Accuracy of 3D Electrical Impedance Tomography for Assessing Lung Ventilation
Accuracy of Three-Dimensional Electrical Impedance Tomography for Assessing Dynamic Regional Lung Ventilation Validated by Four-Dimensional CT: A Prospective Observational Study
This prospective observational study will evaluate how accurately three-dimensional electrical impedance tomography (3D-EIT) measures the distribution of air in the lungs during breathing when compared with four-dimensional computed tomography (4D-CT).
3D-EIT is a noninvasive, radiation-free imaging method that measures changes in electrical impedance across the chest and can provide continuous information about regional lung ventilation. In this study, 4D-CT will be used as the reference method because it can provide detailed images of changes in the lungs throughout the breathing cycle.
Approximately 50 adults, including healthy volunteers and hospitalized patients who are able to undergo both examinations, will be enrolled at a single center. EIT and CT data will be collected at the same time during stable breathing. Researchers will compare the two methods to determine how closely their measurements of ventilation in different lung regions agree. The study will also explore whether EIT can detect changes in regional ventilation under different breathing conditions.
The study will not use the research EIT results to make clinical treatment decisions. The findings may support the future use of EIT as a noninvasive, radiation-free method for monitoring regional lung ventilation.
調査の概要
状態
条件
研究の種類
入学 (推定)
連絡先と場所
研究連絡先
- 名前:Yuxian Wang
- 電話番号:+86 021 6404 1990
- メール:wang.yuxian@zs-hospital.sh.cn
参加基準
適格基準
就学可能な年齢
- 大人
- 高齢者
健康ボランティアの受け入れ
サンプリング方法
調査対象母集団
The study population will include adults aged 18 years or older recruited at the Department of Critical Care Medicine, Zhongshan Hospital, Fudan University. Participants will consist of healthy volunteers and intensive care unit patients for whom CT imaging is considered necessary based on the treating physician's clinical judgment. Eligible participants must be able to undergo simultaneous 4D-CT and 3D-EIT monitoring.
This is a prospective, observational, single-center study. Participation will not alter the participant's routine clinical management, and the research-derived three-dimensional EIT reconstruction results will not be used directly to make clinical treatment decisions.
説明
Inclusion Criteria:
- Age 18 years or older.
- Healthy volunteers or patients in the intensive care unit for whom CT imaging is considered necessary based on the treating physician's clinical judgment.
- Able to cooperate with or undergo simultaneous 4D-CT and 3D-EIT monitoring.
- Able to provide written informed consent personally or through a legally authorized representative, as applicable.
Exclusion Criteria:
- Unable to tolerate or having a contraindication to 4D-CT examination, including pregnancy or severe respiratory distress.
- Extensive chest skin injury or infection, or any condition preventing placement of the EIT electrode belt.
- Body size or chest deformity that may substantially limit the quality of EIT or CT data, including chest circumference outside the compatible range of the EIT device.
- Presence of an implanted electronic device that may interfere with EIT signals, such as a cardiac pacemaker.
- Unable to complete simultaneous 4D-CT and 3D-EIT monitoring because of significant motion, persistent coughing, poor cooperation, or other technical or clinical reasons.
- Any other condition that, in the investigator's judgment, makes the individual unsuitable for participation in the study.
研究計画
研究はどのように設計されていますか?
デザインの詳細
コホートと介入
グループ/コホート |
介入・治療 |
|---|---|
|
Paired 3D-EIT and 4D-CT Assessment Cohort
This single cohort will include approximately 50 adults, including healthy volunteers and hospitalized patients, who are able to undergo synchronous 3D-EIT and 4D-CT data acquisition.
Regional lung ventilation measurements obtained by the two methods during the same breathing cycle will be compared for agreement and accuracy.
Participants will not be assigned to different study groups, and the research EIT results will not be used to direct clinical treatment.
|
Regional lung ventilation data will be collected using a chest electrical impedance tomography system (V200-S).
Electrical impedance changes in the chest will be measured continuously and reconstructed to assess three-dimensional regional ventilation distribution.
EIT data will be acquired synchronously with dynamic computed tomography data and will be used for research analysis only.
The research EIT results will not directly guide clinical treatment decisions.
他の名前:
Dynamic chest computed tomography data will be acquired using an X-ray computed tomography system (uCT 610 Sim) during one or more breathing cycles.
The images will be reconstructed and analyzed to quantify changes in regional lung air volume and ventilation distribution.
The 4D-CT measurements will serve as the reference method for comparison with 3D-EIT measurements.
他の名前:
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Deming and Passing-Bablok Regression Slopes and Intercepts Comparing Regional Ventilation Fraction Measured by 3D-EIT and 4D-CT
時間枠:Periprocedural, during synchronous 3D-EIT and 4D-CT acquisition at the study visit; one respiratory cycle is from the onset of one inspiration to the onset of the next.
|
For each participant, paired regional ventilation fraction measurements obtained by 3D-EIT and 4D-CT during the same synchronized respiratory cycle will be analyzed.
Regional ventilation fraction will be expressed as the percentage of total ventilation within each predefined lung region.
Prespecified regions may include the ventral and dorsal lung regions, dependent and nondependent regions, and the left and right lungs.
Agreement between the two methods will be assessed using Deming regression and Passing-Bablok regression.
Regression slopes, intercepts, and their 95% confidence intervals will be reported.
A slope closer to 1 and an intercept closer to 0 indicate better agreement between the two methods.
|
Periprocedural, during synchronous 3D-EIT and 4D-CT acquisition at the study visit; one respiratory cycle is from the onset of one inspiration to the onset of the next.
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Pearson Correlation Coefficient Between Regional Ventilation Fraction Measurements Obtained by 3D-EIT and 4D-CT
時間枠:Periprocedural, during the synchronous 3D-EIT and 4D-CT acquisition at the study visit
|
Paired regional ventilation fraction measurements obtained from spatially and temporally aligned 3D-EIT and 4D-CT ventilation data will be evaluated using the Pearson correlation coefficient.
Regional ventilation fraction will be expressed as the percentage of total ventilation within each predefined lung region.
Analyses will be performed for predefined regions, including the ventral and dorsal regions, left and right lungs, and dependent and nondependent regions, as applicable.
Pearson correlation coefficients and their 95% confidence intervals will be reported.
|
Periprocedural, during the synchronous 3D-EIT and 4D-CT acquisition at the study visit
|
|
Mean Bias and 95% Limits of Agreement Between Regional Ventilation Fraction Measurements Obtained by 3D-EIT and 4D-CT
時間枠:Periprocedural, during the synchronous 3D-EIT and 4D-CT acquisition at the study visit
|
Agreement between paired regional ventilation fraction measurements obtained by 3D-EIT and 4D-CT will be evaluated using Bland-Altman analysis.
Regional ventilation fraction will be expressed as the percentage of total ventilation within each predefined lung region.
The mean difference between the two methods, corresponding 95% confidence interval, and the lower and upper 95% limits of agreement will be reported.
|
Periprocedural, during the synchronous 3D-EIT and 4D-CT acquisition at the study visit
|
協力者と研究者
研究記録日
主要日程の研究
研究開始 (推定)
一次修了 (推定)
研究の完了 (推定)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
本研究に関する用語
その他の研究ID番号
- B2026-447
個々の参加者データ (IPD) の計画
個々の参加者データ (IPD) を共有する予定はありますか?
医薬品およびデバイス情報、研究文書
米国FDA規制医薬品の研究
米国FDA規制機器製品の研究
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。