Agreement Among Three Methods of Intraoperative Core Temperature Monitoring: Nasopharyngeal Probe, TaiDoc TD-1261 Infrared Tympanic Thermometer, and 3M SpotOn™ Skin Sensor (TEMP-AGREE)
A Prospective Observational Method-Comparison Study of the Agreement Among Nasopharyngeal Temperature Probe, TaiDoc TD-1261 Infrared Tympanic Thermometer, and 3M SpotOn™ Zero-Heat-Flux Skin Sensor for Intraoperative Core Temperature Monitoring in Adult Patients Undergoing Elective Surgery Under General Anesthesia
Keeping a patient's body temperature normal during surgery is an important part of safe anesthetic care. When the body temperature drops below 36°C (called inadvertent perioperative hypothermia), patients face higher risks of bleeding, wound infection, heart problems, and a slower recovery. To prevent these complications, anesthesiologists monitor core body temperature continuously during surgery.
Three monitoring methods are commonly used in clinical practice: (1) a nasopharyngeal probe - a small soft sensor inserted through the nose and placed at the back of the throat, which provides continuous readings and is considered the gold standard; (2) an infrared tympanic thermometer (TaiDoc TD-1261) - a handheld device that measures ear temperature in about one second; and (3) the 3M SpotOn™ sensor - a non-invasive adhesive patch placed on the forehead that continuously estimates core temperature without requiring insertion into any body cavity.
Although all three methods are used interchangeably at Srinagarind Hospital, no study has directly compared all three simultaneously in the same patient to determine how well they agree with each other. If significant differences exist between methods, a clinician who switches from one device to another may unknowingly under-detect or over-estimate hypothermia.
This study will enroll 92 adult patients scheduled for elective surgery under general anesthesia at Srinagarind Hospital, Khon Kaen University. All three temperature sensors will be applied at the same time, and readings will be recorded simultaneously every 15 minutes throughout the operation. The main goal is to determine whether each device gives results close enough to the nasopharyngeal probe to be safely used as an alternative. The study is observational - no medications or procedures will be added to routine care.
研究概览
地位
详细说明
Background: Inadvertent perioperative hypothermia (IPH), defined as a core body temperature < 36.0°C, affects 20-90% of patients undergoing general anesthesia and is associated with coagulopathy, surgical site infection, morbid cardiac events, and prolonged drug metabolism. Accurate intraoperative temperature monitoring is mandated by professional society guidelines.
Study design: Prospective, single-center, triple-method observational agreement study at Srinagarind Hospital, Khon Kaen University, Thailand. All three thermometry modalities are applied simultaneously in each participant. No intervention is assigned; temperature monitoring is additional to standard perioperative care.
Methods: After induction of general anesthesia and tracheal intubation, the nasopharyngeal (NP) probe (reference standard), TaiDoc TD-1261 infrared tympanic thermometer (test method 1), and 3M SpotOn™ zero-heat-flux forehead sensor (test method 2) are placed. Simultaneous temperature readings are recorded at 15-minute intervals from 15 minutes after induction until surgery end. For the TD-1261, two consecutive readings are taken at each time point and averaged; pinna retraction and fresh disposable probe cover are used at every measurement. The SpotOn™ sensor requires a 10-minute equilibration period before the first recorded value.
Statistical analysis: Primary analysis uses repeated-measures Bland-Altman analysis (Myles & Cui, 2007 framework) to estimate bias and limits of agreement (LoA) for each method pair against the NP probe. Pre-specified clinical acceptability criterion: LoA within ±0.5°C (Sessler, 2008). Variance of differences is decomposed into between-subject (σ²_B) and within-subject (σ²_W) components. Secondary analyses: proportional bias assessment, sensitivity/specificity for hypothermia detection (NP < 36.0°C), intraclass correlation coefficient (ICC, two-way mixed-effects model), and multivariable linear mixed-effects regression for clinical covariates. Sample size: 92 patients (Lu et al., 2016 method; 80% power; α = 0.05; SD of differences = 0.30°C for NP vs. TM; LoA acceptability threshold ±0.50°C; 15% dropout allowance).
This study provides the first institution-specific, population-specific evidence on temperature monitoring agreement in Thai surgical patients and constitutes the first published intraoperative validation of the TaiDoc TD-1261 infrared tympanic thermometer.
研究类型
注册 (估计的)
联系人和位置
学习联系方式
- 姓名:Thepakorn Sathitkarnmanee, MD
- 电话号码:66819547622
- 邮箱:thepakorns@gmail.com
研究联系人备份
- 姓名:Sirirat Tribuddharat, MD, PhD
- 电话号码:66816205920
- 邮箱:sirirat.tribuddharat@gmail.com
学习地点
-
-
-
Khon Kaen、泰国、40002
- Faculty of Medicine, Khon Kaen University
-
-
参与标准
资格标准
适合学习的年龄
- 成人
- 年长者
接受健康志愿者
取样方法
研究人群
描述
Inclusion Criteria:
- Age ≥ 18 years
- We will include 3 independent groups of patients receiving general anesthesia with tracheal intubation:
Group 1: Cardiac surgery Group 2: Thoracic surgery Group 3: General surgery, including neurosurgery
- Expected surgical duration ≥ 60 minutes
- ASA Physical Status Classification I- III
- Able to provide written informed consent
Exclusion Criteria:
- Prior nasal surgery, nasal polyps, deviated nasal septum, or significant epistaxis history precluding safe nasopharyngeal probe insertion
- Otitis media, tympanic membrane perforation, significant cerumen impaction, or external auditory canal deformity that would prevent reliable infrared tympanic temperature measurement
- Facial or forehead skin lesions, burns, or known adhesive hypersensitivity at the planned 3M SpotOn™ application site, craniotomy, and head and neck surgery.
- Baseline core temperature < 35.0°C or > 38.5°C at preoperative assessment
- Planned use of neuraxial anesthesia as the primary or adjunct technique
- Emergency procedures, pregnancy, or procedures involving deliberate hypothermia or cardiopulmonary bypass.
学习计划
研究是如何设计的?
设计细节
队列和干预
团体/队列 |
|---|
|
Surgical Patients - All Three Monitoring Methods
Adult patients (age ≥ 18 years) undergoing elective surgery under general anesthesia with tracheal intubation at Srinagarind Hospital and Queen Sirikit Heart Center of the Northeast, Khon Kaen University, with an expected surgical duration ≥ 60 minutes and ASA Physical Status I-III.
All participants have all three temperature monitoring devices (NP probe, TaiDoc TD-1261 IR tympanic thermometer, and 3M SpotOn™ ZHF sensor) applied simultaneously following induction of anesthesia.
|
研究衡量的是什么?
主要结果指标
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Limits of Agreement Between Each Test Method and the Nasopharyngeal Probe
大体时间:From 15 minutes after anesthesia induction until end of surgery; measurements recorded at 15-minute intervals (estimated total observation window: 60-195 minutes per participant)
|
The 95% limits of agreement (LoA = Bias ± 1.96 × SD_total) between (i) the TaiDoc TD-1261 infrared tympanic thermometer and the nasopharyngeal probe, and (ii) the 3M SpotOn™ sensor and the nasopharyngeal probe, estimated by repeated-measures Bland-Altman analysis.
Pre-specified clinical acceptability criterion: LoA must be contained within ±0.5°C with both 95% CI limits of the upper and lower LoA falling within this range.
|
From 15 minutes after anesthesia induction until end of surgery; measurements recorded at 15-minute intervals (estimated total observation window: 60-195 minutes per participant)
|
次要结果测量
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Systematic Bias (Mean Difference) of Each Test Method vs. NP Probe
大体时间:From 15 minutes after anesthesia induction until surgery end (15-minute intervals)
|
Mean difference (°C) with 95% CI between each test method and the NP probe across all paired observations, estimated from the repeated-measures Bland-Altman model
|
From 15 minutes after anesthesia induction until surgery end (15-minute intervals)
|
|
Proportional Bias Assessment
大体时间:From 15 minutes after anesthesia induction until surgery end
|
Linear regression of the difference (test - reference) on the mean of both methods, with cluster-robust standard errors; Pitman's test for correlation between sum and difference
|
From 15 minutes after anesthesia induction until surgery end
|
|
Sensitivity and Specificity for Hypothermia Detection
大体时间:From 15 minutes after anesthesia induction until surgery end
|
Sensitivity, specificity, PPV, and NPV of each test method for detecting intraoperative hypothermia defined as NP temperature < 36.0°C, with AUROC estimated by DeLong's method
|
From 15 minutes after anesthesia induction until surgery end
|
|
Intraclass Correlation Coefficient
大体时间:From 15 minutes after anesthesia induction until surgery end
|
ICC (two-way mixed-effects model, absolute agreement) with 95% CI for each method pair
|
From 15 minutes after anesthesia induction until surgery end
|
|
Percentage of Paired Measurements Within ±0.5°C
大体时间:From 15 minutes after anesthesia induction until surgery end
|
Percentage of all paired simultaneous readings in which the test method falls within ±0.5°C of the simultaneous NP reference reading
|
From 15 minutes after anesthesia induction until surgery end
|
|
Clinical Covariate Association With Method Difference
大体时间:Baseline characteristics collected before surgery; temperature data from 15 minutes post-induction until surgery end
|
Association between clinical covariates (age, sex, BMI, ASA status, surgical duration, anesthetic technique, active warming use, ambient OR temperature, baseline core temperature) and the absolute method difference, estimated by multivariable linear mixed-effects regression
|
Baseline characteristics collected before surgery; temperature data from 15 minutes post-induction until surgery end
|
|
Incidence of Device-Related Adverse Events
大体时间:From device placement until surgery end
|
Rate of epistaxis (NP probe), aural discomfort or canal injury (TD-1261), and adhesive skin reaction (SpotOn™) during the monitoring period
|
From device placement until surgery end
|
合作者和调查者
出版物和有用的链接
一般刊物
- Sessler DI. Perioperative thermoregulation and heat balance. Lancet. 2016 Jun 25;387(10038):2655-2664. doi: 10.1016/S0140-6736(15)00981-2. Epub 2016 Jan 8.
- Torossian A. Thermal management during anaesthesia and thermoregulation standards for the prevention of inadvertent perioperative hypothermia. Best Pract Res Clin Anaesthesiol. 2008 Dec;22(4):659-68. doi: 10.1016/j.bpa.2008.07.006.
- Sessler DI. Temperature monitoring and perioperative thermoregulation. Anesthesiology. 2008 Aug;109(2):318-38. doi: 10.1097/ALN.0b013e31817f6d76.
- Lu MJ, Zhong WH, Liu YX, Miao HZ, Li YC, Ji MH. Sample Size for Assessing Agreement between Two Methods of Measurement by Bland-Altman Method. Int J Biostat. 2016 Nov 1;12(2):/j/ijb.2016.12.issue-2/ijb-2015-0039/ijb-2015-0039.xml. doi: 10.1515/ijb-2015-0039.
- Matsukawa T, Sessler DI, Sessler AM, Schroeder M, Ozaki M, Kurz A, Cheng C. Heat flow and distribution during induction of general anesthesia. Anesthesiology. 1995 Mar;82(3):662-73. doi: 10.1097/00000542-199503000-00008.
- Cork RC, Vaughan RW, Humphrey LS. Precision and accuracy of intraoperative temperature monitoring. Anesth Analg. 1983 Feb;62(2):211-4.
- Wang J, Liang H, Tian C, Rong G, Shao X, Ran C. Agreement of zero-heat-flux thermometry compared with infrared tympanic temperature monitoring in adults undergoing major surgery. Eur J Med Res. 2025 Jan 30;30(1):60. doi: 10.1186/s40001-025-02317-9.
- Iden T, Horn EP, Bein B, Bohm R, Beese J, Hocker J. Intraoperative temperature monitoring with zero heat flux technology (3M SpotOn sensor) in comparison with sublingual and nasopharyngeal temperature: An observational study. Eur J Anaesthesiol. 2015 Jun;32(6):387-91. doi: 10.1097/EJA.0000000000000232.
- Wan L, Shen PY, Zhang SX, Wang LZ. Agreement of infrared ear temperature with nasopharyngeal temperature and diagnostic performance on hypothermia in general anesthetized patients. J Chin Med Assoc. 2022 Nov 1;85(11):1093-1097. doi: 10.1097/JCMA.0000000000000770. Epub 2022 Jul 6.
研究记录日期
研究主要日期
学习开始 (估计的)
初级完成 (估计的)
研究完成 (估计的)
研究注册日期
首次提交
首先提交符合 QC 标准的
首次发布 (实际的)
研究记录更新
最后更新发布 (实际的)
上次提交的符合 QC 标准的更新
最后验证
更多信息
与本研究相关的术语
关键字
其他研究编号
- HE691250
计划个人参与者数据 (IPD)
计划共享个人参与者数据 (IPD)?
IPD 计划说明
IPD 共享时间框架
IPD 共享访问标准
IPD 共享支持信息类型
- 研究方案
药物和器械信息、研究文件
研究美国 FDA 监管的药品
研究美国 FDA 监管的设备产品
此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.