このページは自動翻訳されたものであり、翻訳の正確性は保証されていません。を参照してください。 英語版 ソーステキスト用。

Comparative Study of the Tomey FT-1000, and Goldmann Applanation Tonometer (GAT)

2026年9月3日 更新者:Tomey Corporation

This study is intended to evaluate the agreement between Intraocular pressure(IOP) measurements obtained with the FT-1000 non-contact tonometer and a GAT to confirm the feasibility of the measurement procedures.

This is a prospective comparative study that will be conducted at one clinical site in the United States. Subjects who sign an informed consent form and fulfill all inclusion and exclusion criteria will undergo FT-1000 measurements followed by Goldmann applanation tonometry during a single study visit.

The primary analysis will be performed using one preselected study eye from each subject. Measurements from the fellow eye, when available, may be summarized as exploratory analyses but will not contribute to the primary effectiveness analysis.

In general, descriptive statistics (n, mean, standard deviation, median, minimum and maximum) will be used to summarize continuous variables. Frequencies and percentages will be used to summarize categorical variables. The number of accepted scans and the total number of scans evaluated will be summarized by scan type, device, and subject population.

Agreement between the FT-1000 and the Goldmann applanation tonometer will be evaluated using the study eye measurements by calculating the mean difference, standard deviation (SD), 95% limits of agreement (LOA), Bland-Altman plots, Deming regression, and Pearson correlation coefficient.

Exploratory analyses may also be performed using fellow-eye measurements.

調査の概要

状態

まだ募集していません

条件

研究の種類

観察的

入学 (推定)

30

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

  • 大人
  • 高齢者

健康ボランティアの受け入れ

はい

サンプリング方法

非確率サンプル

調査対象母集団

Low Intraocular pressure(IOP) (22 years old or older) Middle Intraocular pressure(IOP)(22 years old or older) High Intraocular pressure(IOP)(22 years old or older)

説明

Inclusion Criteria:

  1. Male or female subjects 22 years of age or older who have full legal capacity to volunteer on the date the informed consent is signed.
  2. Subjects who are able to understand and follow the study procedures and instructions provided by the clinical staff.
  3. Subjects who voluntarily provide written informed consent.
  4. Subjects who satisfy all study eligibility requirements and are able to undergo both FT-1000 and Goldmann applanation tonometry measurements during the study visit.

Exclusion Criteria:

  1. Subjects with only one functional eye.
  2. Subjects with poor or eccentric fixation in either eye.
  3. Subjects with a history of corneal surgery, including corneal laser or refractive surgery (e.g., LASIK).
  4. Subjects with corneal scarring, keratoconus, or any other clinically significant corneal pathology.
  5. Subjects with an active corneal or conjunctival infection or inflammation in either eye.
  6. Subjects with microphthalmos or buphthalmos.
  7. Subjects who are unable or unwilling to remove contact lenses before study measurements.
  8. Subjects with clinically significant dry eye disease that, in the investigator's opinion, may affect tonometry measurements.
  9. Subjects with blepharospasm (lid squeezing) or nystagmus that may interfere with accurate measurements.
  10. Subjects with abnormal central corneal thickness (CCT).
  11. Subjects with any condition or circumstance that, in the opinion of the investigator, could place the subject at significant risk, interfere with study participation, or confound interpretation of the study results.

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

コホートと介入

グループ/コホート
介入・治療
Low Intraocular pressure(IOP) Group: 7-16 mmHg
Group with IOP of 7-16 mmHg; 10 subjects.
Intraocular pressure measurement by non-contact tonometer and applanation tonometer.
Middle Intraocular pressure(IOP) Group: >16 to <23 mmHg
Group with IOP of >16 to <23 mmHg; 10 subjects.
Intraocular pressure measurement by non-contact tonometer and applanation tonometer.
High Intraocular pressure(IOP)Group: ≥23 mmHg
Group with IOP of ≥23 mmHg; 10 subjects.
Intraocular pressure measurement by non-contact tonometer and applanation tonometer.

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Safety of the test devices
時間枠:Through study completion, approximately 3 days
Any adverse events observed or reported during the study visit will be assessed and documented.
Through study completion, approximately 3 days
Agreement of the Intraocular pressure (IOP) mmHg between the test devices and predicate device
時間枠:Through study completion, approximately 3 days

Agreement of the Intraocular pressure (IOP) mmHg between the FT-1000 and the Goldmann Applanation Tonometer(GAT) will be evaluated using the study eye measurements by calculating the mean difference, standard deviation (SD), 95% limits of agreement (LOA), Bland-Altman plots, Deming regression, and Pearson correlation coefficient.

Exploratory analyses may also be performed using fellow-eye measurements.

Through study completion, approximately 3 days

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (推定)

2026年9月24日

一次修了 (推定)

2026年9月28日

研究の完了 (推定)

2026年9月28日

試験登録日

最初に提出

2026年8月31日

QC基準を満たした最初の提出物

2026年8月31日

最初の投稿 (実際)

2026年9月4日

学習記録の更新

投稿された最後の更新 (実際)

2026年9月9日

QC基準を満たした最後の更新が送信されました

2026年9月3日

最終確認日

2026年8月1日

詳しくは

本研究に関する用語

追加の関連 MeSH 用語

その他の研究ID番号

  • TomeyFT-1000-001

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

はい

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する