Measuring Force During Subpial Resection Procedure Using a Novel Ex Vivo Calf Brain Model With Integrated Sensor

2026年8月12日 更新者:Rolando Del Maestro、McGill University

Neurosurgery is a high-stakes surgical specialty where errors can result in significant morbidity. The amount of force applied simultaneously on the brain with multiple different instruments during complex neurosurgical procedures is a critical safety metric that, to the investigators' knowledge, has not been previously measured in a realistic operative environment.

The investigators have therefore developed a simulation platform integrating an ex vivo calf brain and a 3D-printed skull model attached to a force sensor capable of capturing real-time forces applied to the brain. A case series study will be conducted to evaluate the pattern of force applied. Medical students, neurosurgical residents, and staff neurosurgeons from McGill University will be recruited to perform subpial resections using our ex vivo calf brain simulation platform. The forces applied by the microscissors, bipolar forceps, and ultrasonic aspirator onto the brain will be captured. This study aims to demonstrate the spectrum of force applied during a neurosurgical procedure using an ex vivo calf brain model.

調査の概要

状態

完了

条件

詳細な説明

Background and Rationale: The subpial resection technique is a complex neurosurgical procedure that is essential for brain tumor and epilepsy surgery. However, opportunities for trainees to gain hands-on experience with this procedure without risk to patient safety are limited. The force applied on the brain during tool-tissue contact is a critical factor impacting patient safety. While virtual simulation platforms enable force monitoring, ex vivo animal brains cannot readily capture quantitative data. The investigators have therefore developed a prototype of a simulation platform that integrates a force sensor and a calf brain and can accurately detect real-time forces during simulated subpial resections. This study seeks to gauge the force spectrum using this ex vivo calf brain force detection system.

Hypothesis: The simulation platform will accurately measure forces applied by novice, intermediate, and expert participants with surgical instruments on an ex vivo calf brain during simulated subpial resection procedures.

Co-Primary Outcomes: To assess the individual and combined instrumental force applications during subpial resection procedures in an ex vivo calf brain model.

Secondary Outcome: To differentiate levels of expertise of trial participants.

Setting: Montreal Neurological Institute and Hospital, McGill University, Montreal, Quebec, Canada.

Participants:

Neurosurgeons (experts): Staff neurosurgeons at McGill University specializing in oncology, epilepsy, pediatric, or vascular neurosurgery

Neurosurgical residents (trainees): Neurosurgical residents at McGill University

Medical students (novices): Students enrolled in medical school at McGill University

Design: A cross-sectional case series study.

Methodology: This study was approved by the McGill University Health Centre Research Ethics Board, Neurosciences-Psychiatry. A case series study will be conducted where medical students, neurosurgical residents, and staff neurosurgeons from McGill University will be recruited to perform subpial resection tasks on an ex vivo calf brain model. Calf brains will be placed in a 3D-printed skull model to mimic a realistic human operative environment. The 3D-printed skull model was prototyped from a publicly available CT scan obtained from Embodi3D, an open-access medical imaging repository, and reconstructed in 3D using the open-source software 3D Slicer version 5.10.0. A 6-degree-of-freedom force/torque sensor (Nano17 IP68, ATI Industrial Automation Inc., North Carolina, USA) and a USB data acquisition board (NI-6210, National Instruments Inc., Texas, USA) will be used to measure real-time forces applied to the brain. Participants will use microscissors, bipolar forceps, and a SONOPET ultrasonic aspirator (Stryker, Portage, Michigan, USA) to perform simulated procedures. The tasks will be performed using an OPMI pico surgical microscope (ZEISS, Jena, Germany) and video-recorded using a Blackfly S GigE embedded microscope camera (FLIR, Wilsonville, Oregon, USA), allowing a broader instrument view for evaluation of intraoperative and postoperative performance.

Study Procedure: Upon arrival, participants will read and sign an informed consent form. They will then fill out a pre-trial questionnaire assessing their demographic characteristics (e.g., sex, gender, age, institutional affiliation, etc.). Each participant will receive standardized instructions on instrument use and function and be presented with an image outlining the location of pial cuts and the subpial resections on the ex vivo calf brain in front of them. Participants will adjust the operating microscope according to their preferences. The subpial resection experiment will be divided into three subsequent tasks. The goal is to obtain a noise-free and specific force measurement for each of the major instruments/maneuvers in a standard subpial resection. The first task entails using microscissors to create a 2-cm pial cut in ten different gyri across a calf brain hemisphere (n = 10). The time allocated for this task is 30 seconds for each cut. The second task involves utilizing bipolar forceps to grasp and lift the edge of the pia where cuts are made with the microscissors in task 1. A repetitive lift and hold maneuver will be repeated five times assuming a right to left direction across the pial incision line in 5-second intervals using the left hand. This will be repeated across each of the ten different pial cuts (n = 50). The time allocated for this task is maximum of one minute for each pial defect. Finally, an ultrasonic aspirator in the dominant (right) hand along with the bipolar in the opposite hand will be employed to perform subpial resection at three out of the ten pial cuts originally made in task 1 (n = 3). Each of the three spots would undergo subpial resection for an allocated time of 3 minutes across separate recordings.

Significance: This study will provide the first insights into a realistic operative environment where it will be possible to monitor the force applied on the brain.

研究の種類

観察的

入学 (実際)

11

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究場所

    • Quebec
      • Montreal、Quebec、カナダ、H3A 2B4
        • Neuro Imaging and Surgical Technologies Lab, Department of Neurology and Neurosurgery, Montreal Neurological Institute and Hospital, McGill University

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

  • 大人
  • 高齢者

健康ボランティアの受け入れ

はい

サンプリング方法

非確率サンプル

調査対象母集団

Staff neurosurgeons, neurosurgical residents, and medical students from McGill University

説明

Inclusion Criteria:

  • Right-handed medical students, neurosurgical residents, and staff neurosurgeons from McGill University

Exclusion Criteria:

  • Left-handed individuals

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Forces applied to the brain during each simulated subpial resection procedure step
時間枠:1 day of study
Throughout each simulated ex vivo subpial resection procedure, forces applied to the brain during tool-tissue contact will be measured by the force/torque sensor in grams and Newtons.
1 day of study

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

捜査官

  • 主任研究者:Rolando F. Del Maestro, MD, PhD、Neurosurgical Simulation and Artificial Intelligence Learning Centre, McGill University
  • スタディディレクター:Amir Hooshiar, PhD、Surgical Performance Enhancement and Robotics Centre, McGill University
  • スタディディレクター:D. Louis Collins, PhD、Neuro Imaging and Surgical Technologies Lab, McGill University

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (実際)

2026年6月3日

一次修了 (実際)

2026年7月8日

研究の完了 (実際)

2026年7月8日

試験登録日

最初に提出

2026年6月10日

QC基準を満たした最初の提出物

2026年6月10日

最初の投稿 (実際)

2026年6月16日

学習記録の更新

投稿された最後の更新 (実際)

2026年8月14日

QC基準を満たした最後の更新が送信されました

2026年8月12日

最終確認日

2026年8月1日

詳しくは

本研究に関する用語

その他の研究ID番号

  • 2018 4395, Trial 1

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

はい

IPD 共有サポート情報タイプ

  • STUDY_PROTOCOL
  • SAP
  • ICF
  • ANALYTIC_CODE
  • CSR

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する