Evaluation of the Clinical Benefit of a Transmitter for Contralateral Routing of Signals (CROS)
調査の概要
詳細な説明
Unilateral Hearing Loss (UHL) can be defined as any degree of permanent hearing loss on one ear with normal hearing in the opposite ear. Unilateral hearing loss can be debilitating associated with audiological, psychosocial and educational challenges. Audiological challenges include the reduced ability to localize sounds, reduced awareness on the unaidable side and difficulties hearing in noise or at a distance.
One solution for unaidable UHL is Contralateral Routing of Signals (CROS) and Bilateral Contralateral routing of signals (BiCROS) using a so- called CROS transmitter together with a hearing aid.
The rationale for this clinical investigation is to collect clinical data with a rechargeable CROS transmitter to evaluate the benefits of a BiCROS fitting compared to the alternative treatment option, the monaural hearing aid fitting, and to no treatment in noisy listening situations.
研究の種類
入学 (実際)
段階
- 適用できない
連絡先と場所
研究場所
-
-
Zürich
-
Stäfa、Zürich、スイス、8712
- Sonova AG
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- hearing impaired adults, with & w/o hearing aids (experiences)
- good written and spoken (Swiss) German language skills
- ability to fill in a questionnaire conscientiously
- informed consent as documented by signature
- asymmetric hearing loss: one unaidable ear (min. N6 hearing loss) / the other ear with N3 or N4 hearing loss
- availability of Smartphone
- availability of TV or Tablet
- ability to operate and handle the investigational device
- ability to describe listening impressions/experiences and the use of the hearing aid
- Ability to produce a reliable hearing test result
Exclusion Criteria:
- Contraindications to the medical device:
- aidable hearing loss on the intended CROS side
- acute tinnitus (in either one ear and occurrence less than three months)
- anatomical deformity of the CROS ear (e.g. absence of the auricle)
- Known hypersensitivity or allergy to materials of the investigational device
- any auricle deformities, that prevents secure placement of the (test) devices
- no willingness to test the investigational device system
- no willingness to test the investigational device system for at least 8h a day
- Hint for psychological problems from investigators point of view
- reported fluctuating hearing that could influence the test result
- reported symptoms of vertigo and dizziness by participants
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:他の
- 割り当て:なし
- 介入モデル:単一グループの割り当て
- マスキング:なし(オープンラベル)
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
実験的:Speech Intelligibility with CROS device
The focus of this study is on Speech Intelligibility (SI), evaluated by the Oldenburg Sentence test (OLSA), which measures a speech recognition threshold (SRT) in dB SNR (signal to noise ratio). Therefore each participant will perform the tests with the experimental rechargeable CROS transmitter (CROS) in different interventions, like comparison to monaural fitting and unaided condition. All participants will perform the same tests. The order of the intervention in the speech test is randomized, but will be performed in the same visit by each participant. |
治療はありません。つまり、参加者はCROSデバイスおよび/または補聴器を装着していません。 補助のない状態は、制御介入として機能します。
Each participant will be fitted with the experimental CROS device on the poorer, unaidable ear and with a compatible hearing device on the other ear.
Each participant will be fitted with a hearing aid on the better hearing ear and no device on the poorer, unaidable ear. Monaural hearing aid fitting acts as control intervention. Each participant will be fitted with the experimental CROS device on the poorer, unaidable ear and with a compatible device on the other ear. The difference to the BiCROS condition is the setup of the microphones in the CROS device and the compatible hearing device. This test condition is only tested for speech intelligibility in diffuse noise and acts as control intervention. |
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Speech Intelligibility (SI) in noise (fix direction): BiCROS fitting vs. monaural hearing aid fitting
時間枠:2 weeks
|
The primary objective of this clinical investigation is to evaluate whether a BiCROS fitting (better ear fitted with a hearing device, unaidable ear supplied with a CROS device) improves speech intelligibility in noise compared to a monaural fitting (just the aidable ear is supplied by a hearing aid), for adults with an asymmetric hearing loss, measured with the Oldenburg Sentence Test (OLSA). Results of the speech test will be the SRT (speech recognition threshold) in dB SNR (signal to noise ratio). The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A). The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A). |
2 weeks
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
SI in noise (fix direction): BiCROS fitting vs. unaided
時間枠:2 week
|
A secondary objective of this clinical investigation is to evaluate whether a BiCROS fitting improves speech intelligibility in noise compared to an unaided condition, for adults with an asymmetric hearing loss, measured with the Oldenburg Sentence Test (OLSA). Results of the speech test will be the SRT (dB SNR). The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A). The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A). |
2 week
|
|
SI in diffuse noise: BiCROS fitting vs. monaural fitting
時間枠:2 weeks
|
A secondary objective of this clinical investigation is to evaluate whether a BiCROS fitting improves speech intelligibility by at least 1 dB SNR for 50% SRT in diffuse noise, compared to a monaural hearing aid fitting. The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A). The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A). |
2 weeks
|
|
SI in diffuse noise: BiCROS fitting vs. unaided
時間枠:2 weeks
|
A secondary objective of this clinical investigation is to evaluate whether a BiCROS fitting improves speech intelligibility by at least 1 dB SNR for 50% SRT in diffuse noise, compared to an unaided condition. The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A). The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A). |
2 weeks
|
|
SI in diffuse noise: BiCROS vs. CROS baseline fitting
時間枠:2 weeks
|
A secondary objective of this clinical investigation is to evaluate whether a BiCROS fitting improves speech intelligibility in diffuse noise compared with a CROS baseline fitting. CROS baseline means a fitting of an additional program with different hearing instrument microphone setup. The SI will be tested with the OLSA in diffuse noise. The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A). The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A). |
2 weeks
|
|
SI in diffuse noise: CROS baseline fitting vs. unaided
時間枠:2 weeks
|
A secondary objective of this clinical investigation is to evaluate whether a CROS baseline fitting improves speech intelligibility in diffuse noise compared with an unaided condition. The SI will be tested with the OLSA in diffuse noise. The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A). The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A). |
2 weeks
|
|
SI in diffuse noise: CROS baseline fitting vs. monaural fitting
時間枠:2 weeks
|
A secondary objective of this clinical investigation is to evaluate whether a CROS baseline fitting improves speech intelligibility in diffuse noise compared with a monaural fitting. The SI will be tested with the OLSA in diffuse noise. The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A). The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A). |
2 weeks
|
|
SP in diffuse noise: monaural fitting vs. unaided
時間枠:2 weeks
|
A secondary objective of this clinical investigation is to evaluate whether a monaural fitting improves speech intelligibility in diffuse noise compared with an unaided condition. The SI will be tested with the OLSA in diffuse noise. The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A). The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A). |
2 weeks
|
|
Evaluation of Sound quality
時間枠:3 weeks
|
A secondary objective of this study is to evaluate the participant's satisfaction of the sound quality of the experimental CROS device by a subjective rating during a three-week home trial period. Scale: very unsatisfied- unsatisfied- neither nor- satisfied- very satisfied |
3 weeks
|
|
Evaluation of System stability
時間枠:3 weeks
|
A secondary objective of this study is to evaluate the participant's satisfaction of the system stability of the experimental CROS device during a three-week home trial period. The participant shall report, if there are artefacts audible: never/ 1-2 times a week/ 3-6 times a week/ daily |
3 weeks
|
|
Evaluation of streaming
時間枠:3 weeks
|
A secondary objective of this study is to evaluate the participant's satisfaction of the streaming sound quality of the experimental CROS device, by a subjective rating during a three-week home trial period. Scale: very unsatisfied- unsatisfied- neither nor- satisfied- very satisfied |
3 weeks
|
|
Evaluation of listening effort
時間枠:3 weeks
|
A secondary objective is to evaluate the subjective listening effort in noise reported by the participant whilst wearing the experimental CROS device. Scale: not at all - slightly - somewhat - moderately - very - extremely |
3 weeks
|
協力者と研究者
スポンサー
出版物と役立つリンク
一般刊行物
- Bagatto M, DesGeorges J, King A, Kitterick P, Laurnagaray D, Lewis D, Roush P, Sladen DP, Tharpe AM. Consensus practice parameter: audiological assessment and management of unilateral hearing loss in children. Int J Audiol. 2019 Dec;58(12):805-815. doi: 10.1080/14992027.2019.1654620. Epub 2019 Sep 5.
- Snapp HA, Hoffer ME, Spahr A, Rajguru S. Application of Wireless Contralateral Routing of Signal Technology in Unilateral Cochlear Implant Users with Bilateral Profound Hearing Loss. J Am Acad Audiol. 2019 Jul/Aug;30(7):579-589. doi: 10.3766/jaaa.17121. Epub 2018 Dec 13.
- Snapp H. Nonsurgical Management of Single-Sided Deafness: Contralateral Routing of Signal. J Neurol Surg B Skull Base. 2019 Apr;80(2):132-138. doi: 10.1055/s-0039-1677687. Epub 2019 Jan 17.
- Lieu JE, Tye-Murray N, Karzon RK, Piccirillo JF. Unilateral hearing loss is associated with worse speech-language scores in children. Pediatrics. 2010 Jun;125(6):e1348-55. doi: 10.1542/peds.2009-2448. Epub 2010 May 10.
- McKay S, Gravel JS, Tharpe AM. Amplification considerations for children with minimal or mild bilateral hearing loss and unilateral hearing loss. Trends Amplif. 2008 Mar;12(1):43-54. doi: 10.1177/1084713807313570.
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。