IOL Power Formula Prediction Accuracy at the Extremes of Axial Length: A Prospective Outcome Verification Study (FORMULEX-OV)

2026년 7월 21일 업데이트: Pia Veronika Vécsei-Marlovits, Vienna Hospital Association

Clinical Outcomes of Formula-dependent IOL Power Selection at the Extremes of Axial Length: a Cross-sectional Outcome Verification of Seven ESCRS-hosted Formulas and the Ladas Super Formula 2.0 in Short (<22.0 mm) and Long (>26.0 mm) Eyes Using Refractive, Biometric, and Tomographic Assessment

This single-center, prospective, cross-sectional study evaluates how accurately eight modern intraocular lens (IOL) power calculation formulas predicted the postoperative refractive result in eyes with extreme axial length after cataract surgery. Patients previously operated at the study center who had a very short (axial length < 22.0 mm) or very long (> 26.0 mm) eye and received a spherical monofocal IOL are invited to a single, non-invasive outcome visit. At this visit, standardized manifest refraction, optical biometry, corneal tomography, slit-lamp examination and, in long eyes, macular OCT are performed. The measured postoperative spherical equivalent is compared with the residual refraction that each of the eight formulas had predicted for the actually implanted IOL power, separately in short and long eyes. The study aims to identify which formula predicts the postoperative outcome most accurately at the extremes of axial length, where formula choice is most consequential and least well studied.

연구 개요

상태

아직 모집하지 않음

상세 설명

Modern IOL power formulas achieve high refractive predictability in average-length eyes but deteriorate at the extremes of axial length, where effective lens position modelling diverges most between formulas. A preceding retrospective study at this center quantified the formula-dependent variability of IOL power recommendations at the axial length extremes using fully standardized biometric inputs; however, planning discrepancy alone does not indicate which formula's recommendation yields the better clinical outcome. This prospective outcome verification study closes that gap by linking a standardized postoperative assessment to the pre-existing, pseudonymized preoperative planning dataset.

Design: single-center, prospective, cross-sectional, observational (paired within-eye) study. From the retrospective source cohort (approximately 927 short and 769 long eyes; approximately 1,000-1,200 patients; index surgeries January 2022-December 2025), a stratified random sample of 200 patients is drawn (110 short stratum, 90 long stratum; 55:45 ratio) and invited to a single outcome visit. Each eligible eye contributes up to eight paired, formula-specific prediction errors that differ only by the calculation formula, while the postoperative reference (manifest refraction) and all preoperative inputs are held constant. Eyes are stratified a priori into a short (AL < 22.0 mm) and a long (AL > 26.0 mm) stratum and analyzed separately.

Formulas evaluated: Barrett Universal II, Cooke K6, EVO 2.0, Hill-RBF 3.0, Hoffer QST, Kane and PEARL-DGS (via the ESCRS IOL calculator), and the stand-alone Ladas Super Formula 2.0.

Outcome visit (single visit, all non-invasive standard ophthalmic techniques, no study-mandated therapy, medication or surgery): uncorrected and best-corrected distance visual acuity, IOLMaster 700 optical biometry, CASIA2 and Pentacam AXL corneal tomography, autorefraction and standardized subjective manifest refraction (primary outcome measure), slit-lamp examination with posterior capsule opacification grading, and macular OCT (obligatory in the long stratum, recommended in the short stratum).

Primary analysis: for each formula, the spherical equivalent prediction error (SE-PE = measured postoperative spherical equivalent minus the residual spherical equivalent predicted for the actually implanted IOL power step) is computed and the absolute error (|SE-PE|) compared globally across the eight formulas using the Friedman test, with post-hoc pairwise Wilcoxon signed-rank tests (Holm-corrected, 28 formula pairs), separately in each axial length stratum. The primary accuracy metric per formula is the median absolute prediction error.

All data are processed in pseudonymized form and linked via a study ID and laterality, in accordance with the Declaration of Helsinki (2024), ICH-GCP, and the GDPR. The study is reported per STROBE and is registered on ClinicalTrials.gov prior to the start of recruitment.

연구 유형

관찰

등록 (추정된)

200

연락처 및 위치

이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.

연구 연락처

연구 장소

    • State of Vienna
      • Vienna, State of Vienna, 오스트리아, 1130
        • Klinik Hietzing, Augenabteilung
        • 연락하다:

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

  • 어린이
  • 성인
  • 고령자

건강한 자원 봉사자를 받아들입니다

아니

샘플링 방법

비확률 샘플

연구 인구

Adults previously enrolled in the retrospective IOL power formula comparison study at the extremes of axial length at the study center, who underwent cataract surgery with a spherical (non-toric) monofocal IOL (HOYA Vivinex XY1, HOYA Vivinex Impress XY1-EM, or Johnson & Johnson Tecnis 1-Piece DIB00) between January 2022 and December 2025, and whose preoperative axial length was below 22.0 mm or above 26.0 mm. A stratified random sample of 200 patients (110 short-eye, 90 long-eye) is drawn from this source cohort and invited to a single outcome visit.

설명

Inclusion Criteria:

  • - Participation in the retrospective IOL power formula comparison study at the extremes of axial length (pseudonymized study ID assigned; complete preoperative biometry and archived formula outputs for all eight formula pathways)
  • Cataract surgery with a spherical (non-toric) IOL of one of the three reference models (HOYA Vivinex XY1, HOYA Vivinex Impress XY1-EM, or Johnson & Johnson Tecnis 1-Piece DIB00) at the study center between January 2022 and December 2025
  • Preoperative axial length below 22.0 mm (short stratum) or above 26.0 mm (long stratum), measured on IOLMaster 700
  • Documented implanted IOL model and power available from the operative record
  • At least 3 months since cataract surgery at the time of the outcome visit
  • Written informed consent for the outcome visit, including linkage of newly acquired postoperative data with the previously collected, pseudonymized preoperative planning data

Exclusion Criteria:

  • - Interval ocular surgery on the study eye since the original cataract surgery that may affect refraction or IOL position (e.g., corneal refractive surgery, IOL exchange or repositioning, secondary IOL implantation, or any intraocular surgery). Uncomplicated YAG laser capsulotomy without IOL repositioning is not an exclusion criterion
  • Implanted IOL not matching one of the three reference platforms (e.g., toric, multifocal or EDOF IOL, a different monofocal platform, or a sulcus-fixated IOL)
  • Toric IOL implantation
  • Acute ocular pathology preventing reliable refraction or biometric measurement
  • Cognitive or physical limitations precluding standardized subjective manifest refraction
  • Documented intraoperative complication with confirmed or suspected IOL malpositioning (e.g., posterior capsule rupture with vitreous loss, zonular dialysis requiring a capsular tension ring, sulcus placement)
  • Withdrawal of informed consent

Note: macular pathology is not an exclusion criterion; macular status is documented on OCT and analyzed as a covariate.

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

코호트 및 개입

그룹/코호트
개입 / 치료
Short axial length stratum (AL < 22.0 mm)
Pseudophakic eyes with a preoperative axial length below 22.0 mm that received a spherical monofocal IOL. These eyes undergo the standardized postoperative outcome visit; the measured spherical equivalent is compared with each formula's prediction for the implanted IOL power.
A single, non-invasive multimodal assessment: uncorrected and best-corrected distance visual acuity, IOLMaster 700 optical biometry, CASIA2 and Pentacam AXL corneal tomography, autorefraction and subjective manifest refraction, slit-lamp examination with posterior capsule opacification grading, and macular OCT. No therapeutic intervention, medication or surgery is performed.
Long axial length stratum (AL > 26.0 mm)
Pseudophakic eyes with a preoperative axial length above 26.0 mm that received a spherical monofocal IOL. These eyes undergo the standardized postoperative outcome visit (including obligatory macular OCT); the measured spherical equivalent is compared with each formula's prediction for the implanted IOL power.
A single, non-invasive multimodal assessment: uncorrected and best-corrected distance visual acuity, IOLMaster 700 optical biometry, CASIA2 and Pentacam AXL corneal tomography, autorefraction and subjective manifest refraction, slit-lamp examination with posterior capsule opacification grading, and macular OCT. No therapeutic intervention, medication or surgery is performed.

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
Spherical equivalent prediction error (SE-PE) across eight IOL power formulas, compared within each axial length stratum
기간: Single study visit, at least 3 months after cataract surgery
For each eye, SE-PE is calculated as the measured postoperative spherical equivalent (standardized manifest refraction) minus the residual spherical equivalent predicted by each formula for the actually implanted IOL power step. The absolute prediction error (|SE-PE|) is compared globally across the eight formulas (Barrett Universal II, Cooke K6, EVO 2.0, Hill-RBF 3.0, Hoffer QST, Kane, PEARL-DGS, Ladas Super Formula 2.0) using the Friedman test, with post-hoc pairwise Wilcoxon signed-rank tests (Holm-corrected, 28 formula pairs), performed separately in the short (AL < 22.0 mm) and long (AL > 26.0 mm) strata. The primary accuracy metric per formula is the median absolute prediction error (MedAE).
Single study visit, at least 3 months after cataract surgery

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작 (추정된)

2026년 8월 1일

기본 완료 (추정된)

2027년 6월 1일

연구 완료 (추정된)

2027년 7월 1일

연구 등록 날짜

최초 제출

2026년 7월 14일

QC 기준을 충족하는 최초 제출

2026년 7월 21일

처음 게시됨 (실제)

2026년 7월 24일

연구 기록 업데이트

마지막 업데이트 게시됨 (실제)

2026년 7월 24일

QC 기준을 충족하는 마지막 업데이트 제출

2026년 7월 21일

마지막으로 확인됨

2026년 7월 1일

추가 정보

이 연구와 관련된 용어

개별 참가자 데이터(IPD) 계획

개별 참가자 데이터(IPD)를 공유할 계획입니까?

아니요

약물 및 장치 정보, 연구 문서

미국 FDA 규제 의약품 연구

아니

미국 FDA 규제 기기 제품 연구

아니

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .

구독하다