此页面是自动翻译的,不保证翻译的准确性。请参阅 英文版 对于源文本。

UMOX™: A New Device for Preoxygenation

2013年4月25日 更新者:Louis-Philippe Fortier、Maisonneuve-Rosemont Hospital

Background - Preoxygenation is performed before general anesthesia to increase the oxygen reserves of the body and prolong the safe period of apnea without hypoxia. However, it is often impossible to obtain a perfect face mask fit, and optimal end-tidal fractional oxygen concentration (EtO2) are not attained. The investigators designed a new oropharyngeal preoxygenation device, UMOX™, to avoid leaks during preoxygenation. In this study, the investigators compared the efficacy of UMOX™ with that of a conventional face mask.

Methods - In 50 healthy volunteers breathing 100% oxygen from a circle absorber system for a 5 minute-period, the investigators compared preoxygenation using the 1) a conventional mask; 2) UMOX™ without any instruction; 3) UMOX™ with instruction on mouth breathing; and 4) UMOX™ with a nose clip, in random order. Each subject underwent all steps with a 10-minute rest period of room air breathing between each step. Inspired and end-tidal respiratory gases were measured every 10 seconds. EtO2 variation was analyzed using Anova.

研究概览

地位

完全的

条件

详细说明

This study was as a single center, crossover, randomized trial. Following approval from research and ethics committee from our hospital and written consent from healthy subjects recruited, 50 consenting volunteers were enrolled over a three-month period, from: September 29, 2008 to November 5, 2008.

Subjects enrolled were of both genders, smokers and non-smokers, American Society of Anesthesiologists (ASA) physical status I or II, 18 to 65 years old. Nonrecruitment depended on the presence of any craniofacial abnormality, facial hair (beard or moustache) or absence of teeth. Well-controlled non-severe asthma was not a criterion for exclusion.

The trial was conducted in an operating room, using the Dräger Fabius GS (Dräger, Canada), anesthesia machine in a usual set up with the operating table. The investigators used a, patient circuit system consists of 1.5-L lime soda absorber connected to disposable corrugated tubing and a 2-L breathing bag. Gases were analyzed through a Marquette Medical Systems™ SAM module (GE Healthcare), volumes monitored with a GE Respirator Mechanics-M (GE Healthcare) and SpO2 measured with a Nellcor™ (Covidien) pulse oximetry positioned on a finger. Preoxygenation procedures were explained to the subjects and ample time was allowed so they could become familiar with the mask and mouth piece. Set up for the measure was conducted by the same experimenter. Subjects laid supine and preoxygenation was performed with fresh gas flow: 100% oxygen (O2) 10L/minute, administered trough facemask or mouth piece sequentially in a random order. Randomization was established with a random assignment table (ww.randomization.com). The envelope was open as the subject entered the operating room (OR). At the beginning and between each step, the anesthesia circuit was flushed with 100% (O2). Each subject went through all steps in random order, each requiring a 5-min period of: 1) tidal breathing using the facemask (TVFM); 2) tidal volume breathing with the UMOXTM mouthpiece (TVMP); 3) tidal volume breathing with the UMOXTM mouthpiece and verbal instructions to favor mouth breathing (TVMPMB); (4) tidal volume breathing with the UMOXTM mouthpiece and a nose clip (TVMPNC). The experimenter conducting the steps instructed participants to create a seal with their lips around the mouthpiece and readjusted the facemask when the subject signalled a discernable leak. Each step was interrupted by 10 minutes of breathing normal room air to allow for return to original respiratory status. Continuous recording of experimental measures, displayed on the anesthesia monitoring system, was done on digital films (Canon PowershotS5 IS™). For the purpose of our trial, 10 seconds intervals were considered adequate for following preoxygenation trends. End-tidal O2 was measured continuously for 5 min. We have also evaluated the level of satisfaction of the subject with each technique using a 1-4 scale.Participants were asked to asses their comfort by showing the associated number with their fingers. Finally we described the performance of the facemask ventilation technique as reported by the experimenter on a scale of 1 to 4.Primary outcome was the end-tidal O2 at the end of the 5-minutes preoxygenation period. Secondary outcomes included endtidal O2 as a function of time, the number of subjects reaching at least 87% EtO2, subject satisfaction and ease of ventilation.

Based on the few studies available on different preoxygenation technques, the end-tidal oxygen fraction standard deviation can range from 5% to about 20% depending on the device used for preoxygenation3,4,14,15. Assuming a 12% value for standard deviation, sample size calculations were made for Type I and Type II errors of 5% and 20%, respectively (power of 80%). From these numbers and taking account that we would rely on multiple comparisons, the power analysis suggested that 50 subjects would allow us to find a difference between groups around 9%. The 9nvestigators analyzed the EtO2 final value with one-way ANOVA followed by the Tukey's Multiple Comparisons test. Percentage of participants reaching at least 87% EtO2 are described with survival curves compared with Kaplan-Meier log rank test. Tolerance is presented using Kruskal-Wallis with Dunn's Multiple Comparisons test.

研究类型

介入性

注册 (实际的)

49

阶段

  • 不适用

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习地点

    • Quebec
      • Montréal、Quebec、加拿大、H1T 2M4
        • Maisonneuve-Rosemont Hospital

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

18年 至 65年 (成人、年长者)

接受健康志愿者

是的

有资格学习的性别

全部

描述

Inclusion Criteria:

  • Both genders
  • smokers and non-smokers
  • American Society of Anesthesiologists (ASA) physical status I or II
  • 18 to 65 years old.

Exclusion Criteria:

  • Presence of any craniofacial abnormality
  • facial hair (beard or moustache) or absence of teeth.
  • Well-controlled non-severe asthma was not a criterion for exclusion.

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 主要用途:治疗
  • 分配:随机化
  • 介入模型:交叉作业
  • 屏蔽:无(打开标签)

武器和干预

参与者组/臂
干预/治疗
假比较器:Tidal breathing using the facemask
Control
实验性的:Tidal volume breathing with the UMOXTM, device, no instruction
First experimental measure
Device used to perform preoxygenation in the operating room context
其他名称:
  • Umox
实验性的:Tidal volume breathing with the UMOXTM device, instructions
Second experimental measure
Device used to perform preoxygenation in the operating room context
其他名称:
  • Umox
实验性的:Tidal volume breathing with the UMOXTM, device, nose clip
Third experimental measure
Device used to perform preoxygenation in the operating room context
其他名称:
  • Umox

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
The primary outcome measurement was the evolution of endtidal oxygen fraction during a 5 minute preoxygenation procedure, measurements were done continuously, maximum values were recorded
大体时间:The four steps were done sequentially, each step consisted of a 5 minute recording period under specific conditions, followed by a 10 minute pause before the next step
The goal of this study is to compare the efficacy of preoxygenation using the conventional mask with the mouthpiece of the UMOX™ oropharyngeal device. The primary outcome measurement was endtidal oxygen fraction. The investigators also evaluated the respective tolerability of the two preoxygenation instruments in healthy volunteers. The measures were not part of a surgical procedure, participants were volunteers participating in this experimental design. No follow-up step was included in this protocol.
The four steps were done sequentially, each step consisted of a 5 minute recording period under specific conditions, followed by a 10 minute pause before the next step

次要结果测量

结果测量
大体时间
The investigators also evaluated the respective tolerability of the two preoxygenation instruments in healthy volunteers with a short questionnaire
大体时间:The four steps were done sequentially, each step consisted of a 5 minute recording period under specific conditions, followed by a 10 minute pause before the next step
The four steps were done sequentially, each step consisted of a 5 minute recording period under specific conditions, followed by a 10 minute pause before the next step

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

调查人员

  • 首席研究员:Louis P Fortier, MD, MSc、Université de Montréal

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始

2008年9月1日

初级完成 (实际的)

2008年11月1日

研究完成 (实际的)

2008年11月1日

研究注册日期

首次提交

2013年2月27日

首先提交符合 QC 标准的

2013年4月25日

首次发布 (估计)

2013年4月26日

研究记录更新

最后更新发布 (估计)

2013年4月26日

上次提交的符合 QC 标准的更新

2013年4月25日

最后验证

2013年1月1日

更多信息

与本研究相关的术语

其他研究编号

  • 06108

此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.

订阅