Evaluation of the Clinical Benefit of a Transmitter for Contralateral Routing of Signals (CROS)

8. april 2022 oppdatert av: Sonova AG
This study will evaluate the clinical benefit of a transmitter for contralateral routing of signals. The benefit will be evaluated in noisy environments regarding speech intelligibility when the CROS system is adjusted to different microphone settings. Additionally, data regarding overall system stability, crash reboot rate, sound quality and connectivity will be obtained over a period of time to validate the CROS system in combination with smartphone and accessories. This study is a confirmatory study.

Studieoversikt

Detaljert beskrivelse

Unilateral Hearing Loss (UHL) can be defined as any degree of permanent hearing loss on one ear with normal hearing in the opposite ear. Unilateral hearing loss can be debilitating associated with audiological, psychosocial and educational challenges. Audiological challenges include the reduced ability to localize sounds, reduced awareness on the unaidable side and difficulties hearing in noise or at a distance.

One solution for unaidable UHL is Contralateral Routing of Signals (CROS) and Bilateral Contralateral routing of signals (BiCROS) using a so- called CROS transmitter together with a hearing aid.

The rationale for this clinical investigation is to collect clinical data with a rechargeable CROS transmitter to evaluate the benefits of a BiCROS fitting compared to the alternative treatment option, the monaural hearing aid fitting, and to no treatment in noisy listening situations.

Studietype

Intervensjonell

Registrering (Faktiske)

23

Fase

  • Ikke aktuelt

Kontakter og plasseringer

Denne delen inneholder kontaktinformasjon for de som utfører studien, og informasjon om hvor denne studien blir utført.

Studiesteder

    • Zürich
      • Stäfa, Zürich, Sveits, 8712
        • Sonova AG

Deltakelseskriterier

Forskere ser etter personer som passer til en bestemt beskrivelse, kalt kvalifikasjonskriterier. Noen eksempler på disse kriteriene er en persons generelle helsetilstand eller tidligere behandlinger.

Kvalifikasjonskriterier

Alder som er kvalifisert for studier

18 år og eldre (Voksen, Eldre voksen)

Tar imot friske frivillige

Nei

Kjønn som er kvalifisert for studier

Alle

Beskrivelse

Inclusion Criteria:

  • hearing impaired adults, with & w/o hearing aids (experiences)
  • good written and spoken (Swiss) German language skills
  • ability to fill in a questionnaire conscientiously
  • informed consent as documented by signature
  • asymmetric hearing loss: one unaidable ear (min. N6 hearing loss) / the other ear with N3 or N4 hearing loss
  • availability of Smartphone
  • availability of TV or Tablet
  • ability to operate and handle the investigational device
  • ability to describe listening impressions/experiences and the use of the hearing aid
  • Ability to produce a reliable hearing test result

Exclusion Criteria:

  • Contraindications to the medical device:
  • aidable hearing loss on the intended CROS side
  • acute tinnitus (in either one ear and occurrence less than three months)
  • anatomical deformity of the CROS ear (e.g. absence of the auricle)
  • Known hypersensitivity or allergy to materials of the investigational device
  • any auricle deformities, that prevents secure placement of the (test) devices
  • no willingness to test the investigational device system
  • no willingness to test the investigational device system for at least 8h a day
  • Hint for psychological problems from investigators point of view
  • reported fluctuating hearing that could influence the test result
  • reported symptoms of vertigo and dizziness by participants

Studieplan

Denne delen gir detaljer om studieplanen, inkludert hvordan studien er utformet og hva studien måler.

Hvordan er studiet utformet?

Designdetaljer

  • Primært formål: Annen
  • Tildeling: N/A
  • Intervensjonsmodell: Enkeltgruppeoppdrag
  • Masking: Ingen (Open Label)

Våpen og intervensjoner

Deltakergruppe / Arm
Intervensjon / Behandling
Eksperimentell: Speech Intelligibility with CROS device

The focus of this study is on Speech Intelligibility (SI), evaluated by the Oldenburg Sentence test (OLSA), which measures a speech recognition threshold (SRT) in dB SNR (signal to noise ratio).

Therefore each participant will perform the tests with the experimental rechargeable CROS transmitter (CROS) in different interventions, like comparison to monaural fitting and unaided condition.

All participants will perform the same tests. The order of the intervention in the speech test is randomized, but will be performed in the same visit by each participant.

Ingen behandling, det vil si at deltakerne ikke er utstyrt med CROS-apparat og/eller høreapparat.

Uassistert tilstand fungerer som kontrollintervensjon.

Each participant will be fitted with the experimental CROS device on the poorer, unaidable ear and with a compatible hearing device on the other ear.

Each participant will be fitted with a hearing aid on the better hearing ear and no device on the poorer, unaidable ear.

Monaural hearing aid fitting acts as control intervention.

Each participant will be fitted with the experimental CROS device on the poorer, unaidable ear and with a compatible device on the other ear. The difference to the BiCROS condition is the setup of the microphones in the CROS device and the compatible hearing device.

This test condition is only tested for speech intelligibility in diffuse noise and acts as control intervention.

Hva måler studien?

Primære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Speech Intelligibility (SI) in noise (fix direction): BiCROS fitting vs. monaural hearing aid fitting
Tidsramme: 2 weeks

The primary objective of this clinical investigation is to evaluate whether a BiCROS fitting (better ear fitted with a hearing device, unaidable ear supplied with a CROS device) improves speech intelligibility in noise compared to a monaural fitting (just the aidable ear is supplied by a hearing aid), for adults with an asymmetric hearing loss, measured with the Oldenburg Sentence Test (OLSA). Results of the speech test will be the SRT (speech recognition threshold) in dB SNR (signal to noise ratio).

The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A).

The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A).

2 weeks

Sekundære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
SI in noise (fix direction): BiCROS fitting vs. unaided
Tidsramme: 2 week

A secondary objective of this clinical investigation is to evaluate whether a BiCROS fitting improves speech intelligibility in noise compared to an unaided condition, for adults with an asymmetric hearing loss, measured with the Oldenburg Sentence Test (OLSA).

Results of the speech test will be the SRT (dB SNR). The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A).

The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A).

2 week
SI in diffuse noise: BiCROS fitting vs. monaural fitting
Tidsramme: 2 weeks

A secondary objective of this clinical investigation is to evaluate whether a BiCROS fitting improves speech intelligibility by at least 1 dB SNR for 50% SRT in diffuse noise, compared to a monaural hearing aid fitting.

The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A).

The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A).

2 weeks
SI in diffuse noise: BiCROS fitting vs. unaided
Tidsramme: 2 weeks

A secondary objective of this clinical investigation is to evaluate whether a BiCROS fitting improves speech intelligibility by at least 1 dB SNR for 50% SRT in diffuse noise, compared to an unaided condition.

The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A).

The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A).

2 weeks
SI in diffuse noise: BiCROS vs. CROS baseline fitting
Tidsramme: 2 weeks

A secondary objective of this clinical investigation is to evaluate whether a BiCROS fitting improves speech intelligibility in diffuse noise compared with a CROS baseline fitting. CROS baseline means a fitting of an additional program with different hearing instrument microphone setup.

The SI will be tested with the OLSA in diffuse noise. The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A).

The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A).

2 weeks
SI in diffuse noise: CROS baseline fitting vs. unaided
Tidsramme: 2 weeks

A secondary objective of this clinical investigation is to evaluate whether a CROS baseline fitting improves speech intelligibility in diffuse noise compared with an unaided condition.

The SI will be tested with the OLSA in diffuse noise. The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A).

The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A).

2 weeks
SI in diffuse noise: CROS baseline fitting vs. monaural fitting
Tidsramme: 2 weeks

A secondary objective of this clinical investigation is to evaluate whether a CROS baseline fitting improves speech intelligibility in diffuse noise compared with a monaural fitting.

The SI will be tested with the OLSA in diffuse noise. The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A).

The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A).

2 weeks
SP in diffuse noise: monaural fitting vs. unaided
Tidsramme: 2 weeks

A secondary objective of this clinical investigation is to evaluate whether a monaural fitting improves speech intelligibility in diffuse noise compared with an unaided condition.

The SI will be tested with the OLSA in diffuse noise. The target speaker will be presented from 0° azimuth and is adaptive (start level at 65dB (A).

The noise will be presented on the better ear at an azimuth of 90° or 270° (depending on the site of the better ear), at a level of 70 dB (A).

2 weeks
Evaluation of Sound quality
Tidsramme: 3 weeks

A secondary objective of this study is to evaluate the participant's satisfaction of the sound quality of the experimental CROS device by a subjective rating during a three-week home trial period.

Scale: very unsatisfied- unsatisfied- neither nor- satisfied- very satisfied

3 weeks
Evaluation of System stability
Tidsramme: 3 weeks

A secondary objective of this study is to evaluate the participant's satisfaction of the system stability of the experimental CROS device during a three-week home trial period.

The participant shall report, if there are artefacts audible: never/ 1-2 times a week/ 3-6 times a week/ daily

3 weeks
Evaluation of streaming
Tidsramme: 3 weeks

A secondary objective of this study is to evaluate the participant's satisfaction of the streaming sound quality of the experimental CROS device, by a subjective rating during a three-week home trial period.

Scale: very unsatisfied- unsatisfied- neither nor- satisfied- very satisfied

3 weeks
Evaluation of listening effort
Tidsramme: 3 weeks

A secondary objective is to evaluate the subjective listening effort in noise reported by the participant whilst wearing the experimental CROS device.

Scale: not at all - slightly - somewhat - moderately - very - extremely

3 weeks

Samarbeidspartnere og etterforskere

Det er her du vil finne personer og organisasjoner som er involvert i denne studien.

Sponsor

Publikasjoner og nyttige lenker

Den som er ansvarlig for å legge inn informasjon om studien leverer frivillig disse publikasjonene. Disse kan handle om alt relatert til studiet.

Studierekorddatoer

Disse datoene sporer fremdriften for innsending av studieposter og sammendragsresultater til ClinicalTrials.gov. Studieposter og rapporterte resultater gjennomgås av National Library of Medicine (NLM) for å sikre at de oppfyller spesifikke kvalitetskontrollstandarder før de legges ut på det offentlige nettstedet.

Studer hoveddatoer

Studiestart (Faktiske)

26. oktober 2021

Primær fullføring (Faktiske)

14. mars 2022

Studiet fullført (Faktiske)

14. mars 2022

Datoer for studieregistrering

Først innsendt

22. oktober 2021

Først innsendt som oppfylte QC-kriteriene

22. oktober 2021

Først lagt ut (Faktiske)

3. november 2021

Oppdateringer av studieposter

Sist oppdatering lagt ut (Faktiske)

15. april 2022

Siste oppdatering sendt inn som oppfylte QC-kriteriene

8. april 2022

Sist bekreftet

1. november 2021

Mer informasjon

Begreper knyttet til denne studien

Legemiddel- og utstyrsinformasjon, studiedokumenter

Studerer et amerikansk FDA-regulert medikamentprodukt

Nei

Studerer et amerikansk FDA-regulert enhetsprodukt

Nei

Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .