Comparison of Reliability, Time Cost, and Participant Comfort of Lymphedema Assessments in Upper Extremity

2026년 11월 4일 업데이트: Burak Ertürk, Karadeniz Technical University

Comparison of Reliability, Time Cost, and Participant Comfort of Volumetric, Circumference Measurement, LiDAR, and 3D Modeling Methods in Upper Extremity Volume Measurement

This study aims to evaluate and compare the measurement reliability, total time cost (preparation, acquisition, post-processing), and participant comfort of conventional limb volume measurement methods (water displacement volumetry, tape-based circumference measurement) against modern digital techniques (LiDAR-based mobile scanning, stereophotogrammetric 3D optical scanning) in the upper extremity. The evaluation will be conducted on both healthy adults and patients diagnosed with upper extremity lymphedema. Additionally, the impact of assessor experience on digital data post-processing and the sensitivity of the methods to detect small localized volume increases using a simulated edema model (50 mL) will be investigated.

연구 개요

상태

아직 모집하지 않음

상세 설명

Upper extremity lymphedema is a chronic and progressive condition characterized by the accumulation of protein-rich interstitial fluid, leading to tissue fibrosis and structural volume increases. Precise, reliable, and practical limb volume measurement is vital for diagnosis, staging, and therapeutic monitoring. While water displacement volumetry remains the gold standard, it poses practical challenges including cross-contamination risks, prolonged setup, and inapplicability to patients with wounds or severe limitations. Circumference tape measurement with frustum modeling is widely used but suffers from high examiner dependency and geometric approximation errors. Emerging non-contact technologies, such as mobile Light Detection and Ranging (LiDAR) sensors and optical 3D surface scanning, offer rapid, hygienic, and objective alternatives. However, gaps remain regarding their sensitivity to subclinical volume variations, the influence of post-processing operator experience, procedural time costs, and participant burden.

This methodological, comparative, cross-sectional pilot study evaluates upper extremity volume assessment methods across two cohorts: healthy adults (n = 40) and individuals clinically diagnosed with upper extremity lymphedema (n = 24). Measurements are conducted on the dominant upper extremity in healthy participants and on the affected upper extremity in patients with lymphedema, demarcated from the ulnar styloid process to 2-3 cm distal to the axilla.

All participants undergo the following sequentially:

  1. Water Displacement Volumetry: Conducted using a custom overflow tank and precision graduated cylinders as the reference standard. The hand volume is measured first and subtracted from the full-arm immersion volume. Strict decontamination protocols are applied between sessions.
  2. Circumference Measurement: Circumferential measures are taken at 4 cm intervals in a supine position and converted to limb volume using the truncated cone (Kuhnke/Frustum) formula. Measurements are obtained independently by both an experienced and an inexperienced evaluator.
  3. LiDAR-Based Scanning: Acquired using an iPad equipped with Structure Sensor 3 (Structure application). Standardized positioning is utilized with adhesive anatomical landmarks.
  4. Photogrammetric/Optical 3D Scanning: Acquired using the Creality Raptor X 3D scanner (Creality Scan software).

To address operator dependency in digital workflows, raw 3D point-cloud processing, mesh trimming, and volumetric integration are performed under a blinded protocol by both an experienced operator and an inexperienced operator.

In the healthy adult cohort, a simulated localized edema protocol is additionally conducted to determine method sensitivity to small volume fluctuations. A non-absorbent medical silicone pad (calibrated to exactly 50 mL via prior water volumetry) is affixed to the participant's forearm, after which circumference, LiDAR, and 3D optical scanning procedures are repeated.

Comprehensive time-motion analyses are recorded with a digital stopwatch across three distinct phases for each modality: equipment preparation, data acquisition, and post-processing computation. Participant-reported outcomes, including physical discomfort and fatigue, are documented using a Visual Analog Scale (VAS, 0-100 mm) immediately after each test and ranked comparatively at session conclusion.

연구 유형

관찰

등록 (추정된)

64

연락처 및 위치

이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.

연구 연락처

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

18년 (성인, 고령자)

건강한 자원 봉사자를 받아들입니다

예

연구 대상 성별

모두

샘플링 방법

비확률 샘플

연구 인구

Healthy adult volunteers and clinical patients diagnosed with upper extremity lymphedema evaluated at the physical therapy and rehabilitation laboratories.

설명

Inclusion Criteria:

Healthy Volunteers:

  • Aged between 18 and 65 years
  • Voluntarily willing to participate and providing written informed consent
  • Absence of any known musculoskeletal or neuromuscular deformity, active edema, open wounds, or history of major upper extremity surgery

Lymphedema Patients:

  • Aged between 18 and 65 years
  • Voluntarily willing to participate and providing written informed consent
  • Confirmed clinical diagnosis of upper extremity lymphedema
  • Absence of active skin lesions, open ulcerations, acute cellulitis, or lymphorrhea on the affected upper extremity

Exclusion Criteria (Both Groups):

  • Presence of open wounds, acute skin breakdown, lymphorrhea, active cutaneous infection (cellulitis, fungal infection), or severe dermatological lesions on the tested limb
  • Severe cognitive, systemic, or physical limitations preventing standard testing positioning (supine posture or holding the limb steady) or safe water immersion
  • Known severe skin allergy or dermatological hypersensitivity to silicone or adhesive fixation materials (for the healthy cohort undergoing the simulated edema protocol)

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

  • 관찰 모델: COHORT
  • 시간 관점: CROSS_SECTIONAL

코호트 및 개입

그룹/코호트
Healthy Adults

Forty healthy adult volunteers aged 18 to 65 years with no known deformity, edema, open wound, or prior surgical history in the upper extremity. Participants undergo baseline upper extremity volume assessments on their dominant arm using water displacement volumetry, tape-based circumference measurement (performed independently by experienced and inexperienced evaluators), LiDAR-based scanning (Structure Sensor 3), and stereophotogrammetric 3D optical scanning (Creality Raptor X). Additionally, to determine method sensitivity to small volume fluctuations, participants undergo a simulated edema protocol wherein a calibrated 50 mL non-absorbent medical silicone pad is affixed to the forearm, followed by repeat circumference, LiDAR, and 3D optical scans.

Patients with Lymphedema

Twenty-four adult individuals aged 18 to 65 years clinically diagnosed with upper extremity lymphedema, without active skin lesions, open wounds, or lymphorrhea on the affected limb. Participants undergo upper extremity volume measurements on their affected arm using water displacement volumetry, tape-based circumference measurement (performed independently by experienced and inexperienced evaluators), LiDAR-based scanning (Structure Sensor 3), and stereophotogrammetric 3D optical scanning (Creality Raptor X). The simulated edema protocol is not applied to this cohort.

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
Upper Extremity Limb Volume
기간: Assessed during the single cross-sectional experimental testing session (approximately 1 to 2 hours per participant).

Total upper extremity limb volume (measured in milliliters, mL) evaluated across four measurement modalities: water displacement volumetry (gold standard reference), circumference tape measurement (calculated using the truncated cone / Frustum formula), LiDAR-based scanning (Structure Sensor 3), and stereophotogrammetric 3D optical scanning (Creality Raptor X). Measurement agreement and reliability will be determined using the Intraclass Correlation Coefficient (ICC Model 3,1) and Bland-Altman limits of agreement. Digital 3D model post-processing and tape measurements will be evaluated comparatively between experienced and inexperienced assessors.

Assessed during the single cross-sectional experimental testing session (approximately 1 to 2 hours per participant).

2차 결과 측정

결과 측정
측정값 설명
기간
Procedural Time Cost (Process Duration)
기간: Throughout the 4-hour experimental protocol window (from 1 hour prior to participant arrival, during the 2-hour assessment session, through 1 hour following participant departure)

Total time cost (measured in minutes and seconds using a digital stopwatch) required for each modality, tracked across three distinct phases: preparation time (equipment setup, sanitation, and participant positioning), data acquisition time (actual measurement/scanning), and analysis/post-processing time (mathematical formula calculation or digital point-cloud cropping and mesh volume computation).

Throughout the 4-hour experimental protocol window (from 1 hour prior to participant arrival, during the 2-hour assessment session, through 1 hour following participant departure)
Participant-Perceived Discomfort and Fatigue
기간: Immediately after each measurement procedure and at the end of the visit (within 2 hours)

Participant-reported physical discomfort and fatigue assessed immediately following each measurement modality using a Visual Analog Scale (VAS, 0 to 10 cm, where 0 represents 'no discomfort/fatigue' and 10 represents 'extreme discomfort/fatigue'). Overall method preference and willingness to repeat the procedure will be ranked at the conclusion of all tests.

Immediately after each measurement procedure and at the end of the visit (within 2 hours)
Sensitivity to Localized Small Volume Changes (Simulated Edema Absolute Error)
기간: Immediately following baseline measurements during the single study visit (at approximately 1 hour post-baseline)

Absolute measurement error (calculated in milliliters, mL) in detecting an artificial, localized 50 mL volume increase. In the healthy adult cohort, a calibrated, non-absorbent medical silicone pad (50 mL verified via water displacement) is affixed to the forearm. The error is determined as the absolute difference between the volume change detected by circumference tape, LiDAR, and 3D optical scanning and the true 50 mL volume of the pad.

Immediately following baseline measurements during the single study visit (at approximately 1 hour post-baseline)

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

수사관

  • 수석 연구원: Burak Erturk, PT, PhD, Karadeniz Technical University

간행물 및 유용한 링크

연구에 대한 정보 입력을 담당하는 사람이 자발적으로 이러한 간행물을 제공합니다. 이것은 연구와 관련된 모든 것에 관한 것일 수 있습니다.

일반 간행물

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작 (추정된)

2027년 2월 1일

기본 완료 (추정된)

2028년 2월 1일

연구 완료 (추정된)

2028년 6월 1일

연구 등록 날짜

최초 제출

2026년 10월 30일

QC 기준을 충족하는 최초 제출

2026년 11월 4일

처음 게시됨 (실제)

2026년 11월 6일

연구 기록 업데이트

마지막 업데이트 게시됨 (실제)

2026년 11월 6일

QC 기준을 충족하는 마지막 업데이트 제출

2026년 11월 4일

마지막으로 확인됨

2026년 10월 1일

추가 정보

이 연구와 관련된 용어

개별 참가자 데이터(IPD) 계획

개별 참가자 데이터(IPD)를 공유할 계획입니까?

아니요

IPD 계획 설명

Individual participant data will not be shared publicly because this investigation constitutes the foundational baseline phase of a multi-stage longitudinal research project. The acquired normative volumetric, photographic, and point-cloud datasets will be actively utilized as prospective control and benchmark data in subsequent interrelated clinical and technological validation phases.

약물 및 장치 정보, 연구 문서

미국 FDA 규제 의약품 연구

아니

미국 FDA 규제 기기 제품 연구

아니

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .

구독하다