Clinical Validation of AI-Based Quantitative Eye Movement Analysis From Smartphone 9-Gaze Videos (Glandy EOM) (GLANDY-EOM-VAL)
Prospective Multicenter Observational Study for Clinical Validation of AI-Based Quantitative Eye Movement Analysis Using Smartphone-Recorded 9-Gaze Videos
연구 개요
상태
상세 설명
Ocular motility is commonly evaluated using the 9-gaze examination, which relies on clinician observation and semi-quantitative grading. This approach is inherently subjective and can show inter-observer variability and limited sensitivity to subtle changes over time. Objective, quantitative assessment of eye movement remains an unmet clinical need, particularly for monitoring disease progression and treatment response.
Recent advances in artificial intelligence have enabled automated analysis of facial and ocular features from smartphone-recorded videos, offering the potential for reproducible quantitative measurement of eye movement. Glandy EOM is an investigational software-only medical device (SaMD) developed by THYROSCOPE INC. that processes smartphone-recorded 9-gaze videos to generate quantitative eye movement metrics. Glandy EOM has not yet received regulatory approval and is used for research purposes only in this study.
The study is a prospective, multicenter, multinational observational investigation designed to evaluate the clinical validity of Glandy EOM. The study population consists of adults with suspected or confirmed ocular motility abnormalities; healthy volunteers without any clinical indication for 9-gaze examination are not enrolled. Because eligibility is based on clinical suspicion or diagnosis of ocular motility abnormality rather than confirmed disease at enrollment, some participants may ultimately be clinically judged to have no abnormality on formal evaluation; such subjects remain part of the study population. Participants will undergo a single study visit consisting of (1) a standard clinical 9-gaze examination with clinician grading (-4 to +4), (2) a smartphone video recording performed with the assistance of a trained operator using a standardized setup, and (3) a patient self-recorded smartphone video acquired through the study mobile application.
For quantitative reference measurements, independent rater-derived measurements will be obtained from the recorded videos using a calibrated reference approach. A calibration marker of known size will be placed on the glabella during video acquisition to enable real-world scale estimation for the reference measurements. The calibration marker is used only for rater-based reference measurement and is not used by the AI algorithm.
The primary objective is to evaluate agreement between AI-derived quantitative eye movement metrics and rater-derived measurements of corneal center displacement, expressed as distance (mm) and angular deviation. Secondary objectives are to evaluate correlation between AI-derived metrics and clinician-assessed 9-gaze grading, and to evaluate whether AI-derived metrics obtained from patient self-recorded videos are comparable to those obtained from assistant-aided recordings. The investigational device is used solely for video acquisition and analysis and does not influence clinical decision-making during the study.
연구 유형
등록 (추정된)
연락처 및 위치
연구 연락처
- 이름: Jaemin Park
- 전화번호: +82-52-264-4154
- 이메일: jaemin.park@thyroscope.com
연구 장소
-
-
California
-
San Jose, California, 미국, 95134
- 3003 North First Street #221
-
연락하다:
- Jaemin Park
- 전화번호: +82-52-264-4154
- 이메일: jaemin.park@thyroscope.com
-
-
참여기준
자격 기준
공부할 수 있는 나이
- 성인
- 고령자
건강한 자원 봉사자를 받아들입니다
샘플링 방법
연구 인구
Adults of legal age of consent with suspected or confirmed ocular motility abnormalities, recruited at participating ophthalmology sites.
Some enrolled participants may ultimately be clinically judged to have no abnormality on formal evaluation; these subjects remain part of the study population and are not considered healthy volunteers.
설명
Inclusion Criteria:
- Adults who have reached the legal age of consent for research participation in the applicable jurisdiction.
- Subjects with suspected or known ocular motility abnormalities.
- Subjects who are able to understand and follow gaze instructions required for the 9-gaze examination, as judged by the investigator.
- Subjects who are able to undergo facial video recording using a smartphone.
- Subjects who provide written informed consent.
Exclusion Criteria:
- Conditions that prevent reliable detection of ocular landmarks in video recordings (e.g., severe ptosis, corneal opacity, or significant occlusion of the eyes).
- Ocular or facial conditions that, in the opinion of the investigator, may significantly interfere with image-based analysis.
- Subjects who are unable to comply with study procedures.
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
코호트 및 개입
그룹/코호트 |
개입 / 치료 |
|---|---|
|
Ocular motility abnormality cohort (single cohort)
Adults with suspected or confirmed ocular motility abnormalities scheduled for clinical 9-gaze examination at participating ophthalmology sites.
No separate healthy-volunteer control cohort is enrolled; some participants may ultimately be clinically judged to have no abnormality on formal evaluation but remain part of the study population.
|
Glandy EOM is an investigational software-only medical device developed by THYROSCOPE INC. It has not yet received regulatory approval and is used for research purposes only in this study. Glandy EOM processes smartphone-recorded 9-gaze facial videos to generate AI-derived quantitative eye movement metrics:
AI outputs are generated for research analysis and are NOT used to guide real-time clinical decisions during the study. |
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Agreement between AI-derived and rater-derived corneal center displacement (distance, mm)
기간: At study visit (single visit)
|
Agreement between AI-derived and rater-derived measurements of corneal center displacement expressed as distance (mm) across 9 gaze positions, assessed by intraclass correlation coefficient (ICC) and Bland-Altman analysis.
|
At study visit (single visit)
|
|
Agreement between AI-derived and rater-derived angular deviation (degrees)
기간: At study visit (single visit)
|
Agreement between AI-derived and rater-derived measurements of corneal center displacement expressed as angular deviation (degrees) across 9 gaze positions, assessed by ICC and Bland-Altman analysis.
|
At study visit (single visit)
|
|
Linear association and absolute error vs. rater-derived reference
기간: At study visit (single visit)
|
Linear association and absolute error between AI-derived and rater-derived measurements (both distance and angular deviation), assessed by Pearson correlation coefficient and mean absolute error (MAE), respectively.
|
At study visit (single visit)
|
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Correlation between AI-derived metrics (assistant-aided recording) and clinician-assessed 9-gaze grading
기간: At study visit (single visit)
|
Correlation between AI-derived eye movement metrics obtained from assistant-aided smartphone recordings and clinician-assessed 9-gaze grading (-4 to +4), assessed using Spearman correlation.
|
At study visit (single visit)
|
|
Correlation between AI-derived metrics (patient self-recorded) and clinician-assessed 9-gaze grading
기간: At study visit (single visit)
|
Correlation between AI-derived eye movement metrics obtained from patient self-recorded smartphone videos and clinician-assessed 9-gaze grading (-4 to +4), assessed using Spearman correlation.
|
At study visit (single visit)
|
|
Agreement between AI-derived metrics from patient self-recorded vs. assistant-aided recordings
기간: At study visit (single visit)
|
Agreement between AI-derived metrics obtained from patient self-recorded and assistant-aided smartphone videos, assessed using ICC and Bland-Altman analysis.
|
At study visit (single visit)
|
|
Linear association and absolute error: patient self-recorded vs. assistant-aided
기간: At study visit (single visit)
|
Linear association and absolute error between AI-derived metrics from patient self-recorded and assistant-aided smartphone videos, assessed using Pearson correlation coefficient and mean absolute error (MAE).
|
At study visit (single visit)
|
공동 작업자 및 조사자
스폰서
연구 기록 날짜
연구 주요 날짜
연구 시작 (추정된)
기본 완료 (추정된)
연구 완료 (추정된)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
키워드
추가 관련 MeSH 약관
기타 연구 ID 번호
- THYROSCOPE-EOM-2026-01
개별 참가자 데이터(IPD) 계획
개별 참가자 데이터(IPD)를 공유할 계획입니까?
IPD 계획 설명
약물 및 장치 정보, 연구 문서
미국 FDA 규제 의약품 연구
미국 FDA 규제 기기 제품 연구
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .