- ICH GCP
- Registre américain des essais cliniques
- Essai clinique NCT07951281
Performance of Lumetry for Arterial Carbon Dioxide (paCO2) Assessment and Chronic Obstructive Pulmonary Disease (COPD) Screening
Evaluation of the Clinical Performance of Lumetry for paCO2 Assessment and COPD Screening
The purpose of this clinical study is to evaluate how well the lumetry device performs in two areas: estimating the level of carbon dioxide in the arterial blood (paCO2) and screening for chronic obstructive pulmonary disease (COPD).
lumetry is a small handheld device that analyzes exhaled breath. Participants breathe normally into the device for 60 seconds. No forced or maximal breathing maneuvers are required for the lumetry measurement.
Today, the carbon dioxide level in the blood is usually measured with a blood gas analysis, which requires a blood sample. COPD is usually diagnosed with spirometry, a lung function test that requires maximal breathing effort and trained staff.
In this study, adults with COPD of different severity and adults without known lung disease each attend a single study visit. All participants perform one lumetry breath measurement and a standard lung function test (forced spirometry). Participants with COPD also give a small capillary blood sample from the earlobe for blood gas analysis, and some of them additionally have their carbon dioxide level measured through the skin (transcutaneous measurement). By comparing the lumetry results with these established methods, the study assesses how accurately the device estimates the carbon dioxide level in the blood. By comparing the lumetry COPD screening result with spirometry and the confirmed clinical diagnosis, the study assesses how reliably the device distinguishes between people with and without COPD.
The study is conducted at one hospital in Vienna, Austria, and will include 300 participants.
Aperçu de l'étude
Statut
Les conditions
Type d'étude
Inscription (Estimé)
Phase
- N'est pas applicable
Contacts et emplacements
Coordonnées de l'étude
- Nom: Ruth Masih, MSc Biotechnology
- Numéro de téléphone: +4366488262391
- E-mail: ruth.masih@lumetry.at
Sauvegarde des contacts de l'étude
- Nom: Nejla Hidic, BSc Biomedical Engineering
- Numéro de téléphone: +4366488262391
- E-mail: nejla.hidic@lumetry.at
Lieux d'étude
-
-
State of Vienna
-
Vienna, State of Vienna, L'Autriche, A-1210
- Klinik Floridsdorf, Department of Respiratory and Critical Care Medicine
-
Contact:
- Arschang Valipour, Prim. Priv. -Doz. Dr.
- E-mail: arschang.valipour@gesundheitsverbund.at
-
Contact:
- Theresa Klemm, Medical Doctor
- E-mail: theresa.klemm@gesundheitsverbund.at
-
-
Critères de participation
Critère d'éligibilité
Âges éligibles pour étudier
Accepte les volontaires sains
Sexes éligibles pour l'étude
La description
Inclusion Criteria:
- Written informed consent
- Age between 18 and 90 years
- Diagnosed chronic obstructive pulmonary disease (COPD), stages GOLD I to IV, or no known lung function pathologies
- Able to sit upright for the duration of the procedure without exhaustion
- Women of childbearing potential: not breastfeeding and either a negative pregnancy test prior to the start of the investigation or use of a reliable method of contraception (intrauterine device or oral contraceptive pill) throughout the investigation. Women of childbearing potential are defined as women who have not undergone surgical sterilization (hysterectomy or bilateral oophorectomy) and are not postmenopausal (at least 12 months of amenorrhea in a woman over the age of 50 in the absence of other biological or physiological causes).
Exclusion Criteria:
- Considered by the investigator to be inappropriate for inclusion or unlikely to comply with the requirements of the study
- Lung disease other than COPD (e.g., asthma); participants with asthma-COPD overlap may be included
- Previously documented chronic heart failure with an ejection fraction below 40%
- Mental or physical inability to comply with time schedules and study procedures
- Current participation in a clinical study that may affect breathing behaviour (medicinal product or medical device), or participation within 30 days prior to screening
- Prior participation in this clinical investigation
- Possible or confirmed pregnancy, or breastfeeding
Plan d'étude
Comment l'étude est-elle conçue ?
Détails de conception
- Objectif principal: Diagnostique
- Répartition: N / A
- Modèle interventionnel: Affectation à un seul groupe
- Masquage: Aucun (étiquette ouverte)
Armes et Interventions
Groupe de participants / Bras |
Intervention / Traitement |
|---|---|
|
Expérimental: Participants With and Without COPD
Adults with chronic obstructive pulmonary disease (COPD), stratified by severity (50 GOLD I/II, 50 GOLD III, 50 GOLD IV), and 150 adults without known lung function pathologies. All participants perform a single 60-second lumetry measurement of normal resting breathing, followed by forced spirometry. Participants with COPD additionally undergo arterialized capillary blood gas analysis, with a subgroup of 50 participants with COPD (GOLD III and IV) additionally undergoing transcutaneous carbon dioxide monitoring. |
Handheld capnography and spirometry device that estimates the arterial partial pressure of carbon dioxide (paCO2) and provides a COPD screening result from a single 60-second measurement of normal resting breathing. The device also provides the end-tidal carbon dioxide value (petCO2), which serves as comparator for the paCO2 estimate. Capillary blood sample from the arterialized earlobe, collected immediately after the lumetry measurement. Reference standard for paCO2. Performed in participants with COPD only. Non-invasive measurement of carbon dioxide through the skin (tcPCO2), recorded as close in time as possible to the lumetry measurement and blood sampling. Comparator for paCO2 estimation. Performed in a subgroup of 50 participants with COPD. Standard lung function test performed at the end of the study visit in all participants. Together with the documented clinical diagnosis, reference standard for the COPD screening result. |
Que mesure l'étude ?
Principaux critères de jugement
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Difference in Mean Absolute Deviation From Blood Gas paCO2 Between lumetry-Estimated paCO2 and End-Tidal Carbon Dioxide (petCO2)
Délai: Day 1 (single study visit)
|
In participants with chronic obstructive pulmonary disease (COPD), the carbon dioxide level in the arterial blood (paCO2) estimated by lumetry is compared with the value measured in a capillary blood sample (blood gas analysis, reference standard). The average size of the difference, regardless of direction (mean absolute deviation), is calculated for lumetry and for the standard breath-based value (end-tidal carbon dioxide, petCO2). The outcome is the MAD of petCO2 minus the MAD of lumetry, so a positive value means lumetry is closer to the blood gas value. |
Day 1 (single study visit)
|
|
Sensitivity of the lumetry COPD Screening Result
Délai: Day 1 (single study visit)
|
Percentage of participants with COPD, as established by forced spirometry and documented clinical diagnosis (reference standard), who are classified as positive by the lumetry COPD screening result. |
Day 1 (single study visit)
|
|
Specificity of the lumetry COPD Screening Result
Délai: Day 1 (single study visit)
|
Percentage of participants without COPD, as established by forced spirometry and documented clinical diagnosis (reference standard), who are classified as negative by the lumetry COPD screening result. |
Day 1 (single study visit)
|
Mesures de résultats secondaires
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Standard Deviation of the Differences Between lumetry-Estimated paCO2 and Blood Gas paCO2
Délai: Day 1 (single study visit)
|
In participants with chronic obstructive pulmonary disease (COPD), the difference between the lumetry-estimated arterial carbon dioxide level (paCO2) and the paCO2 measured by capillary blood gas analysis (reference standard) is calculated for each participant. The outcome is the standard deviation of these differences, describing the precision of the lumetry estimate. Unit: millimeters of mercury (mmHg). |
Day 1 (single study visit)
|
|
Percentage of lumetry paCO2 Estimates Within 8 mmHg of Blood Gas paCO2
Délai: Day 1 (single study visit)
|
In participants with COPD, the percentage of lumetry paCO2 estimates that differ by 8.0 millimeters of mercury (mmHg) or less in either direction from the paCO2 measured by capillary blood gas analysis (reference standard). |
Day 1 (single study visit)
|
|
Difference in Mean Absolute Deviation From Blood Gas paCO2 Between lumetry-Estimated paCO2 and Transcutaneous Carbon Dioxide (tcPCO2)
Délai: Day 1 (single study visit)
|
In a subgroup of 50 participants with COPD, the average size of the difference, regardless of direction (mean absolute deviation), from the paCO2 measured by capillary blood gas analysis (reference standard) is calculated for lumetry and for transcutaneous carbon dioxide monitoring (tcPCO2). The outcome is the difference between these two averages. Unit: millimeters of mercury (mmHg). |
Day 1 (single study visit)
|
Autres mesures de résultats
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Number of Participants With Adverse Events, Adverse Device Effects, or Device Deficiencies
Délai: Day 1 (single study visit)
|
Number of participants with at least one adverse event, adverse device effect, or device deficiency related to the study procedures, recorded during the study visit. |
Day 1 (single study visit)
|
|
Cross-Validated Sensitivity of a Re-Trained lumetry COPD Screening Algorithm
Délai: Day 1 (single study visit)
|
Exploratory analysis. Sensitivity of a re-trained version of the lumetry machine learning algorithm for identifying participants with chronic obstructive pulmonary disease (COPD), estimated by stratified repeated k-fold cross-validation using the study data. Reference standard: forced spirometry and documented clinical diagnosis. |
Day 1 (single study visit)
|
|
Cross-Validated Specificity of a Re-Trained lumetry COPD Screening Algorithm
Délai: Day 1 (single study visit)
|
Exploratory analysis. Specificity of a re-trained version of the lumetry machine learning algorithm for identifying participants without COPD, estimated by stratified repeated k-fold cross-validation using the study data. Reference standard: forced spirometry and documented clinical diagnosis. |
Day 1 (single study visit)
|
|
Cross-Validated Mean Absolute Deviation of Re-Trained lumetry paCO2 Estimates From Blood Gas paCO2
Délai: Day 1 (single study visit)
|
Exploratory analysis. Average size of the difference, regardless of direction (mean absolute deviation), between the arterial carbon dioxide level (paCO2) estimated by a re-trained version of the lumetry machine learning algorithm and the paCO2 measured by capillary blood gas analysis, estimated by stratified repeated k-fold cross-validation in participants with COPD. Unit: millimeters of mercury (mmHg). |
Day 1 (single study visit)
|
Collaborateurs et enquêteurs
Parrainer
Collaborateurs
Les enquêteurs
- Chercheur principal: Arschang Valipour, Prim. Priv. -Doz. Dr., Klinik Floridsdorf, Department of Respiratory and Critical Care Medicine
Publications et liens utiles
Publications générales
- Klemm et al. Predicting arterial pCO2 using capnographic recordings - A pilot study using a novel breath analysis device. Eur Respir J. 2025; 66(Suppl 69): PA2529. doi:10.1183/13993003.congress-2025.PA2529
Dates d'enregistrement des études
Dates principales de l'étude
Début de l'étude (Estimé)
Achèvement primaire (Estimé)
Achèvement de l'étude (Estimé)
Dates d'inscription aux études
Première soumission
Première soumission répondant aux critères de contrôle qualité
Première publication (Réel)
Mises à jour des dossiers d'étude
Dernière mise à jour publiée (Réel)
Dernière mise à jour soumise répondant aux critères de contrôle qualité
Dernière vérification
Plus d'information
Termes liés à cette étude
Mots clés
Termes MeSH pertinents supplémentaires
- Processus pathologiques
- Maladie chronique
- Attributs de la maladie
- Maladies des voies respiratoires
- Maladies pulmonaires
- Maladies pulmonaires obstructives
- Signes et symptômes respiratoires
- Conditions pathologiques, signes et symptômes
- Signes et symptômes
- Maladie pulmonaire obstructive chronique
- Hypercapnie
- Techniques d'investigation
- Techniques de laboratoire clinique
- Techniques et procédures de diagnostic
- Diagnostic
- Techniques de diagnostic, système respiratoire
- Techniques de diagnostic, cardiovasculaire
- Tests de la fonction cardiaque
- Tests de fonction respiratoire
- Analyse chimique du sang
- Tests de chimie clinique
- Analyse des gaz du sang
- Oxymétrie
- Surveillance des gaz du sang, transcutanée
Autres numéros d'identification d'étude
- lumetry_03
Plan pour les données individuelles des participants (IPD)
Prévoyez-vous de partager les données individuelles des participants (DPI) ?
Informations sur les médicaments et les dispositifs, documents d'étude
Étudie un produit pharmaceutique réglementé par la FDA américaine
Étudie un produit d'appareil réglementé par la FDA américaine
Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .