stenosis performed using standard non-compliant (NC) balloons. Balloons are sized according to a 0.8:1 to 1:1 ratio relative to the reference lumen and inflated below the rated burst pressure. To maintain a strict comparison and eliminate any scoring effect, no parallel guidewire is allowed in the target vessel during inflation.
Conventional non-compliant (NC) PTCA balloon catheter used for standard lesion pre-dilatation according to routine clinical practice, inflated below the rated burst pressure without parallel guidewires in place.
他の名前:
  • Conventional PTCA Balloon
実験的:Naviscore group
Lesion preparation of calcified coronary stenosis performed using the Naviscore scoring balloon catheter before drug-eluting stent implantation. The device is sized according to a 0.8:1 to 1:1 ratio relative to the reference lumen. To maximize the plaque-modification effect of the external nitinol wires, a mandatory mechanical mobilization sequence must be performed: the balloon is inflated to nominal pressure, fully deflated, mobilized within the lesion, and reinflated. This specific sequence must be repeated 3 to 4 times across the target segment.
Percutaneous coronary preparation of calcified coronary stenosis using the rapid exchange Naviscore scoring balloon catheter before drug-eluting stent deployment. The device incorporates an over-the-balloon metallic nitinol element designed to focally concentrate dilation forces, minimizing balloon slippage and scoring the atheromatous plaque at lower pressures to facilitate optimal stent expansion. Sizing is based on a 0.8:1 to 1:1 ratio relative to the reference lumen. Operators must execute a specific sequence of inflating the scoring balloon to nominal pressure, fully deflating it, mobilizing it within the lesion, and reinflating it, repeated three to four times across the target segment to ensure multi-focal plaque disruption before stenting.
他の名前:
  • Naviscore Scoring Balloon

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Final Minimal Stent Area (MSA)
時間枠:Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure.
The minimal stent area (MSA) will be measured in square millimeters by intravascular ultrasound (IVUS) after final procedural optimization. The primary comparison will evaluate the MSA specifically at the exact site of the initial heaviest calcium burden to demonstrate the superiority of the Naviscore scoring balloon over standard non-compliant (NC) balloons.
Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure.

二次結果の測定

結果測定
メジャーの説明
時間枠
Minimal Stent Area (MSA) Across the Entire Stented Segment
時間枠:Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
The minimal stent area (MSA) in square millimeters measured across the entire length of the deployed segment using intravascular ultrasound (IVUS) pullbacks to evaluate overall stent expansion expansion
Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Stent Eccentricity at the Site of the Minimal Stent Area
時間枠:Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Stent eccentricity index calculated as the ratio of the minimum stent diameter over the maximum stent diameter, measured via intravascular ultrasound (IVUS) at the exact cross-section of the minimal stent area segment
Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Stent Eccentricity at the Initial Heaviest Calcium Burden Site
時間枠:Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Stent eccentricity index calculated as the ratio of the minimum stent diameter over the maximum stent diameter, measured via intravascular ultrasound (IVUS) specifically at the cross-section matching the initial heaviest baseline calcium plaque burden
Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Device Success Rate
時間枠:During the index percutaneous coronary intervention (PCI) procedure
Percentage of cases achieving successful delivery of the assigned pre-dilatation balloon, complete inflation with adequate lesion preparation, and the absolute absence of vessel rupture or advanced coronary perforation categorized by the Ellis classification
During the index percutaneous coronary intervention (PCI) procedure
Major Adverse Cardiac Events (MACE) Rate
時間枠:At 1 and 12 months post-procedure
Percentage of participants experiencing Major Adverse Cardiac Events (MACE), defined as a composite clinical endpoint of myocardial infarction, stroke or transient ischemic attack, or cardiac death
At 1 and 12 months post-procedure
Target Lesion Failure (TLF)
時間枠:At 1 and 12 months post-procedure
Percentage of participants experiencing Target Lesion Failure (TLF), defined as a composite of cardiac death, target-vessel myocardial infarction defined according to ARC-2 criteria, or clinically indicated Target Lesion Revascularization
At 1 and 12 months post-procedure
Target Vessel Failure (TVF)
時間枠:At 1 and 12 months post-procedure
Defined as a composite of cardiac death, target-vessel myocardial infarction, or clinically indicated Target Vessel Revascularization (TVR).
At 1 and 12 months post-procedure
Clinically Indicated Target Lesion Revascularization (TLR)
時間枠:At 1 and 12 months post-procedure
Percentage of participants requiring a repeat percutaneous coronary intervention (PCI) or coronary artery bypass graft surgery (CABG) driven by a recurrent narrowing inside the originally treated target lesion segment (including the stent and its five-millimeter proximal and distal borders), which is accompanied by clinical symptoms of ischemia or objective functional evidence of myocardial ischemia.
At 1 and 12 months post-procedure
Clinically Indicated Target Vessel Revascularization (TVR)
時間枠:At 1 and 12 months post-procedure
Percentage of participants requiring any repeat percutaneous coronary intervention (PCI) or coronary artery bypass graft surgery (CABG) in any segment of the originally treated target vessel, driven by angiographic restenosis or disease progression, and accompanied by clinical symptoms of ischemia or objective functional evidence of myocardial ischemia.
At 1 and 12 months post-procedure
Procedural Success Rate
時間枠:During the index percutaneous coronary intervention (PCI) procedure and until hospital discharge (up to 24 hours post-procedure).
Percentage of cases achieving Device Success (successful delivery of the assigned balloon and complete inflation with adequate lesion preparation) combined with a final residual stenosis of less than thirty percent as measured by quantitative coronary angiography (QCA) and a final Thrombolysis in Myocardial Infarction (TIMI) grade 3 flow, without the occurrence of in-hospital Major Adverse Cardiac Events (MACE).
During the index percutaneous coronary intervention (PCI) procedure and until hospital discharge (up to 24 hours post-procedure).
Final Residual Stent Underexpansion
時間枠:At the end of the index percutaneous coronary intervention (PCI) procedure (immediate post-stenting optimization).
Evaluation of the final residual underexpansion of the drug-eluting stent after deployment and high-pressure post-dilatation, measured in millimeters using motion-corrected X-ray stent visualization software (StentBoost or equivalent stent enhancement method) in two orthogonal views without contrast.
At the end of the index percutaneous coronary intervention (PCI) procedure (immediate post-stenting optimization).
Final Stent Diameter Stenosis
時間枠:At the end of the index percutaneous coronary intervention (PCI) procedure.
The final percentage of stent diameter stenosis evaluated by quantitative coronary angiography (QCA) at the index procedure after final stent deployment and post-dilatation optimization.
At the end of the index percutaneous coronary intervention (PCI) procedure.
Rate of Procedural Complications
時間枠:During the index percutaneous coronary intervention (PCI) procedure (immediate intra-procedural period).
The cumulative rate of procedure-related complications during the index procedure, defined as a composite of cardiac death, acute myocardial infarction, emergency coronary artery bypass graft surgery (CABG), coronary artery perforation (Ellis classification), and cardiac tamponade directly related to the use of the Naviscore scoring balloon or the regular non-compliant (NC) balloon.
During the index percutaneous coronary intervention (PCI) procedure (immediate intra-procedural period).
Procedural Cost-Effectiveness
時間枠:During the index percutaneous coronary intervention (PCI) procedure.
Evaluation of the procedural cost-effectiveness of using the Naviscore scoring balloon compared to standard non-compliant (NC) balloons for calcified lesion preparation, calculated by factoring device costs, overall material utilisation during the intervention, procedure duration, and the need for bailout additional plaque-modification tools or complementary high-pressure devices.
During the index percutaneous coronary intervention (PCI) procedure.

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

協力者

捜査官

  • 主任研究者:Stéphane Carlier, MD, PhD、University of Mons
  • 主任研究者:Antonio Serra

出版物と役立つリンク

研究に関する情報を入力する責任者は、自発的にこれらの出版物を提供します。これらは、研究に関連するあらゆるものに関するものである可能性があります。

一般刊行物

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (推定)

2026年7月1日

一次修了 (推定)

2027年6月1日

研究の完了 (推定)

2028年6月1日

試験登録日

最初に提出

2026年6月10日

QC基準を満たした最初の提出物

2026年6月10日

最初の投稿 (実際)

2026年6月16日

学習記録の更新

投稿された最後の更新 (実際)

2026年6月16日

QC基準を満たした最後の更新が送信されました

2026年6月10日

最終確認日

2026年6月1日

詳しくは

本研究に関する用語

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

はい

IPD プランの説明

The individual participant data (IPD) that underlie the results reported in this publication, including the final minimal stent area (MSA) measurements, procedural success data, and relevant clinical adverse event records, will be shared. Supporting documents, including the study protocol, statistical analysis plan, and informed consent forms, will be provided to support the interpretation of the IPD. All data shared will be de-identified to maintain participant confidentiality in accordance with GDPR requirements.

IPD 共有時間枠

Beginning 6 months after the publication of the study results and ending 5 years following the publication date.

IPD 共有アクセス基準

Access is granted to researchers who provide a methodologically sound proposal for secondary analysis. Proposals must be submitted to the Principal Investigators at the University of Mons (UMONS), Belgium. The study's Steering Committee will review each request. Upon approval, data will be shared under a signed data sharing agreement ensuring strict adherence to participant privacy and institutional ethical guidelines.

IPD 共有サポート情報タイプ

  • STUDY_PROTOCOL
  • SAP
  • ICF

医薬品およびデバイス情報、研究文書