此页面是自动翻译的,不保证翻译的准确性。请参阅 英文版 对于源文本。

Esophageal Pressure and Cardiac Output

2026年11月20日 更新者:Jun-Young Park、Asan Medical Center

Dynamic Concordance Analysis Between Cardiac Oscillation Waveform-derived Esophageal Pressure and Cardiac Output Parameters: A Prospective Observational Study

This prospective observational study aims to determine whether a pressure signal recorded through a standard esophageal temperature probe can track changes in cardiac output during general anesthesia.

During surgery, anesthesiologists need to keep the heart adequately filled (preload) to maintain blood flow to vital organs. Central venous pressure, the traditional bedside measure, reflects intravascular volume poorly. Pulmonary capillary wedge pressure, which estimates left atrial pressure, requires a pulmonary artery catheter, an invasive device that is now rarely used. Dynamic indices such as stroke volume variation and pulse pressure variation are reliable only under strict conditions, such as controlled ventilation with an adequate tidal volume, absence of spontaneous breathing, and a regular heart rhythm.

The esophagus lies directly behind the left atrium. When an esophageal probe is filled with saline and connected to a pressure transducer, small pulsatile "cardiac oscillation" waves can be recorded at a certain depth. The investigators hypothesize that this signal reflects pressure changes transmitted from the left atrium and therefore carries real-time information about cardiac filling and output.

Forty adult patients (aged 20 to 79 years, ASA physical status I to III) scheduled for surgery under general anesthesia with invasive arterial pressure and arterial pressure-based cardiac output monitoring will be enrolled after written informed consent. After induction of anesthesia and tracheal intubation, an 18 Fr esophageal temperature probe, which is routinely used for temperature monitoring, will be filled with 7 to 8 mL of sterile saline, connected to a pressure transducer, and advanced until a clear and stable cardiac oscillation waveform with an amplitude of at least 0.5 to 1.0 mmHg is identified. Anesthesia and surgical care will follow standard practice, and the study will not change clinical management. Arterial pressure, heart rate, esophageal pressure, cardiac output, cardiac index, stroke volume, stroke volume index, stroke volume variation, pulse pressure variation, airway pressure, and positive end-expiratory pressure will be recorded continuously and analyzed around hemodynamic events such as fluid administration, bleeding, and vascular clamping.

The primary outcome is the concordance rate between the percentage changes in stroke volume index and in cardiac oscillation-derived esophageal pressure during these events, assessed by four-quadrant plot analysis after excluding small changes (less than 15%). Secondary outcomes are trending ability assessed by polar plot analysis (angular error), agreement between the percentage changes of the two measures assessed by Bland-Altman analysis (percentage error), and the ability of esophageal pressure changes to predict fluid responsiveness (an increase in stroke volume of more than 15%) assessed by receiver operating characteristic curve analysis.

If the esophageal pressure signal reliably tracks changes in cardiac output, it could offer a simple hemodynamic monitoring method that uses equipment already in routine use during general anesthesia, without additional invasive devices.

研究概览

地位

尚未招聘

研究类型

观察性的

注册 (估计的)

40

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习联系方式

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

20年 至 79年 (成人、年长者)

接受健康志愿者

不

有资格学习的性别

全部

取样方法

非概率样本

研究人群

Adult patients aged 20 to 79 years with ASA physical status I to III who are scheduled for surgery under general anesthesia with invasive arterial pressure and arterial pressure-based cardiac output monitoring as part of routine clinical care at Asan Medical Center, Seoul, Republic of Korea. Eligible patients will be approached consecutively before surgery, and those who provide written informed consent will be enrolled.

描述

Inclusion Criteria:

  • Adult patients scheduled for surgery under general anesthesia with invasive arterial pressure and arterial pressure-based cardiac output monitoring
  • Age 20 years or older and younger than 80 years
  • American Society of Anesthesiologists (ASA) physical status I to III
  • Voluntary written informed consent to participate in the study

Exclusion Criteria:

  • Esophageal disease, such as esophageal varices, esophageal diverticulum, esophageal stricture, or esophageal cancer
  • Anatomical variation in the relative position of the esophagus and heart, such as severe chest wall deformity
  • Significant cardiac functional or structural abnormality on echocardiography, including left ventricular ejection fraction less than 40%, moderate or greater mitral or aortic valve disease, or intracardiac shunt
  • Persistent atrial fibrillation or frequent premature ventricular contractions
  • Need for repeated changes in ventilator settings, such as positive end-expiratory pressure or tidal volume, during surgery
  • Surgery requiring the lateral decubitus or prone position
  • Need for insertion of another device into the esophagus, such as a transesophageal echocardiography probe
  • Body mass index of 40 kg/m2 or greater, or less than 17.5 kg/m2
  • Refusal to participate

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 观测模型:COHORT
  • 时间观点:PROSPECTIVE

队列和干预

团体/队列
Adults under general anesthesia with cardiac output monitoring

Adult patients aged 20 to 79 years (ASA physical status I to III) undergoing surgery under general anesthesia with invasive arterial pressure and arterial pressure-based cardiac output monitoring as part of routine care. After tracheal intubation, an 18 Fr esophageal temperature probe routinely used for temperature monitoring is filled with 7 to 8 mL of sterile saline, connected to a pressure transducer, and positioned where a stable cardiac oscillation waveform (amplitude of at least 0.5 to 1.0 mmHg) is observed. Cardiac oscillation-derived esophageal pressure and hemodynamic variables (stroke volume index, cardiac index, arterial pressure, stroke volume variation, pulse pressure variation) are recorded continuously during hemodynamic events such as fluid administration, bleeding, and vascular clamping. Anesthetic and surgical management follow standard practice and are not altered by the study.

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Concordance rate between changes in stroke volume index and cardiac oscillation-derived esophageal pressure
大体时间:Intraoperative: immediately before and 1 minute after the end of each hemodynamic event, from probe placement after tracheal intubation to the end of surgery

For each intraoperative hemodynamic event (e.g., fluid administration, bleeding, vascular clamping), the percentage change from the pre-event value to the post-event value is calculated for stroke volume index (SVI), measured by arterial pressure-based cardiac output monitoring, and for cardiac oscillation-derived esophageal pressure, measured through a saline-filled esophageal temperature probe. Paired changes are plotted on a four-quadrant plot. After excluding pairs with a change of less than 15% (exclusion zone), the concordance rate is calculated as the percentage of paired changes in which both measures change in the same direction (both increase or both decrease).

Intraoperative: immediately before and 1 minute after the end of each hemodynamic event, from probe placement after tracheal intubation to the end of surgery

次要结果测量

结果测量
措施说明
大体时间
Angular error of trending ability assessed by polar plot analysis
大体时间:Intraoperative: immediately before and 1 minute after the end of each hemodynamic event, from probe placement after tracheal intubation to the end of surgery

Paired percentage changes in stroke volume index and cardiac oscillation-derived esophageal pressure during hemodynamic events are converted into polar coordinates. The mean angular bias (degrees) and the radial limits of agreement (degrees) are calculated to quantify the trending ability of esophageal pressure relative to stroke volume index. Pairs within the 15% exclusion zone are excluded.

Intraoperative: immediately before and 1 minute after the end of each hemodynamic event, from probe placement after tracheal intubation to the end of surgery
Percentage error between changes in stroke volume index and esophageal pressure assessed by Bland-Altman analysis
大体时间:Intraoperative: immediately before and 1 minute after the end of each hemodynamic event, from probe placement after tracheal intubation to the end of surgery

Because stroke volume index and esophageal pressure have different units, agreement is assessed using the percentage changes of each measure during hemodynamic events. Bland-Altman analysis is performed to calculate the bias and 95% limits of agreement between the paired percentage changes, and the percentage error is derived from these values.

Intraoperative: immediately before and 1 minute after the end of each hemodynamic event, from probe placement after tracheal intubation to the end of surgery
Area under the receiver operating characteristic curve (AUC) of esophageal pressure change for predicting fluid responsiveness
大体时间:Intraoperative: immediately before and 1 minute after the end of each fluid administration, from probe placement after tracheal intubation to the end of surgery

Fluid responsiveness is defined as an increase in stroke volume of more than 15% after fluid administration. The ability of the percentage change in cardiac oscillation-derived esophageal pressure to discriminate fluid responders from non-responders is assessed by receiver operating characteristic curve analysis. The AUC with its 95% confidence interval and the optimal cut-off value (Youden index) are reported.

Intraoperative: immediately before and 1 minute after the end of each fluid administration, from probe placement after tracheal intubation to the end of surgery

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (估计的)

2026年12月12日

初级完成 (估计的)

2027年7月1日

研究完成 (估计的)

2027年7月1日

研究注册日期

首次提交

2026年11月20日

首先提交符合 QC 标准的

2026年11月20日

首次发布 (实际的)

2026年11月27日

研究记录更新

最后更新发布 (实际的)

2026年11月27日

上次提交的符合 QC 标准的更新

2026年11月20日

最后验证

2026年11月1日

更多信息

与本研究相关的术语

其他研究编号

  • Eso_pr_CO

计划个人参与者数据 (IPD)

计划共享个人参与者数据 (IPD)?

不

药物和器械信息、研究文件

研究美国 FDA 监管的药品

不

研究美国 FDA 监管的设备产品

不

此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.

订阅