In-line Mechanical Insufflation-Exsufflation in the Management of Ventilated Patients

May 8, 2022 updated by: Eliezer be'eri, Alyn Pediatric & Adolescent Rehabilitation Hospital

In-line Mechanical Insufflation-Exsufflation as an Alternative to Invasive Suction for Secretion Management in Ventilated Patients

Catheter suction (CS), the standard method for airway secretion management during mechanical ventilation, is invasive and has significant hemodynamic and traumatic side effects. In-line mechanical insufflation-exsufflation (IL-MIE) is a new, noninvasive technology that clears secretions by cough-simulation, without interrupting ongoing ventilation. It is not known whether IL-MIE can be safely and effectively used as an alternative to CS in ventilated patients. Methods: A randomized, controlled, non-inferiority study comparing a standard protocol of CS, with automatic IL-MIE (CoughSync, Ruxin Medical Systems, Beijing) performed every 30 minutes, with CS added only if needed, in post-operative ventilated patients.

Study Overview

Detailed Description

The study was designed as a randomized, open-label, parallel, non-inferiority, controlled trial. The study was carried out in two Intensive Care Units - the Cardiac Surgery ICU at Anzhen Hospital, and the Critical Care Medicine ICU at Tian Tan Hospital - both in Beijing, China, over a 36 month period between July 2015 and July 2018. The study group included patients aged 18 to 75 years undergoing mechanical ventilation during the immediate recovery period following a cardiac or neuro-surgical procedure.

Subjects were randomized to either a control group, managed for 8 hours with CS whenever the subject showed signs of airway secretion accumulation, as per standard clinical practice for those ICU's, or a study group, managed for 8 hours with automatic IL-MIE treatments (CoughSync, Ruxin Medical Systems Company Ltd, Beijing, China) performed automatically every 30 minutes, and with CS performed only if signs of airway secretion accumulation manifested despite ongoing IL-MIE. IL-MIE was performed using standard IL-MIE parameters (exsufflation pressure = -60 cm H2O, with flutter, 10 coughs/treatment).

Before each CS treatment, 100% oxygen was administered for one minute, regardless of the patient's baseline oxygen requirement, as per standard operating protocol in that ICU. Before IL-MIE treatments, no additional oxygen was administered beyond the patient's baseline oxygen requirement.

Pharmacological management in both groups included analgesia and sedation as routinely used postoperatively in those ICU's.

Demographic information and vital signs were recorded for all subjects. Arterial partial pressure of oxygen (PaO2), arterial oxygen saturation (SaO2 ) and oxygenation index (PaO2/FiO2) were defined as the primary end points of the study, and arterial partial pressure of carbon dioxide (PaCO2), pulseoximetry (SpO2), heart rate (HR), and ventilator parameters (inspired oxygen, tidal volume [Vt], peak inspiratory pressure [PIP], airway plateau pressure [Pplat], and PEEP) as secondary end points.

All primary and secondary end-point data were recorded at baseline (2 minutes prior to starting the trial), and at 5 minutes, 4 hours and 8 hours after commencement of the trial. The number of CS treatments performed on each subject was recorded throughout the trial. Follow up for adverse events was performed during, and 48 hours after completion of, the trial.

Statistical Analysis:

The minimum sample size required to demonstrate non-inferiority, was calculated to be 49 in each cohort, or a total of 98 subjects in total.

Collected data were analyzed with a mixed model with repeated measures (MMRM) considering all observations (2 minutes before commencement of the trial protocol, and at 5 minutes, 4 hours and 8 hours thereafter) and accounting for the baseline value of oxygenation. For the indices of oxygenation derived from blood gas measurements (PaO2, SaO2 , and oxygenation index), non-inferiority was evaluated by comparison to the two-sided 95% confidence interval of the treatment effect in the MMRM model, with non-inferiority between the IL-MIE and control cohorts established if the lower limit of the 95% confidence interval for the intergroup difference in least squares mean for a measured index was found to be higher than the pre-determined non-inferiority margin for that index. For comparison of other quantitative data between groups, a two-sample t-test or Wilcoxon rank sum test was used, based on the data distribution. For between-group comparison of subjects with significant adverse events, a chi-square test was used. The number of CS treatments performed in each cohort was analyzed post-hoc as an exploratory analysis.

Study Type

Interventional

Enrollment (Actual)

120

Phase

  • Not Applicable

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years to 75 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • patients undergoing mechanical ventilation during the immediate recovery period following a cardiac or neuro-surgical procedure
  • Ventilation expected to last more than 8 hours

Exclusion Criteria:

  • acute spinal cord shock
  • recent airway trauma or surgery
  • cardiogenic pulmonary edema or ARDS necessitating ventilation with a peak end expiratory pressure (PEEP) equal to or greater than 6 cmH2O
  • pneumothorax
  • hemoptysis
  • severe ischemic heart disease
  • lung tumors
  • pulmonary Tuberculosis
  • history of lung transplantation
  • pregnant or breastfeeding women

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Active Comparator: Catheter Suction
Patients in this arm were managed for 8 hours with Catheter Suction whenever the subject showed signs of airway secretion accumulation, as per standard clinical practice for those ICU's
catheter suction is the standard, routine method for clearing secretions from the airway of a ventilated subject, by means of inserting a catheter into the endotracheal tube.
Experimental: Mechanical Inexsufflation
Patients in this are were managed for 8 hours with automatic inexsufflation treatments (CoughSync, Ruxin Medical Systems Company Ltd, Beijing, China) performed automatically every 30 minutes, and with Catheter Suction performed only if signs of airway secretion accumulation manifested
In-line mechanical inexsufflation (IL-MIE) is a new method for performing MIE in intubated patients, which overcomes the drawbacks of MIE for ICU use . IL-MIE devices are integrated in-line with the patient's ventilator circuit, and do not themselves perform insufflations. Rather, the regular inspiration provided by the ventilator serves as the insufflation phase of each simulated cough, and the IL-MIE device performs only exsufflation, timing the onset of each exsufflation to the beginning of passive exhalation. Figure 1 demonstrates the setup and mode of operation of an IL-MIE device. The concept of IL-MIE was first developed by one of the authors (EB) in the Department of Respiratory Rehabilitation of ALYN Hospital in Jerusalem, Israel.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
PaO2
Time Frame: change from baseline to 5 minutes
Arterial partial pressure of oxygen
change from baseline to 5 minutes
PaO2
Time Frame: change from baseline to 4 hours
Arterial partial pressure of oxygen
change from baseline to 4 hours
PaO2
Time Frame: change from baseline to 8 hours
Arterial partial pressure of oxygen
change from baseline to 8 hours
SaO2
Time Frame: change from baseline to 5 minutes
arterial oxygen saturation
change from baseline to 5 minutes
SaO2
Time Frame: change from baseline to 4 hours
arterial oxygen saturation
change from baseline to 4 hours
SaO2
Time Frame: change from baseline to 8 hours
arterial oxygen saturation
change from baseline to 8 hours
PaO2/FiO2
Time Frame: change from baseline to 5 minutes
oxygenation index - the ratio of blood oxygen to the percentage of inspired oxygen
change from baseline to 5 minutes
PaO2/FiO2
Time Frame: change from baseline to 4 hours
oxygenation index - the ratio of blood oxygen to the percentage of inspired oxygen
change from baseline to 4 hours
PaO2/FiO2
Time Frame: change from baseline to 8 hours
oxygenation index - the ratio of blood oxygen to the percentage of inspired oxygen
change from baseline to 8 hours

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
PaCO2
Time Frame: change from baseline to 5 minutes
arterial partial pressure of carbon dioxide
change from baseline to 5 minutes
PaCO2
Time Frame: change from baseline to 4 hours
arterial partial pressure of carbon dioxide
change from baseline to 4 hours
PaCO2
Time Frame: change from baseline to 8 hours
arterial partial pressure of carbon dioxide
change from baseline to 8 hours
SpO2
Time Frame: change from baseline to 5 minutes
blood oxygen saturation
change from baseline to 5 minutes
SpO2
Time Frame: change from baseline to 4 hours
blood oxygen saturation
change from baseline to 4 hours
SpO2
Time Frame: change from baseline to 8 hours
blood oxygen saturation
change from baseline to 8 hours
heart rate
Time Frame: change from baseline to 5 minutes
heart rate
change from baseline to 5 minutes
heart rate
Time Frame: change from baseline to 4 hours
heart rate
change from baseline to 4 hours
heart rate
Time Frame: change from baseline to 8 hours
heart rate
change from baseline to 8 hours
inspired oxygen
Time Frame: change from baseline to 5 minutes
percent fraction inspired oxygen
change from baseline to 5 minutes
inspired oxygen
Time Frame: change from baseline to 4 hours
percent fraction inspired oxygen
change from baseline to 4 hours
inspired oxygen
Time Frame: change from baseline to 8 hours
percent fraction inspired oxygen
change from baseline to 8 hours
ventilator tidal volume
Time Frame: change from baseline to 5 minutes
tidal volume in ml
change from baseline to 5 minutes
ventilator tidal volume
Time Frame: change from baseline to 4 hours
tidal volume in ml
change from baseline to 4 hours
ventilator tidal volume
Time Frame: change from baseline to 8 hours
tidal volume in ml
change from baseline to 8 hours
ventilator peak inspiratory pressure
Time Frame: change from baseline to 5 minutes
peak airway pressure in CMH2O
change from baseline to 5 minutes
ventilator peak inspiratory pressure
Time Frame: change from baseline to 4 hours
peak airway pressure in CMH2O
change from baseline to 4 hours
ventilator peak inspiratory pressure
Time Frame: change from baseline to 8 hours
peak airway pressure in CMH2O
change from baseline to 8 hours
ventilator peak airway plateau pressure
Time Frame: change from baseline to 5 minutes
airway plateau pressure in CMH2O
change from baseline to 5 minutes
ventilator peak airway plateau pressure
Time Frame: change from baseline to 4 hours
airway plateau pressure in CMH2O
change from baseline to 4 hours
ventilator peak airway plateau pressure
Time Frame: change from baseline to 8 hours
airway plateau pressure in CMH2O
change from baseline to 8 hours
ventilator peak end expiratory pressure
Time Frame: change from baseline to 5 minutes
airway end expiratory pressure in CMH2O
change from baseline to 5 minutes
ventilator peak end expiratory pressure
Time Frame: change from baseline to 4 hours
airway end expiratory pressure in CMH2O
change from baseline to 4 hours
ventilator peak end expiratory pressure
Time Frame: change from baseline to 8 hours
airway end expiratory pressure in CMH2O
change from baseline to 8 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

July 1, 2015

Primary Completion (Actual)

July 1, 2018

Study Completion (Actual)

July 1, 2018

Study Registration Dates

First Submitted

February 14, 2022

First Submitted That Met QC Criteria

May 4, 2022

First Posted (Actual)

May 9, 2022

Study Record Updates

Last Update Posted (Actual)

May 12, 2022

Last Update Submitted That Met QC Criteria

May 8, 2022

Last Verified

May 1, 2022

More Information

Terms related to this study

Other Study ID Numbers

  • 047-21

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.