Comparison of Reliability, Time Cost, and Participant Comfort of Lymphedema Assessments in Upper Extremity

2026年11月4日 更新者:Burak Ertürk、Karadeniz Technical University

Comparison of Reliability, Time Cost, and Participant Comfort of Volumetric, Circumference Measurement, LiDAR, and 3D Modeling Methods in Upper Extremity Volume Measurement

This study aims to evaluate and compare the measurement reliability, total time cost (preparation, acquisition, post-processing), and participant comfort of conventional limb volume measurement methods (water displacement volumetry, tape-based circumference measurement) against modern digital techniques (LiDAR-based mobile scanning, stereophotogrammetric 3D optical scanning) in the upper extremity. The evaluation will be conducted on both healthy adults and patients diagnosed with upper extremity lymphedema. Additionally, the impact of assessor experience on digital data post-processing and the sensitivity of the methods to detect small localized volume increases using a simulated edema model (50 mL) will be investigated.

調査の概要

状態

まだ募集していません

詳細な説明

Upper extremity lymphedema is a chronic and progressive condition characterized by the accumulation of protein-rich interstitial fluid, leading to tissue fibrosis and structural volume increases. Precise, reliable, and practical limb volume measurement is vital for diagnosis, staging, and therapeutic monitoring. While water displacement volumetry remains the gold standard, it poses practical challenges including cross-contamination risks, prolonged setup, and inapplicability to patients with wounds or severe limitations. Circumference tape measurement with frustum modeling is widely used but suffers from high examiner dependency and geometric approximation errors. Emerging non-contact technologies, such as mobile Light Detection and Ranging (LiDAR) sensors and optical 3D surface scanning, offer rapid, hygienic, and objective alternatives. However, gaps remain regarding their sensitivity to subclinical volume variations, the influence of post-processing operator experience, procedural time costs, and participant burden.

This methodological, comparative, cross-sectional pilot study evaluates upper extremity volume assessment methods across two cohorts: healthy adults (n = 40) and individuals clinically diagnosed with upper extremity lymphedema (n = 24). Measurements are conducted on the dominant upper extremity in healthy participants and on the affected upper extremity in patients with lymphedema, demarcated from the ulnar styloid process to 2-3 cm distal to the axilla.

All participants undergo the following sequentially:

  1. Water Displacement Volumetry: Conducted using a custom overflow tank and precision graduated cylinders as the reference standard. The hand volume is measured first and subtracted from the full-arm immersion volume. Strict decontamination protocols are applied between sessions.
  2. Circumference Measurement: Circumferential measures are taken at 4 cm intervals in a supine position and converted to limb volume using the truncated cone (Kuhnke/Frustum) formula. Measurements are obtained independently by both an experienced and an inexperienced evaluator.
  3. LiDAR-Based Scanning: Acquired using an iPad equipped with Structure Sensor 3 (Structure application). Standardized positioning is utilized with adhesive anatomical landmarks.
  4. Photogrammetric/Optical 3D Scanning: Acquired using the Creality Raptor X 3D scanner (Creality Scan software).

To address operator dependency in digital workflows, raw 3D point-cloud processing, mesh trimming, and volumetric integration are performed under a blinded protocol by both an experienced operator and an inexperienced operator.

In the healthy adult cohort, a simulated localized edema protocol is additionally conducted to determine method sensitivity to small volume fluctuations. A non-absorbent medical silicone pad (calibrated to exactly 50 mL via prior water volumetry) is affixed to the participant's forearm, after which circumference, LiDAR, and 3D optical scanning procedures are repeated.

Comprehensive time-motion analyses are recorded with a digital stopwatch across three distinct phases for each modality: equipment preparation, data acquisition, and post-processing computation. Participant-reported outcomes, including physical discomfort and fatigue, are documented using a Visual Analog Scale (VAS, 0-100 mm) immediately after each test and ranked comparatively at session conclusion.

研究の種類

観察的

入学 (推定)

64

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年~65年 (大人、高齢者)

健康ボランティアの受け入れ

はい

受講資格のある性別

全て

サンプリング方法

非確率サンプル

調査対象母集団

Healthy adult volunteers and clinical patients diagnosed with upper extremity lymphedema evaluated at the physical therapy and rehabilitation laboratories.

説明

Inclusion Criteria:

Healthy Volunteers:

  • Aged between 18 and 65 years
  • Voluntarily willing to participate and providing written informed consent
  • Absence of any known musculoskeletal or neuromuscular deformity, active edema, open wounds, or history of major upper extremity surgery

Lymphedema Patients:

  • Aged between 18 and 65 years
  • Voluntarily willing to participate and providing written informed consent
  • Confirmed clinical diagnosis of upper extremity lymphedema
  • Absence of active skin lesions, open ulcerations, acute cellulitis, or lymphorrhea on the affected upper extremity

Exclusion Criteria (Both Groups):

  • Presence of open wounds, acute skin breakdown, lymphorrhea, active cutaneous infection (cellulitis, fungal infection), or severe dermatological lesions on the tested limb
  • Severe cognitive, systemic, or physical limitations preventing standard testing positioning (supine posture or holding the limb steady) or safe water immersion
  • Known severe skin allergy or dermatological hypersensitivity to silicone or adhesive fixation materials (for the healthy cohort undergoing the simulated edema protocol)

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 観測モデル:COHORT
  • 時間の展望:CROSS_SECTIONAL

コホートと介入

グループ/コホート
Healthy Adults

Forty healthy adult volunteers aged 18 to 65 years with no known deformity, edema, open wound, or prior surgical history in the upper extremity. Participants undergo baseline upper extremity volume assessments on their dominant arm using water displacement volumetry, tape-based circumference measurement (performed independently by experienced and inexperienced evaluators), LiDAR-based scanning (Structure Sensor 3), and stereophotogrammetric 3D optical scanning (Creality Raptor X). Additionally, to determine method sensitivity to small volume fluctuations, participants undergo a simulated edema protocol wherein a calibrated 50 mL non-absorbent medical silicone pad is affixed to the forearm, followed by repeat circumference, LiDAR, and 3D optical scans.

Patients with Lymphedema

Twenty-four adult individuals aged 18 to 65 years clinically diagnosed with upper extremity lymphedema, without active skin lesions, open wounds, or lymphorrhea on the affected limb. Participants undergo upper extremity volume measurements on their affected arm using water displacement volumetry, tape-based circumference measurement (performed independently by experienced and inexperienced evaluators), LiDAR-based scanning (Structure Sensor 3), and stereophotogrammetric 3D optical scanning (Creality Raptor X). The simulated edema protocol is not applied to this cohort.

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Upper Extremity Limb Volume
時間枠:Assessed during the single cross-sectional experimental testing session (approximately 1 to 2 hours per participant).

Total upper extremity limb volume (measured in milliliters, mL) evaluated across four measurement modalities: water displacement volumetry (gold standard reference), circumference tape measurement (calculated using the truncated cone / Frustum formula), LiDAR-based scanning (Structure Sensor 3), and stereophotogrammetric 3D optical scanning (Creality Raptor X). Measurement agreement and reliability will be determined using the Intraclass Correlation Coefficient (ICC Model 3,1) and Bland-Altman limits of agreement. Digital 3D model post-processing and tape measurements will be evaluated comparatively between experienced and inexperienced assessors.

Assessed during the single cross-sectional experimental testing session (approximately 1 to 2 hours per participant).

二次結果の測定

結果測定
メジャーの説明
時間枠
Procedural Time Cost (Process Duration)
時間枠:Throughout the 4-hour experimental protocol window (from 1 hour prior to participant arrival, during the 2-hour assessment session, through 1 hour following participant departure)

Total time cost (measured in minutes and seconds using a digital stopwatch) required for each modality, tracked across three distinct phases: preparation time (equipment setup, sanitation, and participant positioning), data acquisition time (actual measurement/scanning), and analysis/post-processing time (mathematical formula calculation or digital point-cloud cropping and mesh volume computation).

Throughout the 4-hour experimental protocol window (from 1 hour prior to participant arrival, during the 2-hour assessment session, through 1 hour following participant departure)
Participant-Perceived Discomfort and Fatigue
時間枠:Immediately after each measurement procedure and at the end of the visit (within 2 hours)

Participant-reported physical discomfort and fatigue assessed immediately following each measurement modality using a Visual Analog Scale (VAS, 0 to 10 cm, where 0 represents 'no discomfort/fatigue' and 10 represents 'extreme discomfort/fatigue'). Overall method preference and willingness to repeat the procedure will be ranked at the conclusion of all tests.

Immediately after each measurement procedure and at the end of the visit (within 2 hours)
Sensitivity to Localized Small Volume Changes (Simulated Edema Absolute Error)
時間枠:Immediately following baseline measurements during the single study visit (at approximately 1 hour post-baseline)

Absolute measurement error (calculated in milliliters, mL) in detecting an artificial, localized 50 mL volume increase. In the healthy adult cohort, a calibrated, non-absorbent medical silicone pad (50 mL verified via water displacement) is affixed to the forearm. The error is determined as the absolute difference between the volume change detected by circumference tape, LiDAR, and 3D optical scanning and the true 50 mL volume of the pad.

Immediately following baseline measurements during the single study visit (at approximately 1 hour post-baseline)

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

捜査官

  • 主任研究者:Burak Erturk, PT, PhD、Karadeniz Technical University

出版物と役立つリンク

研究に関する情報を入力する責任者は、自発的にこれらの出版物を提供します。これらは、研究に関連するあらゆるものに関するものである可能性があります。

一般刊行物

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (推定)

2027年2月1日

一次修了 (推定)

2028年2月1日

研究の完了 (推定)

2028年6月1日

試験登録日

最初に提出

2026年10月30日

QC基準を満たした最初の提出物

2026年11月4日

最初の投稿 (実際)

2026年11月6日

学習記録の更新

投稿された最後の更新 (実際)

2026年11月6日

QC基準を満たした最後の更新が送信されました

2026年11月4日

最終確認日

2026年10月1日

詳しくは

本研究に関する用語

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

いいえ

IPD プランの説明

Individual participant data will not be shared publicly because this investigation constitutes the foundational baseline phase of a multi-stage longitudinal research project. The acquired normative volumetric, photographic, and point-cloud datasets will be actively utilized as prospective control and benchmark data in subsequent interrelated clinical and technological validation phases.

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。

購読する