此页面是自动翻译的,不保证翻译的准确性。请参阅 英文版 对于源文本。

Evaluation of the Temperature Difference of Medical Gases Between the Heated Inspiratory Limb and the Tracheal Carina in Patients Undergoing Invasive Mechanical Ventilation With Active Humidification: A Crossover Clinical Study.

2026年11月13日 更新者:Riccardo Re、Azienda Sanitaria Universitaria Integrata del Trentino

The aim of this study is to assess the actual temperature drop of medical gases from the distal end of the heated ventilator circuit to the tracheal carina during invasive mechanical ventilation with active humidification. This analysis reflects the investigators' intention to observe the temperature drop under real-world clinical practice conditions, while assessing potential interactions with patient body temperature, ambient temperature, and the size of unheated ventilatory connectors. These findings may provide insight into the appropriate temperature and humidity targets, thereby further optimizing active humidification during invasive mechanical ventilation.

研究概览

详细说明

Study setting: Patients undergoing invasive mechanical ventilation via orotracheal tube, nasotracheal tube, or tracheostomy cannula.

Research Hypothesis: The temperature drop between the distal end of the heated inspiratory circuit and the distal end of the tracheal tube does not exceed 2°C; therefore, heating medical gases above 37°C is not necessary.

Study Objective: 1) To assess the temperature difference between the distal end of the heated inspiratory limb and the patient's tracheal carina. 2) To identify variables that may influence the temperature drop, including connector size (dead space), the presence of a mainstream end-tidal CO₂ sensor, type, dimension and depth of tracheal tube (orotracheal, nasotracheal, tracheostomy), ventilation mode and minute volume, patient body temperature, and ambient temperature.

Study Design: Crossover clincal study Inclusion Criteria: Adult patients admitted to the Intensive Care Unit 1 of Santa Chiara Hospital in Trento, who are receiving invasive mechanical ventilation with active humidification via a tracheal tube and who already have invasive body temperature monitoring in place.

Exclusion Criteria: Patients ventilated with tracheal tubes smaller than size 7 or patients ventilated using double-lumen tubes for selective bronchial intubation.

Protocol: The study protocol involves measuring, in all patients meeting the inclusion criteria, the temperature at the end of the heated ventilator circuit ("Tpatient") and the air temperature at the level of the tracheal carina. Tpatient will be recorded as the direct reading from the humidifier display. Tracheal temperature will be measured by inserting a sterile temperature probe through the ventilator circuit. To obtain an accurate measurement of air temperature without the probe contacting the tracheal mucosa, a graduated probe will be used, calibrated according to the correct insertion depth corresponding to the tracheal tube in use. The temperature recorded by the probe will then be displayed via a dedicated electrical cable on the bedside multiparameter monitor. The heat humidifier used in the study will be the device currently available in ICU 1 of Santa Chiara Hospital in Trento, namely the Dräger Aquapor H300® humidifier. The device will be set in FREE mode, allowing the investigators to freely adjust temperature settings according to the study protocol. For each enrolled patient, two sets of measurements will be performed using different humidification settings. The first set will use a Tpatient of 37°C with a -2°C gradient (Tpatient 37°C - chamber temperature 35°C), while the second set will use a Tpatient of 40°C with a -3°C gradient (Tpatient 40°C - chamber temperature 37°C). The order in which the first or second measurement set is performed will be randomly assigned using a web-based randomization tool (random.org). Before temperature measurements are performed, correct attainment of the target humidifier temperature will be verified. For each observation, the following variables will also be recorded: age and male patient core body temperature, type of airway device, tube size, depth of tube placement, volume of unheated connectors, presence of a mainstream end-tidal CO₂ sensor, ventilation mode, minute ventilation at the time of measurement, and ambient temperature. At the end of the measurements, the probe will be removed and placed on the patient's pillow to measure the ambient temperature in the immediate vicinity of the patient.

Timeframe: The study aims to enroll 50 patients within a maximum period of one year from the first measurement.

Statistical Analysis: During the statistical analysis, the mean, standard deviation, median, and range of the temperature differences between Tpatient and the air temperature at the tracheal carina will be calculated for the two humidification settings (37°C and 40°C). These values will then be compared using the Bland-Altman method. In addition, potential correlations between temperature variations and the study's objective variables will be analysed, including core body temperature, type and size of the airway device, depth of tube placement, volume of unheated connectors, presence of a mainstream end-tidal CO₂ sensor, minute ventilation at the time of measurement, and ambient temperature.

Expected Results: The temperature drop induced by unheated ventilator connectors (catheter mount and tracheal tube/cannula) does not exceed 2°C and is influenced by both ambient temperature and patient body temperature. Accordingly, a heated inspiratory circuit temperature setting above 37°C is not necessary.

研究类型

介入性

注册 (估计的)

50

阶段

  • 不适用

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习联系方式

学习地点

    • Italy
      • Trento、Italy、意大利、38100
        • 招聘中
        • Santa Chiara Hospital of Trento - ASUIT
        • 接触:

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

18年 及以上 (成人、年长者)

接受健康志愿者

不

有资格学习的性别

全部

描述

Inclusion Criteria: Adult patients admitted to the Intensive Care Unit 1 of Santa Chiara Hospital in Trento, who are receiving invasive mechanical ventilation with active humidification via a tracheal tube and who already have invasive body temperature monitoring in place.

Exclusion Criteria: Patients ventilated with tracheal tubes smaller than size 7 or patients ventilated using double-lumen tubes for selective bronchial intubation

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 主要用途:治疗
  • 分配:随机化
  • 介入模型:交叉作业
  • 屏蔽:无(打开标签)

武器和干预

参与者组/臂
干预/治疗
实验性的:Temperature drop 37°C

Measure of the temperarure drop in two different humidification setting: 37°C - 40°C

Evaluation of the temperature drop from the end of the heated ventilatory tube to the tracheal carina

实验性的:Temperature drop 40°C

Evaluation of the temperature drop from the end of the heated ventilatory tube to the tracheal carina

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Temperature drop
大体时间:50 patients or 1 year

Measure the real temperature drop from the end of the heated respiratory tube and the tracheal carina in the patient with mechanical invasive ventilation and active humidification.

50 patients or 1 year

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (实际的)

2026年6月22日

初级完成 (估计的)

2027年6月22日

研究完成 (估计的)

2027年6月22日

研究注册日期

首次提交

2026年11月13日

首先提交符合 QC 标准的

2026年11月13日

首次发布 (实际的)

2026年11月18日

研究记录更新

最后更新发布 (实际的)

2026年11月18日

上次提交的符合 QC 标准的更新

2026年11月13日

最后验证

2026年6月1日

更多信息

与本研究相关的术语

其他研究编号

  • CARINA TEMP

药物和器械信息、研究文件

研究美国 FDA 监管的药品

不

研究美国 FDA 监管的设备产品

不

此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.

订阅