Evaluation of the Temperature Difference of Medical Gases Between the Heated Inspiratory Limb and the Tracheal Carina in Patients Undergoing Invasive Mechanical Ventilation With Active Humidification: A Crossover Clinical Study.

13 november 2026 bijgewerkt door: Riccardo Re, Azienda Sanitaria Universitaria Integrata del Trentino

The aim of this study is to assess the actual temperature drop of medical gases from the distal end of the heated ventilator circuit to the tracheal carina during invasive mechanical ventilation with active humidification. This analysis reflects the investigators' intention to observe the temperature drop under real-world clinical practice conditions, while assessing potential interactions with patient body temperature, ambient temperature, and the size of unheated ventilatory connectors. These findings may provide insight into the appropriate temperature and humidity targets, thereby further optimizing active humidification during invasive mechanical ventilation.

Studie Overzicht

Gedetailleerde beschrijving

Study setting: Patients undergoing invasive mechanical ventilation via orotracheal tube, nasotracheal tube, or tracheostomy cannula.

Research Hypothesis: The temperature drop between the distal end of the heated inspiratory circuit and the distal end of the tracheal tube does not exceed 2°C; therefore, heating medical gases above 37°C is not necessary.

Study Objective: 1) To assess the temperature difference between the distal end of the heated inspiratory limb and the patient's tracheal carina. 2) To identify variables that may influence the temperature drop, including connector size (dead space), the presence of a mainstream end-tidal CO₂ sensor, type, dimension and depth of tracheal tube (orotracheal, nasotracheal, tracheostomy), ventilation mode and minute volume, patient body temperature, and ambient temperature.

Study Design: Crossover clincal study Inclusion Criteria: Adult patients admitted to the Intensive Care Unit 1 of Santa Chiara Hospital in Trento, who are receiving invasive mechanical ventilation with active humidification via a tracheal tube and who already have invasive body temperature monitoring in place.

Exclusion Criteria: Patients ventilated with tracheal tubes smaller than size 7 or patients ventilated using double-lumen tubes for selective bronchial intubation.

Protocol: The study protocol involves measuring, in all patients meeting the inclusion criteria, the temperature at the end of the heated ventilator circuit ("Tpatient") and the air temperature at the level of the tracheal carina. Tpatient will be recorded as the direct reading from the humidifier display. Tracheal temperature will be measured by inserting a sterile temperature probe through the ventilator circuit. To obtain an accurate measurement of air temperature without the probe contacting the tracheal mucosa, a graduated probe will be used, calibrated according to the correct insertion depth corresponding to the tracheal tube in use. The temperature recorded by the probe will then be displayed via a dedicated electrical cable on the bedside multiparameter monitor. The heat humidifier used in the study will be the device currently available in ICU 1 of Santa Chiara Hospital in Trento, namely the Dräger Aquapor H300® humidifier. The device will be set in FREE mode, allowing the investigators to freely adjust temperature settings according to the study protocol. For each enrolled patient, two sets of measurements will be performed using different humidification settings. The first set will use a Tpatient of 37°C with a -2°C gradient (Tpatient 37°C - chamber temperature 35°C), while the second set will use a Tpatient of 40°C with a -3°C gradient (Tpatient 40°C - chamber temperature 37°C). The order in which the first or second measurement set is performed will be randomly assigned using a web-based randomization tool (random.org). Before temperature measurements are performed, correct attainment of the target humidifier temperature will be verified. For each observation, the following variables will also be recorded: age and male patient core body temperature, type of airway device, tube size, depth of tube placement, volume of unheated connectors, presence of a mainstream end-tidal CO₂ sensor, ventilation mode, minute ventilation at the time of measurement, and ambient temperature. At the end of the measurements, the probe will be removed and placed on the patient's pillow to measure the ambient temperature in the immediate vicinity of the patient.

Timeframe: The study aims to enroll 50 patients within a maximum period of one year from the first measurement.

Statistical Analysis: During the statistical analysis, the mean, standard deviation, median, and range of the temperature differences between Tpatient and the air temperature at the tracheal carina will be calculated for the two humidification settings (37°C and 40°C). These values will then be compared using the Bland-Altman method. In addition, potential correlations between temperature variations and the study's objective variables will be analysed, including core body temperature, type and size of the airway device, depth of tube placement, volume of unheated connectors, presence of a mainstream end-tidal CO₂ sensor, minute ventilation at the time of measurement, and ambient temperature.

Expected Results: The temperature drop induced by unheated ventilator connectors (catheter mount and tracheal tube/cannula) does not exceed 2°C and is influenced by both ambient temperature and patient body temperature. Accordingly, a heated inspiratory circuit temperature setting above 37°C is not necessary.

Studietype

Ingrijpend

Inschrijving (Geschat)

50

Fase

  • Niet toepasbaar

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studiecontact

Studie Locaties

    • Italy
      • Trento, Italy, Italië, 38100
        • Werving
        • Santa Chiara Hospital of Trento - ASUIT
        • Contact:

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

18 jaar en ouder (Volwassen, Oudere volwassene)

Accepteert gezonde vrijwilligers

Nee

Geslachten die in aanmerking komen voor studie

Allemaal

Beschrijving

Inclusion Criteria: Adult patients admitted to the Intensive Care Unit 1 of Santa Chiara Hospital in Trento, who are receiving invasive mechanical ventilation with active humidification via a tracheal tube and who already have invasive body temperature monitoring in place.

Exclusion Criteria: Patients ventilated with tracheal tubes smaller than size 7 or patients ventilated using double-lumen tubes for selective bronchial intubation

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Primair doel: Behandeling
  • Toewijzing: Gerandomiseerd
  • Interventioneel model: Crossover-opdracht
  • Masker: Geen (open label)

Wapens en interventies

Deelnemersgroep / Arm
Interventie / Behandeling
Experimenteel: Temperature drop 37°C

Measure of the temperarure drop in two different humidification setting: 37°C - 40°C

Evaluation of the temperature drop from the end of the heated ventilatory tube to the tracheal carina

Experimenteel: Temperature drop 40°C

Evaluation of the temperature drop from the end of the heated ventilatory tube to the tracheal carina

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Temperature drop
Tijdsspanne: 50 patients or 1 year

Measure the real temperature drop from the end of the heated respiratory tube and the tracheal carina in the patient with mechanical invasive ventilation and active humidification.

50 patients or 1 year

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Werkelijk)

22 juni 2026

Primaire voltooiing (Geschat)

22 juni 2027

Studie voltooiing (Geschat)

22 juni 2027

Studieregistratiedata

Eerst ingediend

13 november 2026

Eerst ingediend dat voldeed aan de QC-criteria

13 november 2026

Eerst geplaatst (Werkelijk)

18 november 2026

Updates van studierecords

Laatste update geplaatst (Werkelijk)

18 november 2026

Laatste update ingediend die voldeed aan QC-criteria

13 november 2026

Laatst geverifieerd

1 juni 2026

Meer informatie

Termen gerelateerd aan deze studie

Andere studie-ID-nummers

  • CARINA TEMP

Informatie over medicijnen en apparaten, studiedocumenten

Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel

Nee

Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct

Nee

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .

Abonneren