Correlations Between Subjective and Objective Responses to Cochlear Implant Stimulation (CRSOSIC)
Multimodal Objective Measures Help Cochlear Implant Programming: Cross-sectional and Prospective Study Focusing on the Correlations Between eSRT, Behavioural and Other Objective Responses (eABR, eCAP)
Cochlear implants have been used for several decades, with major technological advances. However, adjustment can be difficult for patients who have to get used to a new sound that is electrical rather than acoustic. In general, adjustment is based on the patient's responses: the patient must inform the technician of their sensations in order to determine the sound level that is comfortable for them (MCL: maximum comfortable level).
The adjustment therefore depends on a subjective assessment. The aim of the adjustments is to make the sound pleasant while increasing the stimulation level to the maximum tolerated level in order to achieve good listening dynamics. Objective indicators would therefore be useful in improving the accuracy of the adjustments. The objective measurements that can be taken are:
- Electrical compound action potentials (eCAP)
- Electrical auditory brainstem response (eABR)
- Electrical stapedius reflex threshold (eSRT).
Study Overview
Status
Intervention / Treatment
Detailed Description
Observational study of adult patients who are going to receive or have received a cochlear implant. In fact, two groups of patients will be studied:
Incident case cohort: patients included in the cohort will be followed for one year after surgery.
Data will be collected at each postoperative follow-up visit (M1, M3, M6, M9, M12) to monitor their progress. Each objective measurement will be taken on at least three electrodes: on the left, right, and central electrodes. Data will be collected at each postoperative follow-up visit (M1, M3, M6, M9, M12) in order to monitor their progress. Each objective measurement will be taken on at least three electrodes: on the basal, median, and apical parts of the electrode array. Stimulation will be configured using the software normally used to adjust the implant.
Responses are collected:
- for eCAPs: via the same software
- for eABRs: using an auditory evoked potential recording device synchronized with the implant stimulations
- for eSRTs: using a tympanometer with the probe inserted into the external auditory canal on the same side or the opposite side to the stimulation. Cross-sectional sample of prevalent cases: The same measurements will be taken in patients who had a cochlear implant some time ago (evaluation more than one year after surgery). The measurements will be taken only once during their annual check-up. This population will enable us to compare data from patients who have recently had an implant with data from a population that already has a minimum of experience and adaptation to the implant.
Study Type
Enrollment (Estimated)
Contacts and Locations
Study Contact
- Name: Antoine PAUL, PhD
- Phone Number: +33145213128
- Email: antoine.paul@aphp.fr
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
- Adult
- Older Adult
Accepts Healthy Volunteers
Sampling Method
Study Population
Description
Inclusion Criteria:
Incident case cohort:
- Age ≥ 18 years (18 years or older)
- Scheduled for first-time unilateral cochlear implantation within the next 3 months
- Informed about the study and did not object (non-opposition)
Cross-sectional sample of prevalent cases:
- Age ≥ 18 years (18 years or older)
- Underwent unilateral cochlear implantation more than 1 year ago
- Has used the implant for more than 1 year (interval between implantation surgery and study inclusion > 1 year)
- Informed about the study and did not object (non-opposition)
Exclusion Criteria:
Exclusion Criteria - applies to both the incident case cohort and the cross-sectional sample of prevalent cases:
- Prior middle or inner ear surgery
- Cognitive impairment or insufficient comprehension of French that would prevent understanding of study procedures or accurate measurement of comfort thresholds during programming
- Peripheral facial palsy
- Bilateral cochlear implantation
- Patient under legal guardianship/court-ordered protection
Additional exclusion - cross-sectional sample of prevalent cases only:
- More than half of the electrodes not inserted (i.e., <50% electrode insertion)
Secondary exclusion criteria - Incident case cohort:
- More than half of the electrodes not inserted (<50% electrode insertion)
- Postoperative peripheral facial nerve palsy following the implantation surgery
Study Plan
How is the study designed?
Design Details
Cohorts and Interventions
Group / Cohort |
Intervention / Treatment |
|---|---|
|
Incident CI Cohort (Prospective Adults)
Adults ≥18 scheduled for first-time unilateral cochlear implantation within 3 months; followed post-op at M1, M3, M6, M9, M12.
Objective measures on ≥3 electrodes (basal, middle, apical): eCAP via CI software; eABR via evoked-potential system; eSRT via tympanometer.
No experimental intervention (routine programming only).
Main interest: associations between objective metrics and behavioral MCL, with key analyses at M6 and M12.
|
Non-invasive electrophysiological recordings obtained during/after routine cochlear-implant programming, using standard clinical equipment: eCAP via CI software, eABR via an evoked-potential system synchronized with the implant, and eSRT via tympanometry.
No investigational device or randomization.
Measurements are collected for observational correlations with behavioral levels.
Schedule: incident cohort at M1, M3, M6, M9, M12; prevalent cohort single annual visit.
|
|
Prevalent CI Cohort (Cross-Sectional Adults)
Adults ≥18 with unilateral cochlear implant use ≥12 months; single annual follow-up visit.
Same objective measures (eCAP, eABR, eSRT) on ≥3 electrodes using standard clinical equipment; no intervention.
Purpose: cross-sectional correlations and comparison with the incident cohort.
|
Non-invasive electrophysiological recordings obtained during/after routine cochlear-implant programming, using standard clinical equipment: eCAP via CI software, eABR via an evoked-potential system synchronized with the implant, and eSRT via tympanometry.
No investigational device or randomization.
Measurements are collected for observational correlations with behavioral levels.
Schedule: incident cohort at M1, M3, M6, M9, M12; prevalent cohort single annual visit.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
The MCL threshold, the objective response threshold (eSRT), the eSRT response
Time Frame: 15 months
|
"The MCL threshold, expressed in CU (charge unit) units, corresponds to the maximum level of stimulation of the implant required to obtain a sound level considered comfortable by the patient. It will be recorded at each post-operative check-up (M1, M3, M6, M9, and M12). The objective response threshold (eSRT), expressed in CU units, corresponds to the minimum level of implant stimulation required to elicit a stapedial reflex. It will be recorded at each post-operative assessment (M1, M3, M6, M9, and M12). The eSRT response will be collected using a tympanometry probe (commonly used to measure ear impedance in patients during a standard hearing assessment) in the patient's external auditory canal, ipsilateral or contralateral to the implanted side. The tympanometer will continuously record ear impedance during electrical stimulation until a significant variation is obtained, indicating the presence of the stapedial reflex." |
15 months
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Correlation between behavioral MCL and eABR threshold
Time Frame: Month 6 and Month 12 post-operative follow-up visits (incident cohort)
|
Pearson/Spearman correlation between Maximum Comfortable Level (MCL, charge units [CU]) and electrically evoked Auditory Brainstem Response (eABR) threshold (CU); measured on ≥3 electrodes (basal, middle, apical) using standard CI programming and evoked-potential equipment.
|
Month 6 and Month 12 post-operative follow-up visits (incident cohort)
|
|
Correlation between behavioral MCL and eCAP threshold
Time Frame: M1, M3, M6, M9, M12 post-operative visits (incident cohort).
|
Pearson/Spearman correlation between MCL (CU) and electrically evoked Compound Action Potential (eCAP) threshold (CU); measured on ≥3 electrodes (basal, middle, apical) via implant software.
|
M1, M3, M6, M9, M12 post-operative visits (incident cohort).
|
|
Correlation between eSRT and eABR thresholds
Time Frame: M6 and M12 post-operative visits (incident cohort).
|
Pearson/Spearman correlation between electrically evoked Stapedius Reflex Threshold (eSRT, CU) and eABR threshold (CU); ≥3 electrodes per visit.
|
M6 and M12 post-operative visits (incident cohort).
|
|
Correlation between eSRT and eCAP thresholds
Time Frame: M1, M3, M6, M9, M12 post-operative visits (incident cohort).
|
Pearson/Spearman correlation between eSRT (CU) and eCAP threshold (CU); ≥3 electrodes per visit.
|
M1, M3, M6, M9, M12 post-operative visits (incident cohort).
|
|
Correlation between eABR and eCAP thresholds
Time Frame: Month 6 and Month 12 after cochlear implantation (incident cohort)
|
Pearson/Spearman correlation between eABR and eCAP thresholds (both in CU); ≥3 electrodes per visit.
|
Month 6 and Month 12 after cochlear implantation (incident cohort)
|
|
Agreement between intraoperative eSRT by tympanometer vs visual stapedius reflex
Time Frame: ntraoperative (single time point during CI surgery).
|
Association between peri-operative eSRT measured objectively with a tympanometer and the surgeon's visual observation of stapedius movement (yes/no).
|
ntraoperative (single time point during CI surgery).
|
|
Probe frequencies eliciting eSRT
Time Frame: Post-operative follow-up visits at Month 1, Month 3, Month 6, Month 9, and Month 12 (incident cohort), and 1 year or more after cochlear implantation (prevalent cohort)
|
Distribution of probe frequencies (226, 678, 1000 Hz) that yield an eSRT response; eSRT level recorded for each tested frequency.
|
Post-operative follow-up visits at Month 1, Month 3, Month 6, Month 9, and Month 12 (incident cohort), and 1 year or more after cochlear implantation (prevalent cohort)
|
|
Longitudinal profile of objective thresholds (eSRT/eABR/eCAP)
Time Frame: From Month 1 to Month 12 after cochlear implantation (incident cohort)
|
Trajectory of eSRT, eABR, and eCAP thresholds over the first post-operative year, summarized graphically and descriptively.
|
From Month 1 to Month 12 after cochlear implantation (incident cohort)
|
|
Cross-sectional associations among MCL and objective measures in prevalent users
Time Frame: At the annual follow-up visit, at least 12 months after cochlear implant surgery.
|
Correlations between MCL and eSRT/eABR/eCAP, and pairwise correlations among objective measures, in adults with ≥12 months of CI use.
|
At the annual follow-up visit, at least 12 months after cochlear implant surgery.
|
|
Between-cohort comparison of correlation coefficients (prevalent vs incident) Description: Comparison of correlation coefficients for each pair (MCL-eSRT, MCL-eABR, MCL-eCAP, eSRT-eABR, eSRT-eCAP, eABR-eCAP) between the prevalent cross-section and the in
Time Frame: Incident cohort: post-operative follow-up visits from Month 1 through Month 12. Prevalent cohort: annual follow-up visit, at least 12 months after cochlear implant surgery.
|
Between-cohort comparison of correlation coefficients (prevalent vs incident) Description: Comparison of correlation coefficients for each pair (MCL-eSRT, MCL-eABR, MCL-eCAP, eSRT-eABR, eSRT-eCAP, eABR-eCAP) between the prevalent cross-section and the incident cohort.
|
Incident cohort: post-operative follow-up visits from Month 1 through Month 12. Prevalent cohort: annual follow-up visit, at least 12 months after cochlear implant surgery.
|
Collaborators and Investigators
Collaborators
Study record dates
Study Major Dates
Study Start (Estimated)
Primary Completion (Estimated)
Study Completion (Estimated)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Keywords
Other Study ID Numbers
- APHP250486
- 2025-A01446-43 (Other Identifier: IDRCB)
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.