IOL Power Formula Prediction Accuracy at the Extremes of Axial Length: A Prospective Outcome Verification Study (FORMULEX-OV)

21 luglio 2026 aggiornato da: Pia Veronika Vécsei-Marlovits, Vienna Hospital Association

Clinical Outcomes of Formula-dependent IOL Power Selection at the Extremes of Axial Length: a Cross-sectional Outcome Verification of Seven ESCRS-hosted Formulas and the Ladas Super Formula 2.0 in Short (<22.0 mm) and Long (>26.0 mm) Eyes Using Refractive, Biometric, and Tomographic Assessment

This single-center, prospective, cross-sectional study evaluates how accurately eight modern intraocular lens (IOL) power calculation formulas predicted the postoperative refractive result in eyes with extreme axial length after cataract surgery. Patients previously operated at the study center who had a very short (axial length < 22.0 mm) or very long (> 26.0 mm) eye and received a spherical monofocal IOL are invited to a single, non-invasive outcome visit. At this visit, standardized manifest refraction, optical biometry, corneal tomography, slit-lamp examination and, in long eyes, macular OCT are performed. The measured postoperative spherical equivalent is compared with the residual refraction that each of the eight formulas had predicted for the actually implanted IOL power, separately in short and long eyes. The study aims to identify which formula predicts the postoperative outcome most accurately at the extremes of axial length, where formula choice is most consequential and least well studied.

Panoramica dello studio

Stato

Non ancora reclutamento

Descrizione dettagliata

Modern IOL power formulas achieve high refractive predictability in average-length eyes but deteriorate at the extremes of axial length, where effective lens position modelling diverges most between formulas. A preceding retrospective study at this center quantified the formula-dependent variability of IOL power recommendations at the axial length extremes using fully standardized biometric inputs; however, planning discrepancy alone does not indicate which formula's recommendation yields the better clinical outcome. This prospective outcome verification study closes that gap by linking a standardized postoperative assessment to the pre-existing, pseudonymized preoperative planning dataset.

Design: single-center, prospective, cross-sectional, observational (paired within-eye) study. From the retrospective source cohort (approximately 927 short and 769 long eyes; approximately 1,000-1,200 patients; index surgeries January 2022-December 2025), a stratified random sample of 200 patients is drawn (110 short stratum, 90 long stratum; 55:45 ratio) and invited to a single outcome visit. Each eligible eye contributes up to eight paired, formula-specific prediction errors that differ only by the calculation formula, while the postoperative reference (manifest refraction) and all preoperative inputs are held constant. Eyes are stratified a priori into a short (AL < 22.0 mm) and a long (AL > 26.0 mm) stratum and analyzed separately.

Formulas evaluated: Barrett Universal II, Cooke K6, EVO 2.0, Hill-RBF 3.0, Hoffer QST, Kane and PEARL-DGS (via the ESCRS IOL calculator), and the stand-alone Ladas Super Formula 2.0.

Outcome visit (single visit, all non-invasive standard ophthalmic techniques, no study-mandated therapy, medication or surgery): uncorrected and best-corrected distance visual acuity, IOLMaster 700 optical biometry, CASIA2 and Pentacam AXL corneal tomography, autorefraction and standardized subjective manifest refraction (primary outcome measure), slit-lamp examination with posterior capsule opacification grading, and macular OCT (obligatory in the long stratum, recommended in the short stratum).

Primary analysis: for each formula, the spherical equivalent prediction error (SE-PE = measured postoperative spherical equivalent minus the residual spherical equivalent predicted for the actually implanted IOL power step) is computed and the absolute error (|SE-PE|) compared globally across the eight formulas using the Friedman test, with post-hoc pairwise Wilcoxon signed-rank tests (Holm-corrected, 28 formula pairs), separately in each axial length stratum. The primary accuracy metric per formula is the median absolute prediction error.

All data are processed in pseudonymized form and linked via a study ID and laterality, in accordance with the Declaration of Helsinki (2024), ICH-GCP, and the GDPR. The study is reported per STROBE and is registered on ClinicalTrials.gov prior to the start of recruitment.

Tipo di studio

Osservativo

Iscrizione (Stimato)

200

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Contatto studio

Luoghi di studio

    • State of Vienna
      • Vienna, State of Vienna, Austria, 1130
        • Klinik Hietzing, Augenabteilung
        • Contatto:

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

  • Bambino
  • Adulto
  • Adulto più anziano

Accetta volontari sani

No

Metodo di campionamento

Campione non probabilistico

Popolazione di studio

Adults previously enrolled in the retrospective IOL power formula comparison study at the extremes of axial length at the study center, who underwent cataract surgery with a spherical (non-toric) monofocal IOL (HOYA Vivinex XY1, HOYA Vivinex Impress XY1-EM, or Johnson & Johnson Tecnis 1-Piece DIB00) between January 2022 and December 2025, and whose preoperative axial length was below 22.0 mm or above 26.0 mm. A stratified random sample of 200 patients (110 short-eye, 90 long-eye) is drawn from this source cohort and invited to a single outcome visit.

Descrizione

Inclusion Criteria:

  • - Participation in the retrospective IOL power formula comparison study at the extremes of axial length (pseudonymized study ID assigned; complete preoperative biometry and archived formula outputs for all eight formula pathways)
  • Cataract surgery with a spherical (non-toric) IOL of one of the three reference models (HOYA Vivinex XY1, HOYA Vivinex Impress XY1-EM, or Johnson & Johnson Tecnis 1-Piece DIB00) at the study center between January 2022 and December 2025
  • Preoperative axial length below 22.0 mm (short stratum) or above 26.0 mm (long stratum), measured on IOLMaster 700
  • Documented implanted IOL model and power available from the operative record
  • At least 3 months since cataract surgery at the time of the outcome visit
  • Written informed consent for the outcome visit, including linkage of newly acquired postoperative data with the previously collected, pseudonymized preoperative planning data

Exclusion Criteria:

  • - Interval ocular surgery on the study eye since the original cataract surgery that may affect refraction or IOL position (e.g., corneal refractive surgery, IOL exchange or repositioning, secondary IOL implantation, or any intraocular surgery). Uncomplicated YAG laser capsulotomy without IOL repositioning is not an exclusion criterion
  • Implanted IOL not matching one of the three reference platforms (e.g., toric, multifocal or EDOF IOL, a different monofocal platform, or a sulcus-fixated IOL)
  • Toric IOL implantation
  • Acute ocular pathology preventing reliable refraction or biometric measurement
  • Cognitive or physical limitations precluding standardized subjective manifest refraction
  • Documented intraoperative complication with confirmed or suspected IOL malpositioning (e.g., posterior capsule rupture with vitreous loss, zonular dialysis requiring a capsular tension ring, sulcus placement)
  • Withdrawal of informed consent

Note: macular pathology is not an exclusion criterion; macular status is documented on OCT and analyzed as a covariate.

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

Coorti e interventi

Gruppo / Coorte
Intervento / Trattamento
Short axial length stratum (AL < 22.0 mm)
Pseudophakic eyes with a preoperative axial length below 22.0 mm that received a spherical monofocal IOL. These eyes undergo the standardized postoperative outcome visit; the measured spherical equivalent is compared with each formula's prediction for the implanted IOL power.
A single, non-invasive multimodal assessment: uncorrected and best-corrected distance visual acuity, IOLMaster 700 optical biometry, CASIA2 and Pentacam AXL corneal tomography, autorefraction and subjective manifest refraction, slit-lamp examination with posterior capsule opacification grading, and macular OCT. No therapeutic intervention, medication or surgery is performed.
Long axial length stratum (AL > 26.0 mm)
Pseudophakic eyes with a preoperative axial length above 26.0 mm that received a spherical monofocal IOL. These eyes undergo the standardized postoperative outcome visit (including obligatory macular OCT); the measured spherical equivalent is compared with each formula's prediction for the implanted IOL power.
A single, non-invasive multimodal assessment: uncorrected and best-corrected distance visual acuity, IOLMaster 700 optical biometry, CASIA2 and Pentacam AXL corneal tomography, autorefraction and subjective manifest refraction, slit-lamp examination with posterior capsule opacification grading, and macular OCT. No therapeutic intervention, medication or surgery is performed.

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Spherical equivalent prediction error (SE-PE) across eight IOL power formulas, compared within each axial length stratum
Lasso di tempo: Single study visit, at least 3 months after cataract surgery
For each eye, SE-PE is calculated as the measured postoperative spherical equivalent (standardized manifest refraction) minus the residual spherical equivalent predicted by each formula for the actually implanted IOL power step. The absolute prediction error (|SE-PE|) is compared globally across the eight formulas (Barrett Universal II, Cooke K6, EVO 2.0, Hill-RBF 3.0, Hoffer QST, Kane, PEARL-DGS, Ladas Super Formula 2.0) using the Friedman test, with post-hoc pairwise Wilcoxon signed-rank tests (Holm-corrected, 28 formula pairs), performed separately in the short (AL < 22.0 mm) and long (AL > 26.0 mm) strata. The primary accuracy metric per formula is the median absolute prediction error (MedAE).
Single study visit, at least 3 months after cataract surgery

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Stimato)

1 agosto 2026

Completamento primario (Stimato)

1 giugno 2027

Completamento dello studio (Stimato)

1 luglio 2027

Date di iscrizione allo studio

Primo inviato

14 luglio 2026

Primo inviato che soddisfa i criteri di controllo qualità

21 luglio 2026

Primo Inserito (Effettivo)

24 luglio 2026

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

24 luglio 2026

Ultimo aggiornamento inviato che soddisfa i criteri QC

21 luglio 2026

Ultimo verificato

1 luglio 2026

Maggiori informazioni

Termini relativi a questo studio

Altri numeri di identificazione dello studio

  • MA_15_EK_26_128_VK

Piano per i dati dei singoli partecipanti (IPD)

Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?

NO

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

No

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .