UMOX™: A New Device for Preoxygenation

2013년 4월 25일 업데이트: Louis-Philippe Fortier, Maisonneuve-Rosemont Hospital

Background - Preoxygenation is performed before general anesthesia to increase the oxygen reserves of the body and prolong the safe period of apnea without hypoxia. However, it is often impossible to obtain a perfect face mask fit, and optimal end-tidal fractional oxygen concentration (EtO2) are not attained. The investigators designed a new oropharyngeal preoxygenation device, UMOX™, to avoid leaks during preoxygenation. In this study, the investigators compared the efficacy of UMOX™ with that of a conventional face mask.

Methods - In 50 healthy volunteers breathing 100% oxygen from a circle absorber system for a 5 minute-period, the investigators compared preoxygenation using the 1) a conventional mask; 2) UMOX™ without any instruction; 3) UMOX™ with instruction on mouth breathing; and 4) UMOX™ with a nose clip, in random order. Each subject underwent all steps with a 10-minute rest period of room air breathing between each step. Inspired and end-tidal respiratory gases were measured every 10 seconds. EtO2 variation was analyzed using Anova.

연구 개요

상태

완전한

정황

상세 설명

This study was as a single center, crossover, randomized trial. Following approval from research and ethics committee from our hospital and written consent from healthy subjects recruited, 50 consenting volunteers were enrolled over a three-month period, from: September 29, 2008 to November 5, 2008.

Subjects enrolled were of both genders, smokers and non-smokers, American Society of Anesthesiologists (ASA) physical status I or II, 18 to 65 years old. Nonrecruitment depended on the presence of any craniofacial abnormality, facial hair (beard or moustache) or absence of teeth. Well-controlled non-severe asthma was not a criterion for exclusion.

The trial was conducted in an operating room, using the Dräger Fabius GS (Dräger, Canada), anesthesia machine in a usual set up with the operating table. The investigators used a, patient circuit system consists of 1.5-L lime soda absorber connected to disposable corrugated tubing and a 2-L breathing bag. Gases were analyzed through a Marquette Medical Systems™ SAM module (GE Healthcare), volumes monitored with a GE Respirator Mechanics-M (GE Healthcare) and SpO2 measured with a Nellcor™ (Covidien) pulse oximetry positioned on a finger. Preoxygenation procedures were explained to the subjects and ample time was allowed so they could become familiar with the mask and mouth piece. Set up for the measure was conducted by the same experimenter. Subjects laid supine and preoxygenation was performed with fresh gas flow: 100% oxygen (O2) 10L/minute, administered trough facemask or mouth piece sequentially in a random order. Randomization was established with a random assignment table (ww.randomization.com). The envelope was open as the subject entered the operating room (OR). At the beginning and between each step, the anesthesia circuit was flushed with 100% (O2). Each subject went through all steps in random order, each requiring a 5-min period of: 1) tidal breathing using the facemask (TVFM); 2) tidal volume breathing with the UMOXTM mouthpiece (TVMP); 3) tidal volume breathing with the UMOXTM mouthpiece and verbal instructions to favor mouth breathing (TVMPMB); (4) tidal volume breathing with the UMOXTM mouthpiece and a nose clip (TVMPNC). The experimenter conducting the steps instructed participants to create a seal with their lips around the mouthpiece and readjusted the facemask when the subject signalled a discernable leak. Each step was interrupted by 10 minutes of breathing normal room air to allow for return to original respiratory status. Continuous recording of experimental measures, displayed on the anesthesia monitoring system, was done on digital films (Canon PowershotS5 IS™). For the purpose of our trial, 10 seconds intervals were considered adequate for following preoxygenation trends. End-tidal O2 was measured continuously for 5 min. We have also evaluated the level of satisfaction of the subject with each technique using a 1-4 scale.Participants were asked to asses their comfort by showing the associated number with their fingers. Finally we described the performance of the facemask ventilation technique as reported by the experimenter on a scale of 1 to 4.Primary outcome was the end-tidal O2 at the end of the 5-minutes preoxygenation period. Secondary outcomes included endtidal O2 as a function of time, the number of subjects reaching at least 87% EtO2, subject satisfaction and ease of ventilation.

Based on the few studies available on different preoxygenation technques, the end-tidal oxygen fraction standard deviation can range from 5% to about 20% depending on the device used for preoxygenation3,4,14,15. Assuming a 12% value for standard deviation, sample size calculations were made for Type I and Type II errors of 5% and 20%, respectively (power of 80%). From these numbers and taking account that we would rely on multiple comparisons, the power analysis suggested that 50 subjects would allow us to find a difference between groups around 9%. The 9nvestigators analyzed the EtO2 final value with one-way ANOVA followed by the Tukey's Multiple Comparisons test. Percentage of participants reaching at least 87% EtO2 are described with survival curves compared with Kaplan-Meier log rank test. Tolerance is presented using Kruskal-Wallis with Dunn's Multiple Comparisons test.

연구 유형

중재적

등록 (실제)

49

단계

  • 해당 없음

연락처 및 위치

이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.

연구 장소

    • Quebec
      • Montréal, Quebec, 캐나다, H1T 2M4
        • Maisonneuve-Rosemont Hospital

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

18년 (성인, 고령자)

건강한 자원 봉사자를 받아들입니다

연구 대상 성별

모두

설명

Inclusion Criteria:

  • Both genders
  • smokers and non-smokers
  • American Society of Anesthesiologists (ASA) physical status I or II
  • 18 to 65 years old.

Exclusion Criteria:

  • Presence of any craniofacial abnormality
  • facial hair (beard or moustache) or absence of teeth.
  • Well-controlled non-severe asthma was not a criterion for exclusion.

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

  • 주 목적: 치료
  • 할당: 무작위
  • 중재 모델: 크로스오버 할당
  • 마스킹: 없음(오픈 라벨)

무기와 개입

참가자 그룹 / 팔
개입 / 치료
가짜 비교기: Tidal breathing using the facemask
Control
실험적: Tidal volume breathing with the UMOXTM, device, no instruction
First experimental measure
Device used to perform preoxygenation in the operating room context
다른 이름들:
  • Umox
실험적: Tidal volume breathing with the UMOXTM device, instructions
Second experimental measure
Device used to perform preoxygenation in the operating room context
다른 이름들:
  • Umox
실험적: Tidal volume breathing with the UMOXTM, device, nose clip
Third experimental measure
Device used to perform preoxygenation in the operating room context
다른 이름들:
  • Umox

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
The primary outcome measurement was the evolution of endtidal oxygen fraction during a 5 minute preoxygenation procedure, measurements were done continuously, maximum values were recorded
기간: The four steps were done sequentially, each step consisted of a 5 minute recording period under specific conditions, followed by a 10 minute pause before the next step
The goal of this study is to compare the efficacy of preoxygenation using the conventional mask with the mouthpiece of the UMOX™ oropharyngeal device. The primary outcome measurement was endtidal oxygen fraction. The investigators also evaluated the respective tolerability of the two preoxygenation instruments in healthy volunteers. The measures were not part of a surgical procedure, participants were volunteers participating in this experimental design. No follow-up step was included in this protocol.
The four steps were done sequentially, each step consisted of a 5 minute recording period under specific conditions, followed by a 10 minute pause before the next step

2차 결과 측정

결과 측정
기간
The investigators also evaluated the respective tolerability of the two preoxygenation instruments in healthy volunteers with a short questionnaire
기간: The four steps were done sequentially, each step consisted of a 5 minute recording period under specific conditions, followed by a 10 minute pause before the next step
The four steps were done sequentially, each step consisted of a 5 minute recording period under specific conditions, followed by a 10 minute pause before the next step

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

수사관

  • 수석 연구원: Louis P Fortier, MD, MSc, Université de Montréal

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작

2008년 9월 1일

기본 완료 (실제)

2008년 11월 1일

연구 완료 (실제)

2008년 11월 1일

연구 등록 날짜

최초 제출

2013년 2월 27일

QC 기준을 충족하는 최초 제출

2013년 4월 25일

처음 게시됨 (추정)

2013년 4월 26일

연구 기록 업데이트

마지막 업데이트 게시됨 (추정)

2013년 4월 26일

QC 기준을 충족하는 마지막 업데이트 제출

2013년 4월 25일

마지막으로 확인됨

2013년 1월 1일

추가 정보

이 연구와 관련된 용어

기타 연구 ID 번호

  • 06108

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .

구독하다