Posture, Interface, and Peak Cough Flow (PCF P/I)
Body Posture and Interface Effects on Peak Cough Flow in Healthy Adults: A Randomized Crossover Study
Cough is a key airway protective mechanism requiring coordinated neuromuscular function. Peak cough flow (PCF) is a simple, non-invasive measure widely used to assess cough effectiveness and guide clinical decisions. Current ATS/ERS guidelines recommend measuring PCF in a seated position using a mouthpiece with a nose clip; however, this setup is not always feasible in clinical practice. Some patients cannot maintain posture or ensure an adequate mouthpiece seal, making a facial mask a potential alternative. The influence of body position and interface on PCF measurements remains unclear, with limited evidence from randomized crossover studies.
This study aims to evaluate the effects of body position (sitting vs standing) and measurement interface (mouthpiece vs facial mask) on PCF in healthy young adults. In a randomized within-subject crossover design, participants perform PCF measurements under four conditions (Sit-Mask, Stand-Mask, Sit-Mouthpiece, Stand-Mouthpiece) during a single session. The findings will help inform more flexible and standardized PCF assessment procedures in clinical practice.
연구 개요
상태
상세 설명
Cough is an essential protective mechanism that requires intact neuromuscular function and precise coordination across the inspiratory, compressive, and expulsive phases. Peak cough flow (PCF) is a simple, non-invasive measure, widely used to assess cough intensity and guide clinical decisions. Because of its clinical relevance, standardized procedures are essential. Current ATS/ERS recommendations specify seated testing with a mouthpiece and a noseclip (MP). However, this configuration is not always possible or consistently used in practice. Many patients cannot maintain adequate posture or mouthpiece seal, and a facial mask (FM) may be a feasible alternative. Evidence regarding the effects of posture (sitting vs standing) and interface (MP vs FM) on PCF remains inconsistent, with few studies comparing both factors within the same individuals using a randomized crossover design.
Aims To determine the effects of body position (sitting vs standing) and measurement interface (FM vs MP) on PCF in healthy young adults. We hypothesized that both factors would influence PCF.
Methods This randomized within subject crossover study included healthy, non smoking adults aged 20-30 years. Exclusion criteria were chronic disease, top-level athletic status and any signs of obstruction and/or auscultatory noise (wheezing or stridor) before the trial. Each participant completed four randomized conditions during a single session: Sit-Mask, Stand-Mask, Sit-Mouthpiece, and Stand-Mouthpiece. PCF (L/s) was measured using an ultrasonic spirometer (SpiroScout, Ganshorn). For each condition, participants performed three standardized maneuvers consisting of three tidal breaths, a rapid inspiration to total lung capacity, and a maximal cough; the highest PCF per condition was used for analysis. A two factor repeated measures ANOVA evaluated the effects of posture, interface, and their interaction (α = 0.05).
연구 유형
등록 (실제)
단계
- 해당 없음
연락처 및 위치
연구 장소
-
-
-
Brussels, 벨기에, 1180
- Cliniques universitaires Saint-Luc
-
-
참여기준
자격 기준
공부할 수 있는 나이
- 성인
건강한 자원 봉사자를 받아들입니다
설명
Inclusion Criteria:
- Healthy adults aged 20 to 30 years
- Non-smoker
- Self-reported good health status
- Able to perform all four randomized cough conditions
Exclusion Criteria:
- Any chronic disease
- Signs of airway obstruction on clinical examination
- Auscultatory abnormalities (wheezing or stridor) prior to testing
- Incomplete participation in all four conditions
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
- 주 목적: 다른
- 할당: 무작위
- 중재 모델: 크로스오버 할당
- 마스킹: 없음(오픈 라벨)
무기와 개입
참가자 그룹 / 팔 |
개입 / 치료 |
|---|---|
|
활성 비교기: Sit-Mask
|
Peak cough flow measured in the seated position.
The facial mask (V2 Mask™, Hans Rudolph, Inc., Kansas, USA; size fitted to participant) was connected to an ultrasonic spirometer (SpiroScout, Ganshorn) via a PFT filter.
Participants performed three tidal breaths followed by a maximal cough, repeated three times.
Peak cough flow measured in the standing position.
Same device and procedure as Sit-Mask condition.
|
|
활성 비교기: Stand-Mask
|
Peak cough flow measured in the seated position.
The facial mask (V2 Mask™, Hans Rudolph, Inc., Kansas, USA; size fitted to participant) was connected to an ultrasonic spirometer (SpiroScout, Ganshorn) via a PFT filter.
Participants performed three tidal breaths followed by a maximal cough, repeated three times.
Peak cough flow measured in the standing position.
Same device and procedure as Sit-Mask condition.
|
|
활성 비교기: Sit-Mouthpiece
|
Peak cough flow measured in the seated position.
The mouthpiece (Pulmosafe, Lemon Medical GmbH, Hammelburg, Germany) was used with a nose clip and connected to an ultrasonic spirometer (SpiroScout, Ganshorn) via a PFT filter.
Participants performed three tidal breaths followed by a maximal cough, repeated three times.
Peak cough flow measured in the standing position.
Same device and procedure as Sit-Mouthpiece condition.
|
|
활성 비교기: Stand-Mouthpiece
|
Peak cough flow measured in the seated position.
The mouthpiece (Pulmosafe, Lemon Medical GmbH, Hammelburg, Germany) was used with a nose clip and connected to an ultrasonic spirometer (SpiroScout, Ganshorn) via a PFT filter.
Participants performed three tidal breaths followed by a maximal cough, repeated three times.
Peak cough flow measured in the standing position.
Same device and procedure as Sit-Mouthpiece condition.
|
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Peak cough Flow (PCF)
기간: Day 1
|
PCF measurement in L/s
|
Day 1
|
공동 작업자 및 조사자
간행물 및 유용한 링크
일반 간행물
- Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, Crapo R, Enright P, van der Grinten CP, Gustafsson P, Jensen R, Johnson DC, MacIntyre N, McKay R, Navajas D, Pedersen OF, Pellegrino R, Viegi G, Wanger J; ATS/ERS Task Force. Standardisation of spirometry. Eur Respir J. 2005 Aug;26(2):319-38. doi: 10.1183/09031936.05.00034805. No abstract available.
- Brennan M, McDonnell MJ, Duignan N, Gargoum F, Rutherford RM. The use of cough peak flow in the assessment of respiratory function in clinical practice- A narrative literature review. Respir Med. 2022 Mar;193:106740. doi: 10.1016/j.rmed.2022.106740. Epub 2022 Jan 15.
연구 기록 날짜
연구 주요 날짜
연구 시작 (실제)
기본 완료 (실제)
연구 완료 (실제)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
추가 관련 MeSH 약관
기타 연구 ID 번호
- 2023/14JUI/264
개별 참가자 데이터(IPD) 계획
개별 참가자 데이터(IPD)를 공유할 계획입니까?
약물 및 장치 정보, 연구 문서
미국 FDA 규제 의약품 연구
미국 FDA 규제 기기 제품 연구
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .