Real-World Validation of A Minimally Disruptive Ambulatory Blood Pressure Monitoring (ABPM) Huawei Smartwatch in Malaysia (Huawei-SunU02)
Background:
Ambulatory Blood Pressure Monitoring (ABPM) is the gold standard for diagnosing high blood pressure (hypertension) because it tracks blood pressure continuously across day and night. However, standard ABPM requires a cuff that inflates every 15 to 30 minutes during the day and every 30 to 60 minutes at night. These frequent inflations often cause discomfort, interrupt sleep, and cause patients to abandon testing.
Purpose:
This study aims to evaluate whether a new cuff-based Huawei smartwatch can provide a less-disturbing ABPM experience. Researchers will test whether a reduced-frequency measurement schedule (e.g., 7 readings: 4 awake, 3 asleep) can provide the same diagnostic accuracy as a conventional, full 24-hour ABPM protocol (typically 48 readings) while improving patient comfort.
Study Design & Procedures:
Study Population:
Approximately 500 adult participants will be recruited over a 24-month study period, with a targeted subgroup of 50 participants selected for clinical validation. The subgroup will represent normotensive, pre-hypertensive, and hypertensive individuals across diverse age groups and ethnicities.
Monitoring & Comparison: Participants in the validation group will undergo standard 24-hour ABPM at an accredited clinical facility alongside monitoring with the Huawei smartwatch. Data from the smartwatch will be used to compare a reduced-frequency reading schedule against full 24-hour ABPM recordings.
User Feedback: Participants will complete questionnaires assessing comfort, sleep quality, ease of use, and overall preference between the smartwatch and standard ABPM cuffs.
Goal:
The main goal is to determine if a reduced-frequency wearable monitoring protocol can maintain clinical diagnostic accuracy for blood pressure and sleep-dipping patterns while significantly reducing patient burden and sleep disruption.
研究概览
地位
条件
研究类型
注册 (估计的)
联系人和位置
学习地点
-
-
Selangor
-
Subang Jaya、Selangor、马来西亚、47500
- Faculty of Medical and Life Sciences, Sunway University
-
-
参与标准
资格标准
适合学习的年龄
接受健康志愿者
有资格学习的性别
取样方法
研究人群
The study population comprises adult community-dwelling individuals residing in Malaysia recruited via the Huawei Research Platform. From this general screening pool (target N = 500), a targeted clinical validation subgroup (n = 50) will be prospectively selected to undergo dual ABPM and smartwatch evaluation.
To ensure clinical applicability and statistical validity in alignment with ISO 81060-2 and AAMI validation standards, the validation subgroup will be stratified into three distinct cohorts:
- Normotensive Cohort (n = 10): Participants with blood pressure < 135/85 mmHg.
- Undiagnosed Elevated Blood Pressure Cohort (n = 20): Participants with untreated blood pressure > 135/85 mmHg.
- Documented Hypertensive Cohort (n = 20): Participants with documented or self-reported history of history of hypertension.
Selection across all cohorts will maintain balanced demographic representation across age groups, biological sex, and major Malaysian ethnic populations (Malay, Chinese, Indian)
描述
Inclusion Criteria:
- Possesses the specific Huawei wearable required for the study
- Aged ≥18 years
- Malaysian citizen/permanent resident
- Understands English, Malay, or Chinese
- Able to download the required application on a compatible smartphone
- Willing to wear devices continuously
- Able to provide informed consent
Exclusion Criteria:
- Current pregnancy
- Presence of implantable cardiac devices (e.g., permanent pacemakers, •implantable cardioverter-defibrillators [ICDs])
- Uncontrolled cardiac arrhythmias (e.g., active atrial fibrillation, frequent premature ventricular contractions [PVCs])
- Movement disorders or active tremors that interfere with blood pressure measurement or signal acquisition (e.g., severe Parkinson's disease, essential tremor)
- Acute medical conditions (e.g., acute coronary syndromes, stroke/TIA within 3 months, acute decompensated heart failure, severe active infections, or hypertensive emergencies)
- Resting blood pressure >=180/110 mmHg
学习计划
研究是如何设计的?
设计细节
- 观测模型:ECOLOGIC_OR_COMMUNITY
- 时间观点:CROSS_SECTIONAL
队列和干预
团体/队列 |
|---|
|
Normotensive Cohort
Adult participants with normal blood pressure (defined as blood pressure < 135/85 mmHg; n = 10) who undergo simultaneous standard 24-hour ABPM and Huawei smartwatch monitoring. |
|
Undiagnosed Elevated Blood Pressure Cohort
Adult participants with undiagnosed elevated blood pressure (defined as blood pressure > 135/85 mmHg; n = 20) who undergo simultaneous standard 24-hour ABPM and Huawei smartwatch monitoring. |
|
Documented Hypertensive Cohort
Adult participants with documented or self-reported hypertension (n = 20) who undergo simultaneous standard 24-hour ABPM and Huawei smartwatch monitoring. |
|
Screening Population
General adult participants recruited across Malaysia who enroll via the Huawei Research Platform app to provide continuous wearable ambulatory blood pressure and physiological data and complete baseline health assessments over the study period. |
研究衡量的是什么?
主要结果指标
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Mean Difference in Blood Pressure Measured by Reduced-Frequency Smartwatch Schedule Versus Standard 24-Hour Ambulatory Blood Pressure Monitor (ABPM)
大体时间:Month 13 through Month 18 (during clinical ABPM validation testing).
|
Measure Description: Mean blood pressure difference (bias) for 24-hour, daytime, and night-time periods calculated by subtracting reference 24-hour ABPM values (SpaceLabs/Mobil-O-Graph cuff monitor) from reduced-frequency Huawei smartwatch ABPM values (7-measurement schedule). Unit of Measure: mmHg |
Month 13 through Month 18 (during clinical ABPM validation testing).
|
|
Mean Absolute Error (MAE) of Blood Pressure Readings Between Reduced-Frequency Smartwatch Schedule and Standard 24-Hour ABPM
大体时间:Month 13 through Month 18.
|
Measure Description: The average absolute difference between blood pressure readings obtained from the reduced-frequency Huawei smartwatch schedule and the standard 24-hour cuff-based ABPM monitor. Unit of Measure: mmHg |
Month 13 through Month 18.
|
|
Root Mean Square Error (RMSE) of Blood Pressure Readings Between Reduced-Frequency Smartwatch Schedule and Standard 24-Hour ABPM
大体时间:Month 13 through Month 18
|
Measure Description: The root mean square error evaluating agreement between blood pressure values recorded by the reduced-frequency Huawei smartwatch schedule and the standard 24-hour cuff-based ABPM monitor. Unit of Measure: mmHg |
Month 13 through Month 18
|
|
Intraclass Correlation Coefficient (ICC) of Blood Pressure Measurements Between Reduced-Frequency Smartwatch Schedule and Standard 24-Hour ABPM
大体时间:Month 13 through Month 18
|
Measure Description: Agreement reliability calculated using absolute-agreement two-way random-effects ICC comparing reduced-frequency Huawei smartwatch blood pressure estimates against full 24-hour ABPM readings. Unit of Measure: Intraclass Correlation Coefficient (score ranging from -1.0 to +1.0) |
Month 13 through Month 18
|
次要结果测量
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Limits of Agreement (Bland-Altman) for Blood Pressure Readings Between Huawei Smartwatch and Conventional Cuff ABPM
大体时间:Month 13 through Month 18
|
Measure Description: Upper and lower 95% limits of agreement (mean difference +/- 1.96 x standard deviation) calculated via Bland-Altman analysis comparing simultaneous Huawei smartwatch and conventional cuff-based ABPM measurements. Unit of Measure: mmHg |
Month 13 through Month 18
|
|
Participant Comfort Score Evaluated via Post-Monitoring User Experience Survey
大体时间:Immediately following completion of 24-hour monitoring (Month 13 through Month 18)
|
Measure Description: Mean total score for participant-reported physical comfort derived from a standardized 5-point Likert scale questionnaire administered immediately after 24-hour monitoring. Unit of Measure: Score on a scale of 1 to 5 (where 1 = Extreme Discomfort and 5 = Complete Comfort) |
Immediately following completion of 24-hour monitoring (Month 13 through Month 18)
|
|
Number of Participants Indicating Preference for Smartwatch Monitoring Over Conventional Cuff ABPM
大体时间:Immediately following completion of 24-hour monitoring (Month 13 through Month 18)
|
Measure Description: Count of participants who express a preference for smartwatch-based monitoring over traditional arm-cuff ABPM on the post-monitoring user experience survey. Unit of Measure: Participant count |
Immediately following completion of 24-hour monitoring (Month 13 through Month 18)
|
其他结果措施
结果测量 |
措施说明 |
大体时间 |
|---|---|---|
|
Cohen's Kappa Coefficient for Nocturnal Dipping Pattern Classification Concordance
大体时间:Month 16 through Month 20.
|
Measure Description: Inter-rater agreement (Cohen's Kappa) for categorizing participants into nocturnal dipping status categories (dipper vs. non-dipper) comparing reduced-frequency smartwatch schedules against standard full-day ABPM. Unit of Measure: Kappa Statistic (score ranging from -1.0 to +1.0) |
Month 16 through Month 20.
|
|
Minimum Number of Measurements Required to Maintain Diagnostic Accurac
大体时间:Month 16 through Month 20.
|
Measure Description: The minimum threshold count of active blood pressure inflations required per 24-hour period to maintain an ICC >= 0.80 relative to standard 24-hour ABPM. Unit of Measure: Number of blood pressure inflations |
Month 16 through Month 20.
|
合作者和调查者
调查人员
- 首席研究员:Yook Chin Chia、Sunway University
- 首席研究员:Yee How Say, BSc PhD、Sunway University
出版物和有用的链接
一般刊物
- McEvoy JW, McCarthy CP, Bruno RM, Brouwers S, Canavan MD, Ceconi C, Christodorescu RM, Daskalopoulou SS, Ferro CJ, Gerdts E, Hanssen H, Harris J, Lauder L, McManus RJ, Molloy GJ, Rahimi K, Regitz-Zagrosek V, Rossi GP, Sandset EC, Scheenaerts B, Staessen JA, Uchmanowicz I, Volterrani M, Touyz RM; ESC Scientific Document Group. 2024 ESC Guidelines for the management of elevated blood pressure and hypertension. Eur Heart J. 2024 Oct 7;45(38):3912-4018. doi: 10.1093/eurheartj/ehae178. No abstract available.
- Stergiou GS, Menti A, Mariglis D, Kollias A. The quest for accurate wearable blood pressure monitors. Hypertens Res. 2026 Mar;49(3):1025-1029. doi: 10.1038/s41440-025-02410-w. Epub 2025 Nov 5.
- Chia J, Bhatia KS, Mihailidou AS, Kanagaratnam LB. The Role of Ambulatory Blood Pressure Monitoring in Current Clinical Practice. Heart Lung Circ. 2022 Oct;31(10):1333-1340. doi: 10.1016/j.hlc.2022.06.670. Epub 2022 Aug 4.
- Kawaja A, Moosa AS, Lee EKP, Phoon IKY, Ang ATW, Chang ZY, Lim ACA, Yap J, Huang W, Ng DX, Sng MY, Loh HY, Ng CJ. Comparison of patients' acceptance of cuff-based vs wearable 24-hour ambulatory blood pressure monitoring devices: A multi-method study. PLoS One. 2025 Nov 17;20(11):e0336961. doi: 10.1371/journal.pone.0336961. eCollection 2025.
- Stergiou GS, Menti A, Asayama K, Kollias A, Parati G, Schutte AE, Wang J, Aristizabal D, Asmar R, Bilo G, Bothe T, Clark C, Kario K, Li Y, McManus R, Ohkubo T, Palatini P, Picone D, Shimbo D, Tsaganos T, De La Sierra A. Wearable wrist-watch type cuff oscillometric blood pressure monitors: consensus statement by the European Society of Hypertension Working Group on blood pressure monitoring. J Hypertens. 2026 Oct 1;44(10):1721-1731. doi: 10.1097/HJH.0000000000004415. Epub 2026 Aug 26.
- Stergiou GS, Menti A, Asmar R, Bilo G, Ntineri A, Palatini P, Parati G, Shennan A, Wang JG, Sierra A, Kollias A, Schutte AE; STRIDE BP and European Society of Hypertension Working Group on Blood Pressure Monitoring.. Updated practical guide for validation studies of automated cuff blood pressure monitors according to the Universal Standard by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018) and Amendments 1:2020 and 2:2024. J Hypertens. 2026 Oct 20. doi: 10.1097/HJH.0000000000004448. Online ahead of print.
- Stergiou GS, Kario K, Aristizabal D, Beheiry H, Cho MC, Colombari DSA, Itoh H, Jafar TH, Jones E, Li Y, Romero CA, Steckelings UM, Williams B, Borghi C, Charchar F, Doku A, Dorobantu M, Flynn J, Khan N, Kollias A, Kreutz R, Mahmud A, Mancia G, Parati G, Poulter N, Reckelhoff J, Renna N, Shehab A, Schlaich M, Schutte A, Teo BW, Tomaszewski M, Touyz R, Wang J, Weber T, Whelton P, Zhang Y. 2026 International Society of Hypertension Global Hypertension Practice Guidelines. J Hypertens. 2026 Oct 20. doi: 10.1097/HJH.0000000000004451. Online ahead of print.
研究记录日期
研究主要日期
学习开始 (估计的)
初级完成 (估计的)
研究完成 (估计的)
研究注册日期
首次提交
首先提交符合 QC 标准的
首次发布 (实际的)
研究记录更新
最后更新发布 (实际的)
上次提交的符合 QC 标准的更新
最后验证
更多信息
与本研究相关的术语
关键字
其他研究编号
- Huawei-SunU-MinDisABPMStudy
计划个人参与者数据 (IPD)
计划共享个人参与者数据 (IPD)?
IPD 计划说明
De-identified individual participant data (IPD) collected during this study will be shared with qualified external researchers. Specific shared data elements include:
- Demographics & Baseline Characteristics: De-identified participant age, sex, body mass index, and baseline clinical blood pressure measurements.
- Wearable Sensor Data: Continuous physiological streams captured via the Huawei Research Platform, including raw photoplethysmography (PPG) signals, heart rate, heart rate variability, physical activity metrics, and sleep tracking parameters.
- Reference ABPM Data: Matched 24-hour, daytime, and night-time blood pressure and heart rate values, along with derived nocturnal dipping indices, obtained from clinical cuff-based ABPM and reduced-frequency Huawei smartwatch schedules.
- User Experience Data: De-identified survey scores on device comfort, ease of use, sleep disturbance, and monitoring preference.
IPD 共享时间框架
Access Start Date: 6 months after the publication of the primary study results in a peer-reviewed journal.
Access End Date: 5 years after initial publication.
IPD 共享访问标准
Method of Access: Data will be made available via a secure, managed data repository (or directly from the Principal Investigator upon formal request).
Qualified Requesters: Scientific researchers affiliated with recognized academic, clinical, or research institutions.
Request Requirements: Requesters must submit a methodologically sound research proposal and a clear scientific rationale for secondary analysis.
Data Use Agreement (DUA): Execution of a formal Data Use Agreement (DUA) is required to ensure data protection, mandate non-re-identification, and enforce secure handling.
Compliance: All shared files will be fully de-identified in compliance with the Personal Data Protection Act (PDPA) Malaysia and institutional Ethics Committee guidelines.