Evaluation of Calcified Coronary Lesion Preparation With the Naviscore Scoring Balloon (MILOU)

keskiviikko 10. kesäkuuta 2026 päivittänyt: Stephane Carlier, University of Mons

MILOU: Minimize Intra Luminal Obstructive Underexpansion: A Prospective, Multi-Center Randomized Trial for the Evaluation of Calcified Coronary Lesion Preparation With the Naviscore Scoring Balloon

The purpose of this clinical study is to evaluate the effectiveness and safety of a specialized medical device, the Naviscore scoring balloon, in preparing calcified coronary artery narrowings before the implantation of a drug-eluting stent. During percutaneous coronary interventions, the presence of calcified plaques in the heart arteries represents a major challenge because it can prevent stents from expanding fully. When a stent remains under-expanded, it significantly increases the long-term risk of arterial re-narrowing or blood clot formation. To optimize stent expansion, appropriate preparation of the diseased vessel section before stent insertion is a critical phase.

This study is a prospective, multi-center randomized trial designed to test the hypothesis that treating calcified coronary lesions with the Naviscore scoring balloon will achieve a better stent expansion and a larger final minimal stent area compared to standard lesion preparation using regular non-compliant balloons. Eligible participants will be randomized in a one-to-one ratio to one of these two lesion preparation strategies. For all included patients, standard drug-eluting stents will be deployed. The study will use intravascular ultrasound imaging to evaluate the final minimum area of the stent directly inside the treated artery at the site of the highest initial calcium burden. Participant health and clinical outcomes will be monitored for up to twelve months following the procedure.

Tutkimuksen yleiskatsaus

Yksityiskohtainen kuvaus

Percutaneous transluminal coronary angioplasty (PTCA) is the main technique in percutaneous coronary intervention (PCI). It aims to restore coronary blood flow by mechanically enlarging a stenotic arterial segment through controlled balloon inflation within the vessel lumen. The acute luminal gain with balloon angioplasty is the result of multiple mechanisms, ranging from percolation and longitudinal redistribution of atheroma, to plaque fracture, and overstretching of the vessel wall, often accompanied by intimal tearing and medial dissection. While contemporary drug-eluting technologies have significantly improved long-term clinical outcomes, fibrocalcified atheromatous intracoronary plaques still present a significant challenge for stent crossing and optimal deployment.

Research has demonstrated that suboptimal stent expansion and malapposition are significant contributing factors to restenosis and stent thrombosis rates, regardless of the type of stent used. Intravascular imaging registries have identified that twenty to thirty percent of deployed stents remain under-expanded or malapposed in daily clinical practice. Severe coronary calcification can make it difficult to deliver devices and can limit the expansion of stents, which can result in suboptimal improvement in blood flow and an increased risk of adverse events, including target lesion failure. Furthermore, heavily calcified lesions have been found to be associated with higher rates of vessel dissection, perforation and impaired anti-proliferative drug delivery, resulting from the mechanical barrier of the calcium burden.

In order to achieve successful stent deployment, optimal preparation of fibro-calcific lesions is essential and involves a number of plaque modification strategies. Focused force angioplasty is a procedure that targets specific locations of the plaque, focusing inflation forces on the area to induce localized stress and facilitate cracking. It is crucial to ensure that the pre-dilatation balloon is correctly sized to maximise safety by minimising the risk of advanced coronary perforation or severe medial damage. It is imperative that sizing compliance chart data is followed rigorously, as compliant balloons can display an unpredictable non-linear expansion behaviour at higher pressures, whereas non-compliant balloons maintain a stable volume, concentrating their dilating force directly at the calcified site up to their rated burst pressure. Semi-compliant pre-dilatation balloons often expand asymmetrically in resistant lesions, increasing the risk of edge dissections. The sizing and performance of scoring devices and non-compliant balloons remains limited in contemporary clinical literature. This randomised trial is designed to evaluate the post-market clinical performance, safety and comparative effectiveness of the Naviscore scoring balloon catheter manufactured by iVascular.

Opintotyyppi

Interventio

Ilmoittautuminen (Arvioitu)

200

Vaihe

  • Ei sovellettavissa

Yhteystiedot ja paikat

Tässä osiossa on tutkimuksen suorittajien yhteystiedot ja tiedot siitä, missä tämä tutkimus suoritetaan.

Opiskeluyhteys

Opiskelupaikat

      • Bonheiden, Belgia, 2820
        • Imelda
      • Brussels, Belgia, 1070
        • Hôpital Erasme - Cliniques universitaires de Bruxelles
        • Ottaa yhteyttä:
        • Päätutkija:
          • Gabriela Flores Vivian, MD
      • Libramont, Belgia, 6800
        • Hôpital Vivalia
        • Ottaa yhteyttä:
        • Päätutkija:
          • Chadi Ghafari, MD, PhD
      • Liège, Belgia, 4000
        • CHR de la Citadelle
        • Ottaa yhteyttä:
        • Ottaa yhteyttä:
        • Päätutkija:
          • Georges Saad, MD
      • Lodelinsart, Belgia, 6042
        • Humani CHU Charleroi - Chimay
        • Ottaa yhteyttä:
        • Päätutkija:
          • Mohamed Ben Yedder, MD
    • Hainaut
      • Mons, Hainaut, Belgia, 7000
        • CHU - HELORA site Kennedy
        • Ottaa yhteyttä:
        • Päätutkija:
          • Stéphane Carlier, MD, Phd
      • Barcelona, Espanja, 08025
        • Hospital de Sant Pau
        • Ottaa yhteyttä:
        • Ottaa yhteyttä:
          • Elena Jiménez Xarrié
          • Puhelinnumero: +34628839525
        • Päätutkija:
          • Estefanía Fernández-Peregrina, MD
      • Barcelona, Espanja, 08901-08908
      • Cadiz, Espanja, 11009
        • Hospital Puerta del Mar
        • Ottaa yhteyttä:
          • Alejandro Gutiérrez-Barrios, MD
          • Puhelinnumero: +34 620688877
          • Sähköposti: aleklos@hotmail.com
        • Päätutkija:
          • Alejandro Gutiérrez-Barrios, MD
      • Cáceres, Espanja, 10003
        • Hospital San Pedro De Alcantara
        • Ottaa yhteyttä:
        • Päätutkija:
          • Juan Carlos Rama, MD
      • Valencia, Espanja, 46026
        • Hospital La Fe
        • Ottaa yhteyttä:
          • Jorge Sanz Sanchez
          • Puhelinnumero: +34 660360598
          • Sähköposti: sjorge4@gmx.com
        • Ottaa yhteyttä:
        • Päätutkija:
          • Jorge Sanz Sanchez, MD
      • Valladolid, Espanja, 47003
        • Hospital Clínico Universitario de Valladolid
        • Päätutkija:
          • Ignacio J Amat Santos, MD
        • Ottaa yhteyttä:
          • Ignacio J Amat Santos, MD
          • Puhelinnumero: 86571 +34983420000
          • Sähköposti: ijamat@gmail.com
        • Ottaa yhteyttä:
          • Carlos Baladrón Zorita
          • Puhelinnumero: 86571 +34983 42 00 00
          • Sähköposti: cbaladron@icicor.es
      • Luxembourg, Luxemburg, 1210
        • INCCI Haerz Zenter
        • Ottaa yhteyttä:
          • Bruno Pereira, MD
          • Puhelinnumero: + 352 691 86 31 86
          • Sähköposti: dr.pereira@pt.lu
        • Päätutkija:
          • Bruno Pereira, MD
      • Vila Nova de Gaia, Portugali, 4430-999
        • Centro Hospitalar de Vila Nova de Gaia/Espinho
        • Ottaa yhteyttä:
        • Päätutkija:
          • Alberto Rodrigues, MD

Osallistumiskriteerit

Tutkijat etsivät ihmisiä, jotka sopivat tiettyyn kuvaukseen, jota kutsutaan kelpoisuuskriteereiksi. Joitakin esimerkkejä näistä kriteereistä ovat henkilön yleinen terveydentila tai aiemmat hoidot.

Kelpoisuusvaatimukset

Opintokelpoiset iät

  • Aikuinen
  • Vanhempi Aikuinen

Hyväksyy terveitä vapaaehtoisia

Ei

Kuvaus

Inclusion Criteria:

  • Patient written consent is given, ≥18 years.
  • A de novo lesion to be treated in a vessel between 2.5 and 4.0 mm.
  • Moderate to heavily calcified lesions detected by coronary angiography on two orthogonal views, and confirmed by IVUS if the lesion can be crossed with the catheter.

Exclusion Criteria:

  • Patient <18 years old.
  • Pregnant female.
  • Contraindication to dual antiplatelet therapy.
  • Thrombocytopenia (under 100 000).
  • Major surgical intervention planned within one year.
  • Significant left main lesion.
  • Chronic total occlusion.
  • Lesion in a graft.
  • In-stent restenosis lesion.
  • Lesion responsible of a ST elevated Myocardial Infarction (STEMI)

Opintosuunnitelma

Tässä osiossa on tietoja tutkimussuunnitelmasta, mukaan lukien kuinka tutkimus on suunniteltu ja mitä tutkimuksella mitataan.

Miten tutkimus on suunniteltu?

Suunnittelun yksityiskohdat

  • Ensisijainen käyttötarkoitus: Hoito
  • Jako: Satunnaistettu
  • Inventiomalli: Rinnakkaistehtävä
  • Naamiointi: Ei mitään (avoin tarra)

Aseet ja interventiot

Osallistujaryhmä / Arm
Interventio / Hoito
Active Comparator: Standard NC group
Lesion preparation of calcified coronary stenosis performed using standard non-compliant (NC) balloons. Balloons are sized according to a 0.8:1 to 1:1 ratio relative to the reference lumen and inflated below the rated burst pressure. To maintain a strict comparison and eliminate any scoring effect, no parallel guidewire is allowed in the target vessel during inflation.
Conventional non-compliant (NC) PTCA balloon catheter used for standard lesion pre-dilatation according to routine clinical practice, inflated below the rated burst pressure without parallel guidewires in place.
Muut nimet:
  • Conventional PTCA Balloon
Kokeellinen: Naviscore group
Lesion preparation of calcified coronary stenosis performed using the Naviscore scoring balloon catheter before drug-eluting stent implantation. The device is sized according to a 0.8:1 to 1:1 ratio relative to the reference lumen. To maximize the plaque-modification effect of the external nitinol wires, a mandatory mechanical mobilization sequence must be performed: the balloon is inflated to nominal pressure, fully deflated, mobilized within the lesion, and reinflated. This specific sequence must be repeated 3 to 4 times across the target segment.
Percutaneous coronary preparation of calcified coronary stenosis using the rapid exchange Naviscore scoring balloon catheter before drug-eluting stent deployment. The device incorporates an over-the-balloon metallic nitinol element designed to focally concentrate dilation forces, minimizing balloon slippage and scoring the atheromatous plaque at lower pressures to facilitate optimal stent expansion. Sizing is based on a 0.8:1 to 1:1 ratio relative to the reference lumen. Operators must execute a specific sequence of inflating the scoring balloon to nominal pressure, fully deflating it, mobilizing it within the lesion, and reinflating it, repeated three to four times across the target segment to ensure multi-focal plaque disruption before stenting.
Muut nimet:
  • Naviscore Scoring Balloon

Mitä tutkimuksessa mitataan?

Ensisijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Final Minimal Stent Area (MSA)
Aikaikkuna: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure.
The minimal stent area (MSA) will be measured in square millimeters by intravascular ultrasound (IVUS) after final procedural optimization. The primary comparison will evaluate the MSA specifically at the exact site of the initial heaviest calcium burden to demonstrate the superiority of the Naviscore scoring balloon over standard non-compliant (NC) balloons.
Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure.

Toissijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Minimal Stent Area (MSA) Across the Entire Stented Segment
Aikaikkuna: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
The minimal stent area (MSA) in square millimeters measured across the entire length of the deployed segment using intravascular ultrasound (IVUS) pullbacks to evaluate overall stent expansion expansion
Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Stent Eccentricity at the Site of the Minimal Stent Area
Aikaikkuna: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Stent eccentricity index calculated as the ratio of the minimum stent diameter over the maximum stent diameter, measured via intravascular ultrasound (IVUS) at the exact cross-section of the minimal stent area segment
Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Stent Eccentricity at the Initial Heaviest Calcium Burden Site
Aikaikkuna: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Stent eccentricity index calculated as the ratio of the minimum stent diameter over the maximum stent diameter, measured via intravascular ultrasound (IVUS) specifically at the cross-section matching the initial heaviest baseline calcium plaque burden
Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure
Device Success Rate
Aikaikkuna: During the index percutaneous coronary intervention (PCI) procedure
Percentage of cases achieving successful delivery of the assigned pre-dilatation balloon, complete inflation with adequate lesion preparation, and the absolute absence of vessel rupture or advanced coronary perforation categorized by the Ellis classification
During the index percutaneous coronary intervention (PCI) procedure
Major Adverse Cardiac Events (MACE) Rate
Aikaikkuna: At 1 and 12 months post-procedure
Percentage of participants experiencing Major Adverse Cardiac Events (MACE), defined as a composite clinical endpoint of myocardial infarction, stroke or transient ischemic attack, or cardiac death
At 1 and 12 months post-procedure
Target Lesion Failure (TLF)
Aikaikkuna: At 1 and 12 months post-procedure
Percentage of participants experiencing Target Lesion Failure (TLF), defined as a composite of cardiac death, target-vessel myocardial infarction defined according to ARC-2 criteria, or clinically indicated Target Lesion Revascularization
At 1 and 12 months post-procedure
Target Vessel Failure (TVF)
Aikaikkuna: At 1 and 12 months post-procedure
Defined as a composite of cardiac death, target-vessel myocardial infarction, or clinically indicated Target Vessel Revascularization (TVR).
At 1 and 12 months post-procedure
Clinically Indicated Target Lesion Revascularization (TLR)
Aikaikkuna: At 1 and 12 months post-procedure
Percentage of participants requiring a repeat percutaneous coronary intervention (PCI) or coronary artery bypass graft surgery (CABG) driven by a recurrent narrowing inside the originally treated target lesion segment (including the stent and its five-millimeter proximal and distal borders), which is accompanied by clinical symptoms of ischemia or objective functional evidence of myocardial ischemia.
At 1 and 12 months post-procedure
Clinically Indicated Target Vessel Revascularization (TVR)
Aikaikkuna: At 1 and 12 months post-procedure
Percentage of participants requiring any repeat percutaneous coronary intervention (PCI) or coronary artery bypass graft surgery (CABG) in any segment of the originally treated target vessel, driven by angiographic restenosis or disease progression, and accompanied by clinical symptoms of ischemia or objective functional evidence of myocardial ischemia.
At 1 and 12 months post-procedure
Procedural Success Rate
Aikaikkuna: During the index percutaneous coronary intervention (PCI) procedure and until hospital discharge (up to 24 hours post-procedure).
Percentage of cases achieving Device Success (successful delivery of the assigned balloon and complete inflation with adequate lesion preparation) combined with a final residual stenosis of less than thirty percent as measured by quantitative coronary angiography (QCA) and a final Thrombolysis in Myocardial Infarction (TIMI) grade 3 flow, without the occurrence of in-hospital Major Adverse Cardiac Events (MACE).
During the index percutaneous coronary intervention (PCI) procedure and until hospital discharge (up to 24 hours post-procedure).
Final Residual Stent Underexpansion
Aikaikkuna: At the end of the index percutaneous coronary intervention (PCI) procedure (immediate post-stenting optimization).
Evaluation of the final residual underexpansion of the drug-eluting stent after deployment and high-pressure post-dilatation, measured in millimeters using motion-corrected X-ray stent visualization software (StentBoost or equivalent stent enhancement method) in two orthogonal views without contrast.
At the end of the index percutaneous coronary intervention (PCI) procedure (immediate post-stenting optimization).
Final Stent Diameter Stenosis
Aikaikkuna: At the end of the index percutaneous coronary intervention (PCI) procedure.
The final percentage of stent diameter stenosis evaluated by quantitative coronary angiography (QCA) at the index procedure after final stent deployment and post-dilatation optimization.
At the end of the index percutaneous coronary intervention (PCI) procedure.
Rate of Procedural Complications
Aikaikkuna: During the index percutaneous coronary intervention (PCI) procedure (immediate intra-procedural period).
The cumulative rate of procedure-related complications during the index procedure, defined as a composite of cardiac death, acute myocardial infarction, emergency coronary artery bypass graft surgery (CABG), coronary artery perforation (Ellis classification), and cardiac tamponade directly related to the use of the Naviscore scoring balloon or the regular non-compliant (NC) balloon.
During the index percutaneous coronary intervention (PCI) procedure (immediate intra-procedural period).
Procedural Cost-Effectiveness
Aikaikkuna: During the index percutaneous coronary intervention (PCI) procedure.
Evaluation of the procedural cost-effectiveness of using the Naviscore scoring balloon compared to standard non-compliant (NC) balloons for calcified lesion preparation, calculated by factoring device costs, overall material utilisation during the intervention, procedure duration, and the need for bailout additional plaque-modification tools or complementary high-pressure devices.
During the index percutaneous coronary intervention (PCI) procedure.

Yhteistyökumppanit ja tutkijat

Täältä löydät tähän tutkimukseen osallistuvat ihmiset ja organisaatiot.

Yhteistyökumppanit

Tutkijat

  • Päätutkija: Stéphane Carlier, MD, PhD, University of Mons
  • Päätutkija: Antonio Serra

Julkaisuja ja hyödyllisiä linkkejä

Tutkimusta koskevien tietojen syöttämisestä vastaava henkilö toimittaa nämä julkaisut vapaaehtoisesti. Nämä voivat koskea mitä tahansa tutkimukseen liittyvää.

Yleiset julkaisut

Opintojen ennätyspäivät

Nämä päivämäärät seuraavat ClinicalTrials.gov-sivustolle lähetettyjen tutkimustietueiden ja yhteenvetojen edistymistä. National Library of Medicine (NLM) tarkistaa tutkimustiedot ja raportoidut tulokset varmistaakseen, että ne täyttävät tietyt laadunvalvontastandardit, ennen kuin ne julkaistaan ​​julkisella verkkosivustolla.

Opi tärkeimmät päivämäärät

Opiskelun aloitus (Arvioitu)

Keskiviikko 1. heinäkuuta 2026

Ensisijainen valmistuminen (Arvioitu)

Tiistai 1. kesäkuuta 2027

Opintojen valmistuminen (Arvioitu)

Torstai 1. kesäkuuta 2028

Opintoihin ilmoittautumispäivät

Ensimmäinen lähetetty

Keskiviikko 10. kesäkuuta 2026

Ensimmäinen toimitettu, joka täytti QC-kriteerit

Keskiviikko 10. kesäkuuta 2026

Ensimmäinen Lähetetty (Todellinen)

Tiistai 16. kesäkuuta 2026

Tutkimustietojen päivitykset

Viimeisin päivitys julkaistu (Todellinen)

Tiistai 16. kesäkuuta 2026

Viimeisin lähetetty päivitys, joka täytti QC-kriteerit

Keskiviikko 10. kesäkuuta 2026

Viimeksi vahvistettu

Maanantai 1. kesäkuuta 2026

Lisää tietoa

Tähän tutkimukseen liittyvät termit

Yksittäisten osallistujien tietojen suunnitelma (IPD)

Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?

JOO

IPD-suunnitelman kuvaus

The individual participant data (IPD) that underlie the results reported in this publication, including the final minimal stent area (MSA) measurements, procedural success data, and relevant clinical adverse event records, will be shared. Supporting documents, including the study protocol, statistical analysis plan, and informed consent forms, will be provided to support the interpretation of the IPD. All data shared will be de-identified to maintain participant confidentiality in accordance with GDPR requirements.

IPD-jaon aikakehys

Beginning 6 months after the publication of the study results and ending 5 years following the publication date.

IPD-jaon käyttöoikeuskriteerit

Access is granted to researchers who provide a methodologically sound proposal for secondary analysis. Proposals must be submitted to the Principal Investigators at the University of Mons (UMONS), Belgium. The study's Steering Committee will review each request. Upon approval, data will be shared under a signed data sharing agreement ensuring strict adherence to participant privacy and institutional ethical guidelines.

IPD-jakamista tukeva tietotyyppi

  • STUDY_PROTOCOL
  • MAHLA
  • ICF

Lääke- ja laitetiedot, tutkimusasiakirjat

Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta

Ei

Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta

Ei

Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .