A Study Comparing Two Mandibular Advancement Devices in the Treatment of Sleep Apnea (MRA OSA)
A Randomised, Controlled Cross Over Study Comparing Two Mandibular Advancement Devices in the Treatment of Mild-to-moderate Obstructive Sleep Apnea
The goal of this randomised, controlled cross-over study is to compare two mandibular repositioning devices (MRA) already on the market in the treatment of mild-to-moderate obstructive sleep apnea.
Patiënt will test the first MRA for 3 months, after a wash out period of 2 weeks they will test the second MRA for 3 months. It will be randomised what type of MRA will be tested first and second.
Study Overview
Status
Conditions
Intervention / Treatment
Detailed Description
A mandibular repositioning appliance (MRA) is the primary treatment option for patients with mild to moderate obstructive sleep apnea (OSA), as well as for those with severe OSA who are intolerant to continuous positive airway pressure (CPAP) therapy or for whom surgical interventions are not viable. MRAs function by advancing the mandible, thereby increasing the volume of the upper airway and reducing its tendency to collapse during sleep. Various types of MRAs are available, and multiple classification systems can be employed to distinguish between these different types.
Mandibular repositioning appliances (MRAs) can be either custom-made or universal. Custom-made devices, which are typically more expensive, are fabricated in a dental laboratory using dental impressions. In contrast, prefabricated devices, often thermoplastic MRAs, are generally less costly and readily available over the counter.
Mandibular repositioning appliances (MRAs) can be classified as non-titratable, also known as monoblock devices, which have a fixed position, or titratable devices, which allow for gradual adjustment of the mandibular position until the desired therapeutic effect is achieved. Current guidelines recommend the use of titratable MRAs.
Titratable mandibular repositioning appliances (MRAs) can be further categorized into two types: midline traction devices, where the upper and lower components are connected at the front of the device, and bilateral thrust devices, where the upper and lower components are linked in the lateral or (pre-)molar regions.
Within the latter category, various designs exist, including devices that are adjustable on both sides using Hyrax screws and 3D-printed devices equipped with a series of lower components that allow for adjustments in mandibular position.
An example of an MRA adjustable with Hyrax screws is the SomnoDent® device (SomnoMed, Denton, Texas, USA). An example of a device utilizing a series of lower components for titration is the OrthoApnea appliance (ORTHOAPNEA S.L., Málaga, Spain), which features a specific attachment mechanism.
Although studies have already been published comparing different types of "bilateral thrust" devices, no studies to date have directly compared the categories mentioned above. The hypothesis of the investigators is that both appliances have similar effectiveness in the treatment of mild-to-moderate obstructive sleep apnea.
Primary objective:
Improvement of AHI, measured by polygraphy before treatment and after treatment with each MRA.
Secondary objective;
Peripheral oxygen saturation, measured by polygraphy before treatment and after treatment with each MRA
- Average saturation
- Lowest saturation
- Oxygen desaturation index >3% (events/h)
- Total duration with saturation < 90%
- Epworth Sleepiness Scale, taken before treatment and after treatment with each MRA
- Patient experience questionnaire, taken before treatment and after treatment with each MRA
- Patient final choice of preferred MRA
Endpoints:
The investigators assess the objective improvement in OSA by comparing polygraphy results taken before and after treatment with each MRA. To evaluate the subjective effects of the MRA, the investigators compare the scores from the Epworth Sleepiness Scale. The participants subjective experience with each MRA type is measured using a patient experience questionnaire, along with noting the final preference for device type of the participants.
Study design:
- Randomized, controlled cross-over study
- Study population: Patients referred to the Department of Oral and Maxillofacial Surgery at ZOL Genk for Mandibular Repositioning Appliance (MRA) therapy, aged 18 years or older diagnosed with an Apnea-Hypopnea Index (AHI) between 5 and 30 events per hour.
Inclusion criteria:
In order to be eligible to participate in this study, a subject must meet all of the following criteria:
- Provide signed and dated informed consent
- Males or females aged 18 and above
- Apnea-Hypopnea Index (AHI) between 5 and 30 events per hour diagnosed by a polysomnography
Exclusion criteria:
A potential subject who meets any of the following criteria will be excluded from participation in this study:
- Previous surgery on the upper airways
- Prior treatment of OSA (CPAP or MRA)
- Presence of unstable cardiovascular conditions, neurological, mental, or psychiatric disorders
- Insufficient dental or periodontal condition for a mandibular repositioning appliance (MRA)
- Presence of temporomandibular joint dysfunction
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Genk, Belgium, 3600
- Ziekenhuis Oost-Limburg
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
- Adult
- Older Adult
Accepts Healthy Volunteers
Description
In order to be eligible to participate in this study, a subject must meet all of the following criteria:
- Provide signed and dated informed consent
- Males or females aged 18 and above
- Apnea-Hypopnea Index (AHI) between 5 and 30 events per hour diagnosed by a polysomnography
A potential subject who meets any of the following criteria will be excluded from participation in this study:
- Previous surgery on the upper airways
- Prior treatment of OSA (CPAP or MRA)
- Presence of unstable cardiovascular conditions, neurological, mental, or psychiatric disorders
- Insufficient dental or periodontal condition for a mandibular repositioning appliance (MRA)
- Presence of temporomandibular joint dysfunction
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: Randomized
- Interventional Model: Crossover Assignment
- Masking: Single
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: MRA1MRA2
Use of MRA device 1 for 3 months, use of MRA device 2 for 3 months
|
Epworth Sleepiness Scale o Johns, Murray W. "A New Method for Measuring Daytime Sleepiness: The Epworth Sleepiness Scale." Sleep, vol. 14, no. 6, 1991, pp. 540-545. Patient experience questionnaire o Ng, E.T., Perez-Garcia, A., en Lagravère-Vich, M.O. "Development and Initial Validation of a Questionnaire to Measure Patient Experience with Oral Appliance Therapy." Journal of Clinical Sleep Medicine, vol. 19, nr. 8, 1 aug. 2023, pp. 1437-1445. doi: 10.5664/jcsm.10562. PMID: 37082817; PMCID: PMC10394373
Patient final choice of preferred MRA
Polygraphy after 3 months of use of each appliance: an unattended home sleep study at the end of the first treatment period, and in-hospital polygraphy at the end of the second treatment period.
|
|
Experimental: MRA2MRA1
Use of MRA device 2 for 3 months, use of MRA device 1 for 3 months
|
Epworth Sleepiness Scale o Johns, Murray W. "A New Method for Measuring Daytime Sleepiness: The Epworth Sleepiness Scale." Sleep, vol. 14, no. 6, 1991, pp. 540-545. Patient experience questionnaire o Ng, E.T., Perez-Garcia, A., en Lagravère-Vich, M.O. "Development and Initial Validation of a Questionnaire to Measure Patient Experience with Oral Appliance Therapy." Journal of Clinical Sleep Medicine, vol. 19, nr. 8, 1 aug. 2023, pp. 1437-1445. doi: 10.5664/jcsm.10562. PMID: 37082817; PMCID: PMC10394373
Patient final choice of preferred MRA
Polygraphy after 3 months of use of each appliance: an unattended home sleep study at the end of the first treatment period, and in-hospital polygraphy at the end of the second treatment period.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
AHI
Time Frame: 3 monthts after use of first MRA (mandibular advancement device), 3 months after use of second MRA (mandibular advancement device)
|
Improvement of AHI (apnea-hypopnea index), measured by polygraphy before treatment and after treatment with each MRA.
The goal of the mandibular advancement device is to reduce the number of apneas and hypopneas per hour, apneas are periods when a person stops breathing and hypopneas are instances where airflow is blocked, causing shallow breathing.
This is measured by a polygraphy-test overnight at the patiënts home.
|
3 monthts after use of first MRA (mandibular advancement device), 3 months after use of second MRA (mandibular advancement device)
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Peripheral oxygen saturation 1
Time Frame: 3 monthts after use of first MRA, 3 months after use of second MRA
|
Measured by polygraphy
Oxygen saturation in percent is the fraction of oxygen-saturated haemoglobin relative to total haemoglobin (unsaturated + saturated) in the blood * 100% |
3 monthts after use of first MRA, 3 months after use of second MRA
|
|
Peripheral oxygen saturation 2
Time Frame: 3 monthts after use of first MRA, 3 months after use of second MRA
|
Measured by polygraphy o Oxygen desaturation index >3% (events/h) Desaturation episodes are defined as a decrease in the mean oxygen saturation of ≥3% for more than 10 seconds per hour |
3 monthts after use of first MRA, 3 months after use of second MRA
|
|
Peripheral oxygen saturation 3
Time Frame: 3 monthts after use of first MRA, 3 months after use of second MRA
|
Measured by polygraphy o Total duration with saturation < 90% The percentage of cumulative time with oxygen saturation below 90% in total sleep time. |
3 monthts after use of first MRA, 3 months after use of second MRA
|
|
ESS
Time Frame: 3 monthts after use of first MRA, 3 months after use of second MRA
|
Epworth Sleepiness Scale (questionaire) The minimum value is 0 and the maximum value is 24. A higher score means a worse outcome, indicating more excessive daytime sleepiness. The Epworth Sleepiness Scale is scored by adding up the individual scores for each of the eight questions. Each question is rated on a scale of 0 to 3, with 0 being "no chance of dozing" and 3 being "high chance of dozing." ter daytime sleepiness. Score range Interpretation 0-5 Lower normal daytime sleepiness 6-10 Normal daytime sleepiness 11-12 Mild excessive daytime sleepiness 13-15 Moderate excessive daytime sleepiness 16-24 Severe excessive daytime sleepiness |
3 monthts after use of first MRA, 3 months after use of second MRA
|
|
Final choice
Time Frame: after the use of each MRA for 3 months
|
Patient final choice of preferred MRA. At the final visit, after using each MRA (type 1 and type 2) for three months (with a two-week washout period in between), the patient will choose their preferred device. The final choice will be noted down after oral questioning. |
after the use of each MRA for 3 months
|
|
Patient experience questionnaire
Time Frame: 3 months after use of the first MRA, 3 months after use of the second MRA
|
Patient experience with each appliance, assessed after 3 months of use.
The appliance-design domain of the Oral Appliance Therapy Patient Experience Questionnaire (OATPEQ; Ng et al., J Clin Sleep Med 2023) was used, translated into Dutch.
It comprises 13 statements, each scored from 0 to 4, giving a total score ranging from 0 to 52.
A higher score indicates a more positive patient experience.
|
3 months after use of the first MRA, 3 months after use of the second MRA
|
Collaborators and Investigators
Sponsor
Publications and helpful links
General Publications
- Ramar K, Dort LC, Katz SG, Lettieri CJ, Harrod CG, Thomas SM, Chervin RD. Clinical Practice Guideline for the Treatment of Obstructive Sleep Apnea and Snoring with Oral Appliance Therapy: An Update for 2015. J Clin Sleep Med. 2015 Jul 15;11(7):773-827. doi: 10.5664/jcsm.4858.
- Johns MW. A new method for measuring daytime sleepiness: the Epworth sleepiness scale. Sleep. 1991 Dec;14(6):540-5. doi: 10.1093/sleep/14.6.540.
- Ng ET, Perez-Garcia A, Lagravere-Vich MO. Development and initial validation of a questionnaire to measure patient experience with oral appliance therapy. J Clin Sleep Med. 2023 Aug 1;19(8):1437-1445. doi: 10.5664/jcsm.10562.
- Zhou J, Liu YH. A randomised titrated crossover study comparing two oral appliances in the treatment for mild to moderate obstructive sleep apnoea/hypopnoea syndrome. J Oral Rehabil. 2012 Dec;39(12):914-22. doi: 10.1111/joor.12006. Epub 2012 Sep 27.
- Vezina JP, Blumen MB, Buchet I, Hausser-Hauw C, Chabolle F. Does propulsion mechanism influence the long-term side effects of oral appliances in the treatment of sleep-disordered breathing? Chest. 2011 Nov;140(5):1184-1191. doi: 10.1378/chest.10-3123. Epub 2011 May 26.
- Shi X, Lobbezoo F, Chen H, Rosenmoller BRAM, Berkhout E, de Lange J, Aarab G. Comparisons of the effects of two types of titratable mandibular advancement devices on respiratory parameters and upper airway dimensions in patients with obstructive sleep apnea: a randomized controlled trial. Clin Oral Investig. 2023 May;27(5):2013-2025. doi: 10.1007/s00784-023-04945-z. Epub 2023 Mar 17.
- Lawton HM, Battagel JM, Kotecha B. A comparison of the Twin Block and Herbst mandibular advancement splints in the treatment of patients with obstructive sleep apnoea: a prospective study. Eur J Orthod. 2005 Feb;27(1):82-90. doi: 10.1093/ejo/cjh067.
- Uniken Venema JAM, Rosenmoller BRAM, de Vries N, de Lange J, Aarab G, Lobbezoo F, Hoekema A. Mandibular advancement device design: A systematic review on outcomes in obstructive sleep apnea treatment. Sleep Med Rev. 2021 Dec;60:101557. doi: 10.1016/j.smrv.2021.101557. Epub 2021 Oct 1.
- Schmidt-Nowara W, Lowe A, Wiegand L, Cartwright R, Perez-Guerra F, Menn S. Oral appliances for the treatment of snoring and obstructive sleep apnea: a review. Sleep. 1995 Jul;18(6):501-10. doi: 10.1093/sleep/18.6.501.
- Dieltjens M, Vanderveken O. Oral Appliances in Obstructive Sleep Apnea. Healthcare (Basel). 2019 Nov 8;7(4):141. doi: 10.3390/healthcare7040141.
- Almeida FR, Lowe AA. Principles of oral appliance therapy for the management of snoring and sleep disordered breathing. Oral Maxillofac Surg Clin North Am. 2009 Nov;21(4):413-20. doi: 10.1016/j.coms.2009.07.002.
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- Z-2025015
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.