Accuracy of 3D Electrical Impedance Tomography for Assessing Lung Ventilation
Accuracy of Three-Dimensional Electrical Impedance Tomography for Assessing Dynamic Regional Lung Ventilation Validated by Four-Dimensional CT: A Prospective Observational Study
This prospective observational study will evaluate how accurately three-dimensional electrical impedance tomography (3D-EIT) measures the distribution of air in the lungs during breathing when compared with four-dimensional computed tomography (4D-CT).
3D-EIT is a noninvasive, radiation-free imaging method that measures changes in electrical impedance across the chest and can provide continuous information about regional lung ventilation. In this study, 4D-CT will be used as the reference method because it can provide detailed images of changes in the lungs throughout the breathing cycle.
Approximately 50 adults, including healthy volunteers and hospitalized patients who are able to undergo both examinations, will be enrolled at a single center. EIT and CT data will be collected at the same time during stable breathing. Researchers will compare the two methods to determine how closely their measurements of ventilation in different lung regions agree. The study will also explore whether EIT can detect changes in regional ventilation under different breathing conditions.
The study will not use the research EIT results to make clinical treatment decisions. The findings may support the future use of EIT as a noninvasive, radiation-free method for monitoring regional lung ventilation.
Studie Overzicht
Toestand
Conditie
Studietype
Inschrijving (Geschat)
Contacten en locaties
Studiecontact
- Naam: Yuxian Wang
- Telefoonnummer: +86 021 6404 1990
- E-mail: wang.yuxian@zs-hospital.sh.cn
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
- Volwassen
- Oudere volwassene
Accepteert gezonde vrijwilligers
Bemonsteringsmethode
Studie Bevolking
The study population will include adults aged 18 years or older recruited at the Department of Critical Care Medicine, Zhongshan Hospital, Fudan University. Participants will consist of healthy volunteers and intensive care unit patients for whom CT imaging is considered necessary based on the treating physician's clinical judgment. Eligible participants must be able to undergo simultaneous 4D-CT and 3D-EIT monitoring.
This is a prospective, observational, single-center study. Participation will not alter the participant's routine clinical management, and the research-derived three-dimensional EIT reconstruction results will not be used directly to make clinical treatment decisions.
Beschrijving
Inclusion Criteria:
- Age 18 years or older.
- Healthy volunteers or patients in the intensive care unit for whom CT imaging is considered necessary based on the treating physician's clinical judgment.
- Able to cooperate with or undergo simultaneous 4D-CT and 3D-EIT monitoring.
- Able to provide written informed consent personally or through a legally authorized representative, as applicable.
Exclusion Criteria:
- Unable to tolerate or having a contraindication to 4D-CT examination, including pregnancy or severe respiratory distress.
- Extensive chest skin injury or infection, or any condition preventing placement of the EIT electrode belt.
- Body size or chest deformity that may substantially limit the quality of EIT or CT data, including chest circumference outside the compatible range of the EIT device.
- Presence of an implanted electronic device that may interfere with EIT signals, such as a cardiac pacemaker.
- Unable to complete simultaneous 4D-CT and 3D-EIT monitoring because of significant motion, persistent coughing, poor cooperation, or other technical or clinical reasons.
- Any other condition that, in the investigator's judgment, makes the individual unsuitable for participation in the study.
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
Cohorten en interventies
Groep / Cohort |
Interventie / Behandeling |
|---|---|
|
Paired 3D-EIT and 4D-CT Assessment Cohort
This single cohort will include approximately 50 adults, including healthy volunteers and hospitalized patients, who are able to undergo synchronous 3D-EIT and 4D-CT data acquisition.
Regional lung ventilation measurements obtained by the two methods during the same breathing cycle will be compared for agreement and accuracy.
Participants will not be assigned to different study groups, and the research EIT results will not be used to direct clinical treatment.
|
Regional lung ventilation data will be collected using a chest electrical impedance tomography system (V200-S).
Electrical impedance changes in the chest will be measured continuously and reconstructed to assess three-dimensional regional ventilation distribution.
EIT data will be acquired synchronously with dynamic computed tomography data and will be used for research analysis only.
The research EIT results will not directly guide clinical treatment decisions.
Andere namen:
Dynamic chest computed tomography data will be acquired using an X-ray computed tomography system (uCT 610 Sim) during one or more breathing cycles.
The images will be reconstructed and analyzed to quantify changes in regional lung air volume and ventilation distribution.
The 4D-CT measurements will serve as the reference method for comparison with 3D-EIT measurements.
Andere namen:
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Deming and Passing-Bablok Regression Slopes and Intercepts Comparing Regional Ventilation Fraction Measured by 3D-EIT and 4D-CT
Tijdsspanne: Periprocedural, during synchronous 3D-EIT and 4D-CT acquisition at the study visit; one respiratory cycle is from the onset of one inspiration to the onset of the next.
|
For each participant, paired regional ventilation fraction measurements obtained by 3D-EIT and 4D-CT during the same synchronized respiratory cycle will be analyzed.
Regional ventilation fraction will be expressed as the percentage of total ventilation within each predefined lung region.
Prespecified regions may include the ventral and dorsal lung regions, dependent and nondependent regions, and the left and right lungs.
Agreement between the two methods will be assessed using Deming regression and Passing-Bablok regression.
Regression slopes, intercepts, and their 95% confidence intervals will be reported.
A slope closer to 1 and an intercept closer to 0 indicate better agreement between the two methods.
|
Periprocedural, during synchronous 3D-EIT and 4D-CT acquisition at the study visit; one respiratory cycle is from the onset of one inspiration to the onset of the next.
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Pearson Correlation Coefficient Between Regional Ventilation Fraction Measurements Obtained by 3D-EIT and 4D-CT
Tijdsspanne: Periprocedural, during the synchronous 3D-EIT and 4D-CT acquisition at the study visit
|
Paired regional ventilation fraction measurements obtained from spatially and temporally aligned 3D-EIT and 4D-CT ventilation data will be evaluated using the Pearson correlation coefficient.
Regional ventilation fraction will be expressed as the percentage of total ventilation within each predefined lung region.
Analyses will be performed for predefined regions, including the ventral and dorsal regions, left and right lungs, and dependent and nondependent regions, as applicable.
Pearson correlation coefficients and their 95% confidence intervals will be reported.
|
Periprocedural, during the synchronous 3D-EIT and 4D-CT acquisition at the study visit
|
|
Mean Bias and 95% Limits of Agreement Between Regional Ventilation Fraction Measurements Obtained by 3D-EIT and 4D-CT
Tijdsspanne: Periprocedural, during the synchronous 3D-EIT and 4D-CT acquisition at the study visit
|
Agreement between paired regional ventilation fraction measurements obtained by 3D-EIT and 4D-CT will be evaluated using Bland-Altman analysis.
Regional ventilation fraction will be expressed as the percentage of total ventilation within each predefined lung region.
The mean difference between the two methods, corresponding 95% confidence interval, and the lower and upper 95% limits of agreement will be reported.
|
Periprocedural, during the synchronous 3D-EIT and 4D-CT acquisition at the study visit
|
Medewerkers en onderzoekers
Sponsor
Studie record data
Bestudeer belangrijke data
Studie start (Geschat)
Primaire voltooiing (Geschat)
Studie voltooiing (Geschat)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Werkelijk)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Andere studie-ID-nummers
- B2026-447
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .