Comparative Study of the Tomey FT-1000, and Goldmann Applanation Tonometer (GAT)
This study is intended to evaluate the agreement between Intraocular pressure(IOP) measurements obtained with the FT-1000 non-contact tonometer and a GAT to confirm the feasibility of the measurement procedures.
This is a prospective comparative study that will be conducted at one clinical site in the United States. Subjects who sign an informed consent form and fulfill all inclusion and exclusion criteria will undergo FT-1000 measurements followed by Goldmann applanation tonometry during a single study visit.
The primary analysis will be performed using one preselected study eye from each subject. Measurements from the fellow eye, when available, may be summarized as exploratory analyses but will not contribute to the primary effectiveness analysis.
In general, descriptive statistics (n, mean, standard deviation, median, minimum and maximum) will be used to summarize continuous variables. Frequencies and percentages will be used to summarize categorical variables. The number of accepted scans and the total number of scans evaluated will be summarized by scan type, device, and subject population.
Agreement between the FT-1000 and the Goldmann applanation tonometer will be evaluated using the study eye measurements by calculating the mean difference, standard deviation (SD), 95% limits of agreement (LOA), Bland-Altman plots, Deming regression, and Pearson correlation coefficient.
Exploratory analyses may also be performed using fellow-eye measurements.
연구 개요
연구 유형
등록 (추정된)
참여기준
자격 기준
공부할 수 있는 나이
- 성인
- 고령자
건강한 자원 봉사자를 받아들입니다
샘플링 방법
연구 인구
설명
Inclusion Criteria:
- Male or female subjects 22 years of age or older who have full legal capacity to volunteer on the date the informed consent is signed.
- Subjects who are able to understand and follow the study procedures and instructions provided by the clinical staff.
- Subjects who voluntarily provide written informed consent.
- Subjects who satisfy all study eligibility requirements and are able to undergo both FT-1000 and Goldmann applanation tonometry measurements during the study visit.
Exclusion Criteria:
- Subjects with only one functional eye.
- Subjects with poor or eccentric fixation in either eye.
- Subjects with a history of corneal surgery, including corneal laser or refractive surgery (e.g., LASIK).
- Subjects with corneal scarring, keratoconus, or any other clinically significant corneal pathology.
- Subjects with an active corneal or conjunctival infection or inflammation in either eye.
- Subjects with microphthalmos or buphthalmos.
- Subjects who are unable or unwilling to remove contact lenses before study measurements.
- Subjects with clinically significant dry eye disease that, in the investigator's opinion, may affect tonometry measurements.
- Subjects with blepharospasm (lid squeezing) or nystagmus that may interfere with accurate measurements.
- Subjects with abnormal central corneal thickness (CCT).
- Subjects with any condition or circumstance that, in the opinion of the investigator, could place the subject at significant risk, interfere with study participation, or confound interpretation of the study results.
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
코호트 및 개입
그룹/코호트 |
개입 / 치료 |
|---|---|
|
Low Intraocular pressure(IOP) Group: 7-16 mmHg
Group with IOP of 7-16 mmHg; 10 subjects.
|
Intraocular pressure measurement by non-contact tonometer and applanation tonometer.
|
|
Middle Intraocular pressure(IOP) Group: >16 to <23 mmHg
Group with IOP of >16 to <23 mmHg; 10 subjects.
|
Intraocular pressure measurement by non-contact tonometer and applanation tonometer.
|
|
High Intraocular pressure(IOP)Group: ≥23 mmHg
Group with IOP of ≥23 mmHg; 10 subjects.
|
Intraocular pressure measurement by non-contact tonometer and applanation tonometer.
|
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Safety of the test devices
기간: Through study completion, approximately 3 days
|
Any adverse events observed or reported during the study visit will be assessed and documented.
|
Through study completion, approximately 3 days
|
|
Agreement of the Intraocular pressure (IOP) mmHg between the test devices and predicate device
기간: Through study completion, approximately 3 days
|
Agreement of the Intraocular pressure (IOP) mmHg between the FT-1000 and the Goldmann Applanation Tonometer(GAT) will be evaluated using the study eye measurements by calculating the mean difference, standard deviation (SD), 95% limits of agreement (LOA), Bland-Altman plots, Deming regression, and Pearson correlation coefficient. Exploratory analyses may also be performed using fellow-eye measurements. |
Through study completion, approximately 3 days
|
공동 작업자 및 조사자
연구 기록 날짜
연구 주요 날짜
연구 시작 (추정된)
기본 완료 (추정된)
연구 완료 (추정된)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .