Automated ABPI in Peripheral Arterial Disease a Pilot Study (AUTO-PAD)

September 30, 2026 updated by: Hull University Teaching Hospitals NHS Trust

Assessing the Diagnostic Accuracy of AUTOmated Ankle-Brachial Pressure Index Devices Across the Spectrum of Peripheral Arterial Disease (AUTO-PAD)- A Pilot Study

The aim of this pilot study is to investigate the diagnostic accuracy of two automated ABPI devices across the spectrum of PAD (asymptomatic PAD, suspected symptomatic PAD [IC; including mild moderate and severe IC based on the Rutherford classification (21) and CLTI] and post-bypass surveillance). A second aim is to assess their useability and tolerability through semi-structured interviews with patients and clinicians. This means the research will be split into two workstreams.

Study Overview

Study Type

Observational

Enrollment (Estimated)

400

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Study Contact Backup

Study Locations

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

Yes

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

Group 1: Asymptomatic PAD:

Patients identified by the GP (PIC sites). Records screened to identify 'at risk' of PAD with >1 cardiovascular disease risk factor

Group 2: Suspected symptomatic PAD:

2a. Patients with suspected IC referred to dedicated nurse-led claudication clinic.

Group 2b. Patients with suspected CLTI referred to dedicated urgent vascular (HOT) clinic.

Group 3: Patients undergoing surveillance seen in dedicated clinics by vascular scientists

Description

Inclusion Criteria:

Group 1: Asymptomatic

  • Adults aged 18 and over
  • At risk of PAD with ≥1 cardiovascular risk factor
  • No previous diagnosis of PAD
  • English speaking and able to understand test protocols

Group 2a: Suspected PAD (intermittent claudication)

  • Adults aged 18 and over
  • Referred and booked into the nurse-led claudication clinic due to signs/symptoms in keeping with suspected intermittent claudication
  • English speaking and able to understand test protocols

Group 2b: Suspected PAD (CLTI)

  • Adults aged 18 and over
  • Referred and booked into the vascular HOT clinic due to signs/symptoms in keeping with suspected CLTI
  • English speaking and able to understand test protocols

Group 3: Post-bypass surveillance

  • Adults aged 18 and over
  • History of lower limb bypass surgery, enrolled in and attending the local surveillance programme
  • English speaking and able to understand test protocols

Exclusion Criteria:

  • Patients who are unable to provide informed consent
  • Recent (≤4 weeks) lower-limb surgery
  • Ulcers, casts or bandages that cannot or should not be compressed
  • Patients who have had a therapeutic intervention (stent or graft) which extends to the ankle
  • Patients with known or suspected acute deep vein thrombosis (DVT) or superficial thrombophlebitis
  • Marked oedema in one or both feet
  • Inability to tolerate a supine position
  • Patients with spasms or tremors of any kind
  • Arm arteriovenous fistula
  • Inability to cooperate due to impaired cognition

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Observational Models: COHORT
  • Time Perspectives: CROSS_SECTIONAL

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Group 1 Asymptomatic

Participants with one or more CVD risk factors

The MESI mTablet uses 3CUFFTM technology to detect pulse waveforms to assess ABPI. This involves placing a blood pressure cuff on all 4 limbs at the same time. The two arm cuffs then inflate and simultaneously measure the blood pressure in both brachial arteries. The cuff with the highest systolic pressure than reinflates with both the ankle cuffs simultaneously. The ABPI is then calculated and displayed on the tablet. A ratio of <0.9 will indicate the presence of PAD.

The BlueDop device is a cuffless device which comprises an egg-shaped handheld probe component and an output monitor device. The device measures mean arterial blood pressure through an algorithm. Simply, it calculates the average blood pressure based on the normal rise and fall of pressure during each heartbeat. The device then automatically generates an ankle-brachial pressure index and a corresponding traffic-light classification for each lower-limb artery, based on waveform. This output is interpreted using table 2 below. Essentially an ABPI of 0.8-1.0 with a biphasic or triphasic waveform indicates no disease whereas an ABPI <0.8 and / or a monophasic/non-specific waveform indicates the presence of PAD.

Group 2a Suspected PAD (intermittent claudication)

Partcipants presenting with Suspected PAD (intermittent claudication)

The MESI mTablet uses 3CUFFTM technology to detect pulse waveforms to assess ABPI. This involves placing a blood pressure cuff on all 4 limbs at the same time. The two arm cuffs then inflate and simultaneously measure the blood pressure in both brachial arteries. The cuff with the highest systolic pressure than reinflates with both the ankle cuffs simultaneously. The ABPI is then calculated and displayed on the tablet. A ratio of <0.9 will indicate the presence of PAD.

The BlueDop device is a cuffless device which comprises an egg-shaped handheld probe component and an output monitor device. The device measures mean arterial blood pressure through an algorithm. Simply, it calculates the average blood pressure based on the normal rise and fall of pressure during each heartbeat. The device then automatically generates an ankle-brachial pressure index and a corresponding traffic-light classification for each lower-limb artery, based on waveform. This output is interpreted using table 2 below. Essentially an ABPI of 0.8-1.0 with a biphasic or triphasic waveform indicates no disease whereas an ABPI <0.8 and / or a monophasic/non-specific waveform indicates the presence of PAD.

Group 2b Suspected PAD (CLTI)

Participants presenting with tissue loss or rest pain

The MESI mTablet uses 3CUFFTM technology to detect pulse waveforms to assess ABPI. This involves placing a blood pressure cuff on all 4 limbs at the same time. The two arm cuffs then inflate and simultaneously measure the blood pressure in both brachial arteries. The cuff with the highest systolic pressure than reinflates with both the ankle cuffs simultaneously. The ABPI is then calculated and displayed on the tablet. A ratio of <0.9 will indicate the presence of PAD.

The BlueDop device is a cuffless device which comprises an egg-shaped handheld probe component and an output monitor device. The device measures mean arterial blood pressure through an algorithm. Simply, it calculates the average blood pressure based on the normal rise and fall of pressure during each heartbeat. The device then automatically generates an ankle-brachial pressure index and a corresponding traffic-light classification for each lower-limb artery, based on waveform. This output is interpreted using table 2 below. Essentially an ABPI of 0.8-1.0 with a biphasic or triphasic waveform indicates no disease whereas an ABPI <0.8 and / or a monophasic/non-specific waveform indicates the presence of PAD.

Group 3 Post-bypass surveillance

Participants presenting for post-bypass surveillance

The MESI mTablet uses 3CUFFTM technology to detect pulse waveforms to assess ABPI. This involves placing a blood pressure cuff on all 4 limbs at the same time. The two arm cuffs then inflate and simultaneously measure the blood pressure in both brachial arteries. The cuff with the highest systolic pressure than reinflates with both the ankle cuffs simultaneously. The ABPI is then calculated and displayed on the tablet. A ratio of <0.9 will indicate the presence of PAD.

The BlueDop device is a cuffless device which comprises an egg-shaped handheld probe component and an output monitor device. The device measures mean arterial blood pressure through an algorithm. Simply, it calculates the average blood pressure based on the normal rise and fall of pressure during each heartbeat. The device then automatically generates an ankle-brachial pressure index and a corresponding traffic-light classification for each lower-limb artery, based on waveform. This output is interpreted using table 2 below. Essentially an ABPI of 0.8-1.0 with a biphasic or triphasic waveform indicates no disease whereas an ABPI <0.8 and / or a monophasic/non-specific waveform indicates the presence of PAD.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Diagnostic Accuracy
Time Frame: From enrolment to the end of treatment (1 day)

Diagnostic accuracy assessed via sensitivity, specificity, positive predictor value (PPV) and negative predictor value (NPV).

From enrolment to the end of treatment (1 day)

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Validity assessed via correlation
Time Frame: Data collected at enrolment (1 day)

Diagnostic performance assessed via area under the receiver operating characteristic (AUROC) curve and the assessment time difference, assessed via measurement of the time taken to perform each automated ABPI assessment and the manual ABPI assessment.

Data collected at enrolment (1 day)
Diagnostic performance
Time Frame: Data collected at enrolment (1 day)

Assessed via area under the receiver operating characteristic (AUROC) curve.

Data collected at enrolment (1 day)
Time taken to perform each automated ABPI assessment and the manual ABPI assessment.
Time Frame: Data collected at enrolment (1 day)

Time taken to perform each assessment in seconds

Data collected at enrolment (1 day)
Qualitative Workstream - Assessment of usability and tolerability of automated devices.
Time Frame: Within the 2 year recruitment period

Semi structured interviews analysed using thematic analysis

Within the 2 year recruitment period
Visual Analogue Scale (VAS) score of 0-10
Time Frame: Data collected at point of enrolment (1 day)

VAS used to assess tolerability

Data collected at point of enrolment (1 day)

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

June 24, 2026

Primary Completion (Estimated)

September 1, 2028

Study Completion (Estimated)

September 1, 2028

Study Registration Dates

First Submitted

September 14, 2026

First Submitted That Met QC Criteria

September 24, 2026

First Posted (Actual)

September 30, 2026

Study Record Updates

Last Update Posted (Actual)

October 1, 2026

Last Update Submitted That Met QC Criteria

September 30, 2026

Last Verified

September 1, 2026

More Information

Terms related to this study

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.