Clinical Validation of the Blood Pressure Measuring Device Withings BPM Pro 2 (WIHYP-GP) (WIHYP-GP)
February 17, 2026 updated by: Withings
Clinical Validation Of The Brachial Blood Pressure Measuring Device Withings BPM Pro 2 According To "The Universal Protocol For The Validation Of Blood Pressure Measuring Devices By The Association For The Advancement Of Medical Instrumentation / European Society Of Hypertension / International Organization For Standardization (AAMI/ESH/ISO) (And Its Amendment 1 (2020) And 2 (2024))" In The General Population.
The aim of the study is to assess the accuracy of the automatic oscillometric BP measuring device at the brachial level, the WITHINGS BPM Pro 2, in the general population
Study Overview
Status
Completed
Conditions
Intervention / Treatment
Study Type
Interventional
Enrollment (Actual)
89
Phase
- Not Applicable
Contacts and Locations
This section provides the contact details for those conducting the study, and information on where this study is being conducted.
Study Locations
-
-
-
Yerevan, Armenia, 0002
- N2
-
-
Participation Criteria
Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.
Eligibility Criteria
Ages Eligible for Study
- Adult
- Older Adult
Accepts Healthy Volunteers
Yes
Description
Inclusion Criteria:
- Patient older than 18 years;
- Patient who signed the informed consent form;
- Patient followed-up at site (in-patient or out-patient);
- Patient with arm circumference between 22 cm and 42 cm.
Exclusion Criteria:
- Patient unable to give a consent or understand properly protocol information;
- Patient suffering from arrhythmia;
- Patient with poor quality of Korotkov sounds;
- Patient for whom K5 sounds are absent;
- Patient wearing an implantable electric medical device (pacemaker,…);
- Patient with both upper arms suffering from open wound and/or damaged skin.
Study Plan
This section provides details of the study plan, including how the study is designed and what the study is measuring.
How is the study designed?
Design Details
- Primary Purpose: Device Feasibility
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: All patients
Patients consenting to participate to the study and compliant to the inclusion criteria
|
The patient will be measured alternatively with a reference mercury sphygmomanometer (four times) and the device under study (3 times)
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Blood pressure measurement performance evaluation per measure
Time Frame: The time frame is three months.
|
The blood pressure (systolic and diastolic blood pressure) measurement performance will be established using the mean value of the differences and the standard deviation between the device under study and the reference for each measure.
|
The time frame is three months.
|
|
Blood pressure measurement performance evaluation by subject
Time Frame: The time frame is three months.
|
The blood pressure measurement performance by subject will be established using the mean value of the differences and the standard deviation between the averaged per subject value of blood pressure (systolic and diastolic blood pressure) device under study and its reference.
|
The time frame is three months.
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Safety use of the device
Time Frame: The time frame is three months.
|
The rate of adverse effects.
|
The time frame is three months.
|
Collaborators and Investigators
This is where you will find people and organizations involved with this study.
Sponsor
Study record dates
These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.
Study Major Dates
Study Start (Actual)
April 28, 2025
Primary Completion (Actual)
August 21, 2025
Study Completion (Actual)
November 18, 2025
Study Registration Dates
First Submitted
May 2, 2025
First Submitted That Met QC Criteria
May 2, 2025
First Posted (Actual)
May 5, 2025
Study Record Updates
Last Update Posted (Actual)
February 18, 2026
Last Update Submitted That Met QC Criteria
February 17, 2026
Last Verified
February 1, 2026
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- 2025-A00050-49
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
NO
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
No
Studies a U.S. FDA-regulated device product
No
product manufactured in and exported from the U.S.
No
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.