Performance of Lumetry for Arterial Carbon Dioxide (paCO2) Assessment and Chronic Obstructive Pulmonary Disease (COPD) Screening

16 novembre 2026 mis à jour par: Lumetry Diagnostics GmbH

Evaluation of the Clinical Performance of Lumetry for paCO2 Assessment and COPD Screening

The purpose of this clinical study is to evaluate how well the lumetry device performs in two areas: estimating the level of carbon dioxide in the arterial blood (paCO2) and screening for chronic obstructive pulmonary disease (COPD).

lumetry is a small handheld device that analyzes exhaled breath. Participants breathe normally into the device for 60 seconds. No forced or maximal breathing maneuvers are required for the lumetry measurement.

Today, the carbon dioxide level in the blood is usually measured with a blood gas analysis, which requires a blood sample. COPD is usually diagnosed with spirometry, a lung function test that requires maximal breathing effort and trained staff.

In this study, adults with COPD of different severity and adults without known lung disease each attend a single study visit. All participants perform one lumetry breath measurement and a standard lung function test (forced spirometry). Participants with COPD also give a small capillary blood sample from the earlobe for blood gas analysis, and some of them additionally have their carbon dioxide level measured through the skin (transcutaneous measurement). By comparing the lumetry results with these established methods, the study assesses how accurately the device estimates the carbon dioxide level in the blood. By comparing the lumetry COPD screening result with spirometry and the confirmed clinical diagnosis, the study assesses how reliably the device distinguishes between people with and without COPD.

The study is conducted at one hospital in Vienna, Austria, and will include 300 participants.

Aperçu de l'étude

Type d'étude

Interventionnel

Inscription (Estimé)

300

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Coordonnées de l'étude

  • Nom: Ruth Masih, MSc Biotechnology
  • Numéro de téléphone: +4366488262391
  • E-mail: ruth.masih@lumetry.at

Sauvegarde des contacts de l'étude

  • Nom: Nejla Hidic, BSc Biomedical Engineering
  • Numéro de téléphone: +4366488262391
  • E-mail: nejla.hidic@lumetry.at

Lieux d'étude

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

18 ans à 90 ans (Adulte, Adulte plus âgé)

Accepte les volontaires sains

Oui

Sexes éligibles pour l'étude

Tout

La description

Inclusion Criteria:

  • Written informed consent
  • Age between 18 and 90 years
  • Diagnosed chronic obstructive pulmonary disease (COPD), stages GOLD I to IV, or no known lung function pathologies
  • Able to sit upright for the duration of the procedure without exhaustion
  • Women of childbearing potential: not breastfeeding and either a negative pregnancy test prior to the start of the investigation or use of a reliable method of contraception (intrauterine device or oral contraceptive pill) throughout the investigation. Women of childbearing potential are defined as women who have not undergone surgical sterilization (hysterectomy or bilateral oophorectomy) and are not postmenopausal (at least 12 months of amenorrhea in a woman over the age of 50 in the absence of other biological or physiological causes).

Exclusion Criteria:

  • Considered by the investigator to be inappropriate for inclusion or unlikely to comply with the requirements of the study
  • Lung disease other than COPD (e.g., asthma); participants with asthma-COPD overlap may be included
  • Previously documented chronic heart failure with an ejection fraction below 40%
  • Mental or physical inability to comply with time schedules and study procedures
  • Current participation in a clinical study that may affect breathing behaviour (medicinal product or medical device), or participation within 30 days prior to screening
  • Prior participation in this clinical investigation
  • Possible or confirmed pregnancy, or breastfeeding

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Diagnostique
  • Répartition: N / A
  • Modèle interventionnel: Affectation à un seul groupe
  • Masquage: Aucun (étiquette ouverte)

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Expérimental: Participants With and Without COPD

Adults with chronic obstructive pulmonary disease (COPD), stratified by severity (50 GOLD I/II, 50 GOLD III, 50 GOLD IV), and 150 adults without known lung function pathologies. All participants perform a single 60-second lumetry measurement of normal resting breathing, followed by forced spirometry. Participants with COPD additionally undergo arterialized capillary blood gas analysis, with a subgroup of 50 participants with COPD (GOLD III and IV) additionally undergoing transcutaneous carbon dioxide monitoring.

Handheld capnography and spirometry device that estimates the arterial partial pressure of carbon dioxide (paCO2) and provides a COPD screening result from a single 60-second measurement of normal resting breathing. The device also provides the end-tidal carbon dioxide value (petCO2), which serves as comparator for the paCO2 estimate.

Capillary blood sample from the arterialized earlobe, collected immediately after the lumetry measurement. Reference standard for paCO2. Performed in participants with COPD only.

Non-invasive measurement of carbon dioxide through the skin (tcPCO2), recorded as close in time as possible to the lumetry measurement and blood sampling. Comparator for paCO2 estimation. Performed in a subgroup of 50 participants with COPD.

Standard lung function test performed at the end of the study visit in all participants. Together with the documented clinical diagnosis, reference standard for the COPD screening result.

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Difference in Mean Absolute Deviation From Blood Gas paCO2 Between lumetry-Estimated paCO2 and End-Tidal Carbon Dioxide (petCO2)
Délai: Day 1 (single study visit)

In participants with chronic obstructive pulmonary disease (COPD), the carbon dioxide level in the arterial blood (paCO2) estimated by lumetry is compared with the value measured in a capillary blood sample (blood gas analysis, reference standard). The average size of the difference, regardless of direction (mean absolute deviation), is calculated for lumetry and for the standard breath-based value (end-tidal carbon dioxide, petCO2). The outcome is the MAD of petCO2 minus the MAD of lumetry, so a positive value means lumetry is closer to the blood gas value.

Day 1 (single study visit)
Sensitivity of the lumetry COPD Screening Result
Délai: Day 1 (single study visit)

Percentage of participants with COPD, as established by forced spirometry and documented clinical diagnosis (reference standard), who are classified as positive by the lumetry COPD screening result.

Day 1 (single study visit)
Specificity of the lumetry COPD Screening Result
Délai: Day 1 (single study visit)

Percentage of participants without COPD, as established by forced spirometry and documented clinical diagnosis (reference standard), who are classified as negative by the lumetry COPD screening result.

Day 1 (single study visit)

Mesures de résultats secondaires

Mesure des résultats
Description de la mesure
Délai
Standard Deviation of the Differences Between lumetry-Estimated paCO2 and Blood Gas paCO2
Délai: Day 1 (single study visit)

In participants with chronic obstructive pulmonary disease (COPD), the difference between the lumetry-estimated arterial carbon dioxide level (paCO2) and the paCO2 measured by capillary blood gas analysis (reference standard) is calculated for each participant. The outcome is the standard deviation of these differences, describing the precision of the lumetry estimate. Unit: millimeters of mercury (mmHg).

Day 1 (single study visit)
Percentage of lumetry paCO2 Estimates Within 8 mmHg of Blood Gas paCO2
Délai: Day 1 (single study visit)

In participants with COPD, the percentage of lumetry paCO2 estimates that differ by 8.0 millimeters of mercury (mmHg) or less in either direction from the paCO2 measured by capillary blood gas analysis (reference standard).

Day 1 (single study visit)
Difference in Mean Absolute Deviation From Blood Gas paCO2 Between lumetry-Estimated paCO2 and Transcutaneous Carbon Dioxide (tcPCO2)
Délai: Day 1 (single study visit)

In a subgroup of 50 participants with COPD, the average size of the difference, regardless of direction (mean absolute deviation), from the paCO2 measured by capillary blood gas analysis (reference standard) is calculated for lumetry and for transcutaneous carbon dioxide monitoring (tcPCO2). The outcome is the difference between these two averages. Unit: millimeters of mercury (mmHg).

Day 1 (single study visit)

Autres mesures de résultats

Mesure des résultats
Description de la mesure
Délai
Number of Participants With Adverse Events, Adverse Device Effects, or Device Deficiencies
Délai: Day 1 (single study visit)

Number of participants with at least one adverse event, adverse device effect, or device deficiency related to the study procedures, recorded during the study visit.

Day 1 (single study visit)
Cross-Validated Sensitivity of a Re-Trained lumetry COPD Screening Algorithm
Délai: Day 1 (single study visit)

Exploratory analysis. Sensitivity of a re-trained version of the lumetry machine learning algorithm for identifying participants with chronic obstructive pulmonary disease (COPD), estimated by stratified repeated k-fold cross-validation using the study data. Reference standard: forced spirometry and documented clinical diagnosis.

Day 1 (single study visit)
Cross-Validated Specificity of a Re-Trained lumetry COPD Screening Algorithm
Délai: Day 1 (single study visit)

Exploratory analysis. Specificity of a re-trained version of the lumetry machine learning algorithm for identifying participants without COPD, estimated by stratified repeated k-fold cross-validation using the study data. Reference standard: forced spirometry and documented clinical diagnosis.

Day 1 (single study visit)
Cross-Validated Mean Absolute Deviation of Re-Trained lumetry paCO2 Estimates From Blood Gas paCO2
Délai: Day 1 (single study visit)

Exploratory analysis. Average size of the difference, regardless of direction (mean absolute deviation), between the arterial carbon dioxide level (paCO2) estimated by a re-trained version of the lumetry machine learning algorithm and the paCO2 measured by capillary blood gas analysis, estimated by stratified repeated k-fold cross-validation in participants with COPD. Unit: millimeters of mercury (mmHg).

Day 1 (single study visit)

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Les enquêteurs

  • Chercheur principal: Arschang Valipour, Prim. Priv. -Doz. Dr., Klinik Floridsdorf, Department of Respiratory and Critical Care Medicine

Publications et liens utiles

La personne responsable de la saisie des informations sur l'étude fournit volontairement ces publications. Il peut s'agir de tout ce qui concerne l'étude.

Publications générales

  • Klemm et al. Predicting arterial pCO2 using capnographic recordings - A pilot study using a novel breath analysis device. Eur Respir J. 2025; 66(Suppl 69): PA2529. doi:10.1183/13993003.congress-2025.PA2529

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Estimé)

1 décembre 2026

Achèvement primaire (Estimé)

1 décembre 2027

Achèvement de l'étude (Estimé)

1 décembre 2027

Dates d'inscription aux études

Première soumission

16 novembre 2026

Première soumission répondant aux critères de contrôle qualité

16 novembre 2026

Première publication (Réel)

20 novembre 2026

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

20 novembre 2026

Dernière mise à jour soumise répondant aux critères de contrôle qualité

16 novembre 2026

Dernière vérification

1 novembre 2026

Plus d'information

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .

S'abonner