Finger Versus Earlobe Pulse Oximetry During the 6-Minute Walk Test in Interstitial Lung Disease (OXISITE-ILD)
A Prospective, Multicenter, Observational Method-Comparison Study of Finger Versus Earlobe Pulse Oximetry for Oxygen Saturation During the Six-Minute Walk Test in Patients With Interstitial Lung Disease (OXISITE-ILD)
The OXISITE-ILD study is a prospective, multicenter, observational study designed to evaluate the agreement between finger (digital) and earlobe (auricular) pulse oximetry for measuring oxygen saturation (SpO2) during the six-minute walk test (6MWT) in patients with interstitial lung disease (ILD). In routine clinical practice, exercise SpO2 is usually measured at the finger; however, the finger reading can be unreliable in some patients, and there is currently no recommendation on the best sensor location in ILD. This study compares the two sensor locations, recorded at the same time, and evaluates whether any disagreement changes how exercise desaturation is classified and whether ambulatory oxygen is indicated.
All patients undergoing a 6MWT as part of routine ILD care are included consecutively to ensure a pragmatic, real-world representation of the ILD population. The primary objective is to measure the agreement between the two locations in the lowest SpO2 reached during the test, including the size and direction of any difference. Secondary objectives include the reclassification of patients at the clinical desaturation thresholds, the comparison between autoimmune and non-autoimmune ILD, the rate of invalid readings at each location, and the clinical, vascular and functional factors associated with disagreement.
Přehled studie
Postavení
Podmínky
Intervence / Léčba
Detailní popis
OXISITE-ILD is a prospective, consecutive, multicenter, observational, cross-sectional method-comparison (agreement) study.
Each participant is assessed once, during a single 6MWT performed as part of routine care. SpO2 and heart rate are recorded simultaneously at the finger and at the earlobe with two pulse oximeters, at eight timepoints: baseline, each of minutes 1 to 6, and after a 2-minute recovery. The paired heart rate at each timepoint is used as a signal-quality control, so that agreement is assessed only on simultaneous, valid readings. In some participants, an arterial blood gas or a nailfold capillaroscopy may also be performed to help interpret the difference between locations.
A minimum of 100 participants is planned at the coordinating center, sufficient on its own for the primary objective; recruitment may be extended at the coordinating center and at the collaborating centers as the study requires. The participation of collaborating centers is opportunistic and adds precision, subgroup analyses, and external validity.
Typ studie
Zápis (Odhadovaný)
Kontakty a umístění
Studijní kontakt
- Jméno: Jaume Bordas-Martinez, Ph.D.; M.D.
- Telefonní číslo: 938 42 50 00
- E-mail: jbordas@fphag.org
Studijní místa
-
-
Spain
-
Granollers, Spain, Španělsko, 08402
- Nábor
- Hospital de Granollers
-
-
Kritéria účasti
Kritéria způsobilosti
Věk způsobilý ke studiu
- Dospělý
- Starší dospělý
Přijímá zdravé dobrovolníky
Metoda odběru vzorků
Studijní populace
Popis
Inclusion Criteria:
- Adults aged 18 years or older
- Confirmed diagnosis of interstitial lung disease (ILD) based on clinical, radiological and/or histological criteria
- Newly diagnosed or under follow-up for ILD
- Physically and cognitively able to perform the 6-minute walk test (6MWT)
- Signed informed consent
Exclusion Criteria:
- Inability to obtain a reliable, stable baseline pulse oximetry reading at the finger and/or the earlobe, preventing a valid paired baseline measurement and the calculation of agreement
- ILD exacerbation, decompensated comorbidity, or acute respiratory infection at the time of the test
Studijní plán
Jak je studie koncipována?
Detaily designu
Kohorty a intervence
Skupina / kohorta |
Intervence / Léčba |
|---|---|
|
ILD patients undergoing the 6MWT
Adults with an established diagnosis of interstitial lung disease (ILD) undergoing a six-minute walk test (6MWT) as part of routine care.
Each participant is assessed once, with oxygen saturation (SpO2) and heart rate recorded simultaneously at the finger and at the earlobe.
Pre-specified subgroups for the analysis are autoimmune-associated versus non-autoimmune ILD.
|
During a single 6MWT, SpO2 and heart rate are recorded simultaneously at the finger (digital) and the earlobe (auricular) with two pulse oximeters, at eight timepoints (baseline, minutes 1 to 6, and a 2-minute recovery).
No treatment is administered and routine clinical management is not modified; the procedure adds the simultaneous earlobe measurement for comparison with the finger.
|
Co je měření studie?
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Intraclass correlation coefficient for agreement in minimum SpO2 between sensor locations
Časové okno: Day 1
|
Absolute agreement in minimum SpO2 during the six-minute walk test between the finger and earlobe sensors, quantified with the intraclass correlation coefficient (two-way mixed-effects model, absolute agreement) and its 95% confidence interval.
ICC values are interpreted as below 0.50 poor, 0.50 to 0.75 moderate, 0.75 to 0.90 good, and above 0.90 excellent agreement.
|
Day 1
|
|
Bland-Altman agreement in minimum SpO2 between sensor locations
Časové okno: Day 1
|
Agreement in minimum SpO2 during the six-minute walk test between the finger and earlobe sensors, assessed with Bland-Altman analysis: mean bias (earlobe minus finger) with its 95% confidence interval and 95% limits of agreement, evaluated against a prespecified clinically acceptable difference of 3 percentage points.
The direction of the bias is reported, as the finger is hypothesised to register lower values than the earlobe.
|
Day 1
|
Sekundární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Reclassification of exertional desaturation between sensor locations at SpO2 below 88%
Časové okno: Day 1
|
Proportion of participants whose classification of significant exertional desaturation changes between the finger and earlobe sensors at the threshold of minimum SpO2 below 88% (American Thoracic Society indication threshold for ambulatory oxygen).
Agreement is assessed with the Cohen kappa coefficient (Landis-Koch interpretation) and the difference in proportions with the McNemar test for paired data.
At this threshold the McNemar test and the raw reclassification proportions are the primary read, as kappa can be unstable when prevalence is extreme.
|
Day 1
|
|
Reclassification of exertional desaturation between sensor locations at SpO2 below 90%
Časové okno: Day 1
|
Proportion of participants whose classification of significant exertional desaturation changes between the finger and earlobe sensors at the threshold of minimum SpO2 below 90% (British Thoracic Society indication threshold for ambulatory oxygen).
Agreement is assessed with the Cohen kappa coefficient (Landis-Koch interpretation) and the difference in proportions with the McNemar test for paired data.
|
Day 1
|
|
Reclassification of exertional desaturation between sensor locations at a fall in SpO2 of 4% or more
Časové okno: Day 1
|
Proportion of participants whose classification changes between the finger and earlobe sensors at the prognostic threshold of a fall in minimum SpO2 of 4% or more from baseline.
Agreement is assessed with the Cohen kappa coefficient (Landis-Koch interpretation) and the difference in proportions with the McNemar test for paired data.
|
Day 1
|
|
Agreement in minimum SpO2 by autoimmune versus non-autoimmune ILD
Časové okno: Day 1
|
Comparison of the agreement between sensor locations (intraclass correlation coefficient and Bland-Altman) in autoimmune-associated ILD versus non-autoimmune ILD, with interstitial pneumonia with autoimmune features (IPAF) described separately.
Where sample size permits, agreement is also described, exploratorily and without formal hypothesis testing, by major ILD subtype (for example idiopathic pulmonary fibrosis, hypersensitivity pneumonitis, or connective tissue disease-associated ILD), and by autoimmune subtype (for example systemic sclerosis or rheumatoid arthritis).
|
Day 1
|
|
Rate of invalid or poor-quality readings by sensor location
Časové okno: Day 1
|
Comparison, among enrolled participants, of the rate of invalid or poor-quality (low-signal) readings between the finger and earlobe sensors during the six-minute walk test, across the eight measurement timepoints.
At each timepoint the paired heart rate from the two sensors is used as a signal-quality control: a discordant heart rate flags poor signal or sensor disconnection and identifies invalid SpO2 readings, so that agreement is assessed only on simultaneous, valid readings.
The per-participant difference in the rate of invalid readings between sensor locations is reported with its 95% confidence interval and the McNemar test for paired data.
|
Day 1
|
|
Clinical, vascular and functional predictors of discordance
Časové okno: Day 1
|
Clinical, vascular and functional factors associated with discordance between sensor locations, including ILD type (autoimmune-associated versus other), autoimmune subtype (systemic sclerosis versus other), fibrosing versus non-fibrosing ILD, pathological capillaroscopy, Raynaud phenomenon, skin pigmentation, cardiovascular comorbidity, and pulmonary function (FVC and DLCO).
For each discordance criterion (SpO2 below 88%, SpO2 below 90%, and a fall of 4% or more) a logistic regression model is fitted with the patient as the unit of analysis, complemented by mixed-effects logistic regression (GLMM) with the measurement timepoint as a predictor and a random intercept per participant.
Missing data from incomplete follow-up are handled by maximum likelihood; penalised regression (L1/L2) is considered if events are sparse.
|
Day 1
|
Další výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Agreement in SpO2 at additional timepoints
Časové okno: Day 1
|
Agreement between sensor locations in SpO2 at baseline, maximum, per-minute intervals during the walk, and recovery, assessed with the intraclass correlation coefficient and Bland-Altman analysis.
|
Day 1
|
|
Quantitative paired difference and Passing-Bablok regression between sensor locations
Časové okno: Day 1
|
Exploratory analysis of systematic differences between the finger and earlobe in SpO2.
The paired difference (earlobe minus finger) is summarised with its 95% confidence interval (paired t test, or Wilcoxon signed-rank test with a Hodges-Lehmann interval, as appropriate).
Passing-Bablok regression is used to detect constant and proportional differences between the two locations, with 95% confidence intervals for the intercept and the slope.
|
Day 1
|
|
Agreement of each sensor location with baseline arterial oxygen saturation
Časové okno: Day 1
|
Exploratory comparison of the agreement of finger and of earlobe SpO2 with baseline arterial oxygen saturation, in the subgroup of participants undergoing arterial blood gas sampling.
|
Day 1
|
|
Direct and indirect cost associated with invalid finger readings
Časové okno: Day 1
|
Exploratory description of the direct and indirect costs associated with invalid or poor-quality finger readings.
No prespecified cost model is applied.
|
Day 1
|
Spolupracovníci a vyšetřovatelé
Sponzor
Vyšetřovatelé
- Ředitel studie: Jaume Bordas-Martinez, M.D., Ph.D., Hospital de Granollers
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia (Aktuální)
Primární dokončení (Odhadovaný)
Dokončení studie (Odhadovaný)
Termíny zápisu do studia
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Aktuální)
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Více informací
Termíny související s touto studií
Klíčová slova
- Šestiminutový test chůze
- Systémová skleróza
- Idiopatická plicní fibróza
- Pulzní oxymetrie
- Intersticiální plicní onemocnění
- SpO2
- Nasycení kyslíkem
- Srovnání metod
- Ambulantní kyslík
- Intersticiální plicní onemocnění při onemocnění pojivové tkáně
- Earlobe Oximetry
- Finger Oximetry
- Exercise Desaturation
- Bland-Altman Analysis
Další relevantní podmínky MeSH
Další identifikační čísla studie
- OXISITE-ILD
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .
Klinické studie na Intersticiální plicní onemocnění (ILD)
-
Rohit Aggarwal, MDBoehringer IngelheimDokončenoMyositis Associated Interstitial Lund Disease (MA-ILD)Spojené státy