Clinical Evaluation of NIRSense Tissue Oximetry Sensor Monitoring to Early Detect Cardiogenic Shock

2026年9月13日 更新者:NIRSense Inc.

Clinical Evaluation of Using NIRSense Sensor Patch to Measure Microvascular Perfusion Responses and Predict Decompensation in Cardiogenic Shock Patients

This is a prospective, single-center, observational pilot study of the investigational NIRSense Tissue Oximetry Sensor System (Envello Core), a noninvasive wearable sensor patch that uses near-infrared spectroscopy (NIRS) and photoplethysmography (PPG), combined with an electrocardiography/impedance cardiography (ECG/ICG) sensor, to continuously monitor tissue oxygen saturation (StO2), peripheral oxygen saturation (SpO2), pulse rate, respiratory rate, stroke volume, and related metrics. Up to 20 adults admitted to the Cleveland Clinic Florida cardiac intensive care unit with acute myocardial infarction or acute decompensated heart failure and in Society for Cardiovascular Angiography and Interventions (SCAI) shock stage B, C, or D will wear the sensors on the sternum, forearm, and other locations for approximately 24 to 72 hours during standard care. Sensor data will be compared with standard-of-care hemodynamic and laboratory measurements and with the patient's clinical course to evaluate the feasibility of the device, the correlation of its measurements with established clinical parameters, and its potential to detect deterioration in SCAI shock stage earlier. No treatment decisions are made based on the study device.

調査の概要

詳細な説明

Background: Cardiogenic shock (CS) carries a poor prognosis and is the leading cause of death in patients with acute myocardial infarction. When decompensation occurs there is an imperative to institute treatment rapidly. Standard intensive care unit monitoring relies on intermittent hemodynamic and laboratory measures. Continuous, noninvasive measurement of tissue perfusion and oxygenation could allow earlier recognition of deterioration and earlier intervention.

Device: The NIRSense Tissue Oximetry Sensor System is an investigational, non-significant-risk, noninvasive wearable optical physiological monitoring system. It consists of a sensor patch with multiple red and near-infrared light sources and detectors (NIRS and PPG), a pre-applied medical-grade adhesive, an ECG/ICG sensor patch, and a display unit (tablet running the NIRSense application) that receives data over Bluetooth and stores it. The sensor estimates StO2 in microvascular tissue up to about 2.5 cm below the skin and derives PPG-based metrics including pulse rate, respiratory rate, and perfusion index. The device has not been cleared or approved by the U.S. FDA.

Design: Prospective data collection, single-center observational study. Two populations are targeted: (1) patients with acute myocardial infarction, with or without shock (SCAI B to D), admitted for coronary angiography and subsequent interventions, and (2) patients with acute decompensated heart failure admitted for pulmonary artery catheter insertion and subsequent interventions. Participants are enrolled from the existing Cleveland Clinic Florida patient population in a way that avoids disruption of clinical workflow. After informed consent, sensors are placed on the sternum, forearm, and other locations according to manufacturer specifications and remain in place while the patient is in the cardiac intensive care unit, for approximately 24 to 72 hours. Sensors may be repositioned during catheterization or imaging at the investigator's discretion. Participation is limited to the single standard-of-care admission; there is no study follow-up visit or call.

Data collected: Continuous sensor data (StO2, SpO2, pulse rate, respiratory rate, ECG/ICG-derived stroke volume, PPG-derived metrics) and sensor placement images. Data collected during standard-of-care treatment and shared with the sponsor: demographics, Fitzpatrick skin type, medical history, concomitant medications, physical examination, vital signs, a single 12-lead ECG, blood pressure, heart rate, cardiac output (thermodilution) where obtained, clinically ordered lactate values, SCAI stage, and clinical course. Adverse events and device deficiencies are recorded from enrollment through sensor removal. Clinical staff complete a usability survey after using the device on 5 patients.

Analysis: Descriptive statistics for demographics and baseline characteristics; adverse events summarized by frequency and percentage, severity, and device relatedness. Effectiveness analyses use all evaluable data and compare device measurements with standard-of-care hemodynamic parameters and with SCAI stage trajectory, as specified in the Statistical Analysis Plan. Following an interim analysis of the 20-patient pilot, enrollment may be expanded to 200 patients under a protocol amendment; that expansion is not part of this record.

研究の種類

観察的

入学 (推定)

20

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

研究場所

    • Florida
      • Weston、Florida、アメリカ、33331
        • 募集
        • Cleveland Clinic Florida, Weston Hospital
        • コンタクト:
          • David A. Baran, MD
          • 電話番号:954-659-5290
          • メール:barand@ccf.org
        • コンタクト:
        • 主任研究者:
          • David A. Baran, MD

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

  • 大人
  • 高齢者

健康ボランティアの受け入れ

いいえ

サンプリング方法

非確率サンプル

調査対象母集団

Adults admitted to the Cleveland Clinic Florida cardiac intensive care unit with acute myocardial infarction undergoing coronary angiography, or with acutely decompensated heart failure, who are in SCAI shock stage B, C, or D. Participants are recruited from the existing patient population of the recruitment site.

説明

Inclusion Criteria:

  • Acutely decompensated symptomatic heart failure (HF) documented by: (a) symptoms of HF requiring current treatment; (b) Society for Cardiovascular Angiography and Interventions (SCAI) stage B, C, or D; and (c) hospital admission with HF as the primary or secondary diagnosis; OR acute myocardial infarction (MI) patient undergoing coronary angiography
  • Admitted to the intensive care unit and currently in SCAI stage B, C, or D
  • Age 18 years or older
  • Weight greater than 40 kg

Exclusion Criteria:

  • Prior known severe allergy to medical-grade adhesive or tape (self-reported)
  • Currently participating in an investigational drug or device study that would interfere with the conduct or results of this study. Trials requiring extended follow-up for products that were investigational but have since become commercially available are not considered investigational.
  • The investigator believes the participant is not an appropriate candidate for the study

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

コホートと介入

グループ/コホート
介入・治療
Acute myocardial infarction
Patients with acute myocardial infarction, with or without shock (SCAI stage B to D), admitted for coronary angiography and subsequent interventions. Enrolled at the time of consent for coronary angiography.
Investigational noninvasive wearable sensor patches using near-infrared spectroscopy and photoplethysmography to continuously measure tissue oxygen saturation, SpO2, pulse rate, respiratory rate, and perfusion index, plus an ECG/ICG sensor patch for stroke volume, with a tablet display and recording application. Sensors are applied to the sternum, forearm, and other locations with a medical-grade adhesive and worn for approximately 24 to 72 hours during standard cardiac ICU care. Sensor data are collected for research only and are not used for treatment decisions.
他の名前:
  • Envello Core
  • NIRSense Sensor Oximetry System patch
Acute decompensated heart failure
Patients with acute decompensated heart failure (SCAI stage B to D) admitted for pulmonary artery catheter insertion and subsequent interventions. Consented before or after catheterization.
Investigational noninvasive wearable sensor patches using near-infrared spectroscopy and photoplethysmography to continuously measure tissue oxygen saturation, SpO2, pulse rate, respiratory rate, and perfusion index, plus an ECG/ICG sensor patch for stroke volume, with a tablet display and recording application. Sensors are applied to the sternum, forearm, and other locations with a medical-grade adhesive and worn for approximately 24 to 72 hours during standard cardiac ICU care. Sensor data are collected for research only and are not used for treatment decisions.
他の名前:
  • Envello Core
  • NIRSense Sensor Oximetry System patch

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Correlation between NIRSense sensor measurements and standard-of-care hemodynamic parameters
時間枠:Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Correlation coefficients (Pearson and Spearman) between time-matched NIRSense measurements (tissue oxygen saturation, perfusion index, SpO2, pulse rate, respiratory rate, ICG-derived stroke volume) and standard-of-care clinical measurements (arterial blood pressure, heart rate, pulmonary artery and central venous pressures where a pulmonary artery catheter is present, cardiac output by thermodilution, and lactate).
Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application

二次結果の測定

結果測定
メジャーの説明
時間枠
Feasibility: proportion of intended monitoring time with valid sensor data
時間枠:Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Percentage of the intended monitoring period during which each sensor provided valid StO2 and PPG data, per sensor site, including data recovered from onboard memory after wireless transmission gaps.
Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Patient tolerance: device-related adverse events and device deficiencies
時間枠:From enrollment through sensor removal, approximately 24 to 72 hours
Number and percentage of participants with adverse events, adverse device effects (including skin irritation or redness at the adhesive site), and device deficiencies, and number of participants requesting early sensor removal.
From enrollment through sensor removal, approximately 24 to 72 hours
Association between NIRSense measurement trends and change in SCAI shock stage
時間枠:Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Change in NIRSense measurements (StO2, StO2 variability, perfusion index, stroke volume) relative to the time of clinically documented change in SCAI shock stage, including the lead time between the first sustained change in device metrics and the clinically recognized deterioration.
Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Distribution of NIRSense measurements by SCAI shock stage
時間枠:Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Summary statistics of StO2, perfusion index, and ICG-derived stroke volume at each sensor site by SCAI shock stage (B, C, D, E), to inform development of a predictive model of shock status.
Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application

その他の成果指標

結果測定
メジャーの説明
時間枠
Clinical staff usability feedback
時間枠:After the device has been used on 5 patients (approximately 1 to 6 months after first enrollment)
Responses to a paper survey completed by study staff (principal investigator, nurses, residents, and others involved in the study) on their experience using the sensor system.
After the device has been used on 5 patients (approximately 1 to 6 months after first enrollment)

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

捜査官

  • 主任研究者:David A. Baran, MD、Cleveland Clinic Florida
  • スタディディレクター:Thibault Roumengous, PhD、NIRSense Inc.

出版物と役立つリンク

研究に関する情報を入力する責任者は、自発的にこれらの出版物を提供します。これらは、研究に関連するあらゆるものに関するものである可能性があります。

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (実際)

2025年6月17日

一次修了 (推定)

2027年1月1日

研究の完了 (推定)

2027年8月1日

試験登録日

最初に提出

2026年9月13日

QC基準を満たした最初の提出物

2026年9月13日

最初の投稿 (実際)

2026年9月17日

学習記録の更新

投稿された最後の更新 (実際)

2026年9月17日

QC基準を満たした最後の更新が送信されました

2026年9月13日

最終確認日

2026年9月1日

詳しくは

本研究に関する用語

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

未定

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

はい

米国で製造され、米国から輸出された製品。

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する