An International Multicenter, Multivendor Evaluation of the Free-Running Framework for Cardiac Function (FAST-CMR)
Assessment and Validation of the Established Free-Running Framework for Cardiac Function by Magnetic Resonance Imaging (FAST-CMR): An International Multi-Center, Multi-Vendor Study at 1.5T
This project aims to evaluate a new cardiac MRI technique called the Free-Running Framework (FRF), which could simplify and accelerate the process of acquiring cardiac images. The investigators want to verify whether this method can provide functional heart measurements comparable to those obtained with traditional methods. More specifically, the goal of the study is to compare the measurements obtained with FRF to those obtained with standard sequences, to ensure they match and that this new approach can be reliably used in clinical practice. The FRF technique works differently from standard cardiac MRI. In standard exams, patients are asked to hold their breath several times and small electrodes (ECG) are used to monitor the heartbeat during the scan. These steps are needed to get clearer images of the heart as it moves. With FRF, these steps are no longer necessary: the scan is performed while patients' breath normally and without ECG monitoring. In addition, standard MRI takes multiple 2D slices of the heart, one after another. The FRF method instead captures a 3D image of the entire heart in one go, which can improve the consistency of the exam and reduce errors when doctors analyze the images later. This is all possible because the FRF method records data continuously and then organizes the images afterward based on how participants heart and breathing were moving during the exam. This helps the imager to get clear images of the heart, even without breath-holding or ECG monitoring. This project is aimed at individuals with heart disease who require cardiac MRI exams to monitor their health status (age ≥ 18 years) and are able to clearly understand the instructions provided by the research team. The investigators have already conducted small-scale technical and feasibility studies using FRF. These studies have shown that FRF is easy to use, faster than traditional methods, and provides image quality comparable to standard imaging techniques. The investigators now wish to evaluate its use in a clinical setting. More specifically, the investigators need to verify that FRF provides the same essential diagnostic information as standard techniques, so that it can be reliably used in future patient care. A maximum of 300 participants will be included in the investigation of this MRI technique between 2026 and 2031. This is a multi-center study, conducted internationally across 18 centers.
This project is being carried out in compliance with Swiss legislation. The investigators follow all internationally recognized guidelines. The competent ethics committee has reviewed and approved this project.
Studieoversigt
Status
Betingelser
Detaljeret beskrivelse
Cardiovascular disease remains the leading cause of death in industrialized nations. While a range of diagnostic tools exists for cardiovascular disease detection and monitoring, magnetic resonance imaging (MRI) remains the only modality that enables a safe, non-invasive assessment of the heart without exposure to ionizing radiation. MRI allows for a comprehensive evaluation of cardiac anatomy, function, myocardial tissue characterization, and blood flow quantification, making it a powerful tool for cardiovascular diagnostics.
Despite strong clinical evidence supporting its utility, cardiac MRI (CMR) remains underutilized, primarily due to the length and complexity of a traditional CMR exam. Several factors contribute to the length and complexity of a standard CMR exam. First, a standard CMR protocol relies heavily on two-dimensional (2D) image acquisitions, each requiring manual slice planning by an experienced technologist-a process that is both time consuming and highly dependent on operator expertise. Second, traditional image acquisition requires electrocardiogram (ECG) triggering to synchronize with the cardiac cycle, requiring additional setup and potentially introducing errors if the ECG signal is suboptimal. Finally, most conventional sequences rely on repeated patient breath-holding to minimize respiratory motion artifacts, which can be particularly challenging for individuals with severe cardiovascular disease, congenital anomalies, or limited compliance. Even for highly skilled personnel, this process is timeconsuming and inefficient. Consequently, a significant portion of the patient's time in the scanner is spent on preparation and planning rather than actual image acquisition, leading to prolonged exam durations and increased healthcare costs.
Given these challenges, there is a strong need for simplified, automated, and time-efficient CMR acquisitions. In response to this challenge, the investigators research group has developed an innovative "free-running framework" (FRF)-a set of MRI methods that continuously acquire threedimensional (3D) image data across the entire cardiac cycle and throughout free breathing, irrespective of cardiac or respiratory motion. Unlike conventional CMR sequences that require separate, prospectively planned acquisitions for each imaging plane and time point, FRF employs continuous, self-navigated data acquisition. This eliminates the need for complex slice planning and enables retrospective reconstruction of cardiac motion, ensuring that imaging is both standardized and independent of user expertise. By leveraging advanced motion-resolved reconstruction algorithms developed by the investigators group, the investigators can derive both cardiac and respiratory motion from a single dataset, providing a fully automated, 3D whole-heart imaging approach.
Once diagnosed, patients with cardiac diseases often require lifelong monitoring and repeated imaging assessments to guide treatment decisions and evaluate disease progression. This makes a non-ionizing imaging modality like CMR admirable. Nonetheless, the prolonged scan durations and intricate manual planning associated with traditional CMR limit its accessibility and practical feasibility. By eliminating the need for slice planning and reducing scan complexity, FRF has the potential to significantly improve imaging for cardiac disease patients by:
- Standardizing imaging: producing user-independent results robust to anatomical variations.
- Reducing scan times: without compromising diagnostic information.
- Improving accessibility: enabling easier adoption in centers with less experienced technologists.
The feasibility of FRF has been demonstrated in experimental, pre-clinical, and small observational clinical studies, with no observed adverse effects. This study aims to evaluate its clinical feasibility and efficiency in real-world cardiac disease patients across multiple cardiac institutions, serving as a precursor for larger validation studies and eventual clinical implementation. This study will provide the first systematic clinical evaluation of 3D FRF across multiple cardiac institutions in cardiac disease patients, assessing both its technical feasibility and potential workflow benefits in a real-world setting. In particular, this study will:
- Evaluate whether diagnostic information from FRF matches or exceeds standard CMR.
- Assess the efficiency gains and the impact of automated, self-navigated imaging on scan duration and patient comfort.
If successful, this study will lay the groundwork for future multi-center validation clinical trial studies and eventual clinical integration, addressing a critical gap in CMR accessibility and efficiency.
Undersøgelsestype
Tilmelding (Anslået)
Kontakter og lokationer
Studiekontakt
- Navn: Katerina Eyre, PhD
- Telefonnummer: +41213147513
- E-mail: katerina.eyre@chuv.ch
Undersøgelse Kontakt Backup
- Navn: Matthias Stuber, PhD
- Telefonnummer: +41213147534
- E-mail: matthias.stuber@chuv.ch
Studiesteder
-
-
Victoria
-
Melbourne, Victoria, Australien, 3010
- University of Melbourne
-
Kontakt:
- Ruth Lim, MD
- Telefonnummer: +61 3 9035 5511
- E-mail: ruthplim74@gmail.com
-
-
-
-
Capital District
-
Bogotá, Capital District, Colombia, 111156
- Fundacion Cardioinfantil-LaCardio
-
Kontakt:
- Carlos Guerrero, MD
- Telefonnummer: +57 (601) 744 2000
- E-mail: ceguerrero@lacardio.org
-
-
-
-
London
-
London, London, Det Forenede Kongerige, WC2R2LS
- King's College London
-
Kontakt:
- Amedeo Chiribiri, MD, PhD
- Telefonnummer: +44 20 7188 8259
- E-mail: amedeo.chiribiri@kcl.ac.uk
-
-
-
-
California
-
Los Angeles, California, Forenede Stater, 90095
- University of California, Los Angeles
-
Kontakt:
- Kim-Lien Nguyen, MD
- Telefonnummer: +1 (310) 825-1234
- E-mail: KimLienNguyen@mednet.ucla.edu
-
-
Georgia
-
Atlanta, Georgia, Forenede Stater, 30322
- Emory University
-
Kontakt:
- John Oshinski, PhD
- Telefonnummer: +1 404-727-5894
- E-mail: jnoshin@emory.edu
-
-
Illinois
-
Chicago, Illinois, Forenede Stater, 60611
- Ann & Robert H Lurie Children's Hospital of Chicago
-
Kontakt:
- Joshua D. Robinson, MD
- Telefonnummer: +1 312-227-4638
- E-mail: jdrobinson@luriechildrens.org
-
-
Massachusetts
-
Boston, Massachusetts, Forenede Stater, 02115
- Boston's Children Hospital
-
Kontakt:
- Andrew Powell, MD
- Telefonnummer: +1 617-355-2079
- E-mail: andrew.powell@cardio.chboston.org
-
-
Minnesota
-
Rochester, Minnesota, Forenede Stater, 55902
- Mayo Clinic
-
Kontakt:
- Tim Leiner, MD, PhD
- Telefonnummer: +1 507-284-2511
- E-mail: leiner.tim@mayo.edu
-
-
Missouri
-
St Louis, Missouri, Forenede Stater, 63130
- Washington University in St. Louis
-
Kontakt:
- Pamela Woodard, MD
- Telefonnummer: +1 314-747-3386
- E-mail: woodardp@nospam.wustl.edu
-
-
Ohio
-
Columbus, Ohio, Forenede Stater, 43210
- Ohio State University
-
Kontakt:
- Yuchi Han, MD
- Telefonnummer: +1 614-292-6446
- E-mail: yuchi.han@osumc.edu
-
-
Pennsylvania
-
Philadelphia, Pennsylvania, Forenede Stater, 19104
- Children's Hospital of Philadelphia
-
Kontakt:
- Mark Fogel, MD
- Telefonnummer: +1 215-590-3534
- E-mail: fogel@email.chop.edu
-
-
Texas
-
Dallas, Texas, Forenede Stater, 75390
- UT southwestern Medical Center
-
Kontakt:
- Franz Gerald Greil, MD, PhD
- Telefonnummer: +1 214-648-3111
- E-mail: gerald.greil@utsouthwestern.edu
-
-
-
-
Limburg
-
Maastricht, Limburg, Holland, 6229 HX
- Maastricht University Medical Center
-
Kontakt:
- Robert J. Holtackers, PhD
- Telefonnummer: +31 43 387 49 10
- E-mail: rob.holtackers@mumc.nl
-
-
-
-
Lazio
-
Rome, Lazio, Italien, 00168
- Università Cattolica - Fondazione Policlinico Gemelli IRCCS
-
Kontakt:
- Luigi Natale, MD
- Telefonnummer: +971 56 666 0984
- E-mail: luigi.natale@unicatt.it
-
-
-
-
Mie-ken
-
Tsu, Mie-ken, Japan, 514-8507
- Mie University
-
Kontakt:
- Hajime Sakuma, MD, PhD
- Telefonnummer: +81 59 231 5029
- E-mail: sakuma@med.mie-u.ac.jp
-
-
-
-
Singapore
-
Singapore, Singapore, Singapore, 119275
- National University of Singapore
-
Kontakt:
- Liang Zhong, PhD
- Telefonnummer: +65 67042237
- E-mail: gmszl@nus.edu.sg
-
-
-
-
Western Cape
-
Cape Town, Western Cape, Sydafrika, 7700
- University of Cape Town
-
Kontakt:
- Ntobeko Ntusi, MD, PhD
- Telefonnummer: +27 21 406 6358
- E-mail: ntobeko.ntusi@uct.ac.za
-
-
-
-
North Rhine-Westphalia
-
Bonn, North Rhine-Westphalia, Tyskland, 53113
- University of Bonn
-
Kontakt:
- Julian Luetkens, MD, PhD
- Telefonnummer: +49 228 287 19860
- E-mail: Julian.Luetkens@ukbonn.de
-
Cologne, North Rhine-Westphalia, Tyskland, 50931
- Cologne University Medical Center
-
Kontakt:
- David Maintz, MD, PhD
- Telefonnummer: +49 221 478-82035
- E-mail: david.maintz@uk-koeln.de
-
-
State of Berlin
-
Berlin, State of Berlin, Tyskland, 10117
- Charite - Universitatsmedizin Berlin
-
Kontakt:
- Sebastian Kelle, MD, PhD
- Telefonnummer: +493045931182
- E-mail: sebastian.kelle@dhzc-charite.de
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
- Barn
- Voksen
- Ældre voksen
Tager imod sunde frivillige
Prøveudtagningsmetode
Studiebefolkning
Beskrivelse
Inclusion Criteria:
- Diagnosed with cardiac disease of any complexity and scheduled for a clinical cardiac MRI on a clinical 1.5T MRI scanner
- Able to understand and provide written informed consent or, for minors, ability to provide assent with written consent from a parent/legal representative, per local law and site policy prior to the study
- Ability to undergo MRI without contraindications
- Agrees to be informed in the event of incidental findings
Exclusion Criteria:
- Severe claustrophobia preventing MRI completion
- Presence of non-MRI compatible implants (e.g., pacemakers, certain metallic implants)
- Contraindications to the intravenous contrast agent
- Movement disorders or inability to remain still during the scan
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
Kohorter og interventioner
Gruppe / kohorte |
|---|
|
Patients with congenital heart disease
FAST-CMR will evaluate the technique's performance in complex, heterogeneous anatomy.
|
|
Patients unable to perform breath-holding
including those with advanced heart failure, pulmonary hypertension, or chronic obstructive pulmonary disease (COPD), representing a subgroup where conventional breath-held CMR techniques are often suboptimal or not feasible.
|
|
Other cardiovascular disease patients who are able to perform breath-holding
This group represents the standard patient population for whom conventional CMR is currently optimized.
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
mean paired difference (bias) in left ventricular ejection fraction (LVEF) between 5D FISS-FRF and conventional 2D cine CMR
Tidsramme: Baseline (during study MRI acquisition)
|
estimated with a two-sided 95% confidence interval within each pre-defined patient cohort.
The study is designed to ensure that the half-width of this confidence interval does not exceed 1.25% LVEF per cohort.
|
Baseline (during study MRI acquisition)
|
Sekundære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Scan Efficiency
Tidsramme: Baseline (during study MRI acquisition)
|
|
Baseline (during study MRI acquisition)
|
|
Quantitative Image Quality
Tidsramme: Baseline (during study MRI acquisition)
|
o Contrast ratio and endocardial-to-blood-pool interface sharpness between 5D FISS-FRF and conventional 2D cine.
|
Baseline (during study MRI acquisition)
|
|
Qualitative Image Quality
Tidsramme: Baseline (during study MRI acquisition)
|
o Likert scale (0-4) consensus grading by ≥2 expert CMR readers blinded to images from both techniques.
|
Baseline (during study MRI acquisition)
|
|
Failure Rate
Tidsramme: Baseline (during study MRI acquisition)
|
o Percentage of non-diagnostic or excluded scans.
|
Baseline (during study MRI acquisition)
|
|
Anatomical Coverage
Tidsramme: Baseline (during study MRI acquisition)
|
o Extent of volumetric coverage of the entire heart and related great vessels.
|
Baseline (during study MRI acquisition)
|
|
Multiparametric Capability
Tidsramme: Baseline (during study MRI acquisition)
|
o Presence of additional diagnostic insights from respiratory motion tracking of FRF.
|
Baseline (during study MRI acquisition)
|
|
Clinical Acceptability of LVEF Measurements (±5% Threshold)
Tidsramme: Baseline (during study MRI acquisition)
|
o Exploratory analyses will assess whether observed agreement metrics are compatible with commonly used clinical acceptability thresholds (e.g., ±5% LVEF).
These analyses are hypothesis-generating and intended to inform future confirmatory studies.
|
Baseline (during study MRI acquisition)
|
|
LV mass
Tidsramme: Baseline (during study MRI acquisition)
|
Comparison between 5D FISS-FRF and 2D conventional cine
|
Baseline (during study MRI acquisition)
|
|
Regional wall motion abnormalities
Tidsramme: Baseline (during study MRI acquisition)
|
assessed using a 4-point likert scale per AHA segment (normal, hypokinesia, akinesia, dyskinesia)
|
Baseline (during study MRI acquisition)
|
|
Left and right atrial volumes
Tidsramme: Baseline (during study MRI acquisition)
|
maximum and minimum volumes and total atrial emptying fraction derived from conventional 2D cine long-axis views compared to 5D FISS-FRF
|
Baseline (during study MRI acquisition)
|
|
Agreement of Automated Ventricular Ejection Fraction Measurements (LVEF, RVEF)
Tidsramme: Baseline (during study MRI acquisition)
|
Agreement of automatically post-processed ventricular ejection fraction measurements (LVEF, RVEF) derived from 5D FISS-FRF with (1) standard post-processing of 5D FISS-FRF and (2) standard post-processing of conventional 2D cine cardiac MRI.
Agreement will be assessed using appropriate statistical metrics.
|
Baseline (during study MRI acquisition)
|
|
Agreement of Automated Ventricular Function Measurements (LVEDV, LVESV, LVSV, RVEDV, RVESV, RVSV)
Tidsramme: Baseline (during study MRI acquisition)
|
Agreement of automatically post-processed ventricular function metrics (LVEDV, LVESV, LVSV, RVEDV, RVESV, RVSV) derived from 5D FISS-FRF with (1) standard post-processing of 5D FISS-FRF and (2) standard post-processing of conventional 2D cine cardiac MRI.
Agreement will be assessed using appropriate statistical metrics.
|
Baseline (during study MRI acquisition)
|
|
Feasibility of AI-Based Image Reconstruction
Tidsramme: Baseline (during study MRI acquisition)
|
Percentage of reconstructed datasets that are considered clinically usable based on predefined quality criteria.
|
Baseline (during study MRI acquisition)
|
|
Reconstruction Time
Tidsramme: Baseline (during study MRI acquisition)
|
Comparison of reconstruction time between AI-based reconstruction and standard compressed sensing methods.
|
Baseline (during study MRI acquisition)
|
|
Accuracy of Ventricular Ejection Fraction Measurements
Tidsramme: Baseline (during study MRI acquisition)
|
Comparison of left and right ventricular ejection fraction (LVEF, RVEF) between AI-based and standard reconstructions.
|
Baseline (during study MRI acquisition)
|
|
Accuracy of Ventricular Volume Measurements
Tidsramme: Baseline (during study MRI acquisition)
|
Comparison of left and right ventricular end-diastolic volume (EDV), end-systolic volume (ESV), and stroke volume (SV) between AI-based and standard reconstructions.
|
Baseline (during study MRI acquisition)
|
|
Accuracy of Left Ventricular Mass Measurements
Tidsramme: Baseline (during study MRI acquisition)
|
Comparison of left ventricular mass between AI-based and standard reconstructions.
|
Baseline (during study MRI acquisition)
|
|
Voxel-wise Image Similarity Between Reconstruction Methods
Tidsramme: Baseline (during study MRI acquisition)
|
Image similarity assessed using structural similarity index measure (SSIM), peak signal-to-noise ratio (PSNR), and root mean squared error (RMSE).
|
Baseline (during study MRI acquisition)
|
|
Quantitative Image Quality
Tidsramme: Baseline (during study MRI acquisition)
|
Contrast ratio and endocardial-to-blood-pool interface sharpness between reconstruction methods.
|
Baseline (during study MRI acquisition)
|
|
Qualitative Image Quality
Tidsramme: Baseline (during study MRI acquisition)
|
Likert scale (0-4) consensus grading by ≥2 expert CMR readers blinded to images from both reconstruction methods.
|
Baseline (during study MRI acquisition)
|
Samarbejdspartnere og efterforskere
Sponsor
Efterforskere
- Ledende efterforsker: Matthias Stuber, PhD, CHUV
- Ledende efterforsker: Tim Leiner, MD, PhD, Mayo Clinic
- Ledende efterforsker: Kim-Lien Nguyen, MD, University of California, Los Angeles
Publikationer og nyttige links
Generelle publikationer
- Ogier AC, Baup S, Ilanjian G, Touray A, Rocca A, Banus J, Monton Quesada I, Nicoletti M, Ledoux JB, Richiardi J, Holtackers RJ, Yerly J, Stuber M, Hullin R, Rotzinger D, van Heeswijk RB. Cardiac function assessment with deep-learning-based automatic segmentation of free-running four-dimensional whole-heart cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2025 Dec 24;28(1):102677. doi: 10.1016/j.jocmr.2025.102677. Online ahead of print.
- Koktzoglou I, Edelman RR. Radial fast interrupted steady-state (FISS) magnetic resonance imaging. Magn Reson Med. 2018 Apr;79(4):2077-2086. doi: 10.1002/mrm.26881. Epub 2017 Aug 30.
- Yerly J, Roy CW, Milani B, Eyre K, Raifee MJ, Stuber M. High on sparsity: Interbin compensation of cardiac motion for improved assessment of left-ventricular function using 5D whole-heart MRI. Magn Reson Med. 2025 Mar;93(3):975-992. doi: 10.1002/mrm.30323. Epub 2024 Oct 9.
- Bastiaansen JAM, Piccini D, Di Sopra L, Roy CW, Heerfordt J, Edelman RR, Koktzoglou I, Yerly J, Stuber M. Natively fat-suppressed 5D whole-heart MRI with a radial free-running fast-interrupted steady-state (FISS) sequence at 1.5T and 3T. Magn Reson Med. 2020 Jan;83(1):45-55. doi: 10.1002/mrm.27942. Epub 2019 Aug 27.
- Di Sopra L, Piccini D, Coppo S, Stuber M, Yerly J. An automated approach to fully self-gated free-running cardiac and respiratory motion-resolved 5D whole-heart MRI. Magn Reson Med. 2019 Dec;82(6):2118-2132. doi: 10.1002/mrm.27898. Epub 2019 Jul 18.
Datoer for undersøgelser
Studer store datoer
Studiestart (Anslået)
Primær færdiggørelse (Anslået)
Studieafslutning (Anslået)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Faktiske)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Nøgleord
Yderligere relevante MeSH-vilkår
Andre undersøgelses-id-numre
- 2025-00923
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Studerer et amerikansk FDA-reguleret enhedsprodukt
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .
Kliniske forsøg med Hjertesygdomme
-
Assiut UniversityIkke rekrutterer endnuCardiac CT TOF
-
Alexion Pharmaceuticals, Inc.RekrutteringTransthyretin-type Cardiac AmyloidosisJapan
-
IRCCS Ospedale San RaffaeleIkke rekrutterer endnu
-
Impulse DynamicsRekrutteringEvaluer sikkerheden og anvendeligheden af ODOCOR II Intra-cardiac LeadSpanien, Tyskland, Italien
-
Ye ShengIkke rekrutterer endnuOut-of-Hospital Cardiac Arrest (Simuleret)Kina
-
Abbott Medical DevicesAfsluttetPatienten opfylder ACC/AHA/ESC-retningslinjerne for implanterbar cardioverter-defibrillator (ICD) eller cardiac resynchronization therapy (CRT-D)-enhedDet Forenede Kongerige, Tyskland
-
Reham SameehAssiut UniversityUkendtCardiac Magnetic Resonance Imaging i ikke-iskæmisk kardiomyopati
-
Xiangya Hospital of Central South UniversityRekrutteringHjertestop | Post-Cardiac Arrest CareKina
-
Erzincan Binali Yildirim UniversitesiAfsluttetSkulder dysfunktion | Cardiac Implantable Electronic Device (CIED)Tyrkiet (Türkiye)
-
Milton S. Hershey Medical CenterRekrutteringVenstre ventrikulær dysfunktion | Spinalbedøvelsesmidler, der forårsager uønskede virkninger ved terapeutisk brug | Global Cardiac Wall Motion DysfunktionForenede Stater