Automated ABPI in Peripheral Arterial Disease a Pilot Study (AUTO-PAD)

Assessing the Diagnostic Accuracy of AUTOmated Ankle-Brachial Pressure Index Devices Across the Spectrum of Peripheral Arterial Disease (AUTO-PAD)- A Pilot Study

The aim of this pilot study is to investigate the diagnostic accuracy of two automated ABPI devices across the spectrum of PAD (asymptomatic PAD, suspected symptomatic PAD [IC; including mild moderate and severe IC based on the Rutherford classification (21) and CLTI] and post-bypass surveillance). A second aim is to assess their useability and tolerability through semi-structured interviews with patients and clinicians. This means the research will be split into two workstreams.

調査の概要

研究の種類

観察的

入学 (推定)

400

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

研究連絡先のバックアップ

研究場所

      • Hull、イギリス、HU3 2JZ
        • 募集
        • Hull University Teaching Hospitals NHS Trust
        • コンタクト:

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年歳以上 (大人、高齢者)

健康ボランティアの受け入れ

はい

受講資格のある性別

全て

サンプリング方法

非確率サンプル

調査対象母集団

Group 1: Asymptomatic PAD:

Patients identified by the GP (PIC sites). Records screened to identify 'at risk' of PAD with >1 cardiovascular disease risk factor

Group 2: Suspected symptomatic PAD:

2a. Patients with suspected IC referred to dedicated nurse-led claudication clinic.

Group 2b. Patients with suspected CLTI referred to dedicated urgent vascular (HOT) clinic.

Group 3: Patients undergoing surveillance seen in dedicated clinics by vascular scientists

説明

Inclusion Criteria:

Group 1: Asymptomatic

  • Adults aged 18 and over
  • At risk of PAD with ≥1 cardiovascular risk factor
  • No previous diagnosis of PAD
  • English speaking and able to understand test protocols

Group 2a: Suspected PAD (intermittent claudication)

  • Adults aged 18 and over
  • Referred and booked into the nurse-led claudication clinic due to signs/symptoms in keeping with suspected intermittent claudication
  • English speaking and able to understand test protocols

Group 2b: Suspected PAD (CLTI)

  • Adults aged 18 and over
  • Referred and booked into the vascular HOT clinic due to signs/symptoms in keeping with suspected CLTI
  • English speaking and able to understand test protocols

Group 3: Post-bypass surveillance

  • Adults aged 18 and over
  • History of lower limb bypass surgery, enrolled in and attending the local surveillance programme
  • English speaking and able to understand test protocols

Exclusion Criteria:

  • Patients who are unable to provide informed consent
  • Recent (≤4 weeks) lower-limb surgery
  • Ulcers, casts or bandages that cannot or should not be compressed
  • Patients who have had a therapeutic intervention (stent or graft) which extends to the ankle
  • Patients with known or suspected acute deep vein thrombosis (DVT) or superficial thrombophlebitis
  • Marked oedema in one or both feet
  • Inability to tolerate a supine position
  • Patients with spasms or tremors of any kind
  • Arm arteriovenous fistula
  • Inability to cooperate due to impaired cognition

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 観測モデル:COHORT
  • 時間の展望:CROSS_SECTIONAL

コホートと介入

グループ/コホート
介入・治療
Group 1 Asymptomatic

Participants with one or more CVD risk factors

The MESI mTablet uses 3CUFFTM technology to detect pulse waveforms to assess ABPI. This involves placing a blood pressure cuff on all 4 limbs at the same time. The two arm cuffs then inflate and simultaneously measure the blood pressure in both brachial arteries. The cuff with the highest systolic pressure than reinflates with both the ankle cuffs simultaneously. The ABPI is then calculated and displayed on the tablet. A ratio of <0.9 will indicate the presence of PAD.

The BlueDop device is a cuffless device which comprises an egg-shaped handheld probe component and an output monitor device. The device measures mean arterial blood pressure through an algorithm. Simply, it calculates the average blood pressure based on the normal rise and fall of pressure during each heartbeat. The device then automatically generates an ankle-brachial pressure index and a corresponding traffic-light classification for each lower-limb artery, based on waveform. This output is interpreted using table 2 below. Essentially an ABPI of 0.8-1.0 with a biphasic or triphasic waveform indicates no disease whereas an ABPI <0.8 and / or a monophasic/non-specific waveform indicates the presence of PAD.

Group 2a Suspected PAD (intermittent claudication)

Partcipants presenting with Suspected PAD (intermittent claudication)

The MESI mTablet uses 3CUFFTM technology to detect pulse waveforms to assess ABPI. This involves placing a blood pressure cuff on all 4 limbs at the same time. The two arm cuffs then inflate and simultaneously measure the blood pressure in both brachial arteries. The cuff with the highest systolic pressure than reinflates with both the ankle cuffs simultaneously. The ABPI is then calculated and displayed on the tablet. A ratio of <0.9 will indicate the presence of PAD.

The BlueDop device is a cuffless device which comprises an egg-shaped handheld probe component and an output monitor device. The device measures mean arterial blood pressure through an algorithm. Simply, it calculates the average blood pressure based on the normal rise and fall of pressure during each heartbeat. The device then automatically generates an ankle-brachial pressure index and a corresponding traffic-light classification for each lower-limb artery, based on waveform. This output is interpreted using table 2 below. Essentially an ABPI of 0.8-1.0 with a biphasic or triphasic waveform indicates no disease whereas an ABPI <0.8 and / or a monophasic/non-specific waveform indicates the presence of PAD.

Group 2b Suspected PAD (CLTI)

Participants presenting with tissue loss or rest pain

The MESI mTablet uses 3CUFFTM technology to detect pulse waveforms to assess ABPI. This involves placing a blood pressure cuff on all 4 limbs at the same time. The two arm cuffs then inflate and simultaneously measure the blood pressure in both brachial arteries. The cuff with the highest systolic pressure than reinflates with both the ankle cuffs simultaneously. The ABPI is then calculated and displayed on the tablet. A ratio of <0.9 will indicate the presence of PAD.

The BlueDop device is a cuffless device which comprises an egg-shaped handheld probe component and an output monitor device. The device measures mean arterial blood pressure through an algorithm. Simply, it calculates the average blood pressure based on the normal rise and fall of pressure during each heartbeat. The device then automatically generates an ankle-brachial pressure index and a corresponding traffic-light classification for each lower-limb artery, based on waveform. This output is interpreted using table 2 below. Essentially an ABPI of 0.8-1.0 with a biphasic or triphasic waveform indicates no disease whereas an ABPI <0.8 and / or a monophasic/non-specific waveform indicates the presence of PAD.

Group 3 Post-bypass surveillance

Participants presenting for post-bypass surveillance

The MESI mTablet uses 3CUFFTM technology to detect pulse waveforms to assess ABPI. This involves placing a blood pressure cuff on all 4 limbs at the same time. The two arm cuffs then inflate and simultaneously measure the blood pressure in both brachial arteries. The cuff with the highest systolic pressure than reinflates with both the ankle cuffs simultaneously. The ABPI is then calculated and displayed on the tablet. A ratio of <0.9 will indicate the presence of PAD.

The BlueDop device is a cuffless device which comprises an egg-shaped handheld probe component and an output monitor device. The device measures mean arterial blood pressure through an algorithm. Simply, it calculates the average blood pressure based on the normal rise and fall of pressure during each heartbeat. The device then automatically generates an ankle-brachial pressure index and a corresponding traffic-light classification for each lower-limb artery, based on waveform. This output is interpreted using table 2 below. Essentially an ABPI of 0.8-1.0 with a biphasic or triphasic waveform indicates no disease whereas an ABPI <0.8 and / or a monophasic/non-specific waveform indicates the presence of PAD.

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Diagnostic Accuracy
時間枠:From enrolment to the end of treatment (1 day)

Diagnostic accuracy assessed via sensitivity, specificity, positive predictor value (PPV) and negative predictor value (NPV).

From enrolment to the end of treatment (1 day)

二次結果の測定

結果測定
メジャーの説明
時間枠
Validity assessed via correlation
時間枠:Data collected at enrolment (1 day)

Diagnostic performance assessed via area under the receiver operating characteristic (AUROC) curve and the assessment time difference, assessed via measurement of the time taken to perform each automated ABPI assessment and the manual ABPI assessment.

Data collected at enrolment (1 day)
Diagnostic performance
時間枠:Data collected at enrolment (1 day)

Assessed via area under the receiver operating characteristic (AUROC) curve.

Data collected at enrolment (1 day)
Time taken to perform each automated ABPI assessment and the manual ABPI assessment.
時間枠:Data collected at enrolment (1 day)

Time taken to perform each assessment in seconds

Data collected at enrolment (1 day)
Qualitative Workstream - Assessment of usability and tolerability of automated devices.
時間枠:Within the 2 year recruitment period

Semi structured interviews analysed using thematic analysis

Within the 2 year recruitment period
Visual Analogue Scale (VAS) score of 0-10
時間枠:Data collected at point of enrolment (1 day)

VAS used to assess tolerability

Data collected at point of enrolment (1 day)

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (実際)

2026年6月24日

一次修了 (推定)

2028年9月1日

研究の完了 (推定)

2028年9月1日

試験登録日

最初に提出

2026年9月14日

QC基準を満たした最初の提出物

2026年9月24日

最初の投稿 (実際)

2026年9月30日

学習記録の更新

投稿された最後の更新 (実際)

2026年10月1日

QC基準を満たした最後の更新が送信されました

2026年9月30日

最終確認日

2026年9月1日

詳しくは

本研究に関する用語

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する